• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/65

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

65 Cards in this Set

  • Front
  • Back
What resulted from the invention of image intensifiers
replaced flourescent screens
increased image brightness
What is the primary function of the flouroscopy
The primary function of fluoroscopy is real-time
imaging to provide visualization of dynamic pro cesses as they occur.
What has a higher radiation exposure rate; flouro or plain film
flouro
What is the exposure rate of typical flouro
45mGy/min
What is the exposure rate of a typical plain film
900mGy/min
What is higher the TOTAL exposure time for plain film or flouro
flouro (x rays occur for a much shorter period of time
What is the total exposure for a 10 minute flouro study of the abdomen
450mGy
what is the exposure total for an abdominal film
3mGy
What is done to compensate for the increased radiation dose
the exposure rate (as mentioned before) is less.
What is a flourescent screen
theis is s a matererial that immediates visible light in response to a stimuli such as an x ray
Does red light have a short or long wavelength
long
What was the purpose of red adaption googles
to allow the flouroscopist to remain adjust to dark room and peform activity outside to retain dark room adaptation
Why were image intensifiers developed
To overcome the deficiencies of viewing the
dim fluorescent screen image, image intensifier
devices were developed and introduced in 1953.
What are the components of a flouroscopy machine
How does the X ray generator work
it allows selection of kVp and mA that is delivered to the x ray tube
What are 3 additional controls added to a flouroscopy machine that are different than what is seen in a typical x ray machine
low continuous tube current
rapid pulsed exposure
automatic brightness control
What are 3 types of X-ray generators used for flouro
single phase
three phase
constant potential
high frequency
What 2 methods are used to energize the x ray tube for flouroscopy
continous exposure
pulsed exposure
What is the current in continous flouroscopy
For continuous fluoroscopy, the genera-
tor provides a steady tube current while the fluo-
roscope is activated.
How fast are the images acquired in continous exposure
. Images are acquired at a rate
of 30 frames per second, resulting in an acquisi-
tion time of 33 msec per image
How is the exposure during a pulsed flouroscopy
For pulsed fluo-
roscopy, the exposure is delivered in short pulses,
3–10 msec in length.
How fast are the images acquired in pulsed exposure
ypically, a pulse rate of 30
pulses per second is used, with some units allow-
ing the selection of lower pulse rates (15 or 7.5
pulses per second)
What is an advantage of pulsed exposure
improved temporal resolution...motion artifact is reduced by shorter acquisition time
What are pulsed exposures particularly useful for
pulsed fluoroscopy useful for examining
rapidly moving structures such as those seen in
cardiovascular applications.
Can pulsed exposure reduce radiation dose
yes, especially when it is set lower
Why is reproducibility so important
Good exposure reproduc-
ibility is critical for fluoroscopic systems equip-
ped with digital subtraction angiography (DSA),
because differences in tube voltage between maskand contrast images can cause incomplete subtraction.
What type of generator produces the best reproducibility
high frequency generators
What type of generators are capable of producing the shortest exposure pulse
constant potential
Do high frequency generators and three phase generators produce as good a short pulse as constant potential
no, slighly longer
What is the function of the automatic brightness control (ABC)
this acts to keep the overall image brightness seen on the monitor at a
constant level as the image intensifier is panned
over body parts of differing thickness and attenuation
How does the ABC work
by automatically adjustion the kVp and the mA settings
What does the X-ray tube do
The x-ray tube converts electrical energy pro-
vided by the generator into an x-ray beam
How does the X-ray tube work
Within the x-ray tube, electrons are produced by
a heated filament and accelerated toward a posi-
tively charged tungsten anode. The interaction of
the electrons with the anode results in the emis-
sion of x rays.
Is the anode postively charged
yes
What is the anode made out of
tungsten
What is the focal spot
the area of the anode struck by electrons
What is desired; large or small focal spot
small (sharper image)
How is the focal spot size adusted
by changing the anode angle
How is the anode angle changed
What is the typical range of an anode angle
7-20 degrees
What is a problem of inherent fluoroscopy (especially with interventional procedures where there is a long exposure time)
heating of the anode
Does the anode need a large heat capacity
yes
What is done to improve the heat capacity of the anode
To improve heat dissi-
pation, high-speed anode rotation may be used
(over 10,000 rpm). In addition, a circulating water or oil heat exchanger with cooling fans is
commonly installed
What is the type of exposure control that can be used for interventional or angiographic procedures
grid-con-
trolled pulsing to produce very short (millisec-
ond) exposures for cine image recording or
pulsed fluoroscopy. In a grid-controlled tube, the
cathode is at a variable negative potential, ca-
pable of pinching the electron flow off and on for short exposure pulses.
How does the grid controlled tube stop the flow of electrons
In a grid-controlled tube, the
cathode is at a variable negative potential, ca-
pable of pinching the electron flow off and on for short exposure pulses.
What is a problem when using maximum FOV in interventional or angiogrphic procedures
2
limits the heat capacity

When a large FOV is
needed to image with a large image intensifier or
film changer, the anode angle must be large
enough to allow adequate coverage without cut-
off of the beam intensity.
How does the filament size and anode angle effect the focal spot size
Diagram on the left (a)
shows a large anode angle, which provides large radiation field cov-
erage and a small effective focal spot size. However, the actual focal
spot track on the anode is narrow, resulting in low heat capacity.
The center diagram (b) illustrates a configuration with the same ef-
fective focal spot size and a small anode angle. This configuration
results in greater heat capacity but small field coverage. To satisfy
the requirements of both large field coverage and large heat capac-
ity, the filament size must be increased, resulting in a larger effec-
tive focal spot size, as shown in c
What is the result of a large anode angle, small filament size on radiation field and effective focal spot
LEFT IMAGE: shows a large anode angle, which provides large radiation field cov-
erage and a small effective focal spot size. However, the actual focal
spot track on the anode is narrow, resulting in low heat capacity
What does a smaller anode angle do to the effective focal spot size
decrease it
What does a large anode angle do to the effective focal spot size
increase is
What does a LARGE anode angle and a large filament size do to the effective focal spot and radiation field size IMAGE ON RIGHT
INCREASE
What happens to heat capacity with a small focal spot
decreases the heat capacity of the tube
What is the function of a collimator
define the shape of the X-ray beam
What are the 2 sets of bades that are present within the collimator
round and rectangular
What is the function of a round collimator
conforms the X-ray beam to a circular FOV
What is the function of the rectangular collimator
this can be used manually during the examination to reduce the size of the X-ray beam
What is the benefit of manual collimation
Collimation reduces
the exposed volume of tissue, resulting in re-
duced scatter production and improved image
contrast.
How does manual collimation reduce radiation to the patient
coning the x-ray
beam to the area of clinical interest will reduce
overall patient dose by minimizing direct expo-
sure and scatter exposure to sensitive organs that
may be adjacent to the beam
After the Xray is generator and moves through the collimator where does it go next
What are the filters of a flouroscopy machine called
countour or wedge filters
What is the function of filters
to provide further beam shaping and collimation
What is another type of filter besides wedge or contour
equilization filters
What is the function of an equilzation filter
Equalization filters re-
duce glare from unattenuated radiation near the
edge of the patient and equalize light exposure to the video camera.
What do filters look like
it operates in the radiographic mode. Tube current is measured in hundreds of mA instead of less than 5 mA, as in image-intensifying fluoroscopy.
it operates in the radiographic mode. Tube current is measured in hundreds of mA instead of less than 5 mA, as in image-intensifying fluoroscopy.