16 Cards in this Set

  • Front
  • Back

Who makes this unit?

BW technologies
What model do we have?
GAMAX 4
What gases do our units detecet?
H2S, CO, O2, and LEL
What Batteries do they use?
Black and Decker, 3.6 volt, Type 2, VP100- vp 110 batteries, Versa pac charger
What is the operation period for the battery?
12 hours with 3-4 hour charge time for 2 batteries at a time.
How often do we calibrate these units?
6months
Whats unique about 604's unit?
It is the only gas monitor with the optional data logger computer chip located on the back side. which allows the downloading of info onto a PC for review.
How do you determine a low battery on the unit?
The unit will not complete its initial self test or the monitor will beep and flash once every 5 seconds and the low battery flashes. Auto. shutdown is indicated by a series of 8 beeps and a flashing battery.
TWA
Time weighted average: The acucumulated value.
How do you check the TWA?
Press the ok and TWA button at the same time
What will pressing the calibrate and the ok button do?
Allow you to access the alarm set points and calibration
How do you turn it off?
Press and hold the green for 5 seconds.
Whos responsability is it to change the batteries on saturday.
Engineers
Sample readings will take __ seconds for every __ ft.
2
Take sample readings every __?
4 ft
What can cause LEL sensor loss?

Sulfer, silicone, lead, phosphates.