when using scvanging sys, In closed interface, what should scavenging bag be?
partially inflated at all times
If scavenging bag in closed interface is continually collapsed or neg pressure valve opens frequ...
flow should be lowered
If scavenging bag in closed interface is fully distended,
flow should be increased
Open interface scavenging system should always be used with what type?
active
when manually ventilating, what does the scavenging bag do?
slightly expand/deflate
which type of interface requires positive pressure relief valve?
closed, so pressure doesn't build up in pt.
which type of system doesn't have valves?
open
Open interface should only be used with
active disposal system
which type of interface has the parallel tubes in the resevoir?
open
does closed system have valves?
yes
which type of interface requires resevoir/does not require resevoir?
closed does not require resevoir (should only be used with active); open does require
A closed interface with only positive pressure relief should only be used with a
passive disposal sys
In closed interface, if active disposal sys. used,
negative pressure relief valve must also be present
T or F: both passive and active disposal systemes can be used with closed interfaces
T
when manually ventillating, what happens to scavenging bag,
slightly inflates, deflates
Open interface should be used with which type of disposal system?(Active, passive, open, closed?)
active
which type of interface has valves?
closed
what is always required with a closed interface? +pressure relief valve,-pressure rel. valve,resevoir, all
positive pressure relief valve
What gas can't be absorbed?N2, CO2, N2O
N2O
In this type, air is recirculated to room
recirculatory
what are the two types of active disposal system?
active, piped
what is most common type of acive disposal system?
piped
T or F: in active disposal systme gasses are "pushed".
F. pulled
Always required for active duct sys. A. neg pressure relief B. valve and res. C. positive pressure D. All
A and B. neg pressure relief and valves and res.
which type of system can use both passive and active disposal sys?
closed
open system requires what type of disposal system?
active
What are the types of passive absorption?
room ventilation system, piping direct to atmosphere, absorption device
in what type of disposal system is pushing happening, not pulling
passive
descrive the absorption device of passive disposal system
activated charcoal
Two types of passive ventilation systems
nonrecirculating, recirculating
T or F: in recirculating system air is recirculating to room
T
Problems w/passive disposal system of piping direct to atmosphere
+/- pressure by wind currents or obstruction, need to ensure system patency, requires special installation
Benefits of passive disposal system of piping direct to atmosphere
easy to use, low maintenance cost
What does efficiency of adsorption device depend on?
agent, flow rate thru canister, and moisture. moisture may reduce efficiency
Overall advantages of passive disposal system
simpler than active, less expensive than active
Disadvantages of passive disposal system over active
not as effect in lower trace gas levels b/c + pressure promotes outward leaks
What type of disposal system has negative pressure in gas-disposal tubing?
Active disposal system
Disadvantages of piped vaccum (active disposal)
inad # of vacuum outlets,inconve outlets,system overload,damage to vac. sys,person.exp, inconv (on/off)
can move large volumes of gas at low neg pressures
active duct sys
uses flow-inducing devices(fans,pumps,blowers, etc)
active disposal system
Disadvantage sof active disposal system over passive disposal system
complex, req. -press. relief and res. capacity, special install., flow inducing device req added energy and maint
advantages of active duct sys over passive
resistance not a prob, wind doesn't affect sys.
Disadvantages of active disposal system over
more complex,expensive, not automatic, (on/off);requires interface to have negative pressure relief