quality assurance deals with _____
people
_________ is more tangible and obvious than QA.
Quality Control
Quality control deals with _____ and _____
instrumentation and equipment
Each new piece of radiologic equipment should be _____ _____ before it is applied clinically.
acceptance tested
T/F

The acceptance testing of new equipment should be performed by only the manufacturer representatives.
False

Should be done by someone other than the manufacturers representative
If the patients ultimate disease or condition does not agree with the radiologists diagnosis this is an issue with __________
outcome analysis
An acceptable Quality Control program consists of what three steps?
acceptance testing
routine performance monitoring
preventative maintenance
QC is principally the responsibility of the _______
medical physicist
Three tools to measure focal spot size?
pinhole camera
Star pattern
Slit camera
The standard tool for measurement of effective focal spot size is the _______
Slit camera
required x-ray beam filtration
2.5 mm Al
it is normally not possible to measure filtration directly so the _______ of the x-ray beam is measured instead.
half value layer (HVL)
Filtration should be evaluated _________, or at any time a change has occurred in the x-ray tube or tube housing.
annually
What is the permitted radiographic collimator misalignment
+/- 2 % SID
most systems today are equipped with ______ collimators.
positive beam limiting
(PBL)
__________ of a imaging system is determined by the focal spot size of the x-ray tube.
spatial resolution
An acceptable alternative to focal spot size measurement is use of a __________ to determine limiting spatial frequency.
line pair test tool
The measured KVP of a system, should be within _____ % of the indicated KVP.
+/- 10 %
Exposure timer accuracy should be within _____ % of the indicated time for exposure times greater than 10 ms
+/- 5 %
Accuracy of _____ % is acceptable for exposure times of 10 ms or less.
+/- 20%
_________ are designed to provide a constant OD regardless of tissue thickness, composition, or failure of the reciprocity law.
AEC
AEC systems are evaluated by exposing an IR through various thicknesses of ______ or ______.
aluminum or acrylic.
insertion of _____ allows one to adequately assess the functioning of the backup timer.
lead filter
If the phototimer fails, the backup timer should terminate the exposure at ______ or ______, whichever comes first.
6 s or 600 mAs
ability of a radiographic unit to produce a constant radiation output for various combinations of mA and exposure time is called _______
exposure linearity
exposure linearity is determined by a ___________________
precision radiation dosimeter
exposure linearity must be within ____ % for adjacent mA stations
+/- 10 %
sequential radiation exposures should be reproducible to within ______
+/- 5 %
exposure reproducibility is measured using a ________
precision radiation dosimeter
What two methods are available to evaluate exposure reproducibility?
1.) 3 exposures, failure is usually kVp related

2.) 10 exposures with constant technique factors
Intensifying screens should be cleaned no less then _______
every other month
viewbox illumination should be analyzed photometrically with an instrument called a _______
photometer
Film illuminator bulb intensity should be at least _______ and should not vary by more than _____ %
1500 cd/m^2

+/- 10%
What is ESE
entrance skin exposure
What's the procedure for replacing film illuminator bulbs?
replace all of them, not just one.

match to the type of bulb used in adjacent illuminators
under normal operation, the ESE rate shall not exceed _____
10 R/min
Whats the allowed ESE for interventional procedures?
20 R/min
unlimited exposure rates are permitted for recorded fluoroscopy, such as ____________
cineradiography
An entrance skin exposure of approx. ______ mR may be assumed for a cassette spot film
200
An entrance skin exposure of approx. ____ mR may be assumed for a photofluorospot
100
Fluoroscopic ABC should be evaluated _____
annually
Quality control measurements for conventional tomography are to ensure the characteristics evaluated remain ______
constant
T/F

Precise performance standards do not exist for conventional tomography
True
In conventional tomography, agreement between the indicated level and the measured level should be within _____ mm.
+/- 5mm
How is section uniformity evaluated in tomography QC?
By imaging a hole in a lead sheet

(image tracing of the hole should be uniform, no variations, no gaps, no overlaps.
incrementing from one tomographic section to the next, the section level should be accurate to within _____
+/- 2mm
Cross over racks should be cleaned _____
daily
How often should the developer temperature be checked?
daily
How often should the wash water be checked?
daily
How often should the sensitometry and densitometry be monitored?
daily
How often should the processor be scheduled for lubrication?
Weekly or Monthly
How often should the entire rack assembly and processing tanks be cleaned?
weekly
How often should the belts, pulleys, and gears be checked?
weekly
refers to routine procedures done weekly/monthly
scheduled maintenance
A planned program of parts replacement at regular intervals. Requires that a part be replaced before it fails.
preventative maintenance
The worst kind of maintenance, Failure in a system that requires immediate repair.
nonscheduled maintenance
the follow-up of patient reports to see that the diagnosis was correct
outcome analysis
measurement used to ensure adequate filtration of the x-ray beam
minimum half-value layer
Whats does CRCPD stand for?
Council of radiation control Program Directors
What does the CRCPD do?
Formulate state radiation regulations

(50 directors of state radiation control programs and staff)
Who prescribed the 10-step program?
TJC ( the joint commission)
What purpose does the 10 step program serve?
10 actions required to help ensure proper patient management
What's CQI stand for?
continuous quality improvement
the part of quality assurance that deals with the performance of imaging systems.
quality control
Name the people on the diagnostic imaging QC team?
Supervising radiologist, QC tech, medical physicist
What is the permitted variation of radiographic reproducibility?
+/- 5 %
What is used to clean intensifying screens?
soft line free cloth
cleaning solution provided by manufacturer
how often is lead apparel checked?
annually
How do we ensure kvp accuracy ?
with the use a of a dedicated kVp meter. Accurate within 1 KVP
What is the unit of luminance of a viewbox?
candela per meter squared or nit
importance of preventative maintenance
reduces nonscheduled maintenance
What tools are used for processor monitoring?
sensitometer, densitometer, thermometer
Which two measurement elements for radiographic systems share a +/- 10 % tolerance?
exposure linearity & calibration of KVP
Which measurement element for radiographic systems has the highest tolerance percentage?
focal spot size @ +/- 50%
As KVp increases Entrance Skin Exposure (mR) __________
Decreases
Some insurance carries pay for services only from facilities with an approved ______ program.
Quality Control
Collimation should be evaluated _________
semi-annually
The star pattern for focal spot size measurement is easy to use, but has limitations for focal spot sizes less than _____
0.3 mm
What does the image of a line-pair test pattern measure when using as an alternative for focal spot size.
limiting spatial frequency
outcome analysis is the __________ responsibility
radiologists
To evaluate the accuracy of KVP, most medical physicists use one of a # of devices that are based on ________ or ______
filtered ion chambers or filtered photodiodes.