Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
48 Cards in this Set
- Front
- Back
Noninvasively estimates the hemoglobin oxgen saturation of arterial blood
|
pulse oximetry
|
|
most pulse oximeters use two wavelengths, ___ (660) and ___ (940nm) at which HbO2 and deoxyhemoglobin (Hb) have different absorption characteristics.
|
red and infrared
|
|
red and infrared light-emitting diodes ___ in the oximeter probe serve as light source.
|
LED
|
|
Bc HbO2 and Hb differ in light avsorption at each wavelenght the amount of red and infrared light transmitted is related to ___ saturation
|
O2
|
|
a photodiode positioned on the opposite side of the probe serves as the ___
|
photodetector
|
|
To identify the oxgyen saturation of the arterial blood only, the divice relies on the pulsatile nature of___
|
arterial flow
|
|
During ___ a new volume of blood enters the arteriolar bed, and light absorp. increase
|
systole
|
|
During ___ absorption decreases to a minimal level
|
diastole
|
|
Is related to the ratio of minimum and maximum absorption at each wavelength
|
oxygen saturation (SpO2)
|
|
Devices vary by manufacturer in the type of
|
LED, photodiode, and microprocessor
|
|
most pulse oximeter probes use ___ ____, sending light through the arterial bed to a photo detector on the opposite side
|
transmittance spectrophotometry
|
|
With____ pulse oximeter sensors place the light source and detector on the same side of the arterial bed.
|
reflectance spectrophotometry
|
|
Pulse oximeter measurements have an accuracy of +4% at SpO2 greater than
|
80%
|
|
To appreciate the implications of the accuracy of pulse oximetry, one must consider the ___ ___ ___
|
HbO2 equlibration curve
|
|
If the pulse oximeter displays a SpO2 of 95% the true saturation could be as low as ___% or as high as ___%
|
91/99
|
|
If the true saturation is 99% the ___ might be very high.
|
PaO2
|
|
A shift in the HbO2 equlib. curve can change the SpO2 although no change in __ has occurred
|
PaO2
|
|
A shift in the HbO2 equlib. curve can change the SpO2 although no change in __ has occurred
|
PaO2
|
|
The Pulse oxymetry provides no direct information about the Paco2 and may not accurately reflect the Pao2 bc of changes in the shape and position of the ___
|
Hbo2 dissociation curve
|
|
Pulse oximetry is accurate when
|
all hemoglobin is either oxyhemoglobin or deoxyhemoglobe, when there are no other absorbers between the LEDs and the detector other than those present during the empirical calibration, & when all the blood that pulsates is arterial blood
|
|
With low perusion there is an ___ in the ratio of venous blood to arterial blood at the measuring site.
|
increase
|
|
Low perfusion is associated with lower pulse amplitude, so the __ ___ ____ has a greater effect when combined with a low signal
|
noise of motion
|
|
Strategies to address issues trlated to motion and low perfusion include
|
averging the saturation data over alonger periond of time and suspending the report of data until clean data are available.
|
|
The pulse oximetry plethysmographic (POP) waveform may display the effect of ____ ___, and therefore the severity of air trapping in obstructive airway disease
|
pulsus paradoxus
|
|
In pts. with obstuctive lung disease and elevated pulsus paradoxus, there is an altered pulse oximetry baseline tracing manifested as the respiratory waveform _____
|
variation
|
|
Pulsus paradoxus is significantly correlated with the degree of respiratory waveform variation of the pulse oximetry tracing and the amount of auto- _____
|
PEEP (positive end -expiratory pressure)
|
|
respiratory variations in POP waveform amplitude have also been shown to be useful in prediction of ___ responsivensess
|
fluid
|
|
POP waveform amplitude is mearsured on a beat to beat basis as the vertical distance between
|
peak and preceding valley trough in the wavefom
|
|
A POP greater than 15% is prdictive of fluid responsiveness in mechanically ventilated patients with
|
circulatory failure
|
|
- law defining the relationship between the concentration of a substance and the amount of light (I) transmitted through it
|
Beer- Lambert law
|
|
assessment of gas at the proximal airway that plots CO2 on the vertical axis and time or volume on the horizontal axis
|
Capnogram
|
|
- noninvasive technique that measures CO2 levels in inspired and expired gas and displays a capnogram
|
Capnography
|
|
numeric display of CO2 measurement taken from the proximal airway
|
Capnometry
|
|
- a medical imging technique in which an image of a part of the body is inferred from surface electrical measurement. Conducting electrodes are attached to the skin, and small alternating currents are applied to the electrodes. The resulting electrical potententials are measured, and the process may be repeated for numerous different configurations of applied current.
|
Electrical impedance tomography (EIT)
|
|
- PCO2 at end exhalation (PETCO)
|
End-tidal PCO2
|
|
- rib cage or abdominal displacements stretch an elastic belt, which changes light transmission through the fibers result
|
Fiberoptic plethysomography
|
|
- method for measuring respiratory rate and excursion using two electrodes placed on the chest wall and then passing a high-frequency and low-ampere AC current between the electrodes on the chest surface
|
Impedance pneumography-
|
|
instrument capable of measuring all respiratory gases, including respiratory and anesthetic gases, breath by breath
|
Mass spectrometer-
|
|
a condition that can occur during the application of pulse oximetry in which the sensor is not correctly positioned. This effect causes incorrect readings when the sensor is not symmetrically placed, causing one wavelength to be overused in the calculations of satruration
|
Penumbra effect
|
|
a value derived from the photoelectric plethysmographic signal of a pulse oximeter and calculated as the ratio of the pulsatile component (arterial compartment) and the nonpulsatile component (other tissues; venous blood, bone, connective tissue) of the light reaching the device’s detector
|
Perfusion index (PI)-
|
|
- type of plethysmography apparatus that replaces wire coils from the elastic belts with a piezoelectric buckle, which encloses a sensor to generate a voltage in response to stretch passed through the end of the belts.
|
Piezoelectric plethysmography
|
|
- abnormal decrease in systolic pressure and pulse wave amplitude during inspiration
|
Pulsus paradoxus
|
|
method used to measure CO2
|
Raman spectroscopy
|
|
Method for indirectly measuring tidal volume; sensors use a circuit of coiled wire woven into an elastic band and excited by an AC current. Inductance results from alternation electrical currents creating magnetic fields around themselves and those changing magnetic fields altering other electrical currents that they encounter
|
Respiratory inductance plethysmography (RIP)-
|
|
- method that identifies substances by their absorption (also called extinction) of specific wavelengths in the electromagnetic spectrum
|
Spectrophotometry
|
|
thermometer that can measure extremely small changes in temperature
|
Thermistor-
|
|
- apparatus that detects bidirectional airflow at the nose and mouth by sensing the temperature difference between inspired room air and exhaled air that has been warmed to body temperature; an active sensor used to measure temperature
|
Thermocouple
|
|
monitoring- a means of respiratory monitoring of blood gases through electrodes applied to the skin
|
Transcutaneous
|