• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/25

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

25 Cards in this Set

  • Front
  • Back
There are 5 standard monitor alarms:
1. O2 supply pressure (on internal piping)
2. ORC (ratio controller)
3. O2 analyzer (MOST important!)
4. Breathing pressure monitor
5. Respiratory volume (TV, MV)
Describe the three tiered alarm prioritization
Alarms appear in order of occurrence--but only highest priority alarm si annunciated
What is a high priority alarm?
High priority--WARNING
Flashing red
Requires immediate attention

(Ex: prolonged apnea)
What is a medium priority alarm?
Med priority--CAUTION
Flashing Yellow
Requires a prompt response

(Ex: apnea, low TV, continuous high pressure)
What is a low priority alarm?
Low priority--ADVISORY
Constant yellow

(Ex: O2 analyzer not calibrated)
Alarms can be silenced for ___ sec, max of ____ sec
60 sec
120 sec Max
What are the three possible results displayed on the machine screen when it is first turned on?
Functional: all tests pass
Conditionally functional: Minor failure
Non-functional: major failure
The O2 supply pressure alarm is activated when O2 is < ____ psi
30 psi

(can be directly from wall or if something is occluding your o2 line)
How do you test that the ORC is working?
Set O2 and N2O at 10LPM, Dec O2 to 3LPM and watch for automatic dec in N2O.
The ORC maintains what ratio?
O2/N2O ratio of 25+- 3%
What is the most important alarm?
O2 analyzer
What are the two types of O2 analyzers?
Galvanic sensor
Paramagnetic sensor
Explain how a Galavanic sensor works?
On top of inspiratory valve, O2 reacts w/electrolyte material and creates a path for electrical current between anode and two cathodes. Current signal picked up and sent to processer for analysis
Explain how a paramagnetic sensor works?
Internal calibration

Alarms default at 20-30%
(set alarm at 30, expose to RA and then calibrate)
If you calibrate an O2 sensor to less than RA, what will the display be?
Display will be HIGHER than actual % O2
If you calibrate the O2 sensor to >21% how will the affect the display?
Display will be LOWER than actual % O2
What is the High Pressure Alarm defaulted at?
50cm H2O

(squeeze bag to make it go above 50 to check alarm)
When the patients PIP < set threshold pressure, what alarm will go off?
Threshold pressure (apnea) alarm.

Warning--PIP<set threshold
Caution--excessive PEEP ?25cm H2O
Advisory--threshold set TOO LOW

*Auto enabled when vent is turned on
How do you correctly set the threshold pressure apnea alarm?
Set 6cm H2O < PIP
When will the continouous (sustained) pressure alarm go off?
When continuous pressure > threshold for > 15 seconds
When does the sub-atmospheric pressure alarm go off?
When pressure is <-10cm H2O

Warning alarm
(Ex: pt breathing w/an obstruction)
Where is the spirometer located? What does it monitor?
Located on expiratory limb of breathing circuit
Defaults MV <3LPM (nml adult is 4-5)
Digital display of TV, BPM, MV
Exhaled TV= TV + _____
Exhaled TV= TV + FGF
When will a caution apnea alarm go off? A warning apnea alarm?
Caution--breath (>80ml) not detected for 15 seconds

Warning--breath (>80ml) not detected for 30 seconds
What three things can set off the volume monitor (spirometer) alarms?
1. Apnea (volume) alarm
2. Low Minute Volume
3. Reverse flow (>20ml thru sensor--alerts you to missing expiratory valve)