anatomically programmed radiography (APR)

radiographic technique chart programmed into x-ray console, anatomy/body habitus indicated

factors that are selected by APR (6)

kVp, mAs, AEC ionization chamber, FSS, tube, bucky

automatic exposure control regulates _ _ without technologist involvement

exposure time

AEC

radiation detection device that measures the quantity of x-rays received by the patient or IR. exposures reaches predetermined values the system terminates exposure. value set at installation based on receptor system and density desired

types of AEC

phototimer, ionization chambers

phototimer is located

under the cassette

ionization chambers are located

under patient, above cassette

phototimber utilize

photomultiplier tube

ionization chambers utilize

a parallel plate ionization chamber, "cells" or "chambers"

minimum reaction time

length of time necessary for AEC to respond to radiation and for generator to terminate exposure

back up time or mAs

protects patient if AEC isn't working at 150% of expected mAs

back up activation (6)

wrong detector selected, wrong tube selected, incorrect tube (bucky alignment), prostheses, lead shield, incorrect back up timer setting

factors that affect AEC

patient position, ion chamber selection, density selector, collimation, image receptor speed, prosthesis, back up timer