Describe the subatmospheric relief valve in the NAD machine
Opens at -0.5 cm H2O, w/ backup valve that opens at -1.8 cmH2O
Describe the subatmospheric relief valve in the Ohmeda machine
Opens at -0.25 cmH2O
How are the subatmospheric relief valves prone to fail?
Lint/obstruction
How is suction adjusted on the scavenger reservoir?
Needle valve adjusted
How often are the pressure relief valves on the scavenger system cleaned?
Q 6 months
What is an open scavenging system?
No valves, has open relief ports to provide positive and negative pressure relief
Describe the cannister of an open scavenging system
Large cannister, gas flows down tube to the bottom, where suction removed gas. Excessive gas flows up cannister, possibly out exhaust ports
Can you tell if an open scavenging system is dumping into the room?
No
What are the 5 standard monitors for the anesthesia machine?
O2 supply pressure, ORC, O2 analyzer, Breathing Pressure monitor, Respiratory Volume
What is the difference between a medium and low priority alarm?
Medium flashes yellow [caution], low [advisory] is constant yellow