Biochemistry Lab Report

Improved Essays
DIAGNOSIS OF ANAEMIA: DETERMINATION OF HAEMOGLOBIN CONCENTRATION
Introduction
Red blood cells contain the protein molecule haemoglobin (Hb); it is made up of four polypeptide chains and contains iron. Its function is to carry oxygen from the lungs to the bodies tissue and to carry carbon dioxide from tissue back to the lungs. Anaemia can be caused when there is a lack of iron in the body which leads to a reduction in the number of red blood cells. Iron is used to carry oxygen in the blood so If you have fewer red blood cells than is normal, your organs and tissues won 't get as much oxygen and this can be damaging (NHS Choices, 2016). Oxygen is used in respiration which provides energy in the form of ATP for things like metabolic reactions
…show more content…
A series of dilutions were prepared by mixing the stock solution was water in a test tube. The concentrations used were 0.139 mg/mL, 0.111 mg/mL, 0.083mg/mL, 0.057mg/mL and 0.028 mg/mL. There were two lots of each concentration made. The spectrophotometer was set at the previously calculate peak wavelength of 409mn and zeroed with deionised water. A cuvette was filled with 0.139 mg/mL and the absorbance was measured, then again with the same concentration from the other test tube and an average was calculated. This was repeated for each concentration. Then a graph was plotted with concentration on the x-axis and the absorbance along the …show more content…
4 shows the absorbance of different concentrations of Hb at 409nm. To calculate the concentration of the solution I used the equation: Final concentration= (stock added (ml) x stock concentration) / total volume (ml). for example, x= (4x0.139)/5 x=0.111

A graph to show the absorbance of different concentrations of Hb at wavelength 409nm

A table to show the absorbance at 409nm wavelength of the blood sample from the female patient.

Discussion
Fig. 1 shows the absorbance of light of Hb at varying wavelengths, from 375nm to 415nm. The table shows the absorbance increases as the wavelength increase, up to 410nm where the absorbance was 0.868. This shows that the peak absorption is close to this wavelength. Fig. 2 shows that the spectrophotometer was increased by 1nm intervals to get an exact reading where the light is absorbed the most, this wavelength can then be used to determine the concentration of Hb in the unknown sample. The highest absorbance was 0.877 at 409nm. Fig. 3 shows this information in a simple and easy to read

Related Documents

  • Improved Essays

    Results show a significant difference between the correspond absorbance and wavelength recordings of the four pigments. Carotene and Xanthophyll absorb the least amount of range within the wavelengths starting at 400 nm and going to 500 nm peaking between an absorption of 0.2 and 0.6. Meaning that Carotene and Xanthophyll only have a range of 100 nm. Chlorophyll a and chlorophyll b absorb the largest range of wavelengths; starting at 400 nm and going to 500 nm. The absorbance then peaks again between 600 nm and 700 nm.…

    • 736 Words
    • 3 Pages
    Improved Essays
  • Superior Essays

    Color Analysis Lab

    • 1098 Words
    • 5 Pages

    The purpose of this lab is to determine the Ka of an indicator within an unknown solution. This can be determined because acid-base indicators are weak bases or acids, and when neutralized they change colors. The color change is measured by the light absorption spectra in the compound which can be used to calculate the concentration. This experiment measures the concentrations by the different pH values which allows the pKa and Ka to be determined. Through the experiment, and the recorded color change with the use of the Spec 20 the unknown Ka of the solution was found.…

    • 1098 Words
    • 5 Pages
    Superior Essays
  • Improved Essays

    From the second part of the example, cell burst is a possible situation can occur to the cell. In this experiment, cell burst, more specifically hemolysis, red blood cell death, lead organelles to spread in the solution and block the absorbance of light in the colorimeter. By comparison of the measured osmolarity and hemolysis between the different solutions provided the data to understand, how a red blood cell react in different osmolarity to maintain…

    • 988 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Demonstration of Cellular Respiration Inside Lima Beans Cells in Vitro Purpose: Cellular respiration is a series of metabolic reactions that take place in the mitochondria and produce chemical energy in the form of ATP by the breakdown of food molecules. There are three processes involved in cellular respiration; glycolysis, the Krebs cycle and the electron transport chain. During this processes, glucose is oxidized by a series of redox reactions and its electrons and hydrogen ions are donated to two electron carriers called NAD+ and FAD.…

    • 776 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Safety Glass is used in every experiment and was always handled carefully, especially when being heated. If a glass object was dropped, the teacher was immediately notified so that the proper cleanup procedure could be done. Hot plates and lamps get hot quickly, so when I was working with them I made sure I did not touch them. In many of the experiments possible eye and skin irritants such as ZnCl₂ as well as toxic materials like HCl are used.…

    • 663 Words
    • 3 Pages
    Improved Essays
  • Improved Essays

    For this portion of the lab, relying on her knowledge of the visible light spectrum and color absorption, the scientist…

    • 843 Words
    • 4 Pages
    Improved Essays
  • Great Essays

    Unknown Sample .91 ppm .9126 Table 1-1: Experimental…

    • 1251 Words
    • 6 Pages
    Great Essays
  • Superior Essays

    Albumin Synthesis

    • 1183 Words
    • 5 Pages

    Through this experiment we measured how fast a chemical reaction occurs, by changing the degradation rate in Albumin when added to different enzymes. Albumin is a protein found in egg white, which is considered to have important storage and nutritional functions. Albumin has also been used in medicine to treat heavy metal intoxication. We ran Albumin through four different conditions to see which would make the Albumin degrade faster. We predicted that the Albumin would degrade with pepsin the fastest since it is essential for digestion of substances in the stomach.…

    • 1183 Words
    • 5 Pages
    Superior Essays
  • Improved Essays

    The optimal wavelength of maximum cranberry juice absorption was determined to be 525 nm. The calibration curve could then be created by measuring the absorbance of known concentrations, 5% to 25%, of cranberry juice at 525 nm, and generating a calibration curve of absorbance vs concentration. The absorbance of the cranberry apple juice was then measured at the wavelength of maximum absorbance. Absorbance is a measure of the amount of light that passes through a sample at a selected wavelength. The absorbance is measured by the spectrophotometer 20 which functions by emitting a visible light source through the sample.…

    • 543 Words
    • 3 Pages
    Improved Essays
  • Great Essays

    Ideally, the sensitivity was calculated to be 50 ppm/Abs units; the slope of the graph is the equivalent of the sensitivity. The observed sensitivity was found to be 44.7 ppm/Abs units. This means that in one increment of absorbance, the concentration is 44.7 ppm or for 0.01 increment of absorbance, the concentration increases by 0.447 ppm. This shows a low sensitivity for the thiocyanate system. The detection limit was also calculated, which was found to be 0.4038 ppm.…

    • 1332 Words
    • 6 Pages
    Great Essays
  • Improved Essays

    Discussion In this study, the Catechol enzyme was studied under the conditions of varying pH, temperature, substrate concentration, and enzyme concentration. In Figure 1, the data suggested that the trend was neither directly nor inversely proportional, but the highest activity rate was at 24°C. Most enzymes denatured at higher temperatures of approximately 40°C, which led to the inability to see any color change (Helms et al., 1998). At lower temperatures, the enzyme was somewhat efficient because molecules move slower at lower temperatures, so enzymes lost productivity.…

    • 753 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Hemoglobin Lab Report

    • 1772 Words
    • 8 Pages

    Light absorbance of the FeSCN2+ was obtained, and use to calculate the concentration of iron using the Beer-Lambert law, which is then used to compute the hemoglobin concentration in the original…

    • 1772 Words
    • 8 Pages
    Improved Essays
  • Improved Essays

    Protein Lab

    • 1039 Words
    • 5 Pages

    The Biuret test is used to prove or disprove this. Many other articles and studies have been done to show that this test is proven to show the amount and sensitivity of protein molecules. This article and study being published by Salle University in the Philippines, tells the reader about the spectrophotometer and how the concentration levels determine its absorption rate and amount.…

    • 1039 Words
    • 5 Pages
    Improved Essays
  • Improved Essays

    The reason we can use absorbance in place of concentration is because of Beer’s Law, which tell us that the absorbance of a solution is directly proportional to its concentration. For this specific experiment there are two reactants so the pseudo-first-oder kinetics method is needed. This can be accomplished by isolating the effects of the concentration of each reactant from the overall reaction rate. Therefore, by running the experiment with hydroxide in large excess, allowed us to view the effect of only the reactant Crystal Violet. By knowing the two different reaction orders for crystal violet and hydroxide, the rate constant “K” was able to be determined.…

    • 703 Words
    • 3 Pages
    Improved Essays
  • Great Essays

    The molar absorption constants were found to be: 358 for the acid form at wavelength 440 nm, 12 for the acidic form at wavelength 590 nm, 64.0 for the basic form at wavelength 440 nm and 558 for the basic form at wavelength 590 nm. Some of the R2 values are off from the ideal value of 1, while some datum were discarded as it did not make sense. Both parts are not very accurate and precise, due to the large range of readings that the spectrophotometer were showing, and the readings never steadied, therefore could not get a very accurate…

    • 1630 Words
    • 7 Pages
    Great Essays