- Front
- Back
General Purpose Caths |
Cournand lehman goodale-lubin multipurpose A1 multipurpose A2 multipurpose B1 multipurpose B2 balloon flotation(swan-ganz) |
|
cournand |
1st cath designed, used for R heart cath, pressure measuring, and blood sampling |
|
lehman |
thin walled, >inside diameter & <stiffness. R heart cath, pressure measuring, blood sampling |
|
goodale-lubin |
open end w/ 2 side holes(bird's eye), pressure measure and peds |
|
multipurpose A1 |
no side holes, pressure/sat measures, coronary artery bypass (RCA) |
|
multipurpose A2 |
oped, 2 side holes, peds studies & coronary angiography |
|
multipurpose B1 |
curve similar to Lehman, keeps stiffness longer than woven dacron, pressure/sat measures |
|
multipurpose B2 |
open end 2 latteraly opposed side holes |
|
balloon flotation (swan-ganz) |
polyvinyl chloride, multi lumened R heart cath w/ latex balloon at distal tip. continuous bedside hemodynamic monitoring of CVP, PA, PCWP, RV, RA & CO. Yellow hub |
|
Angiographic Caths |
Gensini National Institute of Health (NIH) Eppendorf Lehman Pigtail Van Tassel angled Pigtail Grollman Pigtail |
|
gensini |
3 pair of oval holes at 1.5 cm from tip, tapered tip, used for retrograde percutaneous R & L heart, aortic, ventricular, pulmonary, & vena cava angiographic |
|
national institute of health (NIH) |
nylon core, 4-6 side holes w/ gentle curve, used for high volume injections of RV, LV, PA, & AO |
|
eppendorf |
proximal 20cm reinforced w/ nylon, used for high pressure injections |
|
lehman |
thin walled, 4 side holes at 2.5 cm from tip, slight curve tapered to 5Fr, easy placement into LV |
|
pigtail |
polyethylene, tapered tip, coiled, 4-12 non lateral opposed side holes, most common cath for LV, AO, PA grams |
|
van tassel angled pigtail |
8 side holes, 145 or 155 degree angle 7cm from tip, used for LV & AO |
|
grollman pigtail |
4 spireal side holes, used for RV & PA, coil opposite side of van tassel pigtail |
|
Preformed Coronary Caths |
Judkins L & R Amplatz Variation Amplatz Sones Coronary
|
|
Judkins L & R |
no side hole, preformed tip, used for selective coronary angiography via femoral approach, JL used for cannulation of L coronary system, uniques 90 degree curved tips, secondary to 30 & 180 degree |
|
Amplatz |
end hole, for coronary ostia via femoral percutanious |
|
Amplatz Variations |
AL1-normal root AL3-dilated root/ large male AL4-marked dilation of root/ high branched LCA AR1-normal root AR2/3-abnormal position of RCA & bypass grafts |
|
Sones Coronary |
R & L coronary arteris via brachial cutdown I-1.5 in taper II-1.0 in taper III-2.0 in taper All taper to 5.5 Fr
|
|
Coronary Bypass Caths |
open end, femoral R saphenous bypass cath, internal mammory bypass (IMA), L coronary bypass (LCB), R coronary bypass (RCB) |
|
Brachial Artery Cutdown Caths |
Sones Coronary Castillo Soft Tip Coronary |
|
sones coronary |
R & L coronary arteris via brachial cutdown I-1.5 in taper II-1.0 in taper III-2.0 in taper All taper to 5.5 Fr |
|
castillo |
used when Sone's in not successful I-small II-medium III-large, size of AO |
|
Soft tip coronary |
wire braid body, soft deformable tip (bulb like), available in pigtail, judkins,amplatz, and multipurpose, prevents initial trauma, plaque dislodgement, and vessel dissection |
|
Specialized Caths |
Shirly Transvascular Brockenbrough Transseptal Fogarty Transducer Tip |
|
shirly transvascular |
type I-closed, 5 Fr, tapered tip II-open, 6 Fr, tapered tip III-closed, 6 Fr, tapered tip all have 6 side holes, used for retrograde LA &LV |
|
Brockenbrough Transseptal |
teflon, open, 4 side holes, inserted over brockenbrough needle into LA then LV |
|
Fogarty |
vinyl, balloon at distal end on lumen, vascular embolectomy via surgical cutdown or post cutdown |
|
transducer tip |
ultraminiature, semiconductor strain guage pressure transducer at tip, used high fidelity diagnostic hemodynamic pressure monitoring that allows for blood sampling and volume injections |
|
Guide wires |
fixed core terminating 3 cm from tip yellow end-guidewire white end-diagnostic body stiffness: standard, stiff tip configurations: straight, angled, angled LT (long taper) diameters: .018", .025", .035", .038" lengths:80, 150, 180, 260 cm |
|
Hydrophilic guide wire |
enhanced tip flexibility, added kink resistance straight tip configurations: straight, newton, bentsen fixed & moveable core: bentsen short taper, exchange length, super stiff, heparin coated (H) marker wire. J-tip:Rosen, 1.5mm, 3.0mm, 15mm fixed/moveable core: double ended, exchange length, guidewire w/ heperain (IQH) |