  • Front
  • Back

The 400 series stainless steel is used for such instruments as

scissors, forceps, osteotomes, and chisels

Which of the following would be used to hold back tissue or organs?

Retractors

The weakest part of the instrument is/are the

box lock

What is the most difficult part of a hemostat to clean?

Box lock

The final stage in manufacturing a surgical-grade instrument is called

passivation

Scissors with black handles are often referred to as

Super sharp

The layer on a surgical instrument that is created by the passivation process is called

chromium oxide

How is an instrument with tungsten carbide inserts identified

Gold handles

What type of finish on a surgical instrument is used for laser procedures

Ebonized

Which of the following lubricants is recommended for use on surgical instruments?

Water-based

The chromium oxide layer is intended to make an instrument

resistant to corrosion

Which of the following is not part of routine inspection of instruments

Vendor identification number

Unless otherwise directed by the maufacturer, an instrument should be lubricated

Everytime it is used

When testing a scissor for sharpness it is important to

make several cuts in the testing material

When the jaws of a stainless steel (non-tungsten carbide) needle holder become worn, the needle holder must be

discarded

A crack in a hand-held instrument would most likely be found in the

box lock

Cracked insulation on a laparoscopic instrument can cause

electrical shock and burns

Powered equipment should be lubricated

according to the manufacturer's instructions

When labeling a rigid scope for sterilization, which of the following information is not generally required?

Surgeon's name

Black dots in a fiberoptic light cable indicate

there is damage to the glass rode

which of the following is not a method for checking the visual clarity of a rigid endoscope?

Using a leak tester

When marking and identifying instruments, which of the following methods should never be used?

Engraving

The air flow in the preparation and packaging area must be maintained at

positive pressure

The preparation and packaging area must have an air exchange rate of

10 per hour

What is the acceptable temperature range for the preparation and packaging area?

68 F to 73 F (20 C to 23 C)

The ideal humidity level in the preparation and packaging area is

50%

If a non-perforated tray is used for the sterilization of instrument sets, it must be

tilted on the sterilizer rack

The floors in the preparation area should be

Wet-mopped daily

When assembling a set containing heavy and delicate instruments, where should the heavier instruments be placed in the set?

On the bottom of the set

Which of the following can be done to aid in the drying of wrapped instrument sets?

Place a towel in the bottom of the tray

Which of the following should be used to write on count sheets and tray lists?

Permanent, nontoxic ink

Where should the chemical indicator be placed in a wrapped set of instruments

The center of the set

Muti-layer systems should have chemical indicators places

on each level

What is the maximum size and weight of textile pack?

12''12'' x 20''', 12 pounds

Basin sets should be prepared for sterilization with

an absorbent material between each basin

Which of the following is not one of the basic principles of packaging?

The material must be acceptable for all types of sterilization

Sterilant penetration can be inhibited by using

a wrapper that is too large

Approximately how much space should be left between an item and the inside edges of a peel pouch?

1 inch

Which of the following packaging materials is acceptable for steam sterilization?

Nonwoven wraps

Which of the following packaging materials is acceptable for low-temperature gas plasma sterilization?

Tyvek

Tamper evident seals on rigid containers include

latches and locks

When rigid sterilization containers are processed in a steam sterilizer, they must be placed in which of the following positions on the cart?

Flat