Capiox RX25 and SX25 uses microporous polypropylene hollow fibers
|
true
|
|
Capiox RX25 and SX25 can be used in both closed and open system config
|
true
|
|
Capiox RX25 “rated” time
|
6 hours
|
|
Capiox RX 25 comes in left or right outlet port config
|
true
|
|
Maximum gas flow for the Capiox RX 25?
|
20 L/min
|
|
Maximum water pressure at the heat exchanger inlet should not exceed Capiox RX25?
|
42psi
|
|
Maximum blood flow into the cardiotomy filter should not exceed Capiox RX25?
|
|
|
Minimum operating volume in Capiox RX25?
|
200ml
|
|
Maximum operatin volume in Capiox RX25?
|
4L
|
|
Minimum reservoir level?
|
20ml
|
|
Capiox SX25 priming volume?
|
340ml
|
|
Capiox RX25 priming volume?
|
250ml
|
|
Capiox RX25 pore size?
|
47um
|
|
Capiox RX25 cariotomy filter removes particles larger than?
|
20um
|
|
H2O temp in Capiox RX25 should not exceed?
|
42 C
|
|
Capiox RX25 temp difference between blood and H2O?
|
15 C
|
|
Capiox RX25 minimum gas flow when blood is circulated?
|
0.5 L/min
|
|
Capiox RX25 SA?
|
2.5m^2
|
|
Cincinnati sub zero gallons per minute/hr?
|
4 gallons/min 250 gallons/hr
|
|
Cincinnati sub zero, ice?
|
No
|
|
Capiox 125 AF retrograde flow before antegrade?
|
True
|
|
Capiox 125 priming volume?
|
125mls
|
|
Capiox 125 AF pressure drop?
|
36 mmHg
|
|
Capiox 125 AF pore size?
|
37um
|
|
Capiox AF125X surface coating?
|
X-coating, no heparin based
|
|
Capiox AF125 blood flow range?
|
0.5-7 L/min
|
|
Maximum rated pressure at the blood inlet of the AF125X?
|
760 mmHg
|
|
Recommended CO2 flush time for Capiox 125X?
|
5 minutes
|
|
Sorin maintenance how often?
|
Every 12 months
|
|
Sorin SIII contains a UPS device?
|
True
|
|
RED tubing inserts cannot fit what size tubing?
|
1/2 “
|
|
RED tubing inserts can fit what tubing size?
|
¼” and 3/8”
|
|
Maximum rated pressure at the blood inlet to the Capiox RX25?
|
1000 mmHg
|
|
Maximum negative pressure that can be applied to Capiox RX25?
|
-150 mmHg
|
|
Capiox RX25 blood flow should not exceed?
|
7 L/min
|
|
MYOtherm CPG ratio
|
4:1
|
|
MYOtherm CPG blood line
|
1/4
|
|
MYOtherm CPG line
|
1/8
|
|
MYOtherm CPG recommended flow rate
|
Up to 500 ml/min
|
|
MYOtherm CPG filter screen
|
150 microns
|
|
MYOtherm CPG heat exchange volume
|
44 mls
|
|
MYOtherm CPG heat exchange SA
|
0.06 m^2
|
|
MYOtherm CPG H2O pressure limit
|
40 psi
|
|
MYOtherm CPG Blood phase pressure
|
750 mmHg
|
|
MYOtherm CPG max water temp
|
42C
|
|
MYOtherm CPG incline position?
|
10 degrees
|
|
MYOtherm CPG time to recirculate before delievering CPG
|
1-2 minutes
|