ESI |
Electrospray ionization Ionization source |
|
MALDI |
Matrix assisted laser desorption ionization Ionization source |
|
APCI |
Atmospheric Pressure Chemical Ionization Ionization source |
|
APPI |
Atmospheric Pressure Photoionization Ionization source |
|
Quadrapole Mass Analyzer |
4 rods - opposing rod pairs connected Ions oscillate based on m/z - ions with specific m/z get through |
|
Selectivity/Specificity |
Ability to accurately measure one or more analytes Minimal to no interference |
|
Recovery |
% recovery = (extract response / standard response) x 100 High recovery desirable for good sensitivity, but consistency of recovery is more important |
|
Accuracy |
Closeness to true value Deviation of mean from true value < 15% LLOQ (lower limit of quantitation) < 20% |
|
Precision |
Consistency of values Coefficient of variation < 15% LLOQ < 20% |
|
Linearity |
Ability of method to give results directly proportional to amount of analyte R^2 >= 0.999 |
|
Range (linear range) |
Range to upper and lower concentrations which can be detected with precision, accuracy, and linearity |
|
Limit of Detection |
Lowest analyte concentration differentiable from background noise |
|
Limit of Quantitation |
Minimum analyte amount producing acceptable quantitative results |
|
Robustness |
The effect that changes in operational parameters have on other properties of method Closely related to reproducibility |
|
Stability |
Stability under various conditions Freeze-thaw, short-term room temperature, long term, stock solution stability |