Does a medium priority [flashing yellow] require an immediate or prompt response?
|
Prompt
|
|
How long are the alarms silenced for when the silence button is hit?
|
60 to 120 seconds
|
|
What does it mean when the machine self-test reads "fuctional"?
|
All tests pass
|
|
What does it mean when the machine self-test reads "conditionally fuctional"?
|
Minor failure
|
|
What does it mean when the machine self-test reads "non-functional failure"?
|
Major failure
|
|
When is the O2 supply pressure alarm activated?
|
When O2 is < 30 PSI
|
|
What is the purpose of the ORC [oxygen ratio controller]?
|
Maintains O2/N2O at 25+-3% if O2>1 LPM
|
|
How is the ORC tested?
|
Set O2 and N2O at 10L, then turn O2 to 3LPM, watch N2O drop
|
|
What is a galvanic O2 analyzer?
|
O2 diffuses into sensor and reacts w/ electrolyte -> pathway for current to flow b/t anode and two cathodes -> analyzed
|
|
Where is a galvanic O2 analyzer placed?
|
On top of the inspiratory valve on the breathing circuit
|