Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
127 Cards in this Set
- Front
- Back
A radiographer can best minimize the amount of scatter radiation reaching the image receptor by restricting the beam. True or false? |
True |
|
The principal factors that affect the amount of scatter produced are mas and the type of irradiated material. True or false? |
False |
|
When kilovoltage is increased with no other changes in technical factors, fewer scattered photons will result. True or false? |
false |
|
scattered photons from compton interactions are of no use in demonstrating radiographic structures of interest. True or false? |
True |
|
Image quality is improved when scatter reaching the image receptor is reduced. True or false? |
True |
|
Bone absorbs More radiation and procedures more scatter than soft tissue. True or false? |
False |
|
soft tissue absorbs More radiation and procedures more scatter than bone. True or false? |
false |
|
Lead masks may serve as beam restrictors. True or false? |
True |
|
scatter radiation in radiography is produced during? |
Compton interactions |
|
scattered photons will? |
-impair image quality - add density to image receptor exposure - not be considered part of the useful beam |
|
A radiographer can best reduce the amount of scatter radiation from reaching the receptor by? |
-collimating -beam restriction - utilizing a grid |
|
which of the following is placed between the patient and the image receptor to absorb scatter radiation? |
radiographic grid |
|
A key factor that affects the amount of scatter produced is? |
Kilovoltage |
|
As the kilovoltage increases, the percentage of x-rays that undergo _______interactions will increase. |
Compton |
|
As the kilovoltage increases, the percentage of photons that undergo a ______interaction will decrease. |
photoelectric |
|
If an increase in kilovoltage is accompanied by a reduction in mas, to maintain the same exact dose the overall result will be? |
a decrease in the amount of scatter produced. |
|
As the volume of irradiated tissue ____ , the amount of scatter produced____. |
increases; increases |
|
Tue volume of the irradiated tissue increases as? |
field size and patient thickness increase. |
|
The _____ the atomic number of the irradiated material, the ______ the number of photoelectric absorption interactions that occur. |
higher ; higher |
|
The____ the atomic number of the irradiated material, the____ the number of photoelectric absorption interactions will be, and the____ the scatter produced. |
higher; higher; less |
|
When a beam restricting device is used to limit field size, technical factors may need to____ to compensate for the charge in image receptor exposure. |
be increased |
|
All of the following are considered beam restricting devices except? |
beam filter |
|
Penumbra is ___ when the beam restrictor scatter is at a ___ distance from the port. |
reduced; increased |
|
an effective method for reducing scatter radiation production is the? |
15% rule of kvp change |
|
The shutter of a collimator? |
regulate field size, reduce penumbra, reduce stem radiation |
|
When collimating down from a 14x17 field size to a 8x10 field size: |
mas must be increased to maintain radiograohic image density. |
|
The collimation assembly of a radiographic unit? |
employs an upper and lower set of shutter blades |
|
Placing a lead blocker adjacent to an obese patient's anatomy during a lateral lumbar spine radiograph? |
will improve image quality |
|
The incident beam is significatly altered as it passes through the patient. true or false? |
true |
|
the beam emitted from the patient contains diagnostic information. true or false? |
true |
|
subject distortion occurs naturally. true or false? |
true |
|
subject distortion can be reduced by accurately positioning the patient. true or false? |
true |
|
the patient has an impact on all the properties that affect the radiograoh quality. true or false? |
true |
|
the relationship between image receptor exposure and the patient is termed subject density. true or false? |
true |
|
the relationship between radiographic contrast and the patient is termed subject contrast. true or false? |
true |
|
The ____ the body part, the ____ the attenuation. |
thicker;greater |
|
which of the folloeing elements would attenuate a greater percentage of the xray beam? |
calcium |
|
Bone produces less image receptor ___ because it attenuates the xray beam more than soft tissue does. |
density |
|
Foe the same exposure intensity, air produces? |
greater exposure intensity to the receptor than other materials. |
|
Which of the following is the least attenuator of the xray beam? |
pneumonated lung air (air filled) |
|
As subject density ____, image receptor exposure ____. |
increases; decreases |
|
The degree if differential absorption is referred to as? |
subject contrast |
|
The recorded detail of a structure is dependant on? |
- its position within the body - the okacement of the body in the relationship to the film - the size of the part |
|
Distortion is the misrepresentation of the ____ of an object. |
- length - shape - width |
|
The primary source of radiation exposure for occupational persinnel is due to? |
Compton scatter |
|
Clinically speaking from a diagnostic viewpoint, the only useful radiation exposure is that radiation? |
detected by the receptor. |
|
digital radiography replaces? |
conventional film |
|
which type of system uses a two part process to convert incoming x ray photons to an electronic digital signal? |
indirect acquisition |
|
during the two part acquisition system, a ____ converts xray photons to light. |
scintillator |
|
During the two-part acquisition system, a ___ converts light into an electronic digital signal. |
photodetector |
|
An imaging plate includes a ____ layer. |
-phosphor -protective - conductor |
|
A charged couple device is a photo detecor capable of converting? |
visible light to an electrical signal |
|
The latent image will lose about 25% of it's energy in? |
8 hours |
|
The latent image created by CR is processed by a? |
image reader device. |
|
When the CR imaging plate is scanned, the phosphors of the latent image release energy in the form of? |
light |
|
The histogram generated from the image data allows the digital system? |
- to find the useful areas on the IP - to create a graph based on the anatomy selected |
|
A CR artifact caused by insufficient mAs (low exposure) is? |
quantum mottle |
|
The smallest part of a digital image describes: |
pixel |
|
As pixel size decreases, resolution ____. |
increases. |
|
The window level controls _______. |
image brightness |
|
All of the following are quality factors for digital imaging, except? |
distortion |
|
As the sensitivity number (s) increases, the image becomes? |
underexposed |
|
Detective Quantum Efficiency (DQE) is a measure of the sensitivity and accuracy by which the image receptor converts the incoming data to the output device. True or false? |
True |
|
Bit depth is controlling factor of ____ in digital radiography. |
density |
|
The range of possible shades of gray as determined by the manufacturer of a digital imaging system is called? |
Bit depth |
|
Digital imaging plates are more radiosensitive and conventional film. true or false? |
true |
|
Computed radiography utilizes direct acquisition. True or false? |
False |
|
Which of the following is more desirable for image quality? |
High SNR |
|
Digital radiography utilizes the same xray tube as conventional radiography. true or false? |
true |
|
Each cell of an image matrix is called what? |
a pixel |
|
What is the purpose o the conductor layer of the IP? |
Protect the IP from electrostatic shock |
|
As the exposure index number increases, the image becomes: |
underexposed |
|
In digital imaging, the light that comes off the imaging plate when being scanned by a laser proportional to the xray exposure received by the phosphor. true of false? |
true |
|
Which of the following is referred to as the active layer of the CR cassette? |
Phosphor layer. |
|
Direct digital imaging systems use imaging plates. True or false? |
True |
|
In a digital imaging plate a fill factor of 80% indicates that the pixel is _____% pixel detector and _____% image conversion electronics. |
80 and 20 |
|
Digital radiography (DR) is characterized by? |
the use of reusable receptors. |
|
Each binary number is called a? |
bit |
|
An eight-bit word is called a? |
byte |
|
CR systems acquire clinical information by? |
laser scanning a PSP plate |
|
Resolution is controlled by ___ size. |
pixel |
|
Each pixel in a digital display matrix is located according to its? |
electronic address |
|
A matrix of size 2048x1536 would have ____ pixels. |
3,145,728 |
|
If the digital FOV is 1024 and the display matrix is 512, the pixel size would be? |
2 |
|
In digital image processing, the digital data set is arranged into a histograph distribution in order to determine the? |
- exposure index - values of interest (VOI) |
|
The number of shades of gray; ranges from 8 bits to 32 bits. A grayscale bit depth 8-32 equals a range of 1-4 byte of storage that would be required per pixel in the image matrix. A grayscale bit depth of 12 produces 212 gray levels. This represents 4,096 different shades of gray. |
Gray scale bit depth |
|
Typically inside a cassette; used in computed radiography. |
photostimulable storage phosphor imaging plate |
|
The physical distance between pixels; generally measured from center to center. |
pixel pitch |
|
Applied to the data that has the standard contrast for that exam to give the desired image contrast for display. |
look-up table (LUT) |
|
Two-part process involving a scintillator and a photo detector. |
indirect conversion |
|
A point processing operation that changes the contrast and brightness of the image on the monitor. |
windowing |
|
Direct conversion of incoming xray photons to an electronic signal. |
direct conversion |
|
Correction of an exposure that is outside the range from underexposure or overexposure by shifting the histogram to the correct area. |
rescaling |
|
Imaging systems that replace traditional film with a PSP and cassette; the plate must be processed in an image reader system. |
digital radiography (DR) |
|
Imaging systems that replace traditional film with a reusable detector; the detector and reader are a permanent part of a table or wall unit, and therefore a cassette is not needed. |
digital radiography (CR) |
|
Noise on a digital image is the summation of inherent electron system noise and quantum mottle noise. Of these two, the radiographer can improve the noise on an image by? |
increasing the exposure technique |
|
An S value of 200 represents a? |
exposure to the IP of 1 mR |
|
When using CR technology, it is important to remember that? |
- the IP is extra sensitive to exposure after the inital xray exposure -the IP will lose about 25% of its signal in 8 hours if not processed -the IP is very sensitive to scatter, making grids very important |
|
The size of a digital file size is determined by the? |
-bit depth -pixel size -matrix |
|
In CR technology, preprocessing of data includes? |
-histogram analysis -gray scale analysis -exposure data recognition (EDR) |
|
What compound used in flat panel radiography receptors to convert in coming xrays directly into an electrical signal? |
amorphous selenium |
|
Which of these matrix sizes will have the smallest pixels? |
2048x2048 |
|
Which of these matrices will have the best spatial resolution? |
2048x2048 |
|
Increasing the field of view will have what effect on the pixel size? |
decreased pixel size |
|
which of these technologists errors is most likely to result in incorrect histogram rescaling? |
processing a knee radiograph as a chest |
|
durinf the two part acquisition system, a ____ converts xray photons to light. |
scintillator |
|
during the two part acquisition system, a ____ converts light into an electronic digitwl signal. |
photo detector |
|
an imaging plate includes a ____ layer |
-phosphor layer - protective layer - conductor layer |
|
A charged couple device is a photo detector capable of converting? |
visible light to an electrical signal. |
|
The latent image will lose 25% of its energy in? |
8 hours |
|
The latent image created by CR is processed by an? |
image reader device. |
|
When the CR imaging plate is scanned, the phosphors of the latent image release energy in the format of? |
light |
|
The histogram genersted from the imsge data allows the digital system? |
- to find the useful area (densities) on the IP - to create a graph based on the anatomy selected |
|
A CR artifact caused by unsufficient mAs (low exposure) is? |
quantum mottle |
|
The smallest part of a digital image describes: |
pixel |
|
As oixel size decreases, resolution? |
increases |
|
The window level controls? |
image brightness |
|
All of the following are quality factors for digital imaging, except: |
distortion |
|
As the sensitivity number (s) increases, the image becomes: |
underexposed |
|
Detective quantum efficiency (DQE) is a measure of the sensitivity and accuracy by which the image receptor converts the incoming data to the output device. true or false? |
true |
|
Bit depth is a controlling factor of ___ in digital radiography. |
Density |
|
The range of possible shades of gray as determined by the manufacturer of a digital imaging system is called: |
bit depth |
|
Digital plates are more radiosensitive and conventional film. tru of false? |
true |
|
Computed radiography utilizes direct aquisition. true or false? |
false |
|
Which of the following is more desirable for image quality? |
high snr |
|
digital radiography utilizes the same xray tube as conventionao radiography. true or false? |
true |
|
each cell of an image matrix is called what? |
a pixel |