Iron (III)-Tiocyanate Complexs In Equilibrium
Time Duration: The maximum wavelength of the complex was found to be approximately 470 nm. …show more content…
To measure the linearity of the iron(III) thiocyanate system, the 300 ratio was used to optimize the absorbance. The higher ratios provide large absorbance readings, however due to the extra reagent, the maximum wavelength shifts. These shifts reduce the reliability of the data.
As shown to the left (Figure 9), the concentration was plotted against absorbance. The graph shows a linear curve. The correlation coefficient, depicted on the graph, was approximately equal to one. This means that the data follows the linear curve. However, multiple errors can be introduced and lessen the validity on the Beer’s Law …show more content…
Chemical errors can result in not properly determining or making the required solution concentration. Another chemical error can also be side reactions with substances in the solvent. As seen in the anion analysis, solvent composition can have a negative effects on the measure of the absorbance. Errors can also arise from the instrument being used. At low concentrations, the signal read by the spectrophotometer will contain a large amount of noise, meaning that the signal to noise ratio decreases. This causes the spectra to have broader areas. The sensitivity of the iron(III)-thiocyanate system was also determined. Using a series of dilutions, sample concentration were made between 1-4 ppm. The sensitivity was determined by plotting the absorbance against the concentration. As shown below in Figure