- Front
- Back
when using scvanging sys, In closed interface, what should scavenging bag be?
|
partially inflated at all times
|
|
If scavenging bag in closed interface is continually collapsed or neg pressure valve opens frequ...
|
flow should be lowered
|
|
If scavenging bag in closed interface is fully distended,
|
flow should be increased
|
|
Open interface scavenging system should always be used with what type?
|
active
|
|
when manually ventilating, what does the scavenging bag do?
|
slightly expand/deflate
|
|
which type of interface requires positive pressure relief valve?
|
closed, so pressure doesn't build up in pt.
|
|
which type of system doesn't have valves?
|
open
|
|
Open interface should only be used with
|
active disposal system
|
|
which type of interface has the parallel tubes in the resevoir?
|
open
|
|
does closed system have valves?
|
yes
|
|
which type of interface requires resevoir/does not require resevoir?
|
closed does not require resevoir (should only be used with active); open does require
|
|
A closed interface with only positive pressure relief should only be used with a
|
passive disposal sys
|
|
In closed interface, if active disposal sys. used,
|
negative pressure relief valve must also be present
|
|
T or F: both passive and active disposal systemes can be used with closed interfaces
|
T
|
|
when manually ventillating, what happens to scavenging bag,
|
slightly inflates, deflates
|
|
Open interface should be used with which type of disposal system?(Active, passive, open, closed?)
|
active
|
|
which type of interface has valves?
|
closed
|
|
what is always required with a closed interface? +pressure relief valve,-pressure rel. valve,resevoir, all
|
positive pressure relief valve
|
|
What gas can't be absorbed?N2, CO2, N2O
|
N2O
|
|
In this type, air is recirculated to room
|
recirculatory
|
|
what are the two types of active disposal system?
|
active, piped
|
|
what is most common type of acive disposal system?
|
piped
|
|
T or F: in active disposal systme gasses are "pushed".
|
F. pulled
|
|
Always required for active duct sys. A. neg pressure relief B. valve and res. C. positive pressure D. All
|
A and B. neg pressure relief and valves and res.
|
|
which type of system can use both passive and active disposal sys?
|
closed
|
|
open system requires what type of disposal system?
|
active
|
|
What are the types of passive absorption?
|
room ventilation system, piping direct to atmosphere, absorption device
|
|
in what type of disposal system is pushing happening, not pulling
|
passive
|
|
descrive the absorption device of passive disposal system
|
activated charcoal
|
|
Two types of passive ventilation systems
|
nonrecirculating, recirculating
|
|
T or F: in recirculating system air is recirculating to room
|
T
|
|
Problems w/passive disposal system of piping direct to atmosphere
|
+/- pressure by wind currents or obstruction, need to ensure system patency, requires special installation
|
|
Benefits of passive disposal system of piping direct to atmosphere
|
easy to use, low maintenance cost
|
|
What does efficiency of adsorption device depend on?
|
agent, flow rate thru canister, and moisture. moisture may reduce efficiency
|
|
Overall advantages of passive disposal system
|
simpler than active, less expensive than active
|
|
Disadvantages of passive disposal system over active
|
not as effect in lower trace gas levels b/c + pressure promotes outward leaks
|
|
What type of disposal system has negative pressure in gas-disposal tubing?
|
Active disposal system
|
|
Disadvantages of piped vaccum (active disposal)
|
inad # of vacuum outlets,inconve outlets,system overload,damage to vac. sys,person.exp, inconv (on/off)
|
|
can move large volumes of gas at low neg pressures
|
active duct sys
|
|
uses flow-inducing devices(fans,pumps,blowers, etc)
|
active disposal system
|
|
Disadvantage sof active disposal system over passive disposal system
|
complex, req. -press. relief and res. capacity, special install., flow inducing device req added energy and maint
|
|
advantages of active duct sys over passive
|
resistance not a prob, wind doesn't affect sys.
|
|
Disadvantages of active disposal system over
|
more complex,expensive, not automatic, (on/off);requires interface to have negative pressure relief
|