Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
222 Cards in this Set
- Front
- Back
What is the hydrostatic pressure within the arterial circulation at heart level when the patient is standing? |
0 mmHg |
|
What is the hydrostatic pressure within the arterial circulation at head level when the patient is standing? |
- 30 mmHg |
|
What is the hydrostatic pressure within the arterial circulation at waist level when the patient is standing?
|
50 mmHg |
|
What is the hydrostatic pressure within the arterial circulation at knee level when the patient is standing?
|
75 mmHg |
|
What is the hydrostatic pressure within the arterial circulation at ankle level when the patient is standing?
|
100 mmHg |
|
True/False: When standing, hydrostatic pressure depends upon whether the measurement is made above or below the heart level. |
True |
|
What is the arterial pressure measured in the toe of an upright patient with an arterial blood pressure of 140 mmHg (circulatory pressure)? |
240 mmHg circulatory + hydrostatic = measured pressure *140 mmHg + 100 mmHg = 240 mmHg |
|
What is the arterial pressure measured at the knee of an upright patient with an arterial blood pressure of 140 mmHg (circulatory pressure)?
|
215 mmHg circulatory + hydrostatic = measured pressure *140 mmHg + 75 mmHg = 215 mmHg |
|
What is the arterial pressure measured at the waist of an upright patient with an arterial blood pressure of 140 mmHg (circulatory pressure)?
|
190 mmHg circulatory + hydrostatic = measured pressure *140 mmHg + 50 mmHg = 190 mmHg |
|
What is the arterial pressure measured at the upper arm of an upright patient with an arterial blood pressure of 140 mmHg (circulatory pressure)?
|
140 mmHg circulatory + hydrostatic = measured pressure *140mmHg + 0 mmHg = 140 mmHg |
|
What is the arterial pressure measured at the head of an upright patient with an arterial blood pressure of 140 mmHg (circulatory pressure)? |
110 mmHg
circulatory + hydrostatic = measured pressure *-30 mmHg + 140 mmHg = 110 mmHg |
|
Normally, what happens to venous flow in the legs during inspiration? a) increase b) decrease c) no change |
b) decreases |
|
All of the following occur during expiration except? a) venous return to the heart decreases b) venous flow in the legs increases c) abdominal pressure increases d) the diaphragm rises into the thoracic cavity |
c) abdominal pressure increases *during expiration, abdominal pressure decreases |
|
What is the hydrostatic pressure at all locations within the arterial circulation of a patient in supine position? |
0 mmHg
*all arteries are at the same level as the heart in supine position, therefore, the body is 0 mmHg |
|
What is the arterial pressure at the knee of a supine patient with an arterial blood pressure of 140 mmHg? |
140 mmHg *since the knee is at heart level in supine position, the hydrostatic pressure is 0 mmHg circulatory + hydrostatic = measured pressure *140 mmHg + 0 mmHg = 140 mmHg |
|
What is the arterial pressure in the carotid artery of the neck of a supine patient with an arterial blood pressure of 140 mmHg?
|
140 mmHg *since the neck is at the same level as the heart in supine position, the hydrostatic pressure is 0 mmHg circulatory + hydrostatic = measured pressure *140 mmHg + 0 mmHg = 140 mmHg |
|
True/False: Flow is also called volume flow rate. |
True |
|
__________ indicates the volume of blood moving during a particular time. |
Flow (L/min) |
|
__________ measurements answer the question: "how much?" |
Flow (L/min) |
|
_____________ indicates the speed or swiftness of a fluid moving from one location to another. |
Velocity (cm/s) |
|
______________ answers the question: "how fast?" |
Velocity (cm/s) |
|
_______________ flow occurs when blood moves with a variable velocity and blood accelerates and decelerates as a result of cardiac contraction. |
Pulsatile flow *arterial contraction |
|
True/False: Blood accelerates and decelerates as a result of cardiac contraction and respiration. |
True |
|
True/False: Pulsatile flow commonly appears in the venous circulation. |
False *arterial |
|
____________ flow occurs when blood moves with a variable velocity and blood accelerates and decelerates as a result of respiration. |
Phasic flow *venous circulation |
|
True/False: Blood accelerates and decelerates as a result of respiration and cardiac contraction. |
True |
|
True/False: Phasic flow often appears in the venous circulation. |
True |
|
__________ flow occurs when a fluid moves at a constant speed or velocity. |
Steady flow |
|
___________ flow is present in the venous circulation when individuals stop breathing for a brief moment. |
Steady flow |
|
______________ flow is when the flow streamlines are aligned and parallel. |
Laminar flow |
|
______________ flow is characterized by layers of blood that travel at individual speeds. |
Laminar flow |
|
True/False: Laminar flow patterns are commonly found in normal physiologic states. |
True |
|
True/False: Laminar flow patterns are plug flow or parabolic flow. |
True |
|
______________ flow is characterized by chaotic flow patterns in many different directions and at many speeds. |
Turbulent flow |
|
In ______________ flow, the streamlines are often obliterated. |
Turbulent flow |
|
True/False: Turbulent flow profiles are chaotic. |
True |
|
In a stenotic vessel, where is the highest pressure located? a) before the stenotic area b) in the stenotic area c) after the stenotic area |
a) before the stenotic area |
|
In a stenotic vessel, where is the lowest pressure located?
a) before the stenotic area b) in the stenotic area c) after the stenotic area |
b) in the stenotic area |
|
In a stenotic vessel, where is the maximum velocity located?
a) before the stenotic area b) in the stenotic area c) after the stenotic area |
b) in the stenotic area |
|
In a stenotic vessel, where is the most turbulent flow located?
a) before the stenotic area b) in the stenotic area c) after the stenotic area |
c) after the stenotic area (downstream) |
|
True/False: In the circulatory system, the resistance vessels are called arterioles. |
True |
|
Pressure related to the weight of blood pressing on a vessel measured at a height above or below heart level, is called? |
Hydrostatic pressure |
|
True/False: When supine, hydrostatic pressure is zero everywhere since all vessels are at the same level as the heart. |
True |
|
During _______________, the diaphragm descends. Venous return to the heart increases. Venous flow in the legs decreases. |
inspiration |
|
During __________________, the diaphragm ascends. Venous return to the heart decreases. Venous flow in the legs increases. |
expiration |
|
During __________________, the diaphragm moves downward toward the abdomen. |
inspiration |
|
During _____________, thoracic pressure decreases and abdominal pressure increases. |
inspiration |
|
During _______________, venous return to the heart increases and venous flow to the legs decreases. |
inspiration |
|
During _______________, the diaphragm moves upward into the thorax. |
expiration |
|
During _______________, the thoracic pressure increases and the abdominal pressure decreases. |
expiration |
|
During _____________, venous return to the heart decreases and venous flow to the legs increases. |
expiration |
|
True/False: Venous flow in the legs correlates with the movement of the diaphragm. |
True |
|
True/False: Downward movement of the diaphragm (inspiration) decreases venous flow in the legs. |
True |
|
True/False: Upward movement of the diaphragm (expiration) increases venous flow in the legs. |
True |
|
True/False: When venous flow in the leg decreases, venous return to the heart decreases. |
False *venous flow in legs decrease while venous return to the heart increases |
|
True/False: When venous flow in the leg increases, venous return decreases. |
True |
|
True/False: The Doppler principle is used to measure red blood cell velocities. |
True |
|
True/False: Blood cells moving toward the transducer reflect sound with a higher frequency. |
True |
|
True/False: Blood cells moving away from the transducer reflect sound with a higher frequency. |
False *lower frequency |
|
What is the Doppler Equation? |
Doppler shift = 2(blood velocity)(transducer frequency)(cos theta)/propagation speed |
|
True/False:
The Doppler shift is directly related to velocity. |
True *frequency |
|
True/False: When the blood cell velocity doubles, the Doppler frequency doubles. |
True |
|
True/False: The y-axis of a Doppler Spectrum represents Doppler shift (kHz) or velocity. |
True |
|
True/False: The Doppler Shift is directly proportional to the transducer frequency. |
True |
|
True/False: When the transmitted frequency doubles, the Doppler frequency doubles. |
True |
|
True/False: Calculated velocities from measured Doppler Shifts will be identical, regardless of the transducer frequency. |
True |
|
True/False: Doppler Shift is indirectly related to velocity and transducer frequency. |
False *Doppler Shift is directly related to velocity and transducer frequency |
|
The ______________ _________ represents 100% of the true velocity when blood flow is parallel to the sound beam. |
Doppler Shift |
|
True/False: When blood flow is not parallel to the sound beam, meaning there is an angle, Doppler will measure something less than the true velocity. |
True |
|
True/False: Doppler Shift is directly related to cosine theta. |
True |
|
True/False: When the cosine doubles, the Doppler frequency doubles. |
True |
|
True/False: The cosine increases as angles approach zero degrees. |
True |
|
What is the cosine of zero degrees? |
1.0 |
|
What is the cosine of 90 degrees? |
0 |
|
What is the cosine of 180 degrees? |
- 1.0 |
|
True/False: At angles other than zero degrees and 180 degrees, only a portion of the true velocity is measured. |
True |
|
True/False: Blood cells in adjacent arteries and veins typically flow in opposite directions. |
True |
|
True/False: Continuous wave Doppler uses one transducer crystal. |
False *CW uses two crystals. One that transmits. One that receives. |
|
True/False: For Continuous Wave Doppler, all the velocities within the overlap region are measured. |
True |
|
With ______________ wave Doppler, velocities are measured along the entire sound beam. |
Continuous wave Doppler |
|
True/False: A dedicated CW Doppler probe (pedoff) is small and has only two elements. Anatomic imaging cannot be performed. |
True |
|
True/False: Dedicated CW transducers (pedoff) have increased sensitivity that can detect low amplitude reflections and small Doppler shifts. |
True |
|
______________ wave Doppler measures velocity in a small region, the sample volume. |
Pulsed wave Doppler |
|
For ____________ wave Doppler spectrum, velocities are measured from the sample volume. |
Pulsed wave Doppler |
|
True/False: Aliasing artifact appears only with Continous wave Doppler. |
False *aliasing only occurs with Pulsed wave Doppler |
|
The __________ __________ is the highest Doppler frequency or velocity that can be measured without the appearance of aliasing. |
Nyquist Limit |
|
Aliasing appears when the Doppler shift exceeds the ______________ _____________. |
Nyquist limit |
|
Aliasing appears when the _____________ __________ exceeds the Nyquist limit. |
Doppler shift |
|
Aliasing is eliminated by increasing or decreasing the Doppler shift or ________________ _____________. |
Nyquist limit |
|
When there is less aliasing, blood velocity is _____________, the transducer frequency is ____________, and there is a _______ PRF (shallow gate). |
slower blood velocity lower frequency transducer high PRF |
|
When there is more aliasing, blood velocity is ___________, the transducer frequency is ____________, and there is a ______ PRF (deep gate). |
faster blood velocity higher frequency transducer low PRF |
|
A Doppler spectrum with aliasing has a ________ Nyquist limit because the sample volume is __________. |
low Nyquist limit deep sample volume |
|
A Doppler spectrum without aliasing has a ________ Nyquist limit because the sample volume is __________.
|
high Nyquist limit shallow sample volume |
|
True/False: The Doppler shift is directly related to the transducer frequency. |
True |
|
True/False: Lower frequency transducers create lower Doppler shifts that are less likely to exceed the Nyquist limit and alias. |
True |
|
True/False: Eliminating aliasing improves the ability to measure the maximum velocity with Doppler. |
True |
|
True/False: Pulsed wave Doppler exhibits aliasing with high-velocity flow. |
True |
|
True/False: Pulsed wave Doppler never aliases. |
False *CW Doppler never aliases |
|
True/False: By adjusting the scale or creating a shallower view, the Nyquist limit will increase eliminating aliasing. |
True |
|
True/False: Lowering the transducer frequency will decrease the Doppler shift eliminating aliasing. |
True |
|
True/False: By zeroing the baseline shift, aliasing remains but display becomes more appealing. |
True |
|
______________ wave Doppler never aliases, but has range ambiguity. |
Continuous wave Doppler |
|
The _______ _________ of a Doppler spectrum is related to the number of blood cells. |
gray scale |
|
______________ wave Doppler has region overlap and unlimited maximum velocity. |
Continuous wave Doppler |
|
_______________ wave Doppler has range resolution and aliasing. |
Pulsed wave Doppler |
|
_______________ wave Doppler uses a sample volume and has a limited maximum velocity (nyquist limit). |
Pulsed wave Doppler |
|
A _____________ Doppler transducer has at least one crystal and lower sensitivity. |
Pulsed Doppler transducer |
|
A _____________ Doppler transducer has a dampened PZT, low Q-factor, and a wide bandwidth. |
Pulsed Doppler transducer |
|
A ______________ wave Doppler transducer has at least two crystals and higher sensitivity. |
Continuous wave Doppler transducer |
|
A _______________ wave Doppler transducer has an undampened PZT, high Q-factor, and a narrow bandwidth. |
Continuous wave Doppler transducer |
|
______________ wave Doppler has a normal incidence (90 degrees) and higher frequency which improves resolution. |
Pulsed wave Doppler |
|
_______________ wave Doppler is either 0 degrees or 180 degrees and has lower frequency which avoids aliasing. |
Continuous wave Doppler |
|
______________ Doppler measures mean velocity. |
Color Doppler |
|
_______________ Doppler measures peak velocity. |
Spectral Doppler (PW + CW) |
|
True/False: In regards to the Color Map, the color on the top of the scale is towards the transducer, and the color on the bottom is away from the transducer. |
True *normally red is towards, blue is away |
|
True/False: With variance mode of a Color Map scale, the colors change side to side, as well as, up and down. |
True *hint (side to side): L is for left side, left is for laminar flow. Right is turbulent flow. |
|
What does the black area in the center of the Color Map represent? |
zero Doppler shift |
|
Refer to images on pages: |
330 + 331 regarding Color Map |
|
True/False: Color power Doppler is non-directional. |
True |
|
________ ___________ eliminates low frequency Doppler signals near the spectral baseline. |
Wall filter |
|
True/False: Color 'bleeds' into moving anatomy with a high wall filter setting. |
False *low wall filter setting |
|
True/False: With high wall filters, low-velocity reflectors, such as tissue, are without color. |
True |
|
________________ is a special form of 'mirror image' artifact that appears on a spectral display. |
Crosstalk |
|
______________ flow has a well defined, clear window in spectral Doppler. |
Laminar flow |
|
In spectral broadening, the fill-in of the windows, is associated with ______________ flow. |
turbulent flow |
|
_______________ wave Doppler identifies the highest velocity jets anywhere along the length of the ultrasound beam. |
Continuous wave Doppler |
|
_______________ wave Doppler has range ambiguuity, higher sensitivity, and has good temporal resolution. |
Continuous wave Doppler |
|
_______________ wave Doppler accurately identifies the location of flow, has range resolution, and is subject to aliasing. |
Pulsed wave Doppler |
|
Which Doppler mode(s) measure peak velocity? |
PW and CW Doppler |
|
Which Doppler mode provides two-dimensional flow information directly on anatomic images? |
Color-flow Doppler |
|
Which Doppler mode is used with low velocity or small volume blood flow? |
Power mode Doppler |
|
__________-_________ Doppler has reduced temporal resolution due to multiple packets, is based on pulsed ultrasound, and is subject to aliasing. |
Color-flow Doppler |
|
Which Doppler mode measures mean velocity measurements? |
Color Doppler |
|
True/False: Power Mode Doppler has greater sensitivity than Color-flow Doppler and the lowest temporal resolution. |
True |
|
__________-__________ Doppler is subject to flash artifact, not to aliasing. |
Power-mode Doppler |
|
True/False:
Power-mode Doppler does not acquire velocity measurements . |
True |
|
The Doppler effect is presented as a ____________ shift when the source and the receiver are in _______________. |
frequency shift in motion relative to each other |
|
Starting from the same point, the sound source is moving toward the east at 12 mph and the receiver is moving toward the west at 10 mph. The Doppler shift is _________________. (+/-) |
negative *because the source and receiver are moving farther apart |
|
Starting from the same point, the receiver is moving toward the west at 12 mph and the source is moving toward the west at 10 mph. The Doppler shift is _______________. (+/-) |
negative *because the source and receiver are moving farther apart |
|
Doppler shift produces information about _______________. |
velocity |
|
At what angle between the sound beam and the direction of motion will the Doppler shift be the highest? |
0 degrees or 180 degrees *beam and direction should be parallel |
|
At what angle between the sound beam and the direction of motion will the Doppler shift be closest to zero? |
90 degrees *no Doppler frequency exists because the Cosine of 90 degrees is zero |
|
What is the difference between speed and velocity? |
direction *speed has a magnitude. velocity has a magnitude and direction |
|
What is the current method of processing Doppler signals? |
Fast Fourier Transform FFT *method of spectral analysis used for PW and CW |
|
_____________________ is the method of spectral analysis for Color-flow Doppler. |
Autocorrelation |
|
What is the typical range of Doppler shift found in diagnostic imaging examinations? |
between 20 Hz and 20 kHz (in audible range) |
|
The phenomenon where high velocities appear in the opposite direction is called?
|
aliasing |
|
The frequency at which aliasing occurs is called? |
Nyquist limit |
|
The area of interrogation in a pulsed Doppler exam is called? |
sample volume |
|
True/False: The higher the emitted frequency, the more likely a pulsed wave signal is to alias. |
True |
|
True/False: Shallower sample volumes result in more aliasing. |
False |
|
True/False: Only pulsed wave Doppler exams have a sample volume. |
True |
|
An 8 MHz transducer with a pulse repetition frequency of 5,000 Hz is imaging to a depth of 7 cm. What is the Nyquist frequency? a) 4 MHz b) 2.2 kHz c) 2.5 dB d) 3.5 Hz |
b) 2.2 kHz *Nyquist limit is one-half of the PRF, 2.5 kHz or 2,500 Hz |
|
An 8 MHz transducer with a PRF of 5,000 Hz measures a Doppler shift of 7 kHz. The study is repeated with a 4 MHz transducer. What Doppler shift will be measured?
a) 4 Mhz b) 3,500 Hz c) 4 kHz d) 3.5 dB |
b) 3,500 Hz
*measured Doppler shift is directly related to the transducer frequency
*if the transducer frequency is halved, the Doppler shift will be halved |
|
True/False: Doppler shift is directly related to the transducer frequency. |
True |
|
True/False: A successful Doppler exam cannot be performed on a severely anemic patient. |
False *variations in RBC concentration will not affect the ability to successfully perform a Doppler exam |
|
Which technique is used to perform spectral analysis on color Doppler data? |
autocorrelation |
|
Which technique is used to perform spectral analysis on pulsed Doppler data? |
Fast Fourier Transform FFT |
|
Which of the following terms is synonymous with spectral broadening? |
turbulent flow |
|
Which of the following is an advantage of a large packet size?
a) less aliasing b) increased frame rate c) improved lateral resolution d) increased Doppler accuracy |
d) increased Doppler accuracy
*Doppler is more accurate with larger packets |
|
Why are dedicated CW Doppler transducers so sensitive? a) absence of the matching layer b) increased electrical conductance c) absence of backing material d) inclusion of demodulator |
c) absence of backing material |
|
True/False: Continuous wave Doppler does not contain backing material. |
True |
|
True/False:
The lumen of a horizontal vessel evaluated with a color Doppler sector shaped image, contains a single color. |
False
*the vessel will have both the toward and away colors. The center of the lumen will be black. |
|
Color cannot appear when the angle between the direction of flow and sound is _________ degrees to the vessel. |
90 degrees |
|
The amount of color present on a Color-flow Doppler image can be controlled with the ___________-___________ gain setting. |
Color-flow gain |
|
______________ wave Doppler gain settings alter the gray scale content of a spectral display. |
Pulsed wave Doppler |
|
True/False: Increasing the scale of a color Doppler image can eliminate aliasing. |
True |
|
True/False: The wall filter removes the color from slowly moving reflectors such as blood cells and vibrating tissues. |
True |
|
___________ ___________ eliminate low velocity flows near the baseline of a Doppler spectrum. |
Wall filters |
|
True/False: Increasing the velocity scale eliminates aliasing and removes color from low velocities on the image. |
True |
|
Portions of an image that are brighter than surrounding tissues, or tissues that appear brighter than normal, are called? |
Hyperechoic |
|
Portions of an image that are not as bright as surrounding tissues, or tissues that appear less bright than normal, are called? |
Hypoechoic |
|
An extreme form of hypoechoic, meaning entirely without echoes (echo-free), is called? |
Anechoic |
|
_______________ describes structures with equal echo brightness. |
Isoechoic |
|
A portion of tissue or an image that has similar echo characteristics throughout, is called? |
Homogeneous |
|
A portion of tissue or an image that has differing echo characteristics throughout, is called? |
Heterogeneous |
|
Describe the Six Assumptions of the imaging system. |
|
|
True/False: Reverberations resemble a ladder or Venetian blind. |
True |
|
_________________ are created when a sound wave bounces back and forth between two strong reflectors. |
Reverberations (artifact) * The first two reflections are real, the others are artifacts |
|
True/False: Reverberations are multiple, equally spaced reflections. |
True |
|
True/False:
Comet tail artifact appears when a sound wave bounces back and forth between two very closely spaced objects. |
True *form of reverberation artifact |
|
True/False:
Shadows may provide valuable diagnostic information that helps to characterize tissue. |
True
|
|
True/False: Shadowing is entirely unrelated to the speed of sound in a medium. |
True |
|
True/False: Shadows are hypoechoic regions that extend beneath structures with abnormally high attenuation. |
True |
|
True/False: Edge shadows are created as sound beams refract and diverge along the edge of a curved structure. |
True |
|
True/False: Enhancement is a hyperechoic region that extends beneath structures with abnormally low attenuation. |
True |
|
True/False: Enhancement artifact is the opposite of shadowing. |
True |
|
True/False: Enhancement artifact is related to the speed at which sound travels in a medium. |
False *unrelated |
|
True/False: Focal banding is a hyperechoic horizontal region at the depth of the focus. |
True |
|
True/False: The Mirror artifact is always deeper than the true anatomy. |
True |
|
____________ image artifact appears as a replica of a true reflector. |
Mirror image artifact |
|
True/False: Lobe artifact and refraction artifact degrades lateral resolution. |
True |
|
What are the two methods to eliminate lobes? |
subdicing apodization |
|
True/False: Side lobes and grating lobes diminish lateral resolution. |
True |
|
____________ resolution artifact creates one reflection on the image from two closely spaced reflectors. |
Axial Resolution |
|
True/False: Curved or oblique reflectors redirect sound waves in directions other than back toward the transducer. |
True |
|
True/False: High line density results in better Axial Resolution. |
False *better Spatial Resolution |
|
True/False: Low line density degrades Spatial Resolution. |
True |
|
True/False: More lines on a monitor provide better Spatial Resolution. |
True |
|
True/False: Fewer lines result in poor Axial Resolution. |
False *poor Spatial Resolution |
|
True/False: Higher pixel density results in better Axial Resolution. |
False *better Spatial Resolution |
|
True/False: Lower pixel density degrades Spatial Resolution. |
True |
|
When creating an ultrasound image, all of the following assumptions are made except? a) sound travels in a straight line b) sound travels at 1.54 km/s c) the sound beam is extremely thin d) all structures create reflections of equal magnitude |
d) all structures create reflections of equal magnitude |
|
True/False: Reflections from bone are usually strong, whereas reflections from soft tissue are normally weak. |
True |
|
Which statement regarding Axial Resolution is incorrect? a) axial resolution artifact is related to beam diameter b) higher quality images are associated with smaller numbers c) axial resolution may be reported in units of meters d) numerically, axial resolution equals one-half the pulse length e) too few reflectors appear on the image |
a) axial resolution artifact is related to beam diameter *Axial Resolution is related to the length of the pulse NOT beam diameter |
|
True/False: Axial Resolution is related to the length of the pulse. |
True |
|
An ultrasound pulse has a width of 4 mm, a length of 2 mm, and is produced by a transducer 3,000 times per second. What is the best estimate of the systems Axial Resolution? a) 4 mm b) 2 mm c) 1 mm d) 1,500 Hz |
c) 1 mm * axial resolution is equal to approximately one-half of the pulse length. |
|
Which of the following statements is true about lateral resolution? a) it is also called angular, transverse, and depth resolution b) it is reported in units of time c) it is the same at all depths d) it can be improved by focusing |
d) it can be improved by focusing *lateral resolution (mm) depends upon the width of the ultrasound pulse |
|
True/False: Lateral Resolution depends upon the width of the ultrasound pulse. |
True |
|
True/False: Focusing narrows the beam width, improving Lateral Resolution. |
True |
|
All of the following artifacts result in the placement of too many echoes on the image except? a) shadowing b) reverberation c) mirror image d) grating lobes |
a) shadowing |
|
Which of the following determines the spatial resolution of a display? a) number of pixels b) number of lines on display c) number of bits per pixel d) spatial pulse length |
b) number of lines on display *spatial resolution is determined by the number of horizontal lines on the display |
|
True/False: Spatial Resolution is determined by the number of horizontal lines on the display. |
True |
|
Which of the following determines the spatial resolution of a digital display?
a) number of pixels b) number of lines on display c) number of bits per pixel d) spatial pulse length
|
a) number of pixels
*with digital display, spatial resolution is determined by the pixel density |
|
Which artifact produces an image with an incorrect number of reflectors? a) propagation speed error b) multipath c) enhancement d) side lobes |
d) side lobes *side lobe artifact places a second echo on an image from a single structure |
|
True/False: Mirror image artifacts always appear deeper than the true anatomy. |
True |
|
Which artifact is unrelated to the dimensions of an ultrasound pulse? a) lateral resolution b) depth resolution c) slice thickness d) refraction |
d) refraction *refraction is not related to the shape of a sound pulse |
|
True/False: Refraction artifact depends upon a sound wave striking a boundary with oblique incidence when the tissue on either side of the boundary are dissimilar. |
True |
|
True/False: It is common to visualize artifacts when the dimension of the sound beam is larger than the reflector's dimension. |
True *artifacts are likely to occur when the beam is larger than the reflector in the body |
|
Two reflections, one true and one artifact, are displayed on an ultrasound image. In the body, only one anatomic structure is present. The correct reflection and the artifact are found side by side. What is the most likely cause of this artifact? a) mirror image b) grating lobe c) enhancement d) multipath |
b) grating lobe |
|
Two distinct reflections are observed on an image, but they actually arise from a single anatomic structure. The artifact is positioned deeper than the correct reflection. What is the most likely cause of this artifact? a) grating lobe b) side lobe c) refraction d) mirror image |
d) mirror image |
|
Which one of these artifacts does not result from multiple reflections? a) comet tail b) reverberations c) ring down d) enhancement |
d) enhancement |