12 Cards in this Set

  • Front
  • Back
A self-inflating bag with a pressure gauge site will work only if a pressure gauge is connected to the site or if the connection site is
a. left open
b. plugged
b. plugged
A self inflating bag connected to a 100% oxygen source will deliver up to 100% oxygen
a. by itself
b. only when an oxygen reservior is attached
b. only when an oxygen reservior is attached
A self-inflating bag connected to 100% oxygen, but without an oxygen reservior attached to it, delivers only about _______% oxygen.
40%
You are testing a resuscitation bag. When you squeeze the bag, you __________ feel pressure against your hand.
a. should
b. should not
a. should
What number should the pressure gauge read in the illustration at the right when you squeeze the bag?
The pressure gauge should read 30 to 40cm H2O because the pop-off valve is releasing.
List 3 important factors that determine the peak inspiratory pressure delivered from a self-inflating bag.
1.
2.
3.
1. how hard you squeeze the bag
2. any leakthat may be present between the mask and the baby's face
3. the set-pointof the pressure-release valve.
What are the 7 basic parts of a self-inflating bag?
1. Air inlet and attachment site for oxygen reservoir
2. oxygen light
3. Patient outlet
4. Valve assembly
5.Oxygen reservoir
6. Pressure-release valve (pop-off)
7. Pressure gauge
In a self-inflating bag what is the valve assembly and where is it?
It is positioned between the bag and the patient outlet.
When the bag is squeezed the valve opens to allow oxygen/air to the patient. When the bag reinflates the valve closes to prevent the patients air from being re-breathed.
The amount of supplemental oxygen to use during PPV should be determined by _________ _________.
pulse oximetry
Concentrations of oxygen higher than is present in room air are most reliably delivered using a ________ and an ________ __________.
blender
oxygen reservoir
How do you test a self-inflating bag?
block mask with hand;
do you feel pressure?
Can you force the pressure-release valve open?
Does the pressure gauge register 30 to 40cm H2o pressure when the valve opens?
Does the bag reinflate quickly?

How do you control pressure in a self-inflating bag?
How hard you squeeze the bag

Any leak that may be present between the mask and the baby's face

The set-point of the pressure-release valve