Capiox RX25 and SX25 uses microporous polypropylene hollow fibers
true
Capiox RX25 and SX25 can be used in both closed and open system config
true
Capiox RX25 “rated” time
6 hours
Capiox RX 25 comes in left or right outlet port config
true
Maximum gas flow for the Capiox RX 25?
20 L/min
Maximum water pressure at the heat exchanger inlet should not exceed Capiox RX25?
42psi
Maximum blood flow into the cardiotomy filter should not exceed Capiox RX25?
Minimum operating volume in Capiox RX25?
200ml
Maximum operatin volume in Capiox RX25?
4L
Minimum reservoir level?
20ml
Capiox SX25 priming volume?
340ml
Capiox RX25 priming volume?
250ml
Capiox RX25 pore size?
47um
Capiox RX25 cariotomy filter removes particles larger than?
20um
H2O temp in Capiox RX25 should not exceed?
42 C
Capiox RX25 temp difference between blood and H2O?
15 C
Capiox RX25 minimum gas flow when blood is circulated?
0.5 L/min
Capiox RX25 SA?
2.5m^2
Cincinnati sub zero gallons per minute/hr?
4 gallons/min 250 gallons/hr
Cincinnati sub zero, ice?
No
Capiox 125 AF retrograde flow before antegrade?
True
Capiox 125 priming volume?
125mls
Capiox 125 AF pressure drop?
36 mmHg
Capiox 125 AF pore size?
37um
Capiox AF125X surface coating?
X-coating, no heparin based
Capiox AF125 blood flow range?
0.5-7 L/min
Maximum rated pressure at the blood inlet of the AF125X?
760 mmHg
Recommended CO2 flush time for Capiox 125X?
5 minutes
Sorin maintenance how often?
Every 12 months
Sorin SIII contains a UPS device?
True
RED tubing inserts cannot fit what size tubing?
1/2 “
RED tubing inserts can fit what tubing size?
¼” and 3/8”
Maximum rated pressure at the blood inlet to the Capiox RX25?
1000 mmHg
Maximum negative pressure that can be applied to Capiox RX25?
-150 mmHg
Capiox RX25 blood flow should not exceed?
7 L/min
MYOtherm CPG ratio
4:1
MYOtherm CPG blood line
1/4
MYOtherm CPG line
1/8
MYOtherm CPG recommended flow rate
Up to 500 ml/min
MYOtherm CPG filter screen
150 microns
MYOtherm CPG heat exchange volume
44 mls
MYOtherm CPG heat exchange SA
0.06 m^2
MYOtherm CPG H2O pressure limit
40 psi
MYOtherm CPG Blood phase pressure
750 mmHg
MYOtherm CPG max water temp
42C
MYOtherm CPG incline position?
10 degrees
MYOtherm CPG time to recirculate before delievering CPG
1-2 minutes