two types of ventilators
|
1. double circuit, pneumatically driven ventilators
2. piston |
|
electronically and pneumonically controlled, double circuit, vol preset, time cycled, w/ optional pressure limit control
|
drager anesthesia ventilator (AV-E)
|
|
move up during expiration and down during inspriation
|
ascending bellows
|
|
move down during exhalation
|
descending bellows
pt disconnect bellows continue to move up and down |
|
only machine to use descending bellows
|
drager--julian
has disconnect alarm based on ETco2 |
|
frequency rate
|
1-99bpm
|
|
electronically controlled timing device allows driving gas to enter the bellows chamber to compress the bellows and then exit chamber
|
frequency
|
|
tidal volume
|
control located on bellows chamber
setting 200-1400ml |
|
I:E ratio
|
1:2- 1:4.5
|
|
which ventilator does not have ie ratio
|
ohmeda
controlled by changes in flow rate, tv and rr |
|
regulates flow rate of driving gas into bellows chamber
|
inspiratory flow---keep as low as possible ---d/t barotrauma
|
|
beneath bellows or on circle breathing system
|
peep <18cm h2o
|
|
sigh option
|
ohmeda
vol of every 64th breath--give 150% of tv |