• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/83

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

83 Cards in this Set

  • Front
  • Back
murmurs gen
-abnormal 1st and 2nd heart sounds or additional abnormal noises
-turbulent blood through:
1. altered valves
2. abnormal openings: septal defects, stenotic vessels
3. extracardial effects: fibrous pericarditis
grade 1 murmur
nearly imperceptable: maybe in quiet room at PMI
grade 2 murmur
soft, but definite
-audible right over PMI
grade 3 murmur
low to moderate
- audible also at some distance to PMI
grade 4 murmur
loud
-audible all over chest, but not palpable
grade 5 murmur
very loud
-audible throughout chest and palable
grade 6 murmur
audible
-audible even when stethoscope not in contact with chest wall and palpable
stenosis
-undersized valve of vessel
-murmurs created:
1. systole: semilunar valves
2. diastole: AV valves
Insufficiency
-leakage of valve:
1. systole: AV valves
2. diastole: semilunar valves
mitral insufficiency gen
-can be go to pathogenic bacteria that thickened and shortened valvular leaflets
-systolic murmur: due to valve not closing sufficiently and blood flow back into the atrium
mitral insufficiency ECG
- murmur begins during QRS complex (systole) and ends during the T wave (end of systole)
-during diastole the AV valves should be open and murmur dies out
machinery murmur
-timing: systolic + diastolic murmur
-causes:
1. combo of stenosis and insufficiency
2. patent ductus arteriosus
fetal shunts
1. ductus arteriosus
2. foramen ovale
3. ductus venosus
-high resistance through collapsed lung
- low resistance though placenta and DA
- BP of R heart is higher than L
PDA
- at birth, inflation of lung reverses:
1. BP in L/ R heart
2. flow of oxygenated blood thru DA
-duct is sensitive to O2 and closes within days-months
-systolic-diastolic/ machinery murmur
P foramen ovale
-normal close silmultaneously with rise of BP at birth
- hardly diagnosed b/c atrial pressure are too low to generate audible murmurs
causes of conduction system failure
1.cardiac electrical abnormalities
2. cardiac structural abnormalities
cardiac electrical abnormalities
1. excitation disturbances:
failure of SA node
2. conduction disorder:
AV block
cardiac structural abnormalities
1. morphological: hypertrophy of RV
2. functional: cardiac infarction
Types of excitation disturbances
1. nontropic
2. heterotropic
nomotropic excitation dysfunctions
1. sinus arrhythmia
2. sinus tachycardia
3. sinus bradycardia
heterotropic excitation dysfunctions
-escape rhythm
- extra systoles
- shifting pacemaker
- supraventricular tachycardia
- atrial flutter and fibrillation
-ventricular tachycardia, flutter and fibrillation
sinus arrhythmia
-normal P waves preceeding QRST indicate each sequence intiated by SA node
-normal in dogs: vagal tonus, increase HR during inspiration
sinus tachy- and brady-cardia
-excitation and conduction are normal
- SA node is pacemaker but initiates APs at a rate lower/higher than normal
escape rhythmn
-extreme sinus bradycardia: intrinsic rate of AV note exceeds the sinus rate and AV node takes over
- cause failing AV node: then one of the auxillary pacemakers takes over
supraventricular systole
-premature stimulus generated by heterotropic atrial focus
- normal SA stimulus has no effect
- next systole follows at an interval decided by sinus rhythmn
supraventricular extra systole
-3rd depolarization originates from an ectopic supraventricular pacemaker (P wave precedes QRS)
-negative P wave in lead II and lead III indicates the ectopic initiation of the atrial depolarization
shifting pacemaker
-If SA slows down or fails other pacemakers take over
- possible cause very high vagal tone
- alters P wave and PQ segment
- closer the pacemaker center is to AV node: shorter the PQ segment
ventricular extra systole
-premature stimulus generated by heterotropic ventricular focus
- next systole follows at interval dedicated by sinus rhythmn
supraventricular tachycardia
-high rate of APs generated in atrium: ectopic location
- different shapes of P waves indicated several ectopic atrial pacemakers
atrial flutter and fibrillation
-ectopic (heterotopic) generation of APs cause a high atrial rate
-functional failure of both atria
- myocardial infarction alters QRS
- 50kg dog:
1. flutter: 200- 350 bpm
2. fibrillation: 350- 600 bpm
ventricular tachycardia
-absence of regular preceeding P waves indicates R waves are generated by a ventricular pacemaker
sinus tachy- and brady-cardia
-excitation and conduction are normal
- SA node is pacemaker but initiates APs at a rate lower/higher than normal
escape rhythmn
-extreme sinus bradycardia: intrinsic rate of AV note exceeds the sinus rate and AV node takes over
- cause failing AV node: then one of the auxillary pacemakers takes over
supraventricular systole
-premature stimulus generated by heterotropic atrial focus
- normal SA stimulus has no effect
- next systole follows at an interval decided by sinus rhythmn
supraventricular extra systole
-3rd depolarization originates from an ectopic supraventricular pacemaker (P wave precedes QRS)
-negative P wave in lead II and lead III indicates the ectopic initiation of the atrial depolarization
shifting pacemaker
-If SA slows down or fails other pacemakers take over
- possible cause very high vagal tone
- alters P wave and PQ segment
- closer the pacemaker center is to AV node: shorter the PQ segment
ventricular extra systole
-premature stimulus generated by heterotropic ventricular focus
- next systole follows at interval dedicated by sinus rhythmn
ventricular flutter and fibrillation
-multifocal excitation ceases pumping function
supraventricular tachycardia
-high rate of APs generated in atrium: ectopic location
- different shapes of P waves indicated several ectopic atrial pacemakers
atrial flutter and fibrillation
-ectopic (heterotopic) generation of APs cause a high atrial rate
-functional failure of both atria
- myocardial infarction alters QRS
- 50kg dog:
1. flutter: 200- 350 bpm
2. fibrillation: 350- 600 bpm
ventricular tachycardia
-absence of regular preceeding P waves indicates R waves are generated by a ventricular pacemaker
causes of conduction system failure
1. cardiac electrical abnormalities
2. cardiac structural abnormalities
cardiac electrical abnormalities
1. excitation disturbances: eg failure of SA node
2. conduction disorder: eg AV block
cardiac structural abnormalities
1. morphological: eg RV hypertrophy
2. functional: eg cardiac infarction
AV block gen
- on means by an AP can ordinarily pass from atria into ventricle is AV node and bundle of His
- different conditions either decrease or increase the rate of conduction of the impulse through the AV node and bundle of His, or block impulse entirely
AV block causes
1. ischemia
2. compression: scar tissue, calcification
3. inflammation: infection, rheumatic fever
4. extreme vagal tone
degrees of AV blocks
1st: AV conduction abnormally slow
2nd: dropped beats
3rd: no propagation, total block
1st degree AV block
-normally PQ interval decreases with increasing HR and vice versa
- pathologically decreased rate of conduction through AV node and bundle of His prolongs the PQ interval beyond physiological value
2nd degree AV block
- if conduction from atrium to the ventricle fails intermittently then the atrium depolarizes and generates a P wave
- conduction failure creates absent QRST and ventricle doesn't contract
=dropped beat
-entire conduction system can fail intermittently
3rd degree AV block
- connection between atrium and ventricle broken
-ventricle has to generate own APs:
atrium and ventricle beat at different rate:
1. atrium: higher rate from SA
2. ventricle: auxillary pacemaker at lower rate, beats independent of atrial activities
hypertrophy of RV
-more muscle= stronger signal= bigger R wave
- if only one side: timing of depolarization changes, increases depolarization time
- shift of cardiac vector: hypertrophy causes shift to R
myocardial infarction gen
-embolus that enters a coronary a will cause an occlusion due to minute anastomoses
- muscle supplied cannot sustain function= infarcted
ECG of myocardial infarction
1. ischemia causes ST segment depression w or w/o T wave inversion as a result of altered repolarization
2. myocardial injury causes ST segment elevation w or w/o loss of R wave
3. infarction causes deep Q waves as a result of absence of depolarization current from dead tissue and receeding currents from opposite side of the heart
electrical interference of ECG
- causes: poor contact of leads, dry skin, electric equipment nearby
- resembles atrial flutter or fibrillation
trembling patient and ECG
-equipment doesn't distinguish between cardiac and skeletal m contraction= interference
- can be mistaken for atrial flutter
manifestation of heart failure
-failure to pump blood adequately to meets bodily needs
-manifestations:
1. decrease of cardiac output
2. damming of blood behind the L or R heart
3. overloading of heart through increased cardiac output
types of heart failure
1. low output
2. high output
2. high pressure
types of low output heart failure
- total cardiac output falls below the required minimum:
1. compensated
2. decompensated
3. unilateral
compensated heart failure
- type of low output failure
- minimum cardiac output is restored
- cardiac reserve remains much reduced
decompensated heart failure
- type of low output failure
- minimum cardiac output not restored
- compensatory mechanisms continue
-effects turn disasterous/ vicious cycle that is incompatible with life
unilateral heart failure
-type of low output failure
- only one side fails
high output failure
-heart forced to pump more blood than required because of some bypass of normal circulation
- continuous volume overload finally causes a final failure of the heart
high pressure failure
-heart forced to develop more P than normally required due to unphysiologically high resistance in circulation
- consequences of increased workload cause final heart failure
normal relationships btween cardiac output and atrial pressure (preload)
1. increased metabolism increased cardiac output
2. more blood returning to heart increases atrial P (preload)
3. better filling of ventricles= higher end diastolic volume
- cardiac fibers: longer =stronger
maximum cardiac output
-determined by limitations to heart:
1. max HR
2. max stroke volume
minimum cardiac output
-determined by peripheral tissues: blood volume required to maintain metabolism without damage
-example 51 L/ min
immediate countermeasures to heart failure
-CO drops below minimum:
1. blood returning from tissues backs up in atria: increased atrial P
2. sympathetic stimulation: stimulates cardiac m
- increase CO but still deficient
restoration of require cardiac output
1. heart recovers gradually and partially: more strength
2. not meeting required min CO, water and NaCl retention by kidney: increase in blood V= increase in atrial P
- both result in increasing CO= compensation: kidney adaptation stops, cardiac dynamics are normal, but cardiac reserve remains reduced
fluid retention in kidneys
-decrease in BP regarded as missing blood V (an in hemmorhage):
1. kidneys reduced urine production
2. begins after a few minutes of acute failure until normal BP is restored
fluid retention after moderate heart damage
- pumping ability: 40-50% normal
1. better systemic filling: lower resistance, higher preload
2. CO may return to normal values
fluid retention after severe heart damage
-pumping abiltiy: 25-45% normal:
fluid retention - effects
1. overstretching of heart: further weakening
2. perfusion of kidneys too low to balance salt and water intake/output: fluid/salt retention= severe edema (eg lungs) that can lead to death
reparative process following myocardial infarction
-begins immediately:
1. new collateral blood supply begins to penetrate infarcted tissue: m at fringe areas may begin to function again
2. necrotic tissue replaced by CT
3. undamaged m hypertrophy
4. no hyperplasia of m (no new m)
5. scar remains
6. most of recovery in 5-7 weeks
cardiac reserve in different compensated heart failures
-CO meets needs of tissues at expense of increased atrial P
- little cardiac reserve left: further increased of preload causes almost no change in CO ( graph horizontal)
-heavy exercise usually causes immediate return of symptoms
course of decompensated heart failure
-after initial failure:
1. heart doesn't recover sufficiently
2. cannot supply kidneys to meet normal function
3. kidneys retain salt and water
4. increased blood V
5. progressive edema: lung, etc
6. overstretching of heart m: dilation of ventricles
7. death
course of decompensated heart failure
-neither sympathetic strengthening or fluid retention by kidney can restore minimum CO:
1. kidney doesn't stop retaining: increasingly higher blood volume
2. increase in preload overstretches m and dilates ventricles
3. cardiac edema
4. edema and increased preload further weaken heart and decrease CO
5. death due to lower CO followed by weakining in a cycle
unilateral L heart failure
1. blood backs up into pulmonary veins and BP rises eg 18mm (normal 3)
2. to maintain lung perfusion: BP difference of 10 required
3. R heart increases P in pulmonary a to 28
4. mean lung P rises from 8 to 23, forcing more fluid into interstitium=
pulmonary edema
5. if interstitial capacity exceeded, alveolar edema insues which is life-threatening
unilateral heart failure summary
-L more likely because of higher workload
-most R failure occurs with L failure
- R: peripheral edema
- L: lung edema
high output heart failure
-forced to pump more blood than demanded by tissues
- additional volume increases workload, which can result in overload and final heart failure
-causes:
1. arterio-venous shunts
2. low blood flow resistance
PDA
- fetus: lung small and compact, high resistance
- smaller PDA more unlikely to present
-consequences:
1. to meet demands of tissues, L has to pump more blood= larger
2. large opening means that tissue don't get enough blood: reduced exercise ability, organ disfunction, death
ventricular septal defect
-part of LV blood enters RV:
1. LV needs to pump more blood
2. increased LV workload
3. enlargement, thickening, and dilation of LV
tetralogy of fallot
1. persistant R aortic arch
2. ventricular septal defect
3. pulmonary stenosis
4. ventricular dilation: forcing blood through VSD and stenosis
low resistance to blood flow
-normally arterioles regulate blood flow in tissues: controlled by ANS
- beriberi:
1. vit B deficiency: decreased arteriolar tone
2. heart pumps more blood to maintain BP: inc CO
3. edema
4. leads to wobbling and trembling, sometimes death
high P heart failure
-thickening of ventricular wall more on inside: smaller chamber volume
1. compensation by increased HR
2. eventually cannot eject sufficient blood volume
3. high pressure and high rate of contraction: high workload
4. can lead to final heart failure