Describe the subatmospheric relief valve in the NAD machine
|
Opens at -0.5 cm H2O, w/ backup valve that opens at -1.8 cmH2O
|
|
Describe the subatmospheric relief valve in the Ohmeda machine
|
Opens at -0.25 cmH2O
|
|
How are the subatmospheric relief valves prone to fail?
|
Lint/obstruction
|
|
How is suction adjusted on the scavenger reservoir?
|
Needle valve adjusted
|
|
How often are the pressure relief valves on the scavenger system cleaned?
|
Q 6 months
|
|
What is an open scavenging system?
|
No valves, has open relief ports to provide positive and negative pressure relief
|
|
Describe the cannister of an open scavenging system
|
Large cannister, gas flows down tube to the bottom, where suction removed gas. Excessive gas flows up cannister, possibly out exhaust ports
|
|
Can you tell if an open scavenging system is dumping into the room?
|
No
|
|
What are the 5 standard monitors for the anesthesia machine?
|
O2 supply pressure, ORC, O2 analyzer, Breathing Pressure monitor, Respiratory Volume
|
|
What is the difference between a medium and low priority alarm?
|
Medium flashes yellow [caution], low [advisory] is constant yellow
|