Flow towards transducer _______frequency

Increases

Doppler shift or change in Doppler frequency

Back (Definition)

CW Doppler one crystal is continuously transmitting and one crystal is receiving signals. This causes

Range ambiguity

Aliasing

When high velocities appear negative


Occurs with PW Doppler

Nyquist limit

PRF/2

Is color Doppler pulsed or continuous?


2D?

Pulsed

Brightness of the image is improved by increasing

Output


Amplification - increases all signals the same way


Compensation - creates uniform image. Takes depth into account. Does not increase every signal the same amount

If image is too bright

Decrease output gain first

If image is too dark

Increase receiver gain first (increases amplification after returning to receiver)

To increase sensitivity to low flow, nyquist limit should be

Decreased


Also decreases PRF

Define harmonic imaging

Creating an image from sound reflected at twice the fundamental frequency

How is harmonic frequency created?

Non linear behavior of sound causes energy to shift into higher frequencies by multiples of 2. Harmonics are typically created from strong beam in the main axis. Clearer image.

How does decreasing transducer frequency decrease aliasing?

Causes less Doppler shift


Velocity of blood in echo is extrapolated from frequency detected by machine. If there is less Doppler shift for a given high velocity, it can be detected without raising the PRF or nyquist limit.



Decreasing angle (Cos theta) similarly decreases Doppler shift

As imaging depth increases, PRF

Decreases


PRF = velocity/depth

Mechanical Index

Back (Definition)

When using contrast if only the apex of LV is opacified

Attenuating artifact


Too much contrast!


Increase the mechanical index temporarily to destroy some of the bubbles

Swirling artifact with contrast

MI is too high, decrease to improve opacification

What is ideal mechanical index?

Between 0.1 and 1


Promotes nonlinear resonance and harmonic enhancement

How quickly should myocardium perfume with contrast?

Rest: within 5 beats


Stress: within 2 beats

Strain rate

Difference in velocities between 2 points / distance between 2 points