Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
57 Cards in this Set
- Front
- Back
A good radiograph should have (maximum/minimal) projection geometry
|
minimal
|
|
What are two things that cause motion unsharpness?
|
moving BID or moving patient
|
|
What is defined as "how well a boundary between two areas of differing radiodensity is revealed"?
|
sharpness
|
|
What is defined as "how well an image reveals small objects that are close to each other"?
|
resolution
|
|
If a radiograph is very dense, then it is (black/white)
|
black
|
|
The difference between the blacks, whites, and greys is known as what?
|
contrast
|
|
What is the useful range of radiographic density? (# value)
|
0.3 (light) to 3.0 (dark)
|
|
Increasing current will (increase/decrease) density
|
increase
|
|
Increasing voltage will (increase/decrease) density
|
increase
|
|
Increasing the distance of the x-ray source (tungsten block) will (increase/decrease) density
|
decrease
|
|
Increasing the development time will (increase/decrease) density
|
increase (fog)
|
|
What does the acronym SFD stand for?
|
Source-image receptor distance (I'm guessing the "F" means "film" too)
|
|
Beam intensity varies (directly/inversely) to the SQUARE of the SFD
|
inversely
|
|
Shorter distance between film and x-ray results in (higher/lower) density because of a (high/low) beam intensity
|
higher; high
|
|
Doubling the distance does what to the density?
|
1/4th the density
|
|
Halving the distance gives (2x/4x/8x) the density
|
4x
|
|
Underdevelopment produces (high/low) density images
|
low
|
|
(Over/under)development results in fog
|
over
|
|
Higher speed image receptor requires (more/less) mAs to produce a density change compared with slower-speed image receptors
|
less
|
|
Higher-speed intensifying screens require (less/more) mAs
|
less
|
|
True or False, fog has no effect on image contrast
|
false
|
|
Fog leads to (increased/decreased) density and (increased/decreased) contrast
|
increased; decreased
|
|
If you want to change the contrast, you need to change what parameter?
|
kVp (Kontrast - Kvp)
|
|
(High/Low) contrast has many shades of grey
|
low
|
|
Long scale contrast is synonymous with (high/low) contrast
|
low
(Long = Low) |
|
High kVp = (low/high) contrast
|
low
|
|
Low kVp = (low/high) contrast
|
high
|
|
What is the range for high and low kVp?
|
High = 80-90
Low = 60-65 |
|
A radiograph with many shades of grey is considered (long/short) scale
|
Long (remember, many greys is LOW contrast and Long = Low)
|
|
PAs should have generally (higher/lower) contrast whereas bitewings should have generally (higher/lower) contrast
|
lower, higher
|
|
If you increase kVp, you must (increase/decrease) mAs or time to maintain density
|
decrease
|
|
Changes in density by changing mAs has what residual effect on contrast
|
none
|
|
Why will a change in CONTRAST also have an effect on DENSITY?
|
Because to change contrast you must alter the kVp. When you alter the kVp, you also effect the density at the same time.
|
|
You get better resolution and detail with a (small/large) focal pt.
|
small
|
|
When two distinct x-ray waves strike an object and travel in distinct lines, they hit the film in slightly different locations. The shadow-like distortion created by this is called what?
|
Umbra (penumbra is the border of an umbra)
|
|
With similar techniques, objects farther from the image receptor will always magnify (more/less) than objects closer to the image receptor
|
more
|
|
Divergent x-ray photon paths cause (enlargement/reducing)
|
enlargement
|
|
True or False, X-rays originate from an AREA and travel in diverging, but straight, lines that radiate from the source
|
True
|
|
Distance between focal spot and object should be as (long/short) as possible
|
long
|
|
Image receptor (film) should be as (close/far) as possible to the object
|
close
|
|
If you have the image receptor and object very close together, this will help to reduce what?
|
magnification (this is like of you have a flashlight on your hand but the shadow is on a wall and you bring the wall in closer to your hand... now the shadow gets smaller and smaller until the wall is right up to your hand and the shadow and your hand are virtually the same size)
|
|
True or False, it doesn't matter what angle the CR is to the film, as long as it's pointed at the middle
|
False, it should be perpendicular
|
|
Smaller focal spot = (larger/smaller) pneumbra
|
smaller
|
|
Geometric unsharpness can be minimized by having a (large/small) focal spot
|
small
|
|
Why are lingual cusps sharper?
|
They are close to the image receptor
|
|
To decrease magnification, we should (increase/decrease) the source to object distance
|
increase (think about if you hold a flashlight close to your hand, the shadow will be large. Hold the flashlight far from your hand, and the shadow becomes much smaller [less magnified])
|
|
Radiograph has anatomical accuracy when proximal contacts are (open/closed)
|
open
|
|
Radiograph has anatomical accuracy when buccal and lingual cusps of posterior teeth are (separated/superimposed)
|
superimposed
|
|
Radiograph has anatomical accuracy when buccal and lingual portions of alveolar crest are (separated/superimposed)
|
superimposed
|
|
If you are talking about the MANDIBLE, vertical OVERangulation is when the BID is pointed at the tooth from (above/below)
|
below! for the maxilla, it would be above. the plane reverses when considering maxilla and mandible
|
|
Radiograph has anatomical accuracy when the zygomatic arches (are/are not) covering the roots of the maxillary teeth
|
are not
|
|
In a bitewing, you can tell if it is right side up because the arch should slope (upward/downward) in the 3rd molar region
|
upward
|
|
*in class question
The size of the x-ray tube focal spot influences the image: A. density B. contrast C. definition D. distortion |
C. Definition
|
|
*in class question
Which of the following does not control magnification of the object? A. Focal spot-image receptor distance B. Alignment of image receptor, object and BID C. object-image receptor distance D. Cathode size |
D. Cathode size
|
|
*in class question
True or False: Radiographic fog will increase density and decrease contrast |
A. True
|
|
*in class question
Why extending the source-image receptor distance by using a longer BID a necessary adjunct to intraoral paralleling technique? A. To avoid unsharpness of the image B. To avoid shape distortion of the image C. To reduce scatter radiation D. To avoid anatomical superimposition |
A. To avoid unsharpness of image (could also be B)
|
|
*in class question
How do you change from a long scale contrast to a short scale contrast and still maintain the density? A. Decrease kVp and increase mAs B. Increase kVp and decrease mAs C. Decrease kVp and decrease mAs D. Increase kVp and increase mAs |
A. Decrease kVp and increase mAs (Low kVp = high contrast; high contrast = short scale contrast)
|