NPI

New patient implant. All settings will be set to 0 end of OR case

EOS

End of service/generator replacement. Will do same settings as preop, unless battery completely dead then will be 0

Dosing parameters-Normal mode

1) output current


2) signal frequency


3) pulse width


4) signal on time


5) signal off time

Output current range

0-3.5 mA


Therapeutic range: 1.5-2.25mA

Signal frequency range

1-30 hz


Usually 20-30

Pulse width range

130-1000


Usually 250 or 500

Time on range

7-60 seconds


Usually 30

Time off range

0.2 - 180 minutes


Usually 5 minutes

Output current definition

Amount of electrics current delivered during single pulse stimulation


Therapeutic range: 1.5-2.25mA

What is interval increase increment of output current?

0.25 mA to patients tolerability. Do this about every 2 weeks

Pulse width def

Duration of a single pulse

Signal frequency def

Number of pulses per second (hz)

On time

Time device is delivering FULL output current programmed (does not include ramp up or down)

Off tkme

Amount of time between on times (includes ramp up and down periods)

Duty cycle

Amount of time decide delivering stimulation in a 24 hour period

Magnet dosing parameters

Output current (usually 0.25mA above normal mode current), pulse width, on time


On demand so no off time

How many seizures where magnet is effective

> 60%

Reed switch

On generator, magnet opens and closes it. Must be at least 34 gauss

Excessive stimulation

On time greater than off time

How much higher is magnet mode current than autostim current

At least 0.125

Autostim parameters

Output current (middle between normal and magnet), pulse width and on time


Also will be tachycardia detection

Enforced off time

Always equal to the programmed autostim on time

Goal of dosing

Maximize therapeutic effect


Minimize side effects


Common: output 1.5-2.25mA, pulse width 250/500microdeconds, frequency 20/30 Hz

Order to adjust parameter for size effects

1. Pulse width


2. Frequency


3. Output current-last!


Can adjust autostim theshild for battery life or side effects too

Side effects auto stim related

Verify heartbeat detection


Autostim parameters


Threshold for autostim

Heartbeat detection

1-5


Start with 1= least sensitive

Threshold for autostim

20-70%


Default setting is 70= least sensitive