Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
13 Cards in this Set
- Front
- Back
fresh gas decoupling (FGD) and compensation
|
drager--FGD
compensation(ohmeda) |
|
fgf is diverted into reservoir bag by decoupling valve during ?
|
inhalation
|
|
fgf is diverted between ?
|
fgf and ventilator circuit
|
|
advantage of fgd
|
decreased risk of barotrauma esp w/ children
|
|
fgd contain what relief valve
|
negative pressure relief valve if insufficient gas in reservoir bag
|
|
d-lite sensors
|
ohmeda
vol and flow adjusted via computers |
|
delivers preset tv at constant flow and rr
|
volume control (CMV)
|
|
waveform will steadily increase during inspiration
|
volume control
|
|
adult setting for CMV
|
vt 8-10ml/kg
RR 8-12 bpm |
|
preset inspiratory pressure
|
pressure contol
|
|
inspiratory flow is higtest at beginning of inspiration
|
pc
|
|
tv varies inversely with changes in resistance and compliance
|
pc
|
|
operates with preset ie ratio increased rr, shortens inhalation and decreases tv
|
pc
|