Inline shut off valves shall be permitted to be installed to isolate or shut off piping for servicing of individual rooms or areas



T/F

T

New or replacements shut off valve shall have the operator showing when the valve is open or closed



T/F

T

The shut off valve shall be accessible to other than authorized personnel to facilitate periodic testing



T/F

F

Each riser supplied from the main line shall be provided with a shut off valve in the riser adjacent to the main line



TT/F

T

Regard lus of how many zone valve boxes are installed on a lateral branch off a razor, only one service vow but shall be required for each branch


T/F

T

The zone valve shall not be located in a Room with station outlets/inlets that it controls, such that a wall intervenes between the valve and outlets.



T/F

T

All valves accept valves zone valve Box assemblies shall be secured by any of the following means



any of the above



located in secure areas



locked or latched in their operating positions



located above ceiling, but remaining accessible and not obstructed

Any of the above

With the approval of the authority having jurisdiction, and which areas shall the shut off valve as be permitted to be secured to prevent inappropriate access?



Maternity



anesthetizing



critical care



psychiatric or pediatric

Psychiatric or pediatric

New or replacement shut off valves for medical gas pressure piping systems shall be permitted to be any type as long as they meet the following



they have a minimum CV factor



constructed of material suitable



quarter turn to off



all of the above

All of the above

Shut up valves for positive pressure gasses shall



be cleaned for oxygen service by the manufacture



permit in line serviceability



provided with copper tube extensions for brazing



all of the above

All of the above

A shut off valve shall be placed at the immediate connection of each ____ system to the distribution piping system. What is the name given to the valve?



Main



isolation



source



supply

Source

A shut off valve shall be provided in the ____ supply line when the source valve is not accessible from within the building being served. What is the name given to this valve?



Zone



service



supply



main

Main

What is the name given to the shut off valve installed on all lateral branch piping from a main or riser?



Riser



zone



main



service

Service

What valve shall be readily operable from a standing position in the corridor on the same floor they serve?



Zone



main



service



riser

Zone

Immediately outside each vital life support, critical care and anesthetizing location of moderate sedation, deep sedation, or general anesthesia there shall be located a ____ in each medical gas, or vacuum line or both.



Zone valve



relief valve



area



fire extinguisher

Area alarm

There shall be a zone valve immediately outside of anesthetizing location of ____?



moderate sedation



general anesthesia



all or any of the above



deep sedation

All or any of the above

Inline check valve shall not have threaded connections but shall have ____?



in line serviceability



threaded purge port



brazed extensions



all of the above

All of the above

Service valve shall be placed in the branch piping prior to any ____ on that branch



zone valves



valves for future



inline valve



source valve

Zone valve

What valve is permitted to be installed to isolate or shut off piping for servicing of individual rooms or areas?



riser valve



service valve



future valve



inline valve

Inline valve

The main line valve shall be located on which side of the source valve, when the main line valve enters the building



upstream of the source valve



facility side of the source Valve



main line valve upstream of the service valve



downstream of the source valve

Facility side of the source valve

A positive pressure gas valve, 1/2 inch in size minimum CV is what?



21



17



31



60

17

A positive pressure gas valve, 1" is size minimum CV is what?



110



31



60



58

60