Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
122 Cards in this Set
- Front
- Back
MYOCARDIAL CELLS
|
specialized cylindrical cells that relax and contract, changing the shape of the heart.
|
|
WHAT KIND OF PACEMAKER DOES THE HEART HAVE?
|
The heart has an INTRINSIC pacemaker; The heart generates electrical impulses that travel along a specialized conduction pathway.
*This is the hearts AUTOMATICITY* |
|
WHEN WAS THE FIRST PACEMAKER IMPLANTED IN THE USA?
|
In 1960, by Dr. C. Walton Lillehei.
|
|
HOW DOES A PACEMAKER WORK?
|
The electrical circuit in which the battery provides electricity that travels through a conducting wire to the myocardium.
The myocardium stimulates the heart to the beat (capture). |
|
THE PULSE GENERATOR
|
-contains a battery that provides the energy for sending electrical impulses to the heart.
-houses the circuitry that controls pacemaker operations. |
|
PACING THRESHOLD
|
-the amount of energy the output pulse needs (V and ms) to reliably capture the heart (or cause it to contract)
|
|
MYOCARDIAL AND EPICARDIAL LEADS
|
-leads that are directly applied to the heart.
-fixation mechanisms include: epicardial stab-in, myocardial screw in, and sutuure-on. |
|
WHY DO WE NEED TO PACE?
SYMPTOMATIC BRADYCARDIA |
-2* Type II
-CHB/3* Block |
|
WHY DO WE NEED TO PACE?
UNCONTROLLED TACHYCARDIAS |
-A fib
-A flutter -SVT -VT |
|
SYMPTOMATIC BRADYCARDIA
C R A P |
-weakness, fatigue, syncope/pre-syncope, mental confusion, palpitations, shortness of breath, exercise intolerance.
|
|
PACING MEANS?
|
capturing the heart
|
|
SENSING MEANS
|
"seeing" the heart's own intrinsic rhythm (like a periscope). The pacer then decides if it inhibits or triggers an impulse to be sent out or not.
|
|
TYPES OF PACEMAKERS
|
-Transcutaneous: skin pads which are external.
-Temporary: lead or wire is threaded transvenously to the RV with the wire to a power source externally. -Permanent |
|
FIXED RATE
|
set at a rate, no change
|
|
DEMAND
|
fires when needed if heart rate falls BELOW threshold
|
|
UNDER SENSING
|
competition with underlying rhythm
|
|
OVER SENSING
|
detects stimuli other than heart beat
|
|
FAILURE TO CAPTURE
|
(noncapture) pacer does not stimulate heart
*failure to capture and failure to pace mean the same thing. |
|
WHAT IS UNDERSENSING CAUSED BY?
|
-inappropriately programmed sensitivity
-lead dislodgement -lead failure: insulation break; conductor fracture |
|
WHAT IS NONCAPTURE CAUSED BY?
|
-lead dislodgement
-low output -poor connection at connector block -lead failure |
|
LESS COMMON CAUSES OF NONCAPTURE MAY INCLUDE
|
-Twiddler's Syndrome
-Electrolyte abnormalities (ex: hyperkalemia) -Myocardial infarction -Drug therapy -Battery depletion |
|
WHERE IS THE PACER PLACED?
|
on the non-dominant side
|
|
NURSING MANAGEMENT OF PACER
|
-monitor EKG
-Risk for infection -MICRO-SHOCKS: WEAR GLOVES -Electromagnetic interference (EMI) -Pacer system integrity -Documentation -Discontinuing |
|
MONITOR PERMANENT PACER
PRE-OP |
-what to expect
-teaching how to check pulse -card regarding pacer info |
|
MONITOR PERMANENT PACER
POST-OP |
-avoid reaching above head on affected side
-know symptoms of decreased cardiac output |
|
MONITORING PERMANENT PACER
PATIENT TEACHING |
-check pulse daily
-watch for infection -sx of decreased cardiac output -avoid electrical interference -identification -pacer checks -do not get MRI's -tens units will mess it up -cauderized defibrillators |
|
ELECTROMAGNETIC INTERFERENCE (EMI)
|
-interference caused by electromagnetic energy from a source that is outside the body
-electomagnetic fields that affect pacemakers are radio-frequency waves -few sources of EMI that affect pacemakers are found in the home or office, but several exist in hospitals |
|
EMI MAY RESULT IN THE FOLLOWING:
|
-oversensing
-transient mode change -reprogramming |
|
COUNTERSHOCK
"DEFIBRILLATION"...Unsynchronized Documentation |
-rhythm prior to defib
-# of defibs and joule level -rhythm after shocks -assessment of client -post care: monitor VS, EKG, maintain IV -Assess, Assess, Assess |
|
COUNTERSHOCK
"Cardioversion"...Synchronized |
-used for unstable VT, pulse present, unstable PSVT, Afib, and Aflutter
-These people have a pulse -too fast and we need to slow them down so we countershock |
|
HOW THE PACEMAKER COMPONENTS COMBINE W/BODY TISSUES TO FORM A COMPLETE CIRCUIT
|
-pulse generator is the power source or battery
-leads or wires -cathode (negative electrode) -anode(positive electrode) -body tissue |
|
PACEMAKERS CONTAIN WHAT?
|
-mini computers and batteries in a sealed titanium case with an epoxy connector block that links electrically to the inside of the unit.
|
|
WHEN CHECKING THE A PT'S PACING THRESHOLD WHAT DO WE DO?
|
-double that as the "safety margin" to make sure the output pulse is always sufficiently large to capture the heart.
|
|
WHAT HAPPENS WHEN A MAGNET IS PLACED OVER AN IMPLANTED PACEMAKER?
|
-it closes the reed switch and forces the device into magnet mode, which is usually asynchronous pacing.
-when the magnet is removed the device reverts back to normal behavior. |
|
LEADS ARE INSULATED WIRES THAT DO WHAT?
|
-deliver electrical impulses from the pulse generator to the heart.
-sense cardiac depolarization |
|
PASSIVE FIXATION LEADS
|
-the tines become lodged in the trabeculae (fibrous meshwork) of the heart.
|
|
ACTIVE FIXATION
|
-the helix (or screw) extends into the endocardial tissue.
-allows for lead positioning anywhere in the heart's chamber. |
|
MYOCARDIAL AND EPICARDIAL LEADS
|
-leads applied directly to the heart
-fixation mechanisms include: epicardial stab-in, myocardial screw-in, and suture-on. |
|
WHY DO WE NEED TO PACE?
|
-symptomatic bradycardia
-uncontrolled tachicardias |
|
SYMPTOMATIC BRADYCARDIA
|
-2* Type II
-CHB/3* block |
|
UNCONTROLLED TACHYCARDIAS
|
-AFib
-A flutter -SVT -VT |
|
TYPES OF PACEMAKERS
|
-Transcutaneous (skin pads-external)
-Temporary: lead or wire is threaded transvenously to the RV with the wire to a power source expternally -Permanent |
|
FIXED RATE
|
-set at a rate, no change
|
|
DEMAND RATE
|
-fires when needed if heart rate falls below threshold
|
|
SINGLE CHAMBER
|
-RA or RV
|
|
DUAL CHAMBER
|
-Both RA and RV
|
|
NBG CODES
|
-1st letter: chamber paced
-2nd letter: chamber sensed -3rd letter: device response to sensed beat (inhibited or triggered) -4th letter: programmability -5th letter: anti-tachycardia function |
|
CARDIOVERSION RECOVERY WE WATCH FOR?
|
-recover from IV sedation
-monitor rhythm -VS -Return of gag/swallow -medications: cardizem or verapamil |
|
INSTRUCTIONS FOR DISCHARGE
|
-meds
-documentation: rhythm prior to cardioversion, number of cardioversion attempts, rhythm after cardioversion and assessment of the client |
|
PCD DEVICES
(Pace/Cardiovert/Defibrillate) |
-used for sudden cardiac death and refractory tachycardias
-generator/ leads and patches -programmable: monitors and recognizes rhythms, overdrive for tachys, defibrillates lethal arrythmias |
|
ICD
(Internal Cardiac Defibrillators) |
-sense, pace, shock
|
|
ICD'S
|
Will I get shocked? Yep
What will it feel like? Shock! will I know it? Yep |
|
WHAT IS THE TREATMENT FOR 3* HEART BLOCK?
|
-Need a permanent pacer.
-look at cardiac output to see if you need SWAN |
|
INDICATIONS FOR PACING?
|
-2* Type II
-3* Heart Block -Uncontrolled tachys -AF, SVT, VT, |
|
WHAT DOES FAILURE TO CAPTURE MEAN?
|
-pacer does not stimulate the heart...it does not respond
|
|
POST INSERTION CARE OF PACEMAKER
|
-educate
-card -do not elevate arms or head -s/sx of infection -s/sx of decreased cardiac pacer |
|
THINGS TO REMEMBER ABOUT PACERS AND MAGNETS?
|
-it closes the reed
-magnet acts as a switch -can make asynchronous pacing. |
|
TYPES OF PACEMAKERS
|
-ICP
-Temporary -Permanent -Transthoracic |
|
FAILURE TO OUTPUT
|
-pacer does not generate a pacer spike when needed
|
|
DEFIBRILLATION
|
-asynchronization puts heart into temporary asystole
|
|
AED
|
-cues the rescuer
-place pads -analyze -shock if needed |
|
WHAT SHOULD YOU WEAR TO AVOID MICROSHOCKS?
|
-wear gloves when handling temporary unit
|
|
SYNCHRONIZED CARDIOVERSION
|
-machine synchronizes with QRS
-fires when it senses |
|
ICD
|
-defibrillates when needed.
|
|
WHAT IS HEMODYNAMICS?
|
-systemic circulation of blood
|
|
3 PARTS HEMODYNAMICS IS COMPOSED OF?
|
1.) Preload
2.) Contractility 3.) Afterload |
|
PRELOAD
|
-circulatory effects before blood reaches the heart
-VOLUME, VOLUME, VOLUME!!!! MEASURED BY PA PRESSURE AND CVP. |
|
WHAT NEEDS TO BE DONE TO DECREASE CONTRACTILITY
|
-decrease preload
-balance electrolytes -IABP (allows heart to rest) -give Beta Blockers |
|
WHAT NEES TO BE DONE TO INCREASE CONTRACTILITY?
|
-increase preload (gotta get more in there)
-give inotropic drugs -VAD |
|
AFTERLOAD
|
-circulatory effects as blood leaves the heart resistance or pressure that the ventricle must overcome to eject volume
-Aorta Valve Vessel Constriction -Measured by BP -circulation as it leaves the heart' |
|
DECREASED PRELOAD
|
-Is pt hypovolemic?
-Diuretics given? -bleeding? -Vasodilators given? -DRY!!!! |
|
INCREASED PRELOAD
|
-increased fluid administration?
-volume expanders given? -vasoconstrictors given? -blood products given? -WET!!!!!!!! |
|
CONRACTIBILITY
|
-cardiac muscle's capacity to contract
"stretch" |
|
WHAT IS NON-INVASIVE HEMODYNAMIC MONITORING?
|
BP Monitoring
|
|
HOW DO YOU DECREASE AFTERLOAD?
|
-give vasodilators
-give ACE Inhibitors(prils) -Calcium Channel Blockers |
|
HOW DO YOU INCREASE AFTERLOAD?
|
-give vasopressors or vasoconstrictors
-give dopamine |
|
3 COMPONENTS OF HEMODYNAMIC MONITORING EQUIPMENT
|
-Transducer
-Monitor -Fluid-filled catheter, tubing and flush system |
|
Colors for tube flushing
|
Red-Arterial line
Blue-CVP Yellow-PA pressure Orange-Neuro ICP monitoring |
|
WHAT IS THE TRANSDUCER?
|
-the instrument used to sense physiological events and transforms them into electrical signals
|
|
WHAT DOES THE MONITOR DO?
|
-records and provides a display of the original signal
|
|
FLUID FILLED CATHETER
|
-placed in an artery(ABP)
-rigid tubing -<48 in -attached to flush device that controls flow of solution thru the tubing -flush solution in NS or Heparin -bag to a pressure of 300 mmHG -flow rate of flush solution 3-5 mL/hr-additional 9 mLs an hr for their I & O's..to keep lines clear |
|
PHLEBESTATIC AXIS
|
-marked on the patient's chest, is the precise anatomical point of origin of the hemodynamic pressures being measured
-4th ICS |
|
ZERO REFERENCING
|
-the act of standardizing the transducer to obtain accurate measurements
-the act is performed by nurse upon insertion, q4h, or with position change. this is performed to all hemodynamic lines -Zeroing opens the system to the atmosphere and off to patient, press zero negates atmospheric pressure. |
|
ARTERIAL PRESSURE MONITORING
|
-accurate, direct and invasive
-continuous BP readings -used for lab draws -most commonly used hemodynamic technique -perform Allen's test prior to radial insertion -do NOT give any meds thru Art Line! |
|
CENTRAL VENOUS PRESSURE
(CVP, RAP) |
-measures volume returning to heart
-normal value is 2-6 mm Hg *Look at fluid balance *CVP is altered by change in venous tone,blood volume and RV contractility **CVP looks at fluid balance |
|
IF PATIENT HAS INCREASED CENTRAL VENOUS PRESSURE
|
-fluids given?
-hypervolemic? -pulmonary HTN? -Increased ITP (intrathoracic pressure) -Tamponade (extra fluid in the pericardial sac) -WET!!!!!!!! - |
|
IF PATIENT HAS DECREASED CENTRAL VENOUS PRESSURE
|
-hypovolemia
-vasodilation -sepsis -DRY!!!!!! |
|
PULMONARY ARTERY PRESSURE
(PA) |
-measures pressure of blood volume in the lung circulation
-normal value: 15-30/0-5 -indirect method of monitoring LV preload |
|
PA MONITORING
|
-catheter is placed in the pulmonary artery via subclavian or jugular vein
-pt in Trendelenburg position |
|
PA MONITORING
TYPES OF CATHETERS |
-4 lumen (manual CO)
-5 lumen (continuous CO) -SVO2 monitoring "SWAN" |
|
PA MONITORING
|
-RN responsible for monitoring and recording HR and rhythm w/ BP
-inflate and deflate balloon per dr. order -document measurement of catheter length post insertion -ensure portable CXR done post insertion. |
|
WHY USE A CXR POST INSERTION OF PA MONITOR
|
-check placement
-make sure there is no pneumothorax -normal length should be 55-60 sonometers |
|
INSERTION WAVEFORMS
|
-when a PA monitor is inserted, ventricular irritability may show up on EKG...showing some PVC's.
|
|
PULMONARY ARTERY WEDGE PRESSURE
(WEDGE, PAWP,PACWP) |
-measures volume blood returning to LV
-normal values: 5-12 mmHg |
|
IF DECREASED WEDGE
|
-hypovolemia
-vasodilation -RV failure -DRY!!!!!! |
|
IF INCREASED WEDGE
|
-hypervolemic
-vasoconstriction -WET!!!!! |
|
NOTES ON WEDGE PRESSURES
|
-may use PAD as estimation of wedge in pt's with "normal" valve and lung function
-PAWP is checked at end of expiration* -Always ensure that baloon is deflated after wedging catheter! |
|
SYSTEMIC VASCULAR RESISTANCE
(SVR)*vessel size* |
-measures amount of work required by LV to push blood through vessels
-normal value: 800-1200 dynes/sec/cm2 -also looking at viscosity of the blood |
|
MORE NOTES ON SVR
|
-infections
-pneumothorax -ventricular arrythmias -pulmonary infarc...forgetting to deflate the baloon -thrombosis -fistulas |
|
IF DECREASED SVR
(low #'s) |
-vasodilation
-sepsis -acid/base imbalances -DRY!!!!! (give vasoconstrictors...epi, norepi, etc.) |
|
IF INCREASED SVR
(high #'s) |
-vasoconstriction
-hypervolemia -hypothermia (cold and constricted) -WET!!!!!!!! |
|
STROKE VOLUME
CO/HR |
-measures volume ejected by LV with each beat
-normal value: 60-100 mL/beat |
|
IF DECREASED SV
|
-hypovolemic
-decreased contractility -low #'s -DRY!!!!!! |
|
IF INCREASED SV
|
-hypervolemia
-CHF -high #'s -WET!!!!! |
|
CARDIAC OUTPUT
|
-measures volume of blood ejected by the heart per minute (CO=SVxHR)
-normal value: 4-8 LPM -preload, contractility, afterload has to be good |
|
IF DECREASED CO
|
-hypovolemia
-decreased HR -cardiogenic shock...our pump has failed -increased SVR -DRY!!!! |
|
IF INCREASED CO
|
-hypervolemia
-increased HR -Inotrophic drugs -WET!!!!!!!!!!!!!! |
|
CARDIAC INDEX
|
-measures CO adjusted for body size (CI=CO/BSA)
-normal value: 2-4 L/min/m2 |
|
BSA
|
-body surface area
|
|
MIXED VENOUS OXYGEN SATURATION
|
-measures oxygen saturation of Hgb molecules returning to heart
-normal values: 60-80% |
|
IF DECREASED SVO2
|
-decreased ventilation/perfusion
-increased O2 consumption-fever sepsis -hypovolemia -anemia |
|
IF INCREASED SVO2
|
-decreased O2 consumption
-Sepsis -Hypovolemia (cold) |
|
CHARTING
|
-location of line
-site -dressing -zeroing line -where Swan is at in reference to introducer -taped securely -if connected to continous cardiac output monitor -waveforms |
|
DOPAMINE AND LASIX ARE ORDERED. WHAT IS THE RATIONALE FOR THESE DRUGS?
|
-get fluids out, increase pressure with dopamine.
-BP and HR will increase |
|
WHAT NURSING CARE IS ASSOCIATED WITH THE ADMINISTRATION OF DOPAMINE AND LASIX?
|
-I and O's
-heart rate -lung sounds -check K+ levels -telemetry (CVP and arterial line) |
|
IF THE CARDIAC INDEX IS BELOW 2 WHAT DOES THIS MEAN?
|
-patient is in deep shock
|
|
DRUGS USED TO INCREASE PRELOAD
|
VOLUME EXPANDERS
-NS -LR -Hetastarch (Hespan) -Albumin -Blood products |
|
DRUGS USED TO DECREASE PRELOAD
|
DIURETIC AGENTS
-Furosemide (Lasix) -Torsemide (Demadex) VENOUS DILATORS -nitroglycerine -nitroprusside (Nipride) -Morphine sulfate (note that this is an analgesic agent) |
|
DRUGS USED TO INCREASE AFTERLOAD
|
VASOPRESSORS
-Dopamine (at higher doses) -Epinephrine -Norepinepherine -Neosynephrine |
|
DRUGS USED TO DECREASE AFTERLOAD
|
VASODILATORS
-Nitroprusside (Nipride) -Nitroglycerine (at higher doses) -PDE inhibitors -Calcium Channel Blockers -ACE inhibitors -Hydralazine (Apresoline) |
|
DRUGS TO INCREASE CONTRACTILITY
|
POSITIVE INOTROPHIC DRUGS
-Dobutamine -Dopamine -Epinephrine -Norepinephrine -Isoproterenol (Isuprel) -PDE inhibitors -Digoxin |
|
NEGATIVE INOTROPHIC DRUGS
|
-Beta Blockers
-Calcium Channel Blockers |