Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
89 Cards in this Set
- Front
- Back
- 3rd side (hint)
When a charged particle interacts with an atom or molecule in such a way as to remove an electron completely from its orbit, this interaction is referred to as an: |
Ionization event |
|
|
Which region of the gas curve do ionization chambers operate in? |
Saturation region |
|
|
Radiation can be defined as: |
Emission of energy |
|
|
Which detector has the greatest sensitivity for detecting radiation? |
GM meter |
|
|
What are the QC measures for the dose calibrator and when should they be performed? |
Constancy - daily Linearity - quarterly Accuracy - annually Geometry - installation/maintenance |
There are 4: CLAG |
|
Detector efficiency equation |
CPM/DPM x 100 |
|
|
Geometric efficiency equation |
(1/2)(1-cos°) x 100 |
|
|
Which type of interaction is most desirable for Nuclear Medicine imaging purposes? |
Photoelectric absorption |
|
|
The percent full width at half max is often used as a measure of: |
Resolution |
|
|
Which detector provides a reading of the rate of radiation exposure? |
Ionization chamber |
|
|
Intrinsic efficiency can be defined as: |
The fraction of radiation striking the detector that interacts with it. |
|
|
In the GM meter, the size of the pulse is: |
Independent of the amount of energy deposited in the gas by the ionization event |
|
|
What is dead time? |
Time required to process an ionization event. |
|
|
True or false? We use pulse-height spectrometry to analyze background radiations, scatter radiation, and unknown. Nuclides. |
True |
|
|
There are several factors that cause spectral blurring when analyzing an "actual" spectrum. What are some? |
Counting rate Summing events Backscatter peaks Imperfect energy resolution |
|
|
A detector that produces an electrical signal for an interaction, whose amplitude reflects the energy absorbed is said to have: |
Energy resolution |
|
|
Geometric efficiency of a detector depends on two things |
Position/configuration Distance |
|
|
Ionization detectors require calibrations (blank) and are completed by the rad health Dept. |
Annually |
|
|
Due to cost and size restrictions, semiconductor detectors are not often part of the typical Nuclear Medicine clinic. In which setting would you be most likely find this type of detector? |
Intraoperative suite, cardiology clinic, Nuclear research facility, radiochemistry lab |
|
|
Which of the hint statements is not true of the dose calibrator |
Reports in mR/hr for the source |
Used to measure the quantity of radioactive Uses Argon gas as the detector media Reports in mR/hr for the source Can be used to calculate pre/post-calibration activity |
|
Energy linearity can be defined as |
The proportionality between output pulse amplitude and energy absorbed in the detector |
|
|
If an event occurs during the dead time of the preceding event the effect is ignored with no subsequent effects. This is known as a: |
Non paralyzable system |
|
|
When viewing a pulse-height spectrum for Tc-99m, we often see a weakened photopeak around 140 keV and a much larger photopeak around 511 keV, due to coincidence summing events. |
False |
Summing events only happen with isotopes with more than one useful energy |
|
What is the approximate full width at half max? |
20 keV |
|
|
What is the energy resolution percentage equation? |
(Change in E / max photopeak) x 100 |
|
|
The rate of emission per unit time is known as: |
Disintegrations per time (dp(t)) |
|
|
Describe the methods used to perform the linearity quality control measure for the dose calibrator: |
Decay method - use a source of Tc-99m with 50-100 mCi. Plot and graph the activity level each day over a course of 8 days, along with the expected activity level and compare. Should be within 10% Attenuation Method - use a source of Tc-99m with 50 mCi. Use different sizes sleeves of lead to mimic decay. Plot and graph the activity level vs expected level. Should be within 10% |
2 methods |
|
Explain the ionization avalanche process for GM meters: |
An ionization event ejects a photon that causes another ionization event and so on. |
|
|
Because the well counter has nearly 100% geometric and intrinsic efficiencies, what is important to keep in mind when preparing the sample for counting? |
Sample volume/activity low |
|
|
3 components of absorption and scatter |
Nonuniform detection efficiency Simultaneous detection emitted radiations Attenuation and scatter of radiation outside detectors |
|
|
The exponential increase in number of electrons from cathode to anode occurs in the: |
PM Tube |
|
|
A lead bar phantom is used to determine: |
Spatial resolution |
|
|
Energy corrections are applied to the image... |
As scintillation events are acquired. |
|
|
Typically, an increase in resolution results in: |
A decrease in sensitivity |
|
|
Increasing the hole diameter of septa results in: |
Increased sensitivity |
|
|
The resolution of parallel hole collimators is best at: |
The collimator surface |
|
|
As a pinhole collimator is moved farther away from the thyroid, how will it affect the image? |
The gland will appear smaller |
|
|
In quantitating scintillation camera uniformity, the central field of view is defined as the diameter of the useful field of view times: |
0.75 |
|
|
Static frame mode acquisition is the most appropriate type of image acquisition for which of the following procedures (in the hints) |
Thyroid imaging study |
First phase of a 3 phase bone scan Left ventricular ejection-fraction determination Thyroid imaging study Renal function imaging study |
|
If a 20% window is needed, what percentage should fall on each side of the centerline? |
10% |
|
|
Which of the following (hint) will increase the resolution of a "spot"-view bone image obtained with a parallel hole collimator? |
Moving the detector closer to the patient's body |
Increasing the pulse-height analyzer window width Using a high sensitivity collimator instead of a low energy all purpose collimator Using a different gray scale to display the image with sharper contrast Moving the detector closer to the patient's body |
|
Thallium is added to a sodium iodide crystal of a scintillation detector to allow the detector to: |
Enable the crystal to fluoresce at room temperature |
|
|
The best sensitivity of a low energy high resolution collimator is at which distance? |
Distance does not affect sensitivity |
|
|
For a 364 keV gamma ray, the best detection efficiency of a scintillation camera crystal is a (blank - in hints) thick crystal |
5/8 inch |
1/4 inch 3/8 inch 1/2 inch 5/8 inch |
|
Spatial linearity correction is used on scintillation cameras to: |
Shift events to their correct location and improve flood uniformity |
|
|
Spatial resolution is defined as: |
FWHM mm |
|
|
Equation to determine dimensions of each pixel? |
Diameter of FoV/matrix size |
|
|
How to determine pixel coordinate location |
Convert given location to mm, divide by pixel size. |
|
|
Which of the following (hint) matrix sizes and acquisition modes would be most appropriate for a blood flow study if the feet? |
64 x 64 byte |
64 x 64 byte 64 x 64 word 256 x 256 byte 256 x 256 word |
|
A technologist changes the collimator on a gamma camera from a LEAP collimator to a High-resolution collimator. If the same number of counts is acquired, how will the acquisition time change when the High-resolution collimator is used. |
The time will increase |
|
|
The image density for each pixel is determined by: |
The gamma ray counts stored within the pixel |
|
|
If a pixel can contain 1 byte of information, and if the counts per pixel reaches 256 counts, pixel rollover may occur and would result in: |
The pixel value returning to zero before resuming the count |
|
|
List the three corrections that take place within the scintillation camera. |
Energy Nonlinearity Nonuniformity |
|
|
List the four detection event types |
Valid Object scatter Detector scatter Septal penetration |
|
|
True or false: intrinsic resolution worsens as gamma ray energy increases |
True |
|
|
Image contrast can be defined as. |
The ability to tell the difference between different objects |
|
|
What does quantum mottle refer to |
Noise |
|
|
Over time, PM Tube voltages tend to drift due to fluctuations in their high-voltage supply. Which Quality Assurance technique is applied to help maintain their performance? |
Gamma camera tuning |
|
|
Detection efficiency (blanks) as the thickness of the crystal increases. |
Increases |
|
|
The NaI(Tl) crystal material is hygroscopic. This refers to: |
It's ability to absorb moisture from the air. |
|
|
Contrast in an emission computed tomography image determines what? |
Whether two objects have different activity concentrations |
|
|
The 360° orbit is most often used in all of the following (hints) SPECT studies, except: |
Cardiac |
Brain Cardiac Liver Bone |
|
Body contouring or elliptical orbits in SPECT imaging have an advantage over circular orbit because they: |
Reduce the camera-to-patient distance |
|
|
DICOM does what? |
Allows for image transfer no matter the vendor or modality Provides workflow management and communicates the status of procedures Specifies standard formats for information such as: images, patients, studies, etc. |
|
|
Which attenuation correction method guaranteed an attenuation map? |
Transmission |
|
|
Ramp filtered projections are characterized by: |
Edge information enhancement |
|
|
OSEM reconstruction algorithms may be preferred over GBP because: |
OSEM is much faster |
|
|
A low pass filter does what? |
Allows low frequencies to pass and suppressed high higher ones |
|
|
In the reconstruction of SPECT images, segmentation refers to: |
Separating the organ from the bkg or nearby structures. |
|
|
Scatter correction methods can be applied at any time during the reconstruction process, true or false? |
False |
|
|
When the same nonuniformity is back-projected at each of the angles of acquisition, the result is: |
A ring artifact |
|
|
The center of rotation measurement determines what? |
Determines the offset between the camera and projections. |
|
|
Simple back projection results in: |
Blurring artifact |
|
|
Partial-volume effects are caused by: |
Tissue volumes that are smaller than the pixel/voxel size |
|
|
A correction technique for partial-volume effect is to apply a: |
Recovery coefficient |
|
|
When using iterative reconstruction techniques little post reconstructive filtering is required, true or false? |
True |
|
|
Spatial resolution, as a performance characteristic, is different in SPECT imaging than planar imaging in that: |
Both in-plane and axial resolution is reported |
|
|
PACS is utilized in imaging facilities to: |
Store images, view from any workstation, and transfer images among networks |
|
|
Without attenuation correction, myocardial perfusion studies in females may exhibit an artifact in the (blank) wall |
Anterior |
|
|
In a body area with uniform tissue attenuation which mathematical only attenuation correction technique may be applied: |
Chang |
|
|
For a 64 x 64 matrix, the COR value should be less than: |
0.5 pixel |
Matrix is inversely proportional to pixel size |
|
Noise in frequency space is represented by: |
The highest frequencies |
|
|
Windowing in image processing, refers to the ability to |
Change gray or color scale to improve contrast |
|
|
The most common type of volume rendering is done with: |
Maximum intensity projection (MIP) |
|
|
In referring to attenuation correction, the probability of absorption increases as photon energy decreases. True or false |
True |
|
|
Scatter correction typically employs (blank) to correct errors due to Compton scatter. |
Dual energy window |
|
|
When metal artifacts are noted, it is best to perform reconstruction without using attenuation correction techniques..true or false |
True |
|
|
The sinogram is a: |
Common way to represent full projection data in the form of a 2-D matrix |
|
|
How to calculate ejection fraction: |
Normalize: (bkg cts x #object pixel)/bkg pixel N/ED - ED = ED' N/ES - ES = ES' ED' - ES' / ED'
|
|