Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
10 Cards in this Set
- Front
- Back
GAQ Battery
|
3 AA alkaline batteries
14hrs operating time |
|
GAQ Weight
|
12oz
|
|
GAQ Operating Temp
|
-4 to 122 degrees
|
|
Hydrogen Sulfide
|
Range: 0-200ppm
LOW: 10ppm HIGH: 15ppm |
|
Carbon Monoxide (GAQ)
|
Range: 0-1000ppm
LOW: 35ppm HIGH: 200ppm |
|
O2 (GAQ)
|
Range: 0-30%
LOW: 19.5% HIGH: 23.5% |
|
LEL (GAQ)
|
Range: 0-100% or 0-5% (Methane)
LOW: 10% HIGH: 20% |
|
GAQ Alarms
|
Visual Alarm: Red lights
IntelliFlash: Green lights Audible: 95dB Internal vibrator |
|
GAQ Maintenance
|
Wipe down w/ damp cloth
Check batteries Check calibration date |
|
GAQ Low Battery
|
Visual alarm
Audible alarm Battery Icon on screen 15min remaining |