quality control define |
procedures to ensure certain performance standards are met |
|
quality assurance define |
monitoring outcome parameters |
|
what is the standard calibration? |
10 x 10 field size at 100 SSD +/- 2% or +/- 2mm |
|
QA deals with... |
process |
|
QC deals with... |
product |
|
QA is ____________________ process |
preventive |
|
QC is _______________________ process |
corrective |
|
QA makes sure... |
we are doing the right thing the right way |
|
QA makes sure... |
the results of what we've done are what we expected |
|
What is the purpose of control panel and indicator light |
used to indicate operational function |
|
what is the importance of control panel and indicator light |
essential to a correct and safe equipment operation |
|
issues from control panel and indicator |
indicator light failure or a reading is not within acceptable range |
|
what is the purpose of mechanical and electrical safety devices |
used to verify safety devices |
|
mechanical and electrical safety devices should be __________________ meaning they |
fail safe meaning they should operate independently of any other circuit failure |
|
inability to produce _________________ is an inconvenience, but inability to __________________ can cause serious damage |
motion, stop motion |
|
what does the override switch do? and how often is it checked |
no motion will occur unless override switch is activated while primary switch is on checked weekly |
|
what does the emergency off switch do? how often is it checked |
used in emergencies, normally in the on position, requires positive action to turn it off. when activated all motion should cease checked monthly |
|
Main power switch: when is it used? |
only in extreme circumstances, not considered a back up system |
|
primary control switch: examples |
couch controls, hand pendant |
|
what are limit switches and how often is it checked |
it prevents mechanical motion beyond the machine design -checked quarterly |
|
what does the collsion ring do and how often is it checked |
when touch, it stops machine motion, checked daily |
|
what is the centrifugal switch, and how often is it checked |
controls starting circuit of electric motor, checked quarterly |
|
when can the centrifugal switch be reversed |
only when the motor has slowed to a certain speed |
|
what prevents the radiation beam from being turned on if the door is opened or interrupts the beam when opened |
door interlocks |
|
importance of intercoms and monitors and door interlocks |
interrupt radiation beam if patient is in distress or moved |
|
area radiation monitoring devices purpose |
warning lights activated by the electral power, turns of beam |
|
importance of area radiation monitoring devices |
personnel protection: alerts therapist that the beam has shut down at either the appropriate dose of a fault has occured |
|
methodology of the radiation monitoring devices |
pass or fail do not treat and notify physics |
|
purpose of the distance indicator |
to set SSD/SAD treatments to a correct distance |
|
how often is the SSD pointer checked |
weekly |
|
light field, radiation field and colimator purpose |
to determine if the light and radiation fields are congruent and on central axis |
|
importance of the light field, radiation field, and colimator |
light field is used as a guide to the size, shape, position, and directions of the radiation beam |
|
machine dose rate purpose |
setting appropriate MUs should give a desired dose |
|
importance of machine dose rate |
to ensure the patient receives the correct prescribed dose |
|
ways to check machine dose rate |
diodes and online dosimeters |
|
beam penetration quality purpose |
to make sure the desired dose at any depth for a given exposure is dependent on beam penetration |
|
beam penetration importance |
exposure of patient to the correct machine output setting will result in delivery of prescribed dose at depth |
|
beam penetration quality tests |
yellow flagged test red flagged test notify physics immediately |
|
field symmetry define |
half of the beam profile is a mirror image of the other half |
|
field flatness define |
defined as measurement of the output variation along the central 80% of the beam profile |
|
purpose and importance of field flatness and symmetry |
ensure that symmetry and flatness of the radiation beam are within the baseline profile |
|
how to check for field symmetry and flatness |
use tracker daily |
|
mechanical and optical patient alignment devices purpose |
to properly position patient for treatment |
|
methodology for mechanical and optical patient alignment devices |
notify physics if not within specifications |
|
linear scales on treatment tables purpose |
using scales or digital readouts to indicated table height, longtitudinal, and transverse table top motion |
|
how to check linear scales on treatment tables |
test comparing isocenter to some point below isocenter using scale or digital readout |
|
collimator rotation purpose |
defines the radiation field of treatment |
|
why is it important to check the collimator rotation |
to have reproducibility of day to day setups |
|
how to check collimator rotation |
spirit level |
|
gantry rotation puspose |
angular positions be accurately read |
|
importance of gantry rotation |
to deliver proper dose and irradiate desired tumor sites |
|
how to check gantry rotation |
spirit level |
|
purpose of stability of isocenter under collimator and gantry rotation |
production of a geometrically well defined treatment field which is aimed at the same point under varying machine or couch positions |
|
importance of stability of isocenter under collimator and gantry rotation |
there will be a shift of opposed fields relative to one another when using isocentric technique |
|
treatment couch isocenter puspose |
used for specialized treatment plans |
|
importance of the treatment couch isocenter deviation |
will show early indicator of problems |
|
how to correct treatment couch isocenter |
physics will tape piece of paper to psa isocenter while rotating the couch pedastal |
|
control panel and indicator light -frequency -materials |
daily warm up Argus |
|
mechanical and electrical devices -frequency -material |
dependent on test dependent on test |
|
intercom, monitor, and door interlocks -freq -material |
daily argus |
|
radiation monitoring devices -freq -materials |
daily argus |
|
distance indicators -freq -material |
daily (+/- 2mm argus |
|
light field, radiation field, and collimator -freq -material |
daily (+/- 2mm)....monthly in community argus |
|
machine dose rate -freq -material |
-daily (+/- 2%)...3% in community -phantom, diodes, tracker |
|
beam penetration -freq -material |
daily (+/- 2%) tracker |
|
field symmetry and flatness -freq -material |
daily (+/-2%) tracker |
|
mechanical and optical pt alignment devices -freq -material |
daily (+/- 2mm at iso) argus |
|
linear scales on tx table -freq -material |
monthly (+/-2mm) metric ruler |
|
collimator rotation -freq -material |
monthly (+/- 1 degree) spirit level |
|
gantry rotation -freq -material |
monthly (+/- 2%) spirit level |
|
treatment couch isocenter -freq -material |
monthly (+/- 2mm) no special materials |
|
isocenter stability under rotation of gantry and collimator |
monthly (+/- 2mm) split field test |