Does a medium priority [flashing yellow] require an immediate or prompt response?
Prompt
How long are the alarms silenced for when the silence button is hit?
60 to 120 seconds
What does it mean when the machine self-test reads "fuctional"?
All tests pass
What does it mean when the machine self-test reads "conditionally fuctional"?
Minor failure
What does it mean when the machine self-test reads "non-functional failure"?
Major failure
When is the O2 supply pressure alarm activated?
When O2 is < 30 PSI
What is the purpose of the ORC [oxygen ratio controller]?
Maintains O2/N2O at 25+-3% if O2>1 LPM
How is the ORC tested?
Set O2 and N2O at 10L, then turn O2 to 3LPM, watch N2O drop
What is a galvanic O2 analyzer?
O2 diffuses into sensor and reacts w/ electrolyte -> pathway for current to flow b/t anode and two cathodes -> analyzed
Where is a galvanic O2 analyzer placed?
On top of the inspiratory valve on the breathing circuit