Flow towards transducer _______frequency |
Increases |
|
Doppler shift or change in Doppler frequency |
Back (Definition) |
|
CW Doppler one crystal is continuously transmitting and one crystal is receiving signals. This causes |
Range ambiguity |
|
Aliasing |
When high velocities appear negative Occurs with PW Doppler |
|
Nyquist limit |
PRF/2 |
|
Is color Doppler pulsed or continuous? 2D? |
Pulsed |
|
Brightness of the image is improved by increasing |
Output Amplification - increases all signals the same way Compensation - creates uniform image. Takes depth into account. Does not increase every signal the same amount |
|
If image is too bright |
Decrease output gain first |
|
If image is too dark |
Increase receiver gain first (increases amplification after returning to receiver) |
|
To increase sensitivity to low flow, nyquist limit should be |
Decreased Also decreases PRF |
|
Define harmonic imaging |
Creating an image from sound reflected at twice the fundamental frequency |
|
How is harmonic frequency created? |
Non linear behavior of sound causes energy to shift into higher frequencies by multiples of 2. Harmonics are typically created from strong beam in the main axis. Clearer image. |
|
How does decreasing transducer frequency decrease aliasing? |
Causes less Doppler shift Velocity of blood in echo is extrapolated from frequency detected by machine. If there is less Doppler shift for a given high velocity, it can be detected without raising the PRF or nyquist limit.
Decreasing angle (Cos theta) similarly decreases Doppler shift |
|
As imaging depth increases, PRF |
Decreases PRF = velocity/depth |
|
Mechanical Index |
Back (Definition) |
|
When using contrast if only the apex of LV is opacified |
Attenuating artifact Too much contrast! Increase the mechanical index temporarily to destroy some of the bubbles |
|
Swirling artifact with contrast |
MI is too high, decrease to improve opacification |
|
What is ideal mechanical index? |
Between 0.1 and 1 Promotes nonlinear resonance and harmonic enhancement |
|
How quickly should myocardium perfume with contrast? |
Rest: within 5 beats Stress: within 2 beats |
|
Strain rate |
Difference in velocities between 2 points / distance between 2 points |