• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/95

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

95 Cards in this Set

  • Front
  • Back
Localizer Scans
- similar to images acquired w/ conventional radiography
- AP is normal localizer image, w/ laterals used for head & neck CTs
- image quality is poor compared w/ normal x-ray images
Why are localizer scans important?
- includes all anatomy within the scannable range (z direction)
- miscentering can result in artifacts (patient must be centered between x & y axes [isocenter])
- if patient is misaligned, tech can go correct positioning and re-take localizers (decreases amount of radiation for patient)
It is imperative that CT techs input the correct _________ instructions before data acquisition is initiated.
Directional

head first vs. feet first
supine, prone, or decubitus
What can incorrectly inputting directional instructions cause on CT scans?
- mislabeld images
- misdiagnosis, serious medical errors
Z- Axis Coverage
- localizer scans prescribe the location of slices
- beginning & end landmarks found on localizer scans
DFOV & Image Center
- localizers used to select optimal DFOV
- improved selection basis when using AP & lateral localizers
Step-and-Shoot Scanning
- axial scanning, conventional scanning
- table moves to location, tube rotates once within gantry, table moves to next location, repeat
- used for brain studies
- helps reduce halo & windmill artifacts produced by helical scans
- images perpendicular to to z axis & parallel to ever other slice (ex. slices of sausage)
Interscan Delay
Slight pause in step-and-shoot scanning when table is moving to new location

Can add to exam time
Advantage of Step-and-Shoot Images
Data is acquired in true axial plane & requires less manipulation for display

Can be continguous or non-
Cine or Dynamic Methods
Used for biopsy or to observe progression of IV contrast through vascular structures
Disadvantages of Step-and-Shoot Method
Interscan delay time adds to exam length

Can be difficult for exams needing patients to hold their breath

Slice misregistration common in abdomen
Single-Detector Row Systems
- before 1990 all scanners were like this (one slice per gantry rotation of tube)
- entire detector 20 mm wide, common slice thickness = 10 mm
How to calculate area of patient anatomy to be covered w/ single detector systems
Multiply slice increment by numbers of slices

Ex. Chest CT uses continguous, 4 mm slices with 60 slices planned to be acquired. How much anatomy will be covered?
4 mm x 60 = 240 mm
Multidetector Row Systems (MDCT)
- may contain 4 to 64 parallel rows of detector elements
- provides longer & faster z axis coverage per gantry rotation
How is slice thickness determined with MDCT systems
A combo of x-ray beam width & detector configuration

Minimum slice thickness limited by detector width
When are Axial Scans used
- protocols in which acquisition speed isn't a major concern
- optimal resolution required
- can be used to reduce patient dose by using gaps between slices
Helical Scanning (Spiral, Volumetric)
- introduced late 1980s
- continually rotating tube, constant x-ray output, uninterrupted table mvmt
- eliminates interscan delay
Advantages of Helical Scanning
- easy to use with contrast agent
- reduces respiratory misregistration
- reduces motion artifacts from organs
- fast (most scanners have gantry rotation time of 0.3 sec per 360 degree rotation)
What does the slip ring design allow CT scanners to do?
Continually rotate the x-ray tube
How are slices produced w/ Helical scanning
Produced at a slight tilt, like the rungs on a spring (because anatomy isn't acquired @ true axial plane)
Helical Interpolation
- take the slant & blur out of helical images
- uses complex statistical methods to create "normal" images (bases attenuation coefficient of single voxel to that of the others surrounding it)
- can be described as "best guess"
-
As the degree of helical interpolation INCREASES, accuracy of info & image resolution ___________ (increases/decreases)
Decreases
Slice Thickness Blooming
- interpolation methods that result in a scan that's wider than that selected by operator
MDCT Helical Pitch
- describes CT table movement during helical scan
- pitch = table distance/collimation width

Ex. 16-slice scanner, 0.5 mm slice thickness, table moves 12 mm per rotation

Pitch = 12/(16 x 0.5) = 12/8 = 1.5 pitch
When table feed and beam collimation are identical, pitch is....
1
When does scan overlap occur?
When table feed is less than beam collimation (pitch less than 1)
SDCT Pitch
- table speed varies according to slice thickness & gantry rotation speed
- pitch = table distance/slice thickness

Ex. Slice thickness is 5 mm, table travels 10 mm each second (pitch = 2, slant of "springs" more pronounced)
Increasing pitch helps _________ (increase/decrease) speed of exam, but is offset by the ____________ (increase/decrease) in interpolation required.
Increase/increase
Will an increase in SDCT pitch have an effect on radiation dose?
Helps reduce dose to patient
SDCT Scan Coverage Equation
Number of images = (pitch x total acquisition time x 1) / rotation time
SDCT Distance Covered Equation
Amt of Anatomy covered = (pitch x total acquisition time x 1) / (rotation time x slice thickness)
MDCT Scan Coverage Equation
Amt of Anatomy covered = (pitch x total acquisition time x 1)/rotation time x (slice thickness x slices per rotation)
Does helical scanning allow for slice incrementation changes?
- yes; retrospectively
- allows creation of overlapping slices w/o increasing radiation dose to patient
Dual-Source CT
- uses 2 sets of x-ray tubes & 2 corresponding detector arrays in single CT gantry
- primary goal = increase scan speed
- used w/ contrast exams, 1 tube w/ standard energy & 1 tube w/ contrast energy, merge both images into 1
Scan time is determined by...
Slice Thickness & Speed of Tube Rotation
The more slices we can acquire per tube rotation, the _________ (faster/slower) the scan will be
Faster
How many filaments do most CT tubes have?
2 (one for low mA settings, other for high mA settings)

Higher mA settings allow for shorter scan times

Small filaments improve resolution
kVp ranges for CT
80-150 kVp
Uncoupling Effect
- direct relationship between quantity of exposure, # of photons the IR gets, and the amount of density you'll see on the image
- in CT, when not enough photons hit the detector, the image appears grainy or noisy (as opposed to bright white on x-ray image)
Automatic Tube Current Modulation
- software that automatically adjust mAs to fit the specific anatomic region
- 15-40% reduction in dose w/o losing image quality
- uses less mAs on less thick body parts
Slice Thickness & FOV
- SFOV determined area within gantry where raw data is acquired, DFOV determines how much of raw data used to create image
- slice thickness selected for reconstruction can be no smaller than acquired slice thickness / DFOV cannot be larger than SFOV
Scan Geometry - Tube Arc
- 360 degrees (full scan) is most common
- 180 degrees (partial, half-scan) minimum scan required to make an image
- 400 degrees (overscans [used in 4th gen scanners to reduce motion artifact])
Spatial Resolution (Detail Resolution)
Ability to resolve (as separate objects) small, high-contrast objects

DXR has higher than CT
Contrast Resolution
Ability to differentiate btw. objects w/ very similar densities as their background

CT is superior to DXR
How is CT spatial resolution measured
- directly using line pairs phantom (lp/mm)
- data analysis modulation transfer function (MTF)
Point Spread Function (PSF) - CT Contrast Resolution
- describes lack of sharpness that results when point in the object isn't reproduced as a "true" point in image
- results in blurring effect
Line Spread Function (LSF) - CT Contrast Resolution
- describes unsharpness of an imaging system when line or slit object isn't reproduced as a line or slit image; instead it spreads out as measurable distance
Contrast Transfer Function (CTF) - CT Contrast Resolution
- measures contrast response of system using contrast phantom (series of slits, holes & spaces; resultant contrast is diff. in density btw regions of slits)
Modulation Transfer Function (MTF) - CT Contrast Resolution
- measures resolution capabilities of system by breaking down an object into its frequency components (FFT) [ratio of detail seen in image to the detail existing in actual object]
- MTF of 1 means system has reproduced object exactly (MTF of 0 indicates no detail of object is seen in image)
- can be charted to compare object resolution
How many dimensions can spatial resolution be described in?
2: in-plane resolution (resolution in x, y direction) & longitudinal resolution (resolution in z direction)
Factors that produce the highest Spatial Resolution are...
Large matrix, small DFOV, small pixel size, thin slice thickness, sharp filters (algorithms), small focal spot size (controlled by filament), low ratio pitch (higher pitch = more volume averaging), decreased patient motion
On CT images, objects with a _____% contrast variation can be distinguished.
0.5%
__________ plays an important role in low-contrast resolution
Noise (undesirable fluctuation of pixel values in an image of homogenous material) (want high SNR)

Presence of noise = degraded image quality
Factors that produce the highest Contrast Resolution are...
Higher mAs, high SNR, small pixel size (less volume averaging), thin slices (smaller voxels), soft tissue filters (algorithms), smaller patients
Inherent Contrast
Physical properties of the object & its background (other objects in the SFOV that shouldn't be [ex. buttons on shirt, EKG leads])
Temporal Resolution
- related to time, how are the tissues going to change as they're scanned
- controlled by: gantry rotation speed, # of detector channels, & system speed
- reported in ms (= 1 sec)
Temporal Resolution & Cardiac CT Scanning
- segmented reconstruction (gated to ECG so tube only activates when heart is at rest)
- retrospective gating (removes unwanted info acquired during heartbeat)
Why are quality assurance (QA) protocols so important to follow?
- ensures that you provide the best possible service to patients
- ensures the system is providing the best image quality
- can detect potential problems early, which saves money & machine down-time
What's the most important aspect of QA?
Consistency

Tests must be performed on regular basis, must be recorded using a consistent format, & the testing parameter must be within specified guidelines
Daily CT QA Tests
- average CT number of water (mean of uniform water phantom should be +- 2 HU, SD within 10)
- linearity tests
- cross-field uniformity tests
Monthly CT QA Tests
- high & low contrast resolution tests
Annual CT QA Tests
- table indexing (ensures the table has accurate movement scale)
- distance measuring device accuracy (errors less than 2 mm)
- laser light accuracy (errors less than 2 mm)
- pitch & slice thickness QA tests
Slice Thickness Accuracy
- measured using a step-wedge, ramp, or spiral phantom
- provides a standard to compare w/ scanner
What is a water phantom used to measure?
Noise & Uniformity

Speckles on image indicate a variation in density in water phantom (mostly caused by inherent noise within x-ray system)
Linearity
- relationship btw. CT numbers & linear attenuation values of scanned object @ designated kVp value
- measured using phantom filled w/ varying dense objects
How are radiation levels measured for a CT scanner?
- have to be measured by medical physicist
- obtained using standard head & body phantoms w/ pencil ionization chamber
- should be performed annually
What can CT image artifacts be broadly classified as?
Physics-based, Patient-based, Equipment-induced
Beam-Hardening Artifacts
- caused by polychromatic x-ray beams( lower-energy photons are absorbed creating a harder beam); seen as cupping artifacts or dark bands/streaks btw dense objects in image
- fixed by: filtration, correct calibration, beam-hardening correction software
- most often occurs in head, shoulders, & hips
Partial Volume Artifacts
- occurs when more than 1 type of tissue is contained within a voxel
- creates a fuzzy outer border where volume averaging occurred
- fixed by: thin slices
Edge Gradient Effect Artifacts
- results in streaks or shadings arising from irregularly shaped objects w/ pronounced density difference from surrounding structures
- seen in stomach w/ barium
- fixed by: thin slices, low HU oral contrast (instead of barium)
Motion Artifacts
- appear as shading, streaking, blurring, or ghosting
- helical scanning helps decrease the amount of these
- fixed by: adequate instructions for patient to follow, using short scan time
Metallic Artifacts
- create streak artifacts
- fixed by: removing metal (if possible), angling gantry (if irremovable), thin slices
Out-of-Field Artifacts
- caused by anatomy extending outside of SFOV
- appear as shading & streaking
- seen w/ patients too large for gantry or patients improperly centered within gantry
Ring Artifacts
- caused by imperfect detector elements; appear as concentric rings on image
- equipment-type artifact
- fixed by: recalibrating scanner
Tube Arcing
- aka high-voltage arching
- created from electrical surges within x-ray tube
- can be minor or severe; usually produces error message
- requires service call
Spiral & Cone Beam Effect
- ONLY ON HELICAL SCANS
- created by interpolation & reconstruction processes; caused by cone-shaped beam divergence (the more detectors there are, the more problems)
- creates subtle inaccuracies in CT #s, easily misinterpreted as disease
- fixed by: using low pitch
The vertex of the head causes ______________ artifacts on helical CT scans
Windmill
What happens with...improper concentration of contrast media
- looks like white streaks emanating from center
- CT oral contrast should be 1/10th concentration used for conventional radiography
- fixed by: rescheduling exam, cleansing enema
Post-processing reconstruction involves...
Raw data manipulated to create pixels which are used to create an image
Post-processing reformation involves...
Using image data assembled together to produce images in different planes and 3D images
What parameters can be reconstructed restrospectively?
DFOV, image center, reconstruction algorithm, slice incrementation (helical only), image thickness (MDCT only)

Only can happen b/c reconstruction uses raw, not image, data
What does changing slice incrementation do with helical scan data?
Creates overlapping images (useful in MPR & 3D reformations)
- help improve accuracy of reconstruction
Why is MDCT systems' ability to retrospectively change image thickness important?
Scans can be performed with thin slices (i.e. less radiation) and reconstructed into thicker slices for viewing/storing (imaging files are easier to maintain)
Image Reformation (Image Rendering)
- in order for CT scans to be reformatted, all source images must have identical: DFOV, image center, gantry tilt, & have no gaps (uses image data; parameters can't be manipulated)
Multiplanar Reformation (MPR)
- 2D in nature, allows images to keep original CT attenuation values
- created from the volume of information to display information present within a plane of the volume
- can be created in any plane
Curved Planar Reformation (CPR)
- created along center line of tubular organs
- allow for visualization of vascular structures to determine course & patency
3D Reformation
- represent entire scan volume in 1 image
- manipulates & combines CT values to display an image (doesn't keep original info)
- draws an imaginary line from viewer through data volume
3D Reformation - Surface Rendering
- shaded-surface display (SSD); provides good detail of size & shape of structure
- provides no info about internal structures
- only uses voxels on surface of structure
- largely replaced by volume rendering
3D Reformation - Projection Displays
- maximum-intensity projection (MIP)
- uses selected voxels w/ highest display value
- a volume rendering technique using 3D voxels to produce a 2D image; limited display of depth
- useful for angiography of contrast-enhanced vasculature
3D Reformation - Projection Displays (Minimum Intensity Projection [MinIP])
- selects voxels w/ lowest value to display
- works well for low attenuation structures, like those containing air (trachea, bronchi)
3D Reformation - Volume Rendering
- favored 3D image technique (all voxels contribute to final image)
- 3D semi-transparent representation of structure
- color or grayscale; provides depth to adjacent tissues
- good for demonstrating complex 3D anatomy
How to tell the difference between a MIP (maximum-intensity projection) image and a VRT (volume-rendering technique) image
Look for the appearance of depth; shows on VRT images, MIP images are only 2D
3D Reformation - Endoluminal Imaging
- form of volume rendering
- designed to look inside lumen of structure
- virtual CT takes volume info from lumen & allows one to take a fly tour through the structure (think: virtual colonoscopy)
- provides a much more comprehensive exam
Region-of-Interest Editing
- done to remove obscuring structures from 3D image ("clean up" of images)
- software allows this to be done manually, automatically, or a mixture of both
Factors that Degrade Reformatted Images
- segmentation errors (check for auto segmentation)
- image noise
- artifacts (motion, metal, stair-step)