• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/55

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

55 Cards in this Set

  • Front
  • Back

What is quality assurance

Routine, periodic evaluation of an ultrasound system to guarantee optimal image quality

Quality assurance evaluations must be performed

Periodically and routinely


The sonographer is responsible for implementing quality assurance

Four requirements for a quality assurance program

Assessment of system components


Repairs


Preventative maintenance


Record keeping

Several devices enable the sonographer to perform quality

Test under known, define conditions


Use constant instrument settings


Use a phantom with measurable characteristics


Image in an identical environment

Five goals of a quality assurance program

Guarantee proper operation of the system


Detect gradual changes


Minimize downtime


Reduce the number of non-diagnostic exams


Reduce the number of repeat scans

Several devices enabler snugger for to perform quality assurance

Tissue equivalent phantom


Doppler phantom


Beam profile/slice thickness phantom

An objective statement

Unbiased


Not affected by an individual‘s previous experience, preference, or taste


Information is factual

Hey subjective statement

Influence by an individuals experience or believes


Colored by a person‘s perspective or view of reality


Cannot be verified


Information is affected by opinion believe or Sumption

Quality assurance programs should be based on

Objective standards

Tissue equivalent phantoms have ultrasonic fig featuresSimilar to

Soft tissue


They are used to evaluate characteristics such as grayscale and tissue texture, multi focus and adjustable focus phased array transducer’s

Tessio equivalent phantoms are similar to soft tissue in the following ways

Speed of sound


Attenuation


Scattering characteristics


Echogenicity

Slice thickness phantom contains

A diffuse scattering plane that is at an angle to the incident sound Beam


Mimics soft tissue

Thicker slices diminish


When the sound beam is overly thick

Spatial resolution (image detail) and reduce the ability to visualize small, low contrast reflectors


Cystic structures may appear filled in

Where is grayscale evaluated

In tissue equivalent phantom because it contains small scatters that act like soft tissue

Nylon strings may be in bedded in the phantom to produce

Reflections at strategic locations

Flow phantoms are the devices of choice for

Evaluating Doppler systems

Modern Doppler phantoms include

Hey circulation pump which propels fluid through vessels embedded in a tissue equivalent phantom

Other doppler phantoms contain

Vibrating strings or moving belts

Doppler phantoms include

Vibrating string


Moving belt


Flow phantom

Doppler phantoms assist the characteristics of all Doppler modalities including

Pulsed, continuous wave,color, and power mode

Slice thickness determines

Elevational resolution

Slice thickness is most likely to degrade

Image quality because the imaging plane is thicker than either the beam with or the pulse length


Beam with determines lateral resolution


Pulsed length determines axial resolution

Sensitivity refers to

The ability of a system to display low-level echoes

Two forms of sensitivity or

Normal


Maximum

Normal sensitivity settings are those at which

All the pans, solid masses, and cystic structures in the test phantom are accurately displayed


Output power, TGC, and amplification are adjusted to establish normal sensitivity

Normal sensitivity settings should not very

From one routine evaluation to the next

Maximum sensitivity is evaluated with

Output power and amplification of the system set to the maximum practical levels


With the settings eight tissue equivalent phantom is image, and the depth of tissue like texture on the display is measured

Maximum visualization depthIs used to assess

Sensitivity, and should not differ from one routine evaluation to the next

Sensitivity is also assessed win

Hey sonographer adjusts system controls to change echo brightness from barely visible to full brightness

Dead zone is a region close to

The transducer where images are in accurate


It extends from the transducer to the shallowest Deb from which meaningful reflections appear

The accuracy of reflector depth positioning in a mode, M mode, B-mode, and 2D imaging is called

Depth calibration

Horizontal calibration is the systems ability to

Lace air goes in the correct position when the reflectors are perpendicular to the sun beam

Digital calipers should be checked in both

Vertical and horizontal directions

Focus is the depth at which

Intensity is the highest in the beam is the narrowest


Focal zone surrounds the focus

What resolution is excellent in the focal zone because the beam is an arrow

Lateral resolution

Systems with dynamic receive focusing should produce

Narrow reflections over a wide range of dips

Axial resolution is the

Smallest distance at which two Pins position parallel to the sound beam are displayed as two distinct echoes

Axial resolution is evaluated by scanning

Set of successively closer space pins within the Phantom

Lateral resolution is the

Minimum distance at which to side by side pins are displayed at two distinct images

Another method to evaluate letter resolution is to

Measure the width of reflections on the display that are created bypoint Targets in the phantom

Information within the dead zone is

Unreliable and may not be used in diagnostics sitting

With traditional transducers beam diameter varies with

Depth and is narrowest at the focus

Uniformity describes the systems ability to

Display similar reflectors in the phantom with echoes of equal brightness

With proper TGC identical reflectors should have

The same appearance on the monitor, regardless of their depth

Tissue equivalent phantom is used to evaluate

Die mention, texture, and fill in of cysts


This system should accurately display hollow structures as anechoic and solid structures as hyperechoic

Adjusting the systems output power and amplification should produce

Changes in the grayscale display

It is important to compare the relationship between the image on the system screen with the output of all other display devices such as

Remote viewing stations

Dead zone results from

Transducer ringing in the time it takes a system to switch from the transmit to the receive mode

The dead zone is assessed with the

The shallowest series of pins in a test object

With tissue equivalent phantom, the dead zone is the shallowest depth at

Which uniform tissue texture appears


Higher frequency transducers have thinner dead zone then lower frequency

On acoustic standoff, or gel pad, position between the transducer and the patient allows

Accurate imaging of important superficial structures


A 50 cc bag of IV fluid may also be used

An increasingly deeperDead zone may indicate

A cracked Crystal, detached backing material, or longer pulse duration

Registration accuracy is the ability of the system to

Play’s reflections in proper positions well imaging from different orientations

Range accuracy, or Describes

Vertical Deb calibration


This systems accuracy and placing reflectors are correct depths located parallel to the sounds Beam

If differences appear between the ultrasound scan and the actual pin locations in the test object, the error may be caused by

System malfunction


The speed of sound in the phantom is different than 1540 m/s