• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/78

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

78 Cards in this Set

  • Front
  • Back
What is the point in the respiratory track where inspired gas reaches body temperature, ambient pressure, saturated (BTPS) conditions?
isothermic saturation boundry
ISB
What is the primary goal of humidity therapy?
maintain normal physiologic conditions
Indications for warming inspired gases include all of the following EXCEPT:
A. treating a pt whose airways are reactive to cold
B. providing humidification when the upper airway is bypasses
C. treating a pt with a low body temp. (hypothermia)
D. reducing upper airway inflammation or swelling
D. reducing upper airway inflammation or swelling
Administration of dry gas at flows exceeding 4L/min can cause which of the following?
I. structural damage
II. heat loss
III. water loss
all of them
Inhalation of dry gases can do which of the following?
1. increase viscosity of secretions
2. impair mucociliary motility
3. increase airway irritability
all of them
Clinical indications for delivering cool humidified gas include which of the following?
1. postextubation edema
2. upper airway inflammation
3. croup
4. epiglottis
all of them
Factors affecting a humidifier's performance include all of the following EXCEPT:
a. surface area
b. temperature
c. time of contact
d. outlet size
outlet size
What is the most important factor determining a hudmdifier's performance?
temperature
The greater the temp of gas, the________________?
more water vapor it can hold
Gas leaving an unheated humidifier at 10 'C and 100% relative humidity (absolute humidity =9.4 mg/L) would provide what relative humidity at body temp?
about 20%
What is the simpilest way to increase the humidity output of a humidifier?
increase the temp of either the water or the gas
All of the following are true about humidification and contact time EXCEPT:
1. the longer a gas is exposed to water, the greater is the amount of evaporation
2. the rate of gas flow through a humidifier determines contact time
3. low flows decrease and high flows increase relative humidity output
4. bubble humidifier contact time depends on the water column depth
3. low flows decrease and high flows increase relative humidity output
Simple unheated bubble humidifiers are commonly used to humidify gases with what type of systems?
oronasal oxygen delivry
increasing the flow through an unheated bubble humidifier has what effect?
decreasing the water vapor content
Why shouldn't you heat the reserviors of a bubble humidifier used with an oronasal oxygen delivry system?
condensate will obstruct the delivery tubing
the relief vavle on a bubble humidifier serves what function?
it indicates when flow has been interupted

it protects the device from pressure damage
To protect against obstructed or kinked tubing, simple bubble humidifiers incorporate ________?
pressure relief valves
The typical pressure pop-off incorporated into most bubble humidifiers releases pressure above _____ psig?
2 psig
when checking an oxygen delivery system that incorporates a bubble humidifier running at 6 L/min, you occlude the delivery tubing, and the humidifier pressure relief immediatly pops off. What does this indicate?
mormal, leak free system
A design that increases surface area and enhances evaporation by incorporating an absorbant material partially submerged in a water resevior that is surrounded by a heating element best describes what type of humidifier?
wick
All of the following are TRUE regarding a membrane-type humidifier EXCEPT:
1. water and gas are seperated by a hydrophobic membrane
2. the membrane is permeable to water vapor but not to liquid water
3. as with other passover humidifiers, there is no bubbling action
4. a small layer of liquid water remains on both sides of the membrane
4. a small layer of liquid water remains on both sides of the membrane
Advantages of passover humidifiers include include all of the following EXCEPT:
a. they add minimal flow resistance to breathing circuits
b. they do not require heating to maintain body temp, ambient pressure, saturated (BTPS) conditions
c. they do not generate any bacteria-spreading microaerosol
d. they can maintain water vapor saturation at high flows
b. they do not require heating to maintain body temp,
what type of humidifier "traps" the pts body heat and expired water vapor to raise the humidity of inspired gas?
heat-moisture exchanger
Heat-moisture exchanger (HMEs) are mainly used to do what?
warm and humidify gases delivered to the trachea via ventilator circuits
What are these called?
simple condenser
hygroscopic condenser
hydrophobic condenser
Heat-moisture exchangers (HMEs)
All of the following are features of an ideal heat-moisture exchanger (HME) EXCEPT:
a. high compliance
b. minimial dead space
c. 70% or higher efficiency
d. minimal flow resistance
a. high compliance
All of the following are true of HMEs EXCEPT:
a. moisture output falls at high volumes and rates of breathing
b. high inspiratory flows and high FIO2 values can decrease HME efficiency
c. in-use HMEs have little effect on flow resistance to breathing
d. HMEs reduce bacterial colonization of ventilator circuits
c. in-use HMEs have little effect on flow resistence too breathing
For which of the following pts would you select a heated humidifier?
1. pt receiving O2 through a bypasses upper airway
2. pt receiving long-term mechanical ventilation
3. pt receiving O2 through an oronasal mask
1. pt receiving O2 through a bypasses upper airway
2. pt receiving long-term mechanical ventilation
These are necessary features in selecting a __________?

water level readily visible
over-temp protection
auditory and visual alarms
minimal temp overshoot
all of them
What are some potential problems with manually refilled geated humidifier resivoirs?
1.cross-comtamination and infection
2.variable compliance or delivered volume
3. delivery of dry and/or hot gases
all of them
The following are contraindications for heat-moisture exchangers EXCEPT:
a. presence of thick, copious, or bloody secretions
b. presence of large leak around artificial airway
c. when a pt is suffering from hyperpyrexia
d. when a pts minute ventilation exceeds 10 L/min
c. when a pt is suffering from hyperpyrexia
A patient receiving ventilatory support is being provided with humidification using a heat-moisture exchanger. A DR orders a bronchodilator drug administered through a metered-dose inhaler (MDI) via the ventilator circuit. What must be performed to ensure delivery of the drug to the patient?
The HME must be removed from the circuit during MDI use
A pt has been supported by a mech vent using a HME for the last 3 days. Suctioning reveals an increase in the amount and tenacity of secretions. What action is indicatied?
switch the patient to a large volume heated humidifier
Which of the following is FALSE about heated humidifier condensate?
a. it can block or obstruct the delivery circuit
b. it must be treated as contaminated waste
c. it requires that circuits be drained frequently
d. it poses minimal infection risk
d. it poses minimal infection risk
Which of the following are acceptable means to help minimize the problems caused by condensation in heated humidifier circuits?
1. installing water traps in the circuit
2. using a heated wire circuit
3. setting heater temp to 25-28'C
1. installing water traps in the circuit
2. using a heated wire circuit
When using water traps to help minimize the problems caused by condensation in a heated humidifier ventilator circuit, where would you place the traps?
in both the expiratory and inspiratory limbs, at low points in the circuit
All the following are TRUE when heated wire circuits are used with heated humidifiers EXCEPT:
a. the humidifier operates at a higher temp
b. there is less condensate and a reduced need for drainage
c. cost savings (less water and staff time) are realized
d. there is less infection risk for both the patient and caregivers
a. the humidifier operates at a higher temp
Which of the following pts would you recommend bland aerosol therapy administration?
1. pt with upper airway edema
2. pt with a bypassed upper airway
3. pt who must provide a sputum specimen
all of them
which of the following pts would you NOT recommend bland aerosol therapy administration?
1. pt with a history of airway hyperresponsiveness
2. pt with a bypasses upper airway
3. pt with active bronchoconstriction
1. pt with a history of airway hyperresponsiveness

3. pt with active bronchoconstriction
All of the following indicates a potential need for administration of a water or isotonic saline aerosol EXCEPT:
a.stridor or brassy, crouplike cough
b. evidence of increased volume of secretions
c. hoarseness after extubation
d. patient discomfort after bronchoscopy
b. evidence if increased volume of secretions
The particle size produced by an unltrasonic nebulizer depends mainly on _______?
signal frequency
A Dr orders bland water aerosol administration to a patient with an intact upper airway. Which of the following airway appliances could you use to meet this goal?
1. simple O2 mask
2. face tent
3. T-Tube
4. aerosol mask
face tent and aerosol mask
A Dr orders bland water aerosol administration to a patient with a tracheostomy. Which of the following airway devices could you use to meet this goal?
1. tracheostomy mask
2. face tent
3. T-tube
4. aerosol mask
tracheostomy mask and T-Tube
For what should sputum collected by aerosol therapy induction be inspected?
1. color
2. volume
3. odor
4. consistency
all of them
The risk of overhydration with continuous delivery of bland water aerosols is greatest among which patient group?
1. pt with fluid or electrolyte imbalances
2. pts with fever and infection
3. infants and small children
1. pts with fluid or electrolyte imbalances
3. infants and small children
A pt receiving nasal o2 at 3L/Min complains of nasal dryness and irritation. what could be done to make this patient more comfortable?
adding a humidifier to the delivery system
_________ is a suspension of liquid or solid particles in a gas?
aerosol
What is the retention of aerosol particles resulting from contact with the respiratory tract mucosa called?
deposition
what do the following factors affect:
1. size of the particles
2. shape and motion of the particles
3. physical characteristics of the airways
pulmonary deposition of an aerosol
What is the primary mechanism for deposition of large, high-mass particles (greater than 5 um) in the respiratory tract?
inertial impaction
The following will increase aerosol deposition by _______ _________?
- variable or irregular passages
- turbulent gas flow
inertial impaction
where do most aerosol particles in the 5-10 um range deposit?
upper airways
Where do most aerosol particles in the 1-5 um range deposit?
central airways
A __ second breath hold can increase aerosol deposition as much as __% and increase the ratio of aerosol deposited in lung parenchyma to central airways by fourfold?
10 second; 10%
What is the primary fate of aerosol particles that are between 1 and 0.5 um?
most are cleared during exhalation
Where do most aerosol particles that are less than 3 um deposit?
alveoli
A dr wants to deliver a therapuetic aerosol to the upper airway (nose,larynx,trachea). To help ensure maximum deposition in this area, you would select an aerosol generator with an MMAD in what range?
5 to 50 um
A dr wants to deliver a theraputic aerosol to the central and lower airways. To help ensure maximum deposition in this area, you would select and aerosol generator with an MMAD in what range?
2 to 5 um
A dr wants to deliver a therapeutic aerosol to the lung parenchyma (alveolar region). To help ensure maximum deposition in this area, you would select and aerosol generator with an MMAD in what range?
less than 1 um
What is the primary hazard of aerosol drug therapy?
untoward drug reactions
Drug aerosol delivery systems include all of the following EXCEPT:
a. dry powder inhalers
b. small-volume jet nebs
c. metered-dose inhalers
d. spinning disk nebs
d. spinning disk nebs
what does most of the spray generated by the majority of MDI's consist of?
propellant
When using a MDI without a holding chamber or a spacer, the patients should be instructed to fire the device at what point?
immediately after beginning a slow inspiration
Before inspiration and actuation of a MDI, the patient should exhale to what?
functional residual capacity
T/F:
To ensure delivery of the proper drug dosage with a MDI, should you warm the canister to hand or body temp and shake vigorously?
True
The following are beneficial affects of using a holding chamber with a _____?
Reduction in oropharyngeal aerosol deposition
decrease in need for hand-breath coordination
MDI
After actuating a MDI with a holding chamber, what should the patient be instructed to do?
continue to breath through the device for three breaths
Which of the following devices would you select to deliver an aerosolized bronchodilator to a young child?
a. MDI and spacer
b. MDI, holding chamber, and mask
c. MDI and holding chamber
d. dry powder inhaler
b. MDI, holding chamber, and mask
Advantages of the DPI drug delivery systems include all of the following EXCEPT:
a. low relative cost
b. no propellants required
c. no hand-breath coordination
d. unaffected by humidity
d. unaffected by humidity
Which of the following would be correct instrucions for a patient being taught proper use of a dry powder inhaler?
1. place mouth piece 4 cm from mouth
2. exhale slowly to FRC
3. inhale slowly (less than 30 L/min)
4. repeat until dose is used up
2. exhale slowly to FRC
4. repeat until dose is used up
Which groups are NOT recommended to use DPIs?
infants and children under 5
patients with an acute bromchospastic episode
In mouth breathing adult patients, which of the following factors is crucial in determining whether to use a mask or mouthpiece for aerosol drug delivery with a small-volume jet nebulizer?
a, clinician experience
b.drug concentration
c. patient preference and comfort
d. brand of small-volume jet nebulizer
c. patient preference and comfort
Normally, when using a 50-psi flowmeter to drive a small-volume jet nebulizer, what should you set the flow?
6-8 L/min
To decrease the VDS of a small-volume jet nebulizer during drug administration, what should you do?
continue the treatment until the nebulizer begins to sputter
To minimize a patients infection risk between drug treatments with a small-volume jet nebulizer (SVN) what should you do?
rinse the SVN with sterile water, air dry
Which of the following aerosol drug delivery devices would you recommend against using with a toddler or small child?
1. MDI
2. SVN
3. DPI
4. MDI, holding chamber, and mask
MDI and SVN
Possible complications associated with the selection of an aerosol drug delivery device include all of the following EXCEPT:
a. underdosing or overdosing because of improper technique
b. overhydration or fluid imbalances
c. adverse effects of the specific drug agent
d. enviornmental contamination or caregiver exposure
b. overhydration or fluid imbalances
What does this describe:
Give 4 puffs spaced 1-2 minutes apart; repeat up to 12 puffs with continued improvement. The best dose provides maximum subjective relief and the highest peak expiratory flow rate (PEFR) without side effects?
The proper dose response assessment of an MDI bronchodilator