IAHCSMM
|
INTERNATIONAL ASSOCIATION HEALTHCARE CENTRAL SERVICE MATERIALS MANAGEMENT
|
|
DECONTAM
|
NEGATIVE AIR PRESSURE, KEEPS GERMS AND BATERIA FROM GOING TO CLEAN AREA, 60-65 DEGREES, 30-60% HUMIDITY, 10 AIR EXCHANGES PER HOUR |
|
PREP/PACK AREA
|
68-73 DEGREES, 30-60% HUMIDITY, 10 AIR EXCHANGES PER HOUR
|
|
CLEAN STORAGE
|
75 DEGREES OR BELOW LESS THAN 70% HUMIDTY, 4 AIR EXCHANGES PER HOUR |
|
MEDICAL ASEPSIS |
TO CLEAN EQUIPMENT AND HAND WASHING |
|
SURGICAL ASEPSIS |
STERILIZATION PROCESS OF INSTRUMENTS |
|
ASEPSIS
|
ABSENCE OF MICROOGRANISMS THAT COUSE DISEASE |
|
CONDUCTION |
SOLIDS, DRY HEAT, HEAT END TO END |
|
CONVECTION |
LIQUIDS/GAS, STEAM |
|
PREVAC |
DYNAMIC AIR REMOVAL STERILIZERS |
|
"FLASH"
|
STERILIZE AN ITEM/INSTRUMENT FOR IMMEDIATE USE |
|
STEAM STERILIZATON REQUIRES |
CONTACT, TEMPERATURE, TIME & AND MOISTURE |
|
PHASES OF STEAM STERILIZATION, FLASH, AND REGULAR |
1. CONDITIONING PHASE - STEAM ENTERS CHAMBER AS AIR IS REMOVED THROUGH THE DRAIN, PRESSURE INCREASES INSIDE AS THE CHAMBER REACHES STERILIZATION TEMPERATURE 2. EXPOSURE PHASE - 3, 4 OR 10 MINUTES STERILIZATION TIME BEGINS WHEN THE THERMOCOUPLE SENSES THE CONSTANT PRESSURE CYCLE TEMPERATURE 3. EXHAUST AND DRYING - FILTERED AIR ENTERS, STEAM IS REMOVED FROM THE CHAMBER AND THE PRESSURE INSIDE THE CHAMBER RETURNS TO ATMOSPHERIC PRESSURE |
|
CLEANING |
1ST AND MOST IMPORTANT STEP IN THE STERILIZATION PROCESS, BLOOD, SOIL, AND TISSUE REMOVAL |
|
DECONTAMINATION
|
REMOVES, INACTIVATES AND DESTROYS BLOOD BORNE PATHOGENS ON SURGACE |
|
HEPAT
|
LIVER |
|
ARTHRO |
JOINT |
|
COLECTOMY |
REMOVE PART OF THE COLON |