quality control define

procedures to ensure certain performance standards are met

quality assurance define

monitoring outcome parameters

what is the standard calibration?

10 x 10 field size at 100 SSD


+/- 2% or +/- 2mm

QA deals with...

process

QC deals with...

product

QA is ____________________ process

preventive

QC is _______________________ process

corrective

QA makes sure...

we are doing the right thing the right way

QA makes sure...

the results of what we've done are what we expected

What is the purpose of control panel and indicator light

used to indicate operational function

what is the importance of control panel and indicator light

essential to a correct and safe equipment operation

issues from control panel and indicator

indicator light failure or a reading is not within acceptable range

what is the purpose of mechanical and electrical safety devices

used to verify safety devices

mechanical and electrical safety devices should be __________________ meaning they

fail safe meaning they should operate independently of any other circuit failure

inability to produce _________________ is an inconvenience, but inability to __________________ can cause serious damage

motion, stop motion

what does the override switch do? and how often is it checked

no motion will occur unless override switch is activated while primary switch is on


checked weekly

what does the emergency off switch do?


how often is it checked

used in emergencies, normally in the on position, requires positive action to turn it off. when activated all motion should cease


checked monthly

Main power switch: when is it used?

only in extreme circumstances, not considered a back up system

primary control switch: examples

couch controls, hand pendant

what are limit switches and how often is it checked

it prevents mechanical motion beyond the machine design


-checked quarterly

what does the collsion ring do and how often is it checked

when touch, it stops machine motion, checked daily

what is the centrifugal switch, and how often is it checked

controls starting circuit of electric motor, checked quarterly

when can the centrifugal switch be reversed

only when the motor has slowed to a certain speed

what prevents the radiation beam from being turned on if the door is opened or interrupts the beam when opened

door interlocks

importance of intercoms and monitors and door interlocks

interrupt radiation beam if patient is in distress or moved

area radiation monitoring devices purpose

warning lights activated by the electral power, turns of beam

importance of area radiation monitoring devices

personnel protection: alerts therapist that the beam has shut down at either the appropriate dose of a fault has occured

methodology of the radiation monitoring devices

pass or fail


do not treat and notify physics

purpose of the distance indicator

to set SSD/SAD treatments to a correct distance

how often is the SSD pointer checked

weekly

light field, radiation field and colimator purpose

to determine if the light and radiation fields are congruent and on central axis

importance of the light field, radiation field, and colimator

light field is used as a guide to the size, shape, position, and directions of the radiation beam

machine dose rate purpose

setting appropriate MUs should give a desired dose

importance of machine dose rate

to ensure the patient receives the correct prescribed dose

ways to check machine dose rate

diodes and online dosimeters

beam penetration quality purpose

to make sure the desired dose at any depth for a given exposure is dependent on beam penetration

beam penetration importance

exposure of patient to the correct machine output setting will result in delivery of prescribed dose at depth

beam penetration quality tests

yellow flagged test


red flagged test


notify physics immediately

field symmetry define

half of the beam profile is a mirror image of the other half

field flatness define

defined as measurement of the output variation along the central 80% of the beam profile

purpose and importance of field flatness and symmetry

ensure that symmetry and flatness of the radiation beam are within the baseline profile

how to check for field symmetry and flatness

use tracker daily

mechanical and optical patient alignment devices purpose

to properly position patient for treatment

methodology for mechanical and optical patient alignment devices

notify physics if not within specifications

linear scales on treatment tables purpose

using scales or digital readouts to indicated table height, longtitudinal, and transverse table top motion

how to check linear scales on treatment tables

test comparing isocenter to some point below isocenter using scale or digital readout

collimator rotation purpose

defines the radiation field of treatment

why is it important to check the collimator rotation

to have reproducibility of day to day setups

how to check collimator rotation

spirit level

gantry rotation puspose

angular positions be accurately read

importance of gantry rotation

to deliver proper dose and irradiate desired tumor sites

how to check gantry rotation

spirit level

purpose of stability of isocenter under collimator and gantry rotation

production of a geometrically well defined treatment field which is aimed at the same point under varying machine or couch positions

importance of stability of isocenter under collimator and gantry rotation

there will be a shift of opposed fields relative to one another when using isocentric technique

treatment couch isocenter puspose

used for specialized treatment plans

importance of the treatment couch isocenter deviation

will show early indicator of problems

how to correct treatment couch isocenter

physics will tape piece of paper to psa isocenter while rotating the couch pedastal

control panel and indicator light


-frequency


-materials

daily warm up


Argus

mechanical and electrical devices


-frequency


-material

dependent on test


dependent on test

intercom, monitor, and door interlocks


-freq


-material

daily


argus

radiation monitoring devices


-freq


-materials

daily


argus

distance indicators


-freq


-material

daily (+/- 2mm


argus

light field, radiation field, and collimator


-freq


-material

daily (+/- 2mm)....monthly in community


argus

machine dose rate


-freq


-material

-daily (+/- 2%)...3% in community


-phantom, diodes, tracker

beam penetration


-freq


-material

daily (+/- 2%)


tracker

field symmetry and flatness


-freq


-material

daily (+/-2%)


tracker

mechanical and optical pt alignment devices


-freq


-material

daily (+/- 2mm at iso)


argus

linear scales on tx table


-freq


-material

monthly (+/-2mm)


metric ruler

collimator rotation


-freq


-material

monthly (+/- 1 degree)


spirit level

gantry rotation


-freq


-material

monthly (+/- 2%)


spirit level

treatment couch isocenter


-freq


-material

monthly (+/- 2mm)


no special materials

isocenter stability under rotation of gantry and collimator

monthly (+/- 2mm)


split field test