• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/121

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

121 Cards in this Set

  • Front
  • Back
Increased beam filtration __ radiographic contrast
decreases
Additive pathology requires
an increase in kV
X-ray beam intensity changes with a change in SID, the amount of change in intensity is determined by the
inverse square law
A greater change is needed in kV in the __ range to make a desired change in radiographic density
higher
The rule applied to kV to alter radiographic density is called __ rule
15%
A radiograph that is 100% dark as a result of excessive mAs should be repeated after multiplying the mAs by
.5
The rule applied to mAs to alter radiographic density is called the __ rule
30%
To change the SID but maintain the same density on a radiograph, use the
density maintenance formula
If your mAs is 20 and you change from 100 to 400 speed film/screen, what would your new mAs need to be to maintain the same density?
.5 mAs (.25X20)
T/F A pt with astenic body would require the highest exposure settings
FALSE
T/F Destructive pathology alters normal tissue and makes it harder to penetrate
FALSE
Which of the following is not true about the anode heel effect
it is useful for small parts
T/F When using compensating filters, you would put the thinner part of the filter over the thinner part of the anatomy
FALSE
T/F Two identical x-ray generators with identical exposure factors can produce varying results
TRUE
The __ law states that radiographic density produced with a particular mA and time setting can be reproduced using different mA and time setting as long as the mAs remains constant
reciprocity
Name three important film processing factors that affect film density
developing time, chemical concentration, and temperature
High contrast is also known as __ contrast
short scale
Which of the following is true about subject contrast?
Refers to the anatomical part being radiographed, can be affected by pathology, is the result of differential absorption characteristics of tissue and the kV setting (all of the above)
T/F Film contrast is an inherent characteristic of film that is determined by the manufacturer
TRUE
__ is the controlling factor for radiographic contrast
kV
T/F An increase in beam filtration will cause a decrease in radiographic contrast
TRUE
T/F An increase in OID will result in an increase in radiographic contrast
TRUE
The “true” anatomical image is called
umbra
Which of the following is true when referring to penumbra?
It is geometric unsharpness and it is a loss of recorded detail (A&B only)
Which of the following does not influence recorded detail?
KV
Penumbra is most pronounced when using which of the following?
Large focal spot size, long OID, short SID (All of the above)
T/F The FSS is determined by the size of the filament used
TRUE
When an x-ray exposure is made, the electrons in the electron cloud travel to the anode at a speed of __ miles per second
93,000
T/F The faster the screen/film speed, the greater the recorded detail
FALSE
T/F The shorter the SID , the more pronounced the penumbra
TRUE
T/F The shorter the OID, the more pronounced the penumbra
FALSE
Which of the following can cause motion unsharpness
Involuntary motion, tube movement, cassette movement (All of the above)
TO measure recorded detail, the radiographer must use the
line-pair test
Which of the following is a type of distortion seen on a radiograph?
Single distortion and shape distortion (A&B)
Distortion is most influenced by
SID
IF you have an SID of 40inches, and OID of 5 inches, what is the magnification factor?
40-5=35 40/35= 1.14
Image distortion can be caused by which of the following?
Improper alignment of the x-ray tube, improper alignment of the part being radiographed, improper alignment of the film (all of the above)
Penumbra is more pronounced with
an increase in OID
All of the flowing affect magnification except
FSS
All of the following are terms used to describe recorded detail except
minification
All of the flowing describe a type of distortion, except
rectification
A term used to describe penumbra is
image blur
Cardiac motion, peristalsis, spasm tremor and some respiration are types of __ motion
involuntary
T/F Shape distortion should always be avoided
FALSE
Which of the following combinations will produce the least magnified image?
Long SID , short OID
Which of the following can help eliminate motion unsharpness?
Give the pt clear, concise instructions, immobilize the pt, and use a short exposure time (All of the above)
All of the following produce motion unsharpness, except
movement of electrons from cathode to anode during the exposure
A grid that has lead strips situated parallel to the divergent x-ray beam is called a
focused grid
The main advantage of using a grid is
improved contrast
The main advantage of using a moving grid when compared to a stationary grid is
the grid lines are blurred from the image
An alternative method of reducing the amount of scatter reaching the image receptor is called the
air gap technique
Grid ratio describes the
height of the lead strips to the interspace between them
The SID range that can be used with negligible amounts of grid cutoff is called the
focal range
Which of the following grid ratios provide the best scatter cleanup?
16:1
Which of the following grid ratios present the most significant risk of grid cutoff?
16:1
T/F A grid reduces scatter production
FALSE
Canting refers to the
angle of the grid strips
Grid factor refers to
the amount of change in exposure needed to maintain image density
An upside down linear focused grid appears as
a falloff in density only on the peripheral portions of the radiograph
A grid that has two sets of linear lead strips superimposed and running perpendicular to each other is called a __ grid
crossed
All of the following will cause grid cutoff with linear focused grids, except
angling the beam longitudinally with the long axis of the lead strips
T/F Any tilt of linear gird will result in grid cutoff
FALSE
Grids are recommended when radiographing an anatomical part measuring greater than
10cm
Grids are recommended when
high kV is used
All of the following materials can be used to grid interspacing material, except
lead
Which of the following decreases the production of scatter
thin pt
Which of the following will most effectively reduce the amount of scatter reaching the IR?
Large OID
T/F The AEC allows the radiographer to consistently reproduce images having diagnostically acceptable density
TRUE
Which of the following are benefits of an AEC unit?
Decreases pt dosage & prolongs x-ray tube life (A&B only)
The two types of AEC units are
ionization chamber, phototimer
There are usually __ ionization chambers in a unit:
3
The ionization chamber is made up of __ electrons separated by a thin layer of __ and covered by a thin layer of __
2, air, aluminum (fill in the blank)
The __ stores electricity which will discharge when enough electricity has been stored which in turn will terminate the exposure
capacitor (fill in the blank)
T/F Minimum response time is the time that is set by the radiographer when making a radiographic exposure
FALSE
Which does not affect exposure time?
Film speed
Backup timers should never allow the pt to be exposed to more than __ mAs.
600
T/F OID affects density when using AEC
FALSE
T/F Increasing kV & mA can shorten the minimum response time
FALSE
With which of the AEC units listed below des backscatter have the greatest impact on image quality?
Phototimer type
The purpose of the backup timer in an AED is to
protect pt and protect x-ray tube (A&B)
When using an AEC, each incremental density setting corresponds to a __% change in density
25
What are the 2 most important factors affecting film density when using AEC?
Pt position and detector selection
Activation of the backup timer is most often caused by
radiographer
The ionization chambers in an AED are located _______of the cassette
in front
All of the following are advantages of AEC except:
eliminates need for proper positioning
Insufficient collimation __ exposure time when using AEC
increases
Technique charts _____________.
must be developed for each radiographic room in the imaging department, can be based on variable kV and fixed kV (All of the above)
Good technique charts contain the following information:
kV, mA, exposure time
Anatomical parts must be measured at a point
coinciding with the CR placement
It takes a _% change in mAs to effect a noticeable change in radiographic density
30
Advantages of a fixed kV technique chart are
part measured accuracy is less critical than with variable kV technique, because higher kV settings are used , most part thickness can be penetrated (A&B)
The method used to develop a technique chart by relating one anatomical part to another is called
comparative method
In pediatric radiography
always use the shortest practical exposure time to minimize motion unsharpness, always use the highest practical kV to reduce pt dose, and close collimation should be used to minimize pt dose and reduce scatter fog (All of the above)
T/F mAs must be increased when decreasing the x-ray beam size because scatter radiation is reduced
True
IN cast radiography, for wet plaster cast
add 15 kV or triple mAs
When collimating the x-ray beam from a 14 X 17 inch to a 10X12 inch IR, the mAs should be increased by
40%
T/F mAs controls density whereas kV influences density.
True
mAs controls contrast
false
The base equipment needed to develop a technique chart includes all of the following except
thermometer
Name 2 types of technique charts (fill in the blank)
fixed kV, and variable kV
In a variable kV technique chart, every 1 cm increase in part thickness requires a _ kV increase
2
The smallest component of a digital image is the
pixel
The process by which a CR scanner determines the number and orientation of the radiographic views on a CR plate is called
portioned pattern recognition
The window width determines the image
contrast
The post-processing method that subtracts a mask image from the original is called
energy subtraction
All of the following are types of CR image artifacts except
kernel
The latent image in a CR plate is formed by the trapping of crystal electrons in lattice sites called
F centers
When compared to conventional radiographic film, the CR plates exhibit
wider exposure latitude
The measure of the efficiency of a CR system to convert remnant x-rays into useful image signals is called
QDE
The range of gray shades displayed by pixels is called the
dynamic range
The process of rescaling pixel values to optimize contrast is called
contrast enhancement, contrast rescaling, tone scaling (all of the above)
Fading, geometric distortion, and defocusing are problems associated with
monitors
CR can produce diagnostic images, even with overexposures of
500%
The presence of dark bands at the interfaces of structures that differ widely in density describes a/an
halo artifact
All of the following are types of archival media, except
charge-coupled device (CCD)
A typical image plate loses approximately _ of its stored energy within 8 hours after exposure
25%
The type of laser used in CR scanner is
helium-neon
The CR scanner contains a device that converts light energy to electrical energy called a
photomultiplier tube
A graph of signal values that corresponds to a spectrum of pixel values is a/an
histogram
In order to erase a CR plate, it must be exposed to high-intensity
light
Spatial resolution is limited to the
matrix size and number of pixels (A&B)