Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
95 Cards in this Set
- Front
- Back
Localizer Scans
|
- similar to images acquired w/ conventional radiography
- AP is normal localizer image, w/ laterals used for head & neck CTs - image quality is poor compared w/ normal x-ray images |
|
Why are localizer scans important?
|
- includes all anatomy within the scannable range (z direction)
- miscentering can result in artifacts (patient must be centered between x & y axes [isocenter]) - if patient is misaligned, tech can go correct positioning and re-take localizers (decreases amount of radiation for patient) |
|
It is imperative that CT techs input the correct _________ instructions before data acquisition is initiated.
|
Directional
head first vs. feet first supine, prone, or decubitus |
|
What can incorrectly inputting directional instructions cause on CT scans?
|
- mislabeld images
- misdiagnosis, serious medical errors |
|
Z- Axis Coverage
|
- localizer scans prescribe the location of slices
- beginning & end landmarks found on localizer scans |
|
DFOV & Image Center
|
- localizers used to select optimal DFOV
- improved selection basis when using AP & lateral localizers |
|
Step-and-Shoot Scanning
|
- axial scanning, conventional scanning
- table moves to location, tube rotates once within gantry, table moves to next location, repeat - used for brain studies - helps reduce halo & windmill artifacts produced by helical scans - images perpendicular to to z axis & parallel to ever other slice (ex. slices of sausage) |
|
Interscan Delay
|
Slight pause in step-and-shoot scanning when table is moving to new location
Can add to exam time |
|
Advantage of Step-and-Shoot Images
|
Data is acquired in true axial plane & requires less manipulation for display
Can be continguous or non- |
|
Cine or Dynamic Methods
|
Used for biopsy or to observe progression of IV contrast through vascular structures
|
|
Disadvantages of Step-and-Shoot Method
|
Interscan delay time adds to exam length
Can be difficult for exams needing patients to hold their breath Slice misregistration common in abdomen |
|
Single-Detector Row Systems
|
- before 1990 all scanners were like this (one slice per gantry rotation of tube)
- entire detector 20 mm wide, common slice thickness = 10 mm |
|
How to calculate area of patient anatomy to be covered w/ single detector systems
|
Multiply slice increment by numbers of slices
Ex. Chest CT uses continguous, 4 mm slices with 60 slices planned to be acquired. How much anatomy will be covered? 4 mm x 60 = 240 mm |
|
Multidetector Row Systems (MDCT)
|
- may contain 4 to 64 parallel rows of detector elements
- provides longer & faster z axis coverage per gantry rotation |
|
How is slice thickness determined with MDCT systems
|
A combo of x-ray beam width & detector configuration
Minimum slice thickness limited by detector width |
|
When are Axial Scans used
|
- protocols in which acquisition speed isn't a major concern
- optimal resolution required - can be used to reduce patient dose by using gaps between slices |
|
Helical Scanning (Spiral, Volumetric)
|
- introduced late 1980s
- continually rotating tube, constant x-ray output, uninterrupted table mvmt - eliminates interscan delay |
|
Advantages of Helical Scanning
|
- easy to use with contrast agent
- reduces respiratory misregistration - reduces motion artifacts from organs - fast (most scanners have gantry rotation time of 0.3 sec per 360 degree rotation) |
|
What does the slip ring design allow CT scanners to do?
|
Continually rotate the x-ray tube
|
|
How are slices produced w/ Helical scanning
|
Produced at a slight tilt, like the rungs on a spring (because anatomy isn't acquired @ true axial plane)
|
|
Helical Interpolation
|
- take the slant & blur out of helical images
- uses complex statistical methods to create "normal" images (bases attenuation coefficient of single voxel to that of the others surrounding it) - can be described as "best guess" - |
|
As the degree of helical interpolation INCREASES, accuracy of info & image resolution ___________ (increases/decreases)
|
Decreases
|
|
Slice Thickness Blooming
|
- interpolation methods that result in a scan that's wider than that selected by operator
|
|
MDCT Helical Pitch
|
- describes CT table movement during helical scan
- pitch = table distance/collimation width Ex. 16-slice scanner, 0.5 mm slice thickness, table moves 12 mm per rotation Pitch = 12/(16 x 0.5) = 12/8 = 1.5 pitch |
|
When table feed and beam collimation are identical, pitch is....
|
1
|
|
When does scan overlap occur?
|
When table feed is less than beam collimation (pitch less than 1)
|
|
SDCT Pitch
|
- table speed varies according to slice thickness & gantry rotation speed
- pitch = table distance/slice thickness Ex. Slice thickness is 5 mm, table travels 10 mm each second (pitch = 2, slant of "springs" more pronounced) |
|
Increasing pitch helps _________ (increase/decrease) speed of exam, but is offset by the ____________ (increase/decrease) in interpolation required.
|
Increase/increase
|
|
Will an increase in SDCT pitch have an effect on radiation dose?
|
Helps reduce dose to patient
|
|
SDCT Scan Coverage Equation
|
Number of images = (pitch x total acquisition time x 1) / rotation time
|
|
SDCT Distance Covered Equation
|
Amt of Anatomy covered = (pitch x total acquisition time x 1) / (rotation time x slice thickness)
|
|
MDCT Scan Coverage Equation
|
Amt of Anatomy covered = (pitch x total acquisition time x 1)/rotation time x (slice thickness x slices per rotation)
|
|
Does helical scanning allow for slice incrementation changes?
|
- yes; retrospectively
- allows creation of overlapping slices w/o increasing radiation dose to patient |
|
Dual-Source CT
|
- uses 2 sets of x-ray tubes & 2 corresponding detector arrays in single CT gantry
- primary goal = increase scan speed - used w/ contrast exams, 1 tube w/ standard energy & 1 tube w/ contrast energy, merge both images into 1 |
|
Scan time is determined by...
|
Slice Thickness & Speed of Tube Rotation
|
|
The more slices we can acquire per tube rotation, the _________ (faster/slower) the scan will be
|
Faster
|
|
How many filaments do most CT tubes have?
|
2 (one for low mA settings, other for high mA settings)
Higher mA settings allow for shorter scan times Small filaments improve resolution |
|
kVp ranges for CT
|
80-150 kVp
|
|
Uncoupling Effect
|
- direct relationship between quantity of exposure, # of photons the IR gets, and the amount of density you'll see on the image
- in CT, when not enough photons hit the detector, the image appears grainy or noisy (as opposed to bright white on x-ray image) |
|
Automatic Tube Current Modulation
|
- software that automatically adjust mAs to fit the specific anatomic region
- 15-40% reduction in dose w/o losing image quality - uses less mAs on less thick body parts |
|
Slice Thickness & FOV
|
- SFOV determined area within gantry where raw data is acquired, DFOV determines how much of raw data used to create image
- slice thickness selected for reconstruction can be no smaller than acquired slice thickness / DFOV cannot be larger than SFOV |
|
Scan Geometry - Tube Arc
|
- 360 degrees (full scan) is most common
- 180 degrees (partial, half-scan) minimum scan required to make an image - 400 degrees (overscans [used in 4th gen scanners to reduce motion artifact]) |
|
Spatial Resolution (Detail Resolution)
|
Ability to resolve (as separate objects) small, high-contrast objects
DXR has higher than CT |
|
Contrast Resolution
|
Ability to differentiate btw. objects w/ very similar densities as their background
CT is superior to DXR |
|
How is CT spatial resolution measured
|
- directly using line pairs phantom (lp/mm)
- data analysis modulation transfer function (MTF) |
|
Point Spread Function (PSF) - CT Contrast Resolution
|
- describes lack of sharpness that results when point in the object isn't reproduced as a "true" point in image
- results in blurring effect |
|
Line Spread Function (LSF) - CT Contrast Resolution
|
- describes unsharpness of an imaging system when line or slit object isn't reproduced as a line or slit image; instead it spreads out as measurable distance
|
|
Contrast Transfer Function (CTF) - CT Contrast Resolution
|
- measures contrast response of system using contrast phantom (series of slits, holes & spaces; resultant contrast is diff. in density btw regions of slits)
|
|
Modulation Transfer Function (MTF) - CT Contrast Resolution
|
- measures resolution capabilities of system by breaking down an object into its frequency components (FFT) [ratio of detail seen in image to the detail existing in actual object]
- MTF of 1 means system has reproduced object exactly (MTF of 0 indicates no detail of object is seen in image) - can be charted to compare object resolution |
|
How many dimensions can spatial resolution be described in?
|
2: in-plane resolution (resolution in x, y direction) & longitudinal resolution (resolution in z direction)
|
|
Factors that produce the highest Spatial Resolution are...
|
Large matrix, small DFOV, small pixel size, thin slice thickness, sharp filters (algorithms), small focal spot size (controlled by filament), low ratio pitch (higher pitch = more volume averaging), decreased patient motion
|
|
On CT images, objects with a _____% contrast variation can be distinguished.
|
0.5%
|
|
__________ plays an important role in low-contrast resolution
|
Noise (undesirable fluctuation of pixel values in an image of homogenous material) (want high SNR)
Presence of noise = degraded image quality |
|
Factors that produce the highest Contrast Resolution are...
|
Higher mAs, high SNR, small pixel size (less volume averaging), thin slices (smaller voxels), soft tissue filters (algorithms), smaller patients
|
|
Inherent Contrast
|
Physical properties of the object & its background (other objects in the SFOV that shouldn't be [ex. buttons on shirt, EKG leads])
|
|
Temporal Resolution
|
- related to time, how are the tissues going to change as they're scanned
- controlled by: gantry rotation speed, # of detector channels, & system speed - reported in ms (= 1 sec) |
|
Temporal Resolution & Cardiac CT Scanning
|
- segmented reconstruction (gated to ECG so tube only activates when heart is at rest)
- retrospective gating (removes unwanted info acquired during heartbeat) |
|
Why are quality assurance (QA) protocols so important to follow?
|
- ensures that you provide the best possible service to patients
- ensures the system is providing the best image quality - can detect potential problems early, which saves money & machine down-time |
|
What's the most important aspect of QA?
|
Consistency
Tests must be performed on regular basis, must be recorded using a consistent format, & the testing parameter must be within specified guidelines |
|
Daily CT QA Tests
|
- average CT number of water (mean of uniform water phantom should be +- 2 HU, SD within 10)
- linearity tests - cross-field uniformity tests |
|
Monthly CT QA Tests
|
- high & low contrast resolution tests
|
|
Annual CT QA Tests
|
- table indexing (ensures the table has accurate movement scale)
- distance measuring device accuracy (errors less than 2 mm) - laser light accuracy (errors less than 2 mm) - pitch & slice thickness QA tests |
|
Slice Thickness Accuracy
|
- measured using a step-wedge, ramp, or spiral phantom
- provides a standard to compare w/ scanner |
|
What is a water phantom used to measure?
|
Noise & Uniformity
Speckles on image indicate a variation in density in water phantom (mostly caused by inherent noise within x-ray system) |
|
Linearity
|
- relationship btw. CT numbers & linear attenuation values of scanned object @ designated kVp value
- measured using phantom filled w/ varying dense objects |
|
How are radiation levels measured for a CT scanner?
|
- have to be measured by medical physicist
- obtained using standard head & body phantoms w/ pencil ionization chamber - should be performed annually |
|
What can CT image artifacts be broadly classified as?
|
Physics-based, Patient-based, Equipment-induced
|
|
Beam-Hardening Artifacts
|
- caused by polychromatic x-ray beams( lower-energy photons are absorbed creating a harder beam); seen as cupping artifacts or dark bands/streaks btw dense objects in image
- fixed by: filtration, correct calibration, beam-hardening correction software - most often occurs in head, shoulders, & hips |
|
Partial Volume Artifacts
|
- occurs when more than 1 type of tissue is contained within a voxel
- creates a fuzzy outer border where volume averaging occurred - fixed by: thin slices |
|
Edge Gradient Effect Artifacts
|
- results in streaks or shadings arising from irregularly shaped objects w/ pronounced density difference from surrounding structures
- seen in stomach w/ barium - fixed by: thin slices, low HU oral contrast (instead of barium) |
|
Motion Artifacts
|
- appear as shading, streaking, blurring, or ghosting
- helical scanning helps decrease the amount of these - fixed by: adequate instructions for patient to follow, using short scan time |
|
Metallic Artifacts
|
- create streak artifacts
- fixed by: removing metal (if possible), angling gantry (if irremovable), thin slices |
|
Out-of-Field Artifacts
|
- caused by anatomy extending outside of SFOV
- appear as shading & streaking - seen w/ patients too large for gantry or patients improperly centered within gantry |
|
Ring Artifacts
|
- caused by imperfect detector elements; appear as concentric rings on image
- equipment-type artifact - fixed by: recalibrating scanner |
|
Tube Arcing
|
- aka high-voltage arching
- created from electrical surges within x-ray tube - can be minor or severe; usually produces error message - requires service call |
|
Spiral & Cone Beam Effect
|
- ONLY ON HELICAL SCANS
- created by interpolation & reconstruction processes; caused by cone-shaped beam divergence (the more detectors there are, the more problems) - creates subtle inaccuracies in CT #s, easily misinterpreted as disease - fixed by: using low pitch |
|
The vertex of the head causes ______________ artifacts on helical CT scans
|
Windmill
|
|
What happens with...improper concentration of contrast media
|
- looks like white streaks emanating from center
- CT oral contrast should be 1/10th concentration used for conventional radiography - fixed by: rescheduling exam, cleansing enema |
|
Post-processing reconstruction involves...
|
Raw data manipulated to create pixels which are used to create an image
|
|
Post-processing reformation involves...
|
Using image data assembled together to produce images in different planes and 3D images
|
|
What parameters can be reconstructed restrospectively?
|
DFOV, image center, reconstruction algorithm, slice incrementation (helical only), image thickness (MDCT only)
Only can happen b/c reconstruction uses raw, not image, data |
|
What does changing slice incrementation do with helical scan data?
|
Creates overlapping images (useful in MPR & 3D reformations)
- help improve accuracy of reconstruction |
|
Why is MDCT systems' ability to retrospectively change image thickness important?
|
Scans can be performed with thin slices (i.e. less radiation) and reconstructed into thicker slices for viewing/storing (imaging files are easier to maintain)
|
|
Image Reformation (Image Rendering)
|
- in order for CT scans to be reformatted, all source images must have identical: DFOV, image center, gantry tilt, & have no gaps (uses image data; parameters can't be manipulated)
|
|
Multiplanar Reformation (MPR)
|
- 2D in nature, allows images to keep original CT attenuation values
- created from the volume of information to display information present within a plane of the volume - can be created in any plane |
|
Curved Planar Reformation (CPR)
|
- created along center line of tubular organs
- allow for visualization of vascular structures to determine course & patency |
|
3D Reformation
|
- represent entire scan volume in 1 image
- manipulates & combines CT values to display an image (doesn't keep original info) - draws an imaginary line from viewer through data volume |
|
3D Reformation - Surface Rendering
|
- shaded-surface display (SSD); provides good detail of size & shape of structure
- provides no info about internal structures - only uses voxels on surface of structure - largely replaced by volume rendering |
|
3D Reformation - Projection Displays
|
- maximum-intensity projection (MIP)
- uses selected voxels w/ highest display value - a volume rendering technique using 3D voxels to produce a 2D image; limited display of depth - useful for angiography of contrast-enhanced vasculature |
|
3D Reformation - Projection Displays (Minimum Intensity Projection [MinIP])
|
- selects voxels w/ lowest value to display
- works well for low attenuation structures, like those containing air (trachea, bronchi) |
|
3D Reformation - Volume Rendering
|
- favored 3D image technique (all voxels contribute to final image)
- 3D semi-transparent representation of structure - color or grayscale; provides depth to adjacent tissues - good for demonstrating complex 3D anatomy |
|
How to tell the difference between a MIP (maximum-intensity projection) image and a VRT (volume-rendering technique) image
|
Look for the appearance of depth; shows on VRT images, MIP images are only 2D
|
|
3D Reformation - Endoluminal Imaging
|
- form of volume rendering
- designed to look inside lumen of structure - virtual CT takes volume info from lumen & allows one to take a fly tour through the structure (think: virtual colonoscopy) - provides a much more comprehensive exam |
|
Region-of-Interest Editing
|
- done to remove obscuring structures from 3D image ("clean up" of images)
- software allows this to be done manually, automatically, or a mixture of both |
|
Factors that Degrade Reformatted Images
|
- segmentation errors (check for auto segmentation)
- image noise - artifacts (motion, metal, stair-step) |