• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/104

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

104 Cards in this Set

  • Front
  • Back
Scatter radiation is produced during a _________ interaction
Compton (no use in demonstrating structures of interest)
Do we worry about coherent scatter in diagnostic radiography?
No because diagnostic energy range is 50-100 keV and the energy required for coherent is 10 keV
Scatter radiation will reduce contrast by adding...
More shades of gray
Why does increasing scatter increase the density?
Each photon that hits the IR creates density; a single high energy photon can cause multiple interactions, causing more density
As scatter increases, radiographic density _____________ (increases/decreases) and there is a __________ (increase /decrease) in contrast
Increases; decrease (more grays)
As scatter decreases, radiographic density _____________ (increases/decreases) and there is a __________ (increase /decrease) in contrast
Decreases; increase (fewer grays)
If we increase kV and leave all other technical factors constant, what will happen?
Overall increase in density b/c more photons are exiting the patient & striking the film
What kind of interactions increase with an increase in kVp?
Compton interactions (outer shell electron)
As tissue volume increases, the amount of scatter radiation __________ (increases/decreases)
Increases
What are the 2 reasons to reduce your field size to only the area of interest?
1) Controls amount of scatter produced
2) Reduce patient dose
Which interaction is preferred for diagnostic imaging, Compton or Photoelectric?
Photoelectric
As the volume of the irradiated tissue increases, which type of interaction increases?
Compton Scatter
As the atomic number of the irradiated material increases, which type of interaction increases?
Compton
An increase in _______ (mAs/kVp) would need to occur when the field size is collimated small.
mAs
Aperture Diaphragm
- simplest type of beam-restricting device
- flat piece of lead w/ an opening (aperture) to allow radiation to pass through
- opening can be any shape; size of opening isn't adjustable
- inexpensive
In an aperture diaphragm, the field size to be radiated is controlled by the _____
SID (increasing SID will increase field size to be radiated)
Penumbra
- "fuzzy area" out toward the edges of the primary beam
- the less we have, the better our image detail will be ("use the smallest focal spot")
Aperture Diaphragm effect upon Penumbra
Little impact
Cylinder effect upon Penumbra
Great effect
Cones/Cylinders
- circular aperture diaphragm w/ metal extensions (cones are flared @ ends)
- don't have much of an effect on DR as they do on film
- cylinders are constant in diameter, adjustable in length
Which, a cone or a cylinder, is better at reducing penumbra?
Cylinder
Adjustable Cylinders
- attach beneath & on the outside of tube housing
- as cylinder length increases the amount of radiation reaching patient/film decreases
- increasing cylinder length will decrease the field size
- off-focus radiation will be reduced
Equation for Determining Cone Field Size
Projected Image Size = (SID x lower diameter of cone)/distance from focal spot to bottom of cone
What's the most frequently used method for beam restriction?
Collimation
What do the bottom shutters of the collimator control?
Reduce penumbra along periphery of beam
What do the upper shutters of the collimator control?
Attenuate most of the off-focus radiation from reaching film
Collimator Test
- a test done to check accurate alignment of the light source & the x-ray beam
- US Federal Standard = 2% SID error (ex. 40" SID x 2% = .8" allowable error)
Automatic Collimators
- positive beam limitation (PBL) device
- adjusts dimension of x-ray beam to IR placed in bucky
- will allow you to collimate smaller, but not bigger than the field size of the IR
What does the mirror in a collimator act as?
A filter that absorbs low energy x-ray photons to 1 mm/Al eq

(Added Filtration)
What is the total filtration goal standard?
2.5 mm of aluminum equivalent (mm/Al eq)
Ancillary Devices
- lead blockers are pliable lead sheets that prevent radiation from striking the film cassette
- lead masks cut to specific shape & size
What's the purpose of a grid?
To improve the contrast of the image by absorbing the patient-generated scatter radiation (Compton) before it strikes the film
When are grids used?
1) Body parts larger than 10 cm
2) Procedures using 60 kVp or higher
What are the vertical strips in a grid designed to do?
Absorb scatter radiation (lead) RADIOPAQUE
What are the interspace materials in a grid designed to do?
Allow remnant photons to strike the film (aluminum or plastic fiber) RADIOLUCENT
How will the grid strips be running in a rectangular grid?
With the long axis of the rectangle (used for portables)
3 Expectations about Outcome of Image when Using Grids
1) Higher contrast
2) Reduction in density (fewer photons hit IR)
3) Grid artifacts
What is the grid ratio?
The relationship of the HEIGHT of the lead strips to the DISTANCE BETWEEN the strips

GR = h / D
What does the thickness of the lead strips in a grid represent in the grid ratio?
Nothing (irrelevant in terms of calculating grid ratio)
A(n) __________ (proportional/inverse) relationship exists between the distance between the lead strips & grid ratio WHEN the height of the grid strips is constant
Inverse
High Grid Ratio
- stops scatter radiation better
- higher contrast images
- requires higher adjustment in technique
- attenuate more of the remnant beam (have to increase mAs)
What is the grid frequency?
The number of vertical strip per unit of length/width

Vertical strips will be thinner as the frequency per unit of distance increases
Grid Patterns
- linear (parallel strips of lead)
- focused grids [diagnostic]
- cross-hatched or criss-cross (2 linear patterns 90 degrees to each pattern) [mammo]
What causes grid cutoff with a cross-hatched grid?
Tube angulation
Do linear grids allow tube angulation?
Yes; must be parallel w/ lead strips
Focused Grids
- linear-positioned lead strips
- angled similar to angulation of primary beam divergence
- REQUIRES PROPER SID (to reduce scatter & allow remnant photons to pass)
Latitude
- freedom from usual restraints, limitations, or regulations
- high latitude = larger allowable changes
- grid latitude: amount you can be off & still get acceptable image
In focused grids, as grid ratio increases, the latitude for SID __________ (increases/decreases)
Decreases (small error in SID can cause large grid cutoff)
Where are permanent grids located?
- within Potter-Bucky diaphragm, above film location
- grid lines parallel w/ table length, or vertical direction of wall bucky
As grid ratio or frequency increases, the radiographic density ___________ (increases/decreases)
Decreases (more scatter removed along w/ more remnant; less photons striking film)
As grid ratio or frequency increases, the radiographic contrast ___________ (increases/decreases)
Increases (beam less polyenergetic, more homogenous)
Which grid type is used most frequently?
Focused grids
Grid Conversion Factors (GCF)
- "constants" used to determine change in technical factors needed for specific grid ratios
- mAs adjustments must be made when changing grid ratio
GCF Formula
GCF = mAs with grid / mAs without grid

mAs with grid = GCF x mAs without grid

mAs without grid = mAs with grid/GCF
GCF - No grid
mAs increase 1X, GCF = 1
GCF - 5:1 Grid
mAs increase 2X, GCF = 2
GCF - 6:1 Grid
mAs inceases 3X, GCF = 3
GCF - 8:1 Grid
mAs increase 4X, GCF = 4
GCF - 12:1 Grid
mAs increase 5X, GCF = 5
GCF - 16:1 Grid
mAs increase 6X, GCF = 6
Off-Level Grid Error
- occurs when tube is angled across the long axis of the grid strips or grid is unleveled in comparison to tube
- typically happens in portable grids
- appearance: decrease in density
- To Fix: level grid under patient, angle tube parallel w/ length of grid
Off-Center Grid Errors
- Tube isn't centered (CR) w/ the center of the focused grid (tube not centered to center of grid cassette)
- occurs in fixed or portable grids
- decrease in density
Off-Focus Grid Errors
- SID outside acceptable range of focused grid
- occurs w/ fixed or portable grids
- decrease in density at lateral margins of image
Upside-Down Grid Error
- center of grid will allow the x-ray beam to pass unrestricted flipped upside down or not
- on edges, grid strips are angled opposite to angle of divergence
- severe decrease in density @ lateral margins
Does collimation change the diverging angle of the primary beam?
No
Air Gap Technique
- alternate method to use a grid if one isn't available
- use large OID (will reduce amount of scatter reaching the film with the air gap)
- need to increase SID to reduce magnification
- kVp = 70 at least
Reverse-Cassette Technique
- flip cassette so remnant radiation strikes back side of cassette first
- foil acts as filter, stopping Compton radiation
- increase in mAs needed to maintain density
Grids do not CREATE scatter, they __________ scatter
ABSORB (patient creates scatter)
What kind of latitude does a high grid ratio grid have?
Narrow concerning SID (precise) & alignment (centered & leveled)
What size GCF do high grid ratio grids have?
Larger (takes more mAs to get good density)
Who are the forefathers of technique?
Ed C. Jerman (1920s), Arthur Fuchs (1940s)
What's the goal of any exposure system?
Consistency
Exposure System Selection
- often incorporated into pre-programmed control consoles
- suggested starting points
- variables change according to anatomical part thickness
Calipers
- device that measures part thickness
- important QC measure when not using AEC
What are the 2 main methods of part measurement?
- Central ray entrance & exit
- Thickest part
Fixed kVp Systems
- developed by Fuchs during WW2
- reduction in patient dose, provides more info within image, increased consistency in image receptor exposure, shorter exposure times
- lengthens exposure latitude (gives variability in quantity of mAs to try to hit "sweet spot")
- Disadvantage: higher amount of scatter results in reduced image contrast
Variable kVp Systems
- developed by Jerman (1925)
- first exposure "method"
- used when kVp range was limited by equipment generators
- allow for small changes in exposure for small changes in part thickness, higher contrast
- Disadvantage: more patient dose, higher repeat rate, less latitude
Anatomically Programmed Radiography (APR) Systems
- commonly used computerized exposure control
- provides choices of anatomical parts w/ suggested exposure factors
APR Systems & AEC
- selection of exposure time (mAs) eliminated
- must still choose mA, kVp, distance
Establishing Fixed kVp Exposure Systems
- kVp held constant (mAs varies, eliminates effects of multiple variables)
- kVp has effect on contrast, type of interaction, scatter, average energy of beam
- provides consistency, user friendly
Optimal kVp
- maximum kVp that result in images w/ appropriate contrast
- produces lower contrast & minimum patient dose
Establishing a Variable kVp Exposure System
- kVp is varied while mAs is constant
- thicker body part requires higher kVp value; mAs is specified for each body part
- Disadvantage: varying kVp from image to image lessens control control over the variables it controls
Historical Development of Radiographic Practice
- +2 for each additional cm of tissue
- (2 kVp x part cm) + 30 = new kVp
- not accurate (often too low kVp)
- mAs values determined through trial & error
Variable kVp System Flaws
- 15% rule not thoroughly understood (used +/- 10 kVp as compensation for doubling or halving mAs)
- at low kVp, dramatic effect on image; not as profound at high kVp
Establishing Stepped Variable kVp System
- straight variable kVp charts often resulted in too much or too little contrast
- remedied w/ multiple kVp scales
3 Criteria of kVp Scale
1) All film contrast is acceptable to radiologist
2) Small part kVp recommendations provide adequate penetration
3) Large part kVp recommendations limit scatter & fog
mAs Scale
Step 1: acceptable kVp range = 70-84 kVp (for all part thickness)
Step 2: must compensate w/ mAs via 15% rule
Step 3: fine tuning (measure w/ calipers)
Automatic Exposure Controls (AEC)
- devices designed to reproduce images w/ consistent density & contrast regardless of differences in anatomical sizes or textures of same anatomical part
- Goal: reproducibility & consistency btw. patients & technologists
AED (Automatic Exposure Devices)
- phototiming: older method of AEC
- ionization chambers: most current method used
Phototimers
- use a photomultiplier tube to detect the amount of remnant radiation exiting from the patient
- phosphor screen emits light when struck by radiation, light gathered by tube & converted to electric current, charges the capacitor, reaches a preset level, terminates exposure (quick process)
- cannot use a lead-backed cassette w/ these
Ionization Chambers
- placed between the patient & the IR (measures remnant radiation leaving patient)
- filled w/ air; measure amount of ionization that occurs in air within chamber
- terminates exposure when predetermined amount of air has been ionized
- typical 3 chamber configuration seen in clinic
What is a critical skill to have when using ionization chambers?
Critical positioning skills are needed to place the area of interest over the correct ionization chamber
The use of AEC requires the art of ________________
Positioning
What does the AEC do when all 3 chambers are selected?
Averages the signals from the cells; when appropriate density is reached the exposure is terminated

Cell receiving most radiation will contribute the greatest electrical signal
Density Controls
- regulate image density (-2, -1, 0/N, +1, +2)
- DENSITY CONTROLS SHOULD NOT BE USED TO COMPENSATE FOR PATIENT PART THICKNESS OR KVP CHANGES
What will poor patient positioning increase while using AEC?
Repeat image rate
Collimation w/ AEC
- do NOT collimate close to areas of ionization chamber
- too wide of collimation will produce scatter radiation that may undercut the patient
Minimum Response Time
Time necessary for AEC to respond to the ionization & send a signal to terminate exposure (0.001 second)
What's the purpose of a back-up timer?
Prevents extreme over exposure of the patient to radiation

Should be set to 150% of anticipated manual exposure time
Anatomically Programmed Radiography - Cookbook Radiography
Specific "recipes" are used for different sized patients
Using Film/Screen Combos w/ AEC
- most AEC's are calibrated to a particular film/screen speed
- changing the speed will result in undesired high/low density unless setting are changed to accommodate the speed change
Extremity cassettes used in conjunction w/ AEC calibrated for par screens will result in __________________ (overexposed/underexposed) films
Underexposed
Is 100 speed considered a higher or lower speed?q
Lower (less than 400)

Means that sensitivity to radiation is lower, need more radiation to get proper image
AEC Don'ts
DON'T use AEC on extremities such as hands/feet/ankles

DON'T increase mA to increase the density of image

DON'T increase the kV to increase the density of the image