Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
121 Cards in this Set
- Front
- Back
Increased beam filtration __ radiographic contrast
|
decreases
|
|
Additive pathology requires
|
an increase in kV
|
|
X-ray beam intensity changes with a change in SID, the amount of change in intensity is determined by the
|
inverse square law
|
|
A greater change is needed in kV in the __ range to make a desired change in radiographic density
|
higher
|
|
The rule applied to kV to alter radiographic density is called __ rule
|
15%
|
|
A radiograph that is 100% dark as a result of excessive mAs should be repeated after multiplying the mAs by
|
.5
|
|
The rule applied to mAs to alter radiographic density is called the __ rule
|
30%
|
|
To change the SID but maintain the same density on a radiograph, use the
|
density maintenance formula
|
|
If your mAs is 20 and you change from 100 to 400 speed film/screen, what would your new mAs need to be to maintain the same density?
|
.5 mAs (.25X20)
|
|
T/F A pt with astenic body would require the highest exposure settings
|
FALSE
|
|
T/F Destructive pathology alters normal tissue and makes it harder to penetrate
|
FALSE
|
|
Which of the following is not true about the anode heel effect
|
it is useful for small parts
|
|
T/F When using compensating filters, you would put the thinner part of the filter over the thinner part of the anatomy
|
FALSE
|
|
T/F Two identical x-ray generators with identical exposure factors can produce varying results
|
TRUE
|
|
The __ law states that radiographic density produced with a particular mA and time setting can be reproduced using different mA and time setting as long as the mAs remains constant
|
reciprocity
|
|
Name three important film processing factors that affect film density
|
developing time, chemical concentration, and temperature
|
|
High contrast is also known as __ contrast
|
short scale
|
|
Which of the following is true about subject contrast?
|
Refers to the anatomical part being radiographed, can be affected by pathology, is the result of differential absorption characteristics of tissue and the kV setting (all of the above)
|
|
T/F Film contrast is an inherent characteristic of film that is determined by the manufacturer
|
TRUE
|
|
__ is the controlling factor for radiographic contrast
|
kV
|
|
T/F An increase in beam filtration will cause a decrease in radiographic contrast
|
TRUE
|
|
T/F An increase in OID will result in an increase in radiographic contrast
|
TRUE
|
|
The “true” anatomical image is called
|
umbra
|
|
Which of the following is true when referring to penumbra?
|
It is geometric unsharpness and it is a loss of recorded detail (A&B only)
|
|
Which of the following does not influence recorded detail?
|
KV
|
|
Penumbra is most pronounced when using which of the following?
|
Large focal spot size, long OID, short SID (All of the above)
|
|
T/F The FSS is determined by the size of the filament used
|
TRUE
|
|
When an x-ray exposure is made, the electrons in the electron cloud travel to the anode at a speed of __ miles per second
|
93,000
|
|
T/F The faster the screen/film speed, the greater the recorded detail
|
FALSE
|
|
T/F The shorter the SID , the more pronounced the penumbra
|
TRUE
|
|
T/F The shorter the OID, the more pronounced the penumbra
|
FALSE
|
|
Which of the following can cause motion unsharpness
|
Involuntary motion, tube movement, cassette movement (All of the above)
|
|
TO measure recorded detail, the radiographer must use the
|
line-pair test
|
|
Which of the following is a type of distortion seen on a radiograph?
|
Single distortion and shape distortion (A&B)
|
|
Distortion is most influenced by
|
SID
|
|
IF you have an SID of 40inches, and OID of 5 inches, what is the magnification factor?
|
40-5=35 40/35= 1.14
|
|
Image distortion can be caused by which of the following?
|
Improper alignment of the x-ray tube, improper alignment of the part being radiographed, improper alignment of the film (all of the above)
|
|
Penumbra is more pronounced with
|
an increase in OID
|
|
All of the flowing affect magnification except
|
FSS
|
|
All of the following are terms used to describe recorded detail except
|
minification
|
|
All of the flowing describe a type of distortion, except
|
rectification
|
|
A term used to describe penumbra is
|
image blur
|
|
Cardiac motion, peristalsis, spasm tremor and some respiration are types of __ motion
|
involuntary
|
|
T/F Shape distortion should always be avoided
|
FALSE
|
|
Which of the following combinations will produce the least magnified image?
|
Long SID , short OID
|
|
Which of the following can help eliminate motion unsharpness?
|
Give the pt clear, concise instructions, immobilize the pt, and use a short exposure time (All of the above)
|
|
All of the following produce motion unsharpness, except
|
movement of electrons from cathode to anode during the exposure
|
|
A grid that has lead strips situated parallel to the divergent x-ray beam is called a
|
focused grid
|
|
The main advantage of using a grid is
|
improved contrast
|
|
The main advantage of using a moving grid when compared to a stationary grid is
|
the grid lines are blurred from the image
|
|
An alternative method of reducing the amount of scatter reaching the image receptor is called the
|
air gap technique
|
|
Grid ratio describes the
|
height of the lead strips to the interspace between them
|
|
The SID range that can be used with negligible amounts of grid cutoff is called the
|
focal range
|
|
Which of the following grid ratios provide the best scatter cleanup?
|
16:1
|
|
Which of the following grid ratios present the most significant risk of grid cutoff?
|
16:1
|
|
T/F A grid reduces scatter production
|
FALSE
|
|
Canting refers to the
|
angle of the grid strips
|
|
Grid factor refers to
|
the amount of change in exposure needed to maintain image density
|
|
An upside down linear focused grid appears as
|
a falloff in density only on the peripheral portions of the radiograph
|
|
A grid that has two sets of linear lead strips superimposed and running perpendicular to each other is called a __ grid
|
crossed
|
|
All of the following will cause grid cutoff with linear focused grids, except
|
angling the beam longitudinally with the long axis of the lead strips
|
|
T/F Any tilt of linear gird will result in grid cutoff
|
FALSE
|
|
Grids are recommended when radiographing an anatomical part measuring greater than
|
10cm
|
|
Grids are recommended when
|
high kV is used
|
|
All of the following materials can be used to grid interspacing material, except
|
lead
|
|
Which of the following decreases the production of scatter
|
thin pt
|
|
Which of the following will most effectively reduce the amount of scatter reaching the IR?
|
Large OID
|
|
T/F The AEC allows the radiographer to consistently reproduce images having diagnostically acceptable density
|
TRUE
|
|
Which of the following are benefits of an AEC unit?
|
Decreases pt dosage & prolongs x-ray tube life (A&B only)
|
|
The two types of AEC units are
|
ionization chamber, phototimer
|
|
There are usually __ ionization chambers in a unit:
|
3
|
|
The ionization chamber is made up of __ electrons separated by a thin layer of __ and covered by a thin layer of __
|
2, air, aluminum (fill in the blank)
|
|
The __ stores electricity which will discharge when enough electricity has been stored which in turn will terminate the exposure
|
capacitor (fill in the blank)
|
|
T/F Minimum response time is the time that is set by the radiographer when making a radiographic exposure
|
FALSE
|
|
Which does not affect exposure time?
|
Film speed
|
|
Backup timers should never allow the pt to be exposed to more than __ mAs.
|
600
|
|
T/F OID affects density when using AEC
|
FALSE
|
|
T/F Increasing kV & mA can shorten the minimum response time
|
FALSE
|
|
With which of the AEC units listed below des backscatter have the greatest impact on image quality?
|
Phototimer type
|
|
The purpose of the backup timer in an AED is to
|
protect pt and protect x-ray tube (A&B)
|
|
When using an AEC, each incremental density setting corresponds to a __% change in density
|
25
|
|
What are the 2 most important factors affecting film density when using AEC?
|
Pt position and detector selection
|
|
Activation of the backup timer is most often caused by
|
radiographer
|
|
The ionization chambers in an AED are located _______of the cassette
|
in front
|
|
All of the following are advantages of AEC except:
|
eliminates need for proper positioning
|
|
Insufficient collimation __ exposure time when using AEC
|
increases
|
|
Technique charts _____________.
|
must be developed for each radiographic room in the imaging department, can be based on variable kV and fixed kV (All of the above)
|
|
Good technique charts contain the following information:
|
kV, mA, exposure time
|
|
Anatomical parts must be measured at a point
|
coinciding with the CR placement
|
|
It takes a _% change in mAs to effect a noticeable change in radiographic density
|
30
|
|
Advantages of a fixed kV technique chart are
|
part measured accuracy is less critical than with variable kV technique, because higher kV settings are used , most part thickness can be penetrated (A&B)
|
|
The method used to develop a technique chart by relating one anatomical part to another is called
|
comparative method
|
|
In pediatric radiography
|
always use the shortest practical exposure time to minimize motion unsharpness, always use the highest practical kV to reduce pt dose, and close collimation should be used to minimize pt dose and reduce scatter fog (All of the above)
|
|
T/F mAs must be increased when decreasing the x-ray beam size because scatter radiation is reduced
|
True
|
|
IN cast radiography, for wet plaster cast
|
add 15 kV or triple mAs
|
|
When collimating the x-ray beam from a 14 X 17 inch to a 10X12 inch IR, the mAs should be increased by
|
40%
|
|
T/F mAs controls density whereas kV influences density.
|
True
|
|
mAs controls contrast
|
false
|
|
The base equipment needed to develop a technique chart includes all of the following except
|
thermometer
|
|
Name 2 types of technique charts (fill in the blank)
|
fixed kV, and variable kV
|
|
In a variable kV technique chart, every 1 cm increase in part thickness requires a _ kV increase
|
2
|
|
The smallest component of a digital image is the
|
pixel
|
|
The process by which a CR scanner determines the number and orientation of the radiographic views on a CR plate is called
|
portioned pattern recognition
|
|
The window width determines the image
|
contrast
|
|
The post-processing method that subtracts a mask image from the original is called
|
energy subtraction
|
|
All of the following are types of CR image artifacts except
|
kernel
|
|
The latent image in a CR plate is formed by the trapping of crystal electrons in lattice sites called
|
F centers
|
|
When compared to conventional radiographic film, the CR plates exhibit
|
wider exposure latitude
|
|
The measure of the efficiency of a CR system to convert remnant x-rays into useful image signals is called
|
QDE
|
|
The range of gray shades displayed by pixels is called the
|
dynamic range
|
|
The process of rescaling pixel values to optimize contrast is called
|
contrast enhancement, contrast rescaling, tone scaling (all of the above)
|
|
Fading, geometric distortion, and defocusing are problems associated with
|
monitors
|
|
CR can produce diagnostic images, even with overexposures of
|
500%
|
|
The presence of dark bands at the interfaces of structures that differ widely in density describes a/an
|
halo artifact
|
|
All of the following are types of archival media, except
|
charge-coupled device (CCD)
|
|
A typical image plate loses approximately _ of its stored energy within 8 hours after exposure
|
25%
|
|
The type of laser used in CR scanner is
|
helium-neon
|
|
The CR scanner contains a device that converts light energy to electrical energy called a
|
photomultiplier tube
|
|
A graph of signal values that corresponds to a spectrum of pixel values is a/an
|
histogram
|
|
In order to erase a CR plate, it must be exposed to high-intensity
|
light
|
|
Spatial resolution is limited to the
|
matrix size and number of pixels (A&B)
|