Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
154 Cards in this Set
- Front
- Back
Minimum lead equivalency needed for fluoroscopy unit protective curtain
|
0.25mm
|
|
Minimum lead equivalency needed for fluoroscopy unit bucky slot cover
|
0.25mm
|
|
Max exposure rate of a unit without automatic exposure controls
|
5 R/min
|
|
Total filtration permanently required in the useful fluoroscopic beam should not be less than
|
2.5mm A1
|
|
Max cumulative time for continuous exposure is:
|
5 min
|
|
Minimum lead equivalent needed for gonadal shielding
|
0.5mm
|
|
This unit is directly proportional to the exposure delivered to the patient
|
mAs
|
|
Unit for measurement of patient's absorbed dose
|
Rad
|
|
Source-skin-distance minimum
|
12 inches
|
|
If X-ray source is below the table, useful beam exposure must be measured at what distance above the table top
|
30 cm
|
|
Patient exposure rate for 5 min of fluoro time
|
10-25 rads
|
|
Fluoro units operate at this tube current
|
2-5 mA
|
|
Intensity measured at the table top at 80 kvp should not exceed how much R/min for each mA of tube current
|
2.2 R/min
|
|
Table top exposure rates for units without auto exposure control must be measured every
|
3 years
|
|
Dark adaptation for the eye takes how long
|
30 minutes
|
|
Decreased brightness gain at the image periphery is called
|
Vignetting
|
|
Vertical line rate is normally
|
525 lines
|
|
Video disc recorders record at a maximum frequency response rate of
|
30 frames per second
|
|
Output intensity is measured in which unit
|
mR or R
|
|
Target-to-tabletop distance must be greater than and should be greater than...
|
12 and 18 inches
|
|
Normal viewing distance
|
10-15 inches
|
|
Integration time (for brain to recognize an object)
|
0.2 seconds
|
|
Typical X-ray exposure for routine 2minute upper GI series
|
6 roentgens
|
|
Scatter radiation exposure at 1 foot from the pt could be as high as
|
500 mR per hour
|
|
Most scatter occurs at which direction from the primary beam
|
135 degrees
|
|
mA setting for spot filming where a spot film camera is not used
|
150 mA
|
|
Max exposure during 9 month pregnancy for occupational worker
|
500 millirems
|
|
Time for fluoro table to move from horizontal to vertical position
|
20-25 seconds
|
|
Most commonly used video disc frame rate
|
30 frame/sec
|
|
Greatest dose to the pt comes from spot filming cameras using this film size
|
105 mm
|
|
Recognition/integration time of human eye
|
0.2 sec
|
|
Which CCTV system results in least pt exposure
|
Plumicon
|
|
Total permanently fixed filtration of the useful beam must be
|
3 mm aluminum equivalent
|
|
At tube current at 80 kvp, the max intensity shld be
|
2.2 rads per minute
|
|
mA setting for fluoro is typically less than
|
Less than 5 mA
|
|
Predetermined time limit for fluoro beam
|
5 minutes
|
|
Max exposure rate for all combos of tube potential and current, for those with auto exposure control
|
10 roentgens/minute
|
|
Max exposure rate for all combos of tube potential and current, for those without auto exposure control
|
5 roentgens/ minute
|
|
Max whole body dose equivalent fo persons over 18
|
1250 millirems per calendar quarter
|
|
For exposure to pregnant women, risk of malformations is increased at doses above
|
15 rad
|
|
High radiation area could expose whole body in one hour to dose greater than
|
100 millirems
|
|
A radiation area could expose a major portion of the body in one hour to dose greater than
|
5 millirems
|
|
SI unit analogous to the rad is
|
Gray
|
|
Rad is a unit of
|
Absorbed dose (unit)
|
|
SI unit analogous to the rem
|
Sievert (unit)
|
|
Cardiac cath exam results in a bone marrow dose of approximately
|
190 millirads
|
|
5 minute fluoro patient exposure
|
10-30 roentgens
|
|
If skin exposure was 5 roentgens, dose to the fetus
|
100 millirads
|
|
Barium enema exam causes what dose to gonads
|
500 millirads
|
|
Exposure of occupational worker to 5 rems requires state notification within
|
24 hours
|
|
Exposure of 2 rems within a quarter requires state notification within
|
Within 30 days
|
|
Exposure reduction of .25 mm lead
|
97%
|
|
Maximum dose whole body dose exposure per year
|
5 rems per year
|
|
Leakage radiation at a distance of 1 meter from the target cannot exceed
|
100 milliroentgens in 1 hour
|
|
Patient positioning for gallbladder spot films
|
Rpo, upright, inhalation
|
|
Tube current and potential must be recorded how often
|
Weekly
|
|
Size of input: output phosphor diameter
|
Minification gain
|
|
Scattered radiation received is proportional to
|
Dose received by the patient
|
|
Most effective medium for static viewing in terms of reducing patient dose
|
Video disc recorder
|
|
Ratio between the vertical and horizontal lines on the TV monitor
|
Kell factor
|
|
Contrast of the image intensifying tube is measured by
|
conversion factor
|
|
maximum magnification for cinefluorography is done via which method
|
total over framing
|
|
A film badge reports in which units
|
Rems
|
|
Line pair grid is used to measure
|
Resolution - resolving power
|
|
Standard CINE skin exposure in one minute
|
1-5 R/min
|
|
Max tech exposure from barium enema at distance of one foot from the table (per hour)
|
400-500 mR/hr
|
|
Statistical fluctuations of xray photons cause
|
quantum mottle
|
|
This maintains a fixed dose to the input phosphor
|
the AEC (automatic exposure control)
|
|
Ratio of photons that leave the output phosphor to the number of Xray photons that strike the input phosphor
|
Flux gain
|
|
Increasing KVP beyond optimum range can result in
|
Poor subject contrast
|
|
Half value layer determines which characteristic of the Xray
|
Xray quality
|
|
Ability to distinguish separate objects in line pairs per mm
|
Resolving power
|
|
The portion of the image intensifier responsible for magnification
|
electrostatic lenses
|
|
resolution of a 9" image intensifier tube
|
3.5 lp/mm
|
|
light photons intercepting when electrons strike a fluorescent screen forms a...
|
Real image
|
|
xray interaction with contrast media produces what effect
|
photoelectric
|
|
Proper parameters for an overhead fluoroscopy tube?
|
4.5R/min with a water phantom 30cm above tabletop
|
|
Magnification with an image intensifier is achieved at which site
|
electrostatic lens site
|
|
Fluoro with a grid in place affects exposure rates how
|
1.5-2 x higher
|
|
Greatest radiosensitivity of fetus occurs when
|
2-8 weeks gestation
|
|
safest time to expose a woman of childbearing age?
|
Day 1-10
|
|
Primary protective barrier for a dedicated C-arm is what
|
Input phosphor
|
|
Formula for calculating the maximum permissible lifetime dose for a technologist
|
age x 1REM
|
|
Maximum dose rate for a patient undergoing routine fluoro
|
5 R/min
|
|
The quantity of the beam is controlled by which unit of measurement
|
miliamperage
|
|
which filming method gives the best resolution?
|
Spot film radiograph
|
|
Which fluoroscopic method gives patient dose reduction of up to to 95%?
|
video disc recorder
|
|
100erg/gram is equivalent to?
|
1 rad
|
|
Fluoro units with AECs or CINE cameras must have physicist calibration how often
|
every year
|
|
Maximum exposure level while using boost position?
|
20 R/min
|
|
Which imaging modality has the best resolution
|
Radiograph
|
|
Image intensifiers convert xray photons to...?
|
xray photons to light to electrons to light
|
|
central portion of the tube possesses the greatest
|
Resolution
|
|
Compared to the input phosphor, the image output phosphor is...
|
Smaller and brighter
|
|
Input phosphor of 2nd generation tube is made of..
|
Cessium iodide
|
|
How does moving the image intensifier away affect the patient dose
|
Increases the patient dose
|
|
As spot film size increases, what happens to the patient dose?
|
Patient dose increases
|
|
Unit used to specify the quality of an Xray beam
|
HVL
|
|
Visual acuity is defined as
|
the ability of the human eye to perceive fine detail
|
|
cycles per time unit in a TV camera is known as...
|
bandwidth
|
|
types of pick-up tubes
|
plumicon, vidicon, orthicon
|
|
TV camera element which converts light to an electric signal
|
Pickup
|
|
Output intensity is measured in...
|
mR or R
|
|
Brightness ratio of the periphery to the center of the output screen is termed:
|
Contrast
|
|
Mag mode input screen diameter change formula for exposure rate increase
|
(After/before)squared
|
|
How is resolving power measured?
|
Line pairs per milimeter
|
|
Characteristics of contrast
|
High atomic number and low toxicity
|
|
Which factors produce a brighter image at the output phosphor?
|
Flux gain and minification
|
|
What type of TV system has all its signals carried through cables?
|
Closed circuit TV
|
|
Output intensity of a tube is measured in...
|
Miliroentgens
|
|
Electronic intensification occurs when...
|
Electrons are moved at high speed from photocathode to output phosphor
|
|
genetic dose means..
|
Effects exhibited in future offspring of persons who have been irradiated
|
|
Dose effect curve that radiation protection guides are based on...
|
linear nonthreshold curve
|
|
how many rads of acute x-radiation to the eyes cause cataracts?
|
Hundreds of RADs
|
|
A disadvantage of thermoluminescent dosimeters
|
No permanent record is provided
|
|
Rem is a unit of
|
Absorbed dose equivalent
|
|
Rad is the unit of:
|
Absorbed dose
|
|
Half value layer defined
|
The thickness of a specified substance which reduces the exposure rate of the primary beam by one half
|
|
CINE exam typical skin exposure/min
|
2-5 roentgens/minute
|
|
A material placed in the useful primary X-ray beam to absorb preferentially the less penetrating radiations
|
Filter
|
|
Most common material used in thermoluminescent dosimeters
|
lithium fluoride
|
|
Leakage radiation at a distance of 1 meter from the target cannot exceed
|
100 milliroentgens in 1 hour
|
|
Instrument used to expose film to precisely controlled steps of increasing light intensity is:
|
sensitometer
|
|
Device made up of different density filters
|
step-wedge
|
|
Which effect is responsible for producing an image in a barium enema examination?
|
photoelectric effect
|
|
Very radiopaque structures fulling up the center of the imagine field in AEC pediatric fluoroscopy causes
|
Patient exposure to increase to a maximum
|
|
Primary barrier for the Xray beam is the
|
image intensifier
|
|
Beam alignment and image intensifier resolution and contrast should be monitored how often...
|
Monthly
|
|
Due to the phosphor aging, which should be measured periodically to check the contrast
|
conversion
|
|
This is considered a stochastic effect
|
Cancer
|
|
Line pair grid is used to measure
|
Resolution
|
|
Primary xray interaction taking place within the contrast media
|
photoelectric effect
|
|
Maximum mag during CINE is achieved with which filming method
|
total over framing
|
|
Ratio between the vertical and horizontal lines on the TV monitor
|
Kell factor
|
|
The primary protective barrier for a dedicated C-arm unit
|
Input phosphor of the image intensifier
|
|
For fixed fluoroscopic tube, the distance from the source to the tabletop shall be
|
15 inches
|
|
Lead aprons word during fluoroscopic procedures are designed to protect the radiographer from
|
compton scattered radiation
|
|
9" image intensifier tube has an inherent resolution of approx
|
4 lp/mm
|
|
High resolution area is defined as an area where an individual can receive a dose of
|
100 mR/hr
|
|
Light scattering causing less contrast is called?
|
Veiling glare
|
|
When X-rays are directed toward the patient, most X-rays are...?
|
Absorbed by the patient
|
|
annual whole body dose equivalent for persons under 18 years of age
|
0.5 Rem
|
|
Radiation effect in which severity varies with radiation and for which a threshold may occur
|
Non-stochastic effect
|
|
Speed of a camera system is calculated by dividing focal length by lens diameter. Number is called?
|
F-number
|
|
CINE framing due to poor optical coupling
|
Underframing
|
|
How to eliminate quantum noise
|
Increase tube current-mAs
|
|
Vertical resolution of the TV system is the ability to resolve
|
Horizontal lines
|
|
Purpose of viewing a TV system with interlaced horizontal scanning is...
|
Reduce image "flicker"
|
|
Conversion factor is calculated how
|
Light intensity at the output phosphor divided by the dose rate
|
|
Recording head of a video disc recorder uses what to record images
|
laser light and magnetic fields
|
|
Which component of the image intensifier changes the field of view from 9" to 6"
|
electrostatic lens
|
|
How does going from 9" to 6" change patient dose
|
Increases patient radiation dose
|
|
Conversion factor is
|
Radiation absorption by the input phosphor to output phosphor light gain
|
|
type of contrast that's more likely to cause renal effects
|
High Osmolarity
|