Purpose of Grids
|
Absorb (secondary) scatter radiation, thereby improving image contrast
|
|
Grid Materials
|
Radiopaque strips (usually lead)
Radiolucent interspace material (aluminum or plastic fiber) |
|
Grid Ratio
|
= height of lead strip/distance between strips
High grid ratio is best at removing scatter (though it requires greater accuracy and higher patient dose) |
|
Grid Frequency
|
# of lead strips/inch or cm
(60-200 lines/in normal) High frequency shows less grid lines High frequency needs higher technique/patient dose |
|
Lead content
|
Determined by grid ratio and frequency
As lead content increases, better at removing scatter, better contrast High lead content: high grid ratio and low frequency |