• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/140

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

140 Cards in this Set

  • Front
  • Back

What is the function of a grid?

-Designed to improve contrast by absorbing scatter radiation before it reaches the film


-Absorbs scatter radiation produced by x-ray photon interactions with patient

Where is a grid used?

-Place in between the patient and the IR

What are the possible photon interactions?

-Pass thorough unaffected


-Absorbed


-Scatter (interact and change direction)

When is a grid used?

when body part 10cm or greater or technique above 60 kVp

How can you maintain image quality with grid use?

-Increase technical factors


-Increase patient dose

How are grids classified?

-Materials


-Grid ratio


-Grid frequency

What are grids made of?

-Radiopaque strips – usually lead


-Placed side by side


-Absorb scatter (and some primary) radiation


-Radiolucent inter-space material


-Usually aluminum or plastic fiber


-Radiation passes thru easily with only absorbing small amount of

What is grid ratio?

-Ratio of the height of the lead strips to the distance between the strips


-GR = h/D


-h=lead strip height


-D = interspace width

What happens with a higher grid ratio?

-More effective at removing scatter


-Requires greater accuracy in positioning


-Increased possibility of having grid error due to misalignment

What are the available grid ratios?

-5:1, 6:1, 8:1, 10:1,12:1, and 16:1



-most common is 12:1

What is the relationship between distance between lead strips and grid ratio when height of strips remains the same?

Inverse

What are the characteristics of increasing grid ratio?

-Grid lines may be taller


-Higher grid ratio absorbs more radiation


-Grid lines may be closer together


-Higher grid ratio absorbs more radiation

What is grid frequency?

-Number of grid lines per inch



-Higher grid frequency has thinner lead strips

What are the possible grid ranges?

-Grid ranges


-60-196 lines/inch


-25-80 lines/inch


-Most common


-85-103 lines/inch


-33-41 lines/inch

What is total lead content?

-Combine info from grid ratio and grid frequency


-Determines total lead quantity


-Lead content usually greater:


-Higher grid ratio


-Lower grid frequency

What is the result of increased lead content?

-Increased removal of scatter


-Improve contrast


-Increase lead content


-Decrease scatter reaching film


-Increase contrast

What are the types of grid pattern?

Linear



Criss-cross

What are the types of linear grid?

Parallel and Focused

What is a criss-cross grid?

-Two grids sandwiched together


-Strips usually both parallel to each other


-And at right angles to the long dimension of the grid


-Not commonly used

What is a parallel grid?

-All lines parallel to beam


-Primary beam will never be aligned with all the strips


-Only the strips directly under the x-ray tube in true alignment with primary beam


-Not recommended for SID less than 140cm

What is a linear parallel grid?

-Strips don’t coincide with beam divergence


-Get some grid cut-off at lateral edges


-Best used for long SIDs

What is a focused grid?

-Lead strips are tilted progressively as they move away from center


-Lines through each strip converges at a point known as the grid focus


-Lead strips match divergence of the x-ray beam


-X-ray beams would meet at a line called the convergence line


-X-ray tube must be located at the convergence line to eliminate lateral grid cut-off

What is grid radius?

-Distance from grid to the convergence point of a grid


-Coincides with useful SID provided by grid


-Convergence line the source of x-rays

What is focal range?

-SID range most effective for grid


-Distance between grid and x-ray tube for best grid performance

What are the factors when choosing a grid?

-1.Grid conversion factor (dose required)


-2.Selectivity


-3.Contrast improvement ability

What is grid conversion factor?

-Also called Bucky Factor


-Calculates required change in technique when changing from no grid to a grid


-GCF = mAs with grid/mAs without grid

When does GCF increase?

with higher grid ratios and higher kVp

What is the formula for GCF?

-GCF = mAs needed with grid/mAs needed without grid



mAs1/mAs2 = GCF1/GCF2

If not given kVp, or chart, what are the grid conversion factors?

No grid =1


5:1 or 6:1 = 3


8:1 = 4


12:1 = 5


16:1 = 6

What do grid vary in respect to?

-Grid ratio


-Frequency


-Lead content

What are the factors that influence GCF?

-Grid ratio


-As grid ratio increases, absorbs more scatter


-Higher grid ratio = higher bucky factor


-KVp


-As kVp increases, produces more scatter


-Higher kVp – higher bucky factor

How can you measure grid performance?

Selectivity



Contrast Improvement Ability

What is selectivity?

-Amount of the primary beam/scatter a grid absorbs/transmits


-Selectivity = %primary radiation transmitted/% scatter transmitted

How does lead content and scatter affect selectivity?

-The better a grid removes scatter, the greater the degree of selectivity


-Grids with higher lead content have a greater selectivity


-So: increased lead content form higher grid ratios and lower grid frequencies

Which of the following has a higher lead content?


10:1 grid with 10-lines/inch or 10:1 grid with 8 lines/inch

10:1 grid with 8 lines/inch as strips are wider and absorb more radiation

What is contrast improvement ability?

-Measurement of grid performance (contrast)


= K


-K = contrast with grid/contrast without grid

What is contrast improvement ability dependant on?

-KVp


-Field size


-Volume – scatter production


-Higher K#, greater the contrast improvement

What changes were made by Potter-Bucky to moving grids?

-Improvements to grid use


-Realigned strips in 1 direction


-Made strips thinner


-Moves grid during exposure to reduce visibility of grid lines (perpendicular to grid lines)

What are the types of moving grids?

-Reciprocating



-Oscillating

What is a reciprocating grid?

-Motor drives the grid back and forth during the exposure


-Total distance moved is 2-3cm

What is an oscillating grid?

-Oscillating Grid


-Electromagnet pulls the grid to one side and then releases it during the exposure


-Moves in circular motion

What is the air gap technique?

-Alternate to using grid


-Patient placed at greater OID


-Creates an air-gap between patient and IR


-Less scatter reaches IR


-Improves contrast


-Disadvantages


-Reduced sharpness due to increased OID

What air gap is equivalent to a 15:1 grid ratio?

25cm air-gap

What are the types of grid error?

-Off-level/tilted


-Angled against grid lines


-Off-center


-Off-focus (incorrect SID)


-Upside down


-Moire effect

What is grid cut off?

with the use of a parallel grid


-Undesirable or abnormal absorption

What is Off-level or Tilted grid error?

-Grid cut-off with use of tilted grid


-Tube angled across grid lines


-Only error possible with parallel grid


-Resulting image:


-Decrease in density across entire image

What is off-center tube alignment error?

-X-ray tube not centered to center of grid


-Laterally de-centered tube


-Divergence of beam not matched up to focused grid


-Resulting image:


-Decreased density across entire image


-Increasing lateral de-centering increases loss of density

What is tube outside of focal range error?

-Incorrect SID


-Tube is outside grid’s specified focal range


-Resulting image:


-Grid cut-off along edges of image


-Farther from focal range: the greater the cut-off


-More noticeable below focal range


-Higher grid ratios require more precise centering

What is focused grid used inverted error?

-Focused grid used inverted, upside down


-Resulting image:


-Severe peripheral cut-off

What is Moire effect?

-In digital systems with stationary grids


-Grid lines running in same direction as movement of laser beam scanning, imaging plate visible


-To prevent:


-Use high frequency grids


-103 lines/inch or greater

What is beam restriction?

Collimation

What are the factors that influence the amount of scatter produced?

-Kilovoltage


-Volume of material


-Type of material

Kv effects on scatter?

-Controls the penetrability of the beam


-As KV increases more photons go through the patient


-As KV increases, scatter production increases


-At a higher KV there are more Compton interactions that photoelectric interactions


-More scatter leaves the patient


-Good for patient, bad for image

Amount of irradiated material's affect on scatter?

-Affected by the


-Atomic number of the material


-Volume of the material


-The higher the atomic number the less scatter


-Volume is affected by the field size and the patient thickness


-As volume increases the amount of scatter increases

What are the types of beam restrictors?

-Use beam restrictors to control the field size and patient volume



-Aperture diaphragm


-Cones/Cylinders


-Light-beam collimator

What is an aperture diaphragm?

-Flat sheet of metal (lead) with a hole in the middle


-Different diaphragms for different size receptors and different distances

What are the advantages and disadvantages of an aperture diaphragm?

-Advantages


-Low cost, easy to use, simple design



-Disadvantages


-Large penumbra on periphery of image, off-focus radiation, no light

What are cones/cylinders?

-Circular aperture diaphragms with metal extensions


-Cone = gets larger at the end

What are the advantages and disadvantages of cones/cylinders?

-Advantages:


-Inexpensive, easy to use


-Disadvantages


-Fixed field size, flared cones still have large penumbra on periphery of image and off-focus radiation

What are collimators?

-Added filtration by mirror

What are the advantages and disadvantages of collimators?

-Advantages:


-Infinite number of field sizes, light source, reduce penumbra and off-focus radiation


-Disadvantages:


-More complex, more expensive (than a simple style)

What is field light?

-A light shines on a mirror (45 degrees) in the x-ray beam


-Light source and x-ray source are equal distance from mirror


-Crosshair in the center to locate middle of the beam


-Size of field shown on front of collimator

What are do shutters do?

-Regulate the field size


-3 sets of lead plates at right angles to each other


-Upper shutters reduce off-focus radiation


-Lower 2 pairs reduce penumbra on periphery

What are the SC 35 recommendations for collimators?

-Beam must be size of IR or smaller


-Should see edges on all x-ray images


-Use smallest IR possible


-Restrict beam to area of interest


-Irradiating any areas outside of interest = unnecessary radiation


-Avoid irradiation to gonads, female breasts, and thyroid

When does scatter increase?

-KV increases = increases


-Field size/ volume increases = increases


-Thickness of object being imaged/volume increases = increases


When scatter increases contrast decreases

When does image contrast decrease?

-KV increases = decreases


-Field size/ volume increases = decreases


-Thickness of object being imaged/volume increases = decreases


When scatter increases contrast decreases

What is filtration?

-Selectively absorbs low energy photons


-Absorbed by the patient


-Don’t contribute to the image


-“Harden” the beam


-Measured in thickness of aluminum equivalency (AL/Eq)


-Using filtration requires increase in exposure factors, to compensate for decrease in exposure


-Overall decrease in patient dose

What is inherent filtration?

From within the design of the tube and housing


-0.5 – 1.0mm Al/Eq.


-Mirror


-1.0mm Al/Eq.

What is added filtration?

-Outside the tube and housing and before the IR (usually before the collimator)


-Should absorb many low energy photons and few high energy photons


-Collimator


-Contributes 1.0mm Al/Eq.

What is the SC 35 recommendation for an x-ray unit with a maximum kV at 70 or over?

2.5mm Al/Eq.

What is total filtration?

-Sum of the inherent and the added filtration


-Not any compound or compensating filters


-As photon energy increases


-The % of photons attenuated decreases even when filtration is increases

What happens if there is greater than 3.0mm Al filtration?

-A reduction in the skin dose doe not warrant the exposure increase


-Hard on tube

How do we check total filtration?

-Half Value Layer


-The amount of absorbing material that will reduce the intensity of the primary beam to half its original value


-Indirect measurement of the total filtration

What is compound filtration?

-Uses 2 or more materials that compliment each other in absorption ability


-Each layer absorbs the characteristic photons of the previous layer


-K-edge filter


-Used often in radiotherapy

What are compensating filters?

-Compensates for unequal absorption within the patient when part is differing thickness and or/composition


-Decreases dose, as without filter two images may be needed


-Produces a more uniform exposure


-Aluminum or leaded plastic


-Attach to the collimator or under the patient

What is atomically programmed radiography?

-Radiographic technique chart programmed into x-ray console


-Body part indicated with either words or image


-Anatomy and body habitus indicated


-Technologist can override settings when necessary due to patient condition

What is automatic exposure control?

-Regulates exposure time without technologist involvement


-Technologists still control mA and kVp based on patient condition


-A radiation detection device measures the quantity of x-rays received by the patient or IR
-Exposure reaches predetermined values the system terminates exposure


-Value set at installation based on receptor system and density desired

What are the types of AEC?

Phototimer and Ionization chambers

What is a phototimer?

-Older style not commonly in use


-Utilizes photomultiplier tube


-Detector under cassette

What are ionization chambers?

-Utilizes parallel plate ionization chamber – detectors also called cells or chambers


-Referred to as “phototiming” although technology is different


-Detector in front of cassette but under patient

What is minimum reaction time?

-Length of time necessary for AEC to respond to radiation and for generator to terminate exposure


-Min. reaction time in todays machines usually less than 0.001 secs.


-Can’t terminate quickly enough with extremely high speed screen combos

What is back-up time?

-Manually set time to terminate exposure


-Can’t exceed tube limit


-Wastes tube life


-Dramatically increases patient dose


How should back-up time be set?

-Should be set at 150% of anticipated manual exposure mAs


-Above 50 kVp terminates at 600 mAs


-Below 50 kVp terminates at 2000 mAs

What causes back-up activation?

-Wrong detector selected


-Wrong tube selected


-Incorrect tube – bucky alignment


-Tube is not centered to bucky


-Prosthesis


-Lead shield


-Incorrect back-up timer setting


-Set too short and prematurely terminates exposure

What factors affect AEC?

-Patient position


-Ion chamber selection


-Density selector


-Collimation


-Image receptor speed


-Back up timer too short


-Prothesis, FB, or lead shield over chamber

How does patient position affect AEC?

-Density based on tissue directly above chamber


-Must position patient to place adequate tissue above chamber to reach a desired image on receptor


-Positioning must be accurate

How does ion chamber position affect AEC?

-Selection based on tissue and desired density on film


-Ex. Lateral chest – center chamber


-PA chest: Usually both tight and left chambers, density evened by generator


-Shoulders, spines, knees, and abdomen– all have specific chamber selections

How does the density selector affect AEC?

-+3, +2, +1, 0, -1, -2, -3


-Increases or decreases exposure index value


-Steps usually set at a 25% increase or reduction in radiation intensity from preset radiation turn off dose


-Makes the AEC chamber more or less sensitive to radiation

How does collimation affect AEC?

-Too little, scatter will prematurely terminate exposure


-Too much, cut off portion of sensor and exposure will go to back-up to reach set radiation amount

How does image receptor speed affect AEC?

-Calibrated for one speed only


-Faster speed systems result in: Too much exposure used


-Slower speed systems result in: Too little exposure used

When should mobile x-rays be used and why?

should only be used if patient cannot come to department


increased exposure to patient (?) and techs, others

What is a capacitor discharge mobile?

-discharge off high voltage to capacitor supplied by batteries


-at exposure switch causes the rectification circuit to charge capacitor in generate not tube


-at indicate level, capacitor filled


depression of switch discharges power to Xray tube


-Xray photons then produced and released

What are battery operated mobiles?

nickel-cadmium batteries , rechargeable


provides power to Xray tube


DC converter, step up transformer and rectifier to get desired kV


Then applied to tube


same as mobile units at NAIT

How many batteries does a battery operated mobile have?

two sets:


one to drive it


one for X-rays

What are the advantages of a battery operated mobile?

batteries can fail


recharging can take 2 - 12 hour


batteries require maintenance


mA value is pre set

What needs to be paid careful attention to when using a mobile?

positioning


cassette placement


grid use


lower grid ratio


parallel


other peoples whereabouts

Mobile safety

key to turn power on and off


front bumper will stop movement if impacted


clean it off regularly


take precautions against radiation exposure to self and others

Mobile Radiation Safety Chart

-duty to protect everyone, people not required --must leave area


-announce intention to expose


-carry lead aprons


-do not put parts in primary beam (except appropriate patient parts)


-provide gonadal shielding


-maximize your distance from patient during exposure


-avoid repeats

What is an artifact?

-optical density / brightness on an image unrelated t the intended patient anatomy


-may require a repeat or lead to misdiagnosis


-ID is important to correct the error, identify trends and prevent further artifacts from occurring

What are foreign bodies>?

-are not considered artifacts as they pertain to the patients anatomy at the time of the exam


-typically the full foreign body is included on all views

What is a prothesis or fixation?

internal fixtures, external fixation devices and prostheses are also not considered to be artifacts and often must be fully included

What is an implanted device?

implanted devices such as pacemakers, neurostimulators, brea and cochlear implants are also not considered to be artifacts

What are the time periods when artifacts can occur?

exposure and acquisition


processing or post acquisition and display


handling and storage

What is exposure and acquisition artifact

-those related to radio exam itself


-caused by the patient, tech, or the equipment during procedure


-easy to identify and correct


-can be avoided with use of appropriate care and attention

What is a double exposure?

- two images on one cassette

How is scatter (quantum mottle) caused?

-too few radiation photons at image receptor


-grainy, speckly image


-how should you decrease noise - increased mAs but that increases the dose

What is incorrect cassette orientation?

tube side not facing the tube



see cassette materials on image

What is backscatter artifact?

due to lack of lead at back of detector

What is a phantom image?

-occurs because ht machine doesn't erase the cassette properly


-decreased brightness and contrast, see parts of old image

What is a dropout artifact?

DR pixel array failure


small areas where pixel is missing

How do defects or scratches on the imaging plate occur and what do they cause?

- increases image brightness in areas



- scratched IP


Dust in IP cassette or DR detector

What is tube filter failure?

-filters are the assembly of the cry tube becoming loose and mispositioned in the cry beam


-the filters then attenuate the cray beam non uniformly and result in unwanted attenuation pattern


-lines of darker/lighter image

What is processing or post acquisition artifact and display?

-occur after image acquisition for digital systems


includes image professors software


-malfunctions, IRD problems, printing and display artifacts


-dropout artifact, pixels missing i mage, dirt on laser

What is histogram error?

improper image brightness

What is indirect collimation?

not parallel to cassette


brightness issues

What is halo artifact?

dark bands at the structure interface where the brightness is very different

What is Moire artifact?

-from aliasing (sampling errors)


-the lines of the grid coincide with the lines of the image


-the image processor then mistakenly samples


-both the grid signal and the image signal therefore sampling the image signal less than twice per cycle

What is image reader/scanner error?

-memory problems


-digitization problems


-communication problems


-dirt in laser / optics system


-all can cause:


skipped scan lines


missing pixels


distorted images

What do printer errors cause?

laser misalignment



may cause shading (dark areas)



may cause corduroy effect

What is image enhancement artifact?

-from any erroneous post processing


incorrect presets


brightness


contrast


mag


edge enhancement


spatial filtering

What is pixel dropout artifact?

from computer screen irregularities


tiny blank areas (pixels)

What is handling/storage artifact?

these artifacts occur due to the way the IPs are handles and stored



often a result of IPs stored too close to radiation or heat source

What is IP fogging?

-scatter and background radiation


decreased contrast


-digital image receptors are very sensitive to


-radiation fogging


-routine erasure of Ip critical to eliminating the occurrence of this artifact

What is heat blur?

IP exposured to heat before read



image is blurry

What is post-processing ?

-There are many changes that can be made to the presentation of the image data


-There are also many names for some of these functions, sometimes vendor specific


-The digital image displayed is only a small part of the obtained image data; otherwise it would be impossible for us to view it


-So each image is only a “window” on the total data range available

What is windowing?

-The process of selecting a segment of the total pixel value range (dynamic range) and then displaying the pixel values (shades of grey) within that segment


-The technologist can adjust both the center of the window (window level) and the width of the window (window width)


-The combination of these parameters determine the range of pixel values that will be displayed in the image (brightness and contrast)

What is window level?

-Controls image brightness


-Direct = Relationship between window level and image brightness


-Inverse = Relationship between window level and density


-When window level increases, the brightness increases and the density decreases

How does window level work?

-Works by addition or subtraction of each pixel value


-Adapted and displayed for specific exams


-Default Window Level is different for each area of interest


-Information outside chosen window level is lost on image

What is window width?

-Controls image contrast


-Gradational enhancement, contrast enhancement, tone scaling, image compression, image expansion…


-Window width controls contrast by setting the range of grey shades that are visible

Why should you not do any windowing before saving?

Technologist should avoid adjusting the WW/WL prior to sending the image to PACS. Once the WW/WL is adjusted and saved, the wide dynamic range form the original image is lost, leaving only the range that was saved. The radiologist then has a narrower dynamic range to use when evaluating the image.

What is histogram equalization?

Also used to enhance brightness and/or contrast

What is LUT?

(Look-up table) adjustments


-Another way to change brightness and contrast, or invert shades


-The LUT for the image is compared to and adjusted to look similar to an LUT stored in the imaging system processor


-The stored LUTs each can be selected to produce images with different contrast characteristics

What is edge enhancement/sharpening?

-Can enhance or suppress certain image frequencies (spatial resolution adjustments)
-Edge enhancement or sharpening


-High pass filtering


-Increases detail, increases noise


-Amplifies the high frequencies or deletes the low frequencies


-Small structures can be buried in an enhanced edge


-The edges of all structures are enhanced – those of diagnostic value and undesirable edges

What is image smoothing?

-Can enhance or suppress certain image frequencies (spatial resolution adjustments)


-Smoothing


-Low pass filtering


-Amplifies the low frequencies, deletes the high frequencies, or averages pixel values between adjacent pixels


-Decreases noise, decreases detail

What is image stitching?

-Combining multiple images into one

What is image inversion?

-Changing white to black and black to white

What is image subtraction?

-Removing part of the anatomy

What is image magnification?

-Zoom


-Electronic magnification

What is region on interest?

-To quantify pixel values


-For identification

What is background removal or shuttering?

-Removes white background around image


-It is more pleasant to look at without the white borders