What are important considerations when planning a functional processing area?
The height and overall dimensions of the work surface, plus a clear surrounding area
Noncritical"
Instruments, equipment, and materials that do not normally penetrate or contact mucous membranes but which are exposed to splatter, spray, or splashing of blood, or touched by contaminated hands. Items require intermediate-level disinfecton
To identify total resource s required to meet the guidelines of this directive and those of OSHA and the EPA, commands must use the ________
The planning, Programming, and Budeting System (PPBS)
Spray-Wipe-Spray
An acceptable method of cleaning and disinfecting
Review and revise all infection control policies and procedures at least _______
Annually
Biological indicators"
Monitor designed to assess whether sterilization actually occurred. Consists of bacterial endospores impregnated in paper strips or sealed in glass ampules or plastic vials
The safest and provides an effective cleaning process.
Automated Washer Processor
What are the types of Sterilizaton Monitors
Biological Monitors, Dosage indicators, and Process Indicators
What are the determinants of the space requirements for sterilization areas?
Size of and sufficient access for loading, unloading, and servicing the sterilizer.
When do you use Bead Sterilizers?
Only for intra-appointment sterilization of clean metallic instruments
In Chemical Vapor Sterilization a mixture of formaldehyde, water, and other chemicals heated under pressure forms a gas that can achieve sterilization. Sterilization requires ___ minutes at _____F
20 minutes at 270F
The ICO must ensure that infection control functions are addressed at least ________
quarterly
Dental Item Classification"
Dental items are classified as critical, semicritical, or noncritical based on the pathways through which cross contamination may occur and the location and technique of instrument use.
Ultrasonic Cleaning
Safer and more effective than manual scrubbing. Always use baskets in the ultrasonic unit to reduce instrument handling and improve cleaning efficiency.
Bioburden"
The number of micro-organism contaminating an object. AKA bioload or microbial load
True or False. Rinse and scrub contaminated instruments or materials in the patient treatment areas before taking it to the receiving section of the sterilization area.
False. Following the completion of a patient's treatment, take it directly to the receiving section of the sterilization area. Do not rinse, scrub, or unnecessarily handle contaminated instruments in the DTRs.
What kind of gloves should you wear while handling all potentially contaminated items?
Wear heavy, puncture proof gloves.
Semicritical"
Instruments, equipment, and materials that frequently contact mucous membrane but cannot be sterilized due to their design or inability to withstand heat. At minimum, items require high-level disinfection
Dosage Indicators"
Chemical dyes that chnge color when exposed to steam, dry heat, or chemical vapors for a specific period of time. Determines whether the conditions necessary for sterilization have been met.
True or False. Liquid Chemical Sterilization should be treated as high-level disinfectants rather than sterilant.
True
A positive biological monitoring occured
"1. Notify the ICO
Maintain a consolidated sterilization log for each sterilizer which must contain:
Identification number, sterilization date, duration and temperature of sterilization cycles operator's name, biological monitoring results, repair and preventive maintenance dates, and a synopsis of action taken.
Biological monitoring of the sterillization process at least _____
Weekly
For ethylene oxide sterilization, sterilization can be achieved in ___ to___ hours at ____F
2 to 3 hours at 120F. CO should not purchase new ethylene oxide equipment
True or False. Dosage and process indicators are replacements for biological monitoring
False. They are not replacements.
Bowie-Dick Test"
Diagnostic test of a prevacuum sterilizer's ability to remove air from the chamber and prevent air reentrant This is not a sterility assurance test.
Engineering Controls"
Equipment or methods which isolate or remove bloodborne pathogens from the workspace
Sterilizer (gravity displacement type)"
A type of sterilizer in which incoming steam displaces the residual air through a port or drain usually in or near the bottom of the sterilizer chamber. 30 minutes at 121-123C(250-254F) and 15 to 17 pounds per square inch or 15 minutes at 132-135C(270-274F) and 30 to 32 psi
All contaminated, reusable items must be decontaminated by immersion in an EPA-registered disinfectant unless_________
Items are cleaned in an ultrasonic cleaner with an EPA-registered disinfectant which is also approved as an ultrasonic cleaner
Critical Item"
Instruments and materials that penetrate the skin, mucus membranes, or bone. These items must be sterile before use
When receiving and cleaning during sterilization, physical separation from the remainder of the sterilization area is ideal. If physical separation is not obtainable, __________ and _________ are critical
proper outfitting and equipment selection
Work Practice Controls"
Controls that reduce the likelihood of exposure by altering the way one performs a task such as having patients brush their teeth, safe technique of recapping needles
What is the most preferable sterilization?
Heat sterilization
Manual Scrubbing
Time consuming and presents increased potential for contamination injury, but effective for cleaning instruments
True or False. To achieve sterility, moist heat under pressure must come in contact with all surfaces of allitems for the appropiate lenght of time.
True
Process Indicators"
Chemical Dyes that change color upon short exposure to sterilizing conditions
Perform biological monitoring ______
weekly
DTFs must have a central sterilization are because ________
Activity is safer, provides more efficient use of materials and personnel, and standardizes execution and monitoring procedures.
True or false. Liquid chemical disinfectant or sterilants cannot be biologically monitored
True
How do you sterilize Rubber products?
"Nonheat stable - ethylene oxide or chemical sterilization
Perform the _________ test weekly when using prevacuum sterilizers
Bowie-Dick test
True or false. In steam heat sterilization, do not allow sterile instruments and packs to cool before moving to storage areas.
False. Placing warm, wrpped, sterilized items on cool surfaces can induce condensate formation and result in contamination.
Chemical indicator"
Chemical dyes used to determine whether the conditions required for sterilization are met. AKA chemical monitor, dosage indicator, or process indicator
Sterilizer (prevacuum type)"
A type of sterilizer which relies on one or more pressure and vacuum exursions at the beginning or end of the cycle.Requires shorter cycle times at 132-135C
Who appoints the Infectious Control Officer?
Co and OIC
In dry heat sterilization, the typical dry heat cycle is ____ minutes at _____F, plus the time required tto bring the load up to sterilization temperature.
90 minutes at 320-345F
The preferred method for cleaning Burs and Diamonds is _______
Dry heat for 90 minutes at 320-345F. Clean in ultrasonic cleaner before sterilizing.
Biological Monitor"
A bacterial endospore test designed to assess whether sterilization has actually occurred. AKA biological indicator or biological spore test
What is Sterile Sterility?
Free from all living micro organisms.
What are Engineering Controls?
Equipment or methods which remove bloodbore pathogens from the workplace.
"Who directs continous review and upgrade of the sterilization and infection control syllabus at the Naval School of Dental Assisting and technology (NSDAT) for dental ""A"" school?"
BUMED (MED-54).
What is Saturated Steam Sterlization?
Steam heat under pressure for sufficient length of time to kill all forms of micro organisms
Dental Item are classified as?
Critical, Semi Critical and Non critical.
15 minutes at?"
"132-135 C 270-274 F
What is Bioburden?
"Number of Micro-organisms contaminating an object.
What is a Chemical indicator?
"Chemical dyes used to determine whether sterlization has been met. Also know as
Semicritical Items are?
Materials that frequently contact mucous membranes but cannot be sterlized due to their inability to withstand heat.
What is a Prevacuum type Sterilizer?
Sterilizer which relies on one or more pressure and vacuum excursions at the beginning or end of the cycle.
Infection Control function are addressed how often?
At least Quarterly.
Inital Infection control training is done how often after initial training?
Annually.
Non Critical Items are?
Materials that do not normally penetrate or contact mucous membranes but are exposed to splatter, spray or splashing of blood.
What is a Unit Dose?
Quantity of materials or supplies required to treat a single patient.
What is Chemical disinfection?
The destruction or inhibition of most viruses and bacteria while in their growth phase.
Asepsis is?
Process of preventing the access of micro-organisms
Commmands must document infection control training with?
Names of persons attending and conduction. Dates and a summary of the contents of the training.
How long are infection control manuals maintained?
For at least three years.
Critical Items are?
Materials that penetrate the skin, mucous membrane or bone.
What is a Gravity Displacement type sterilizer?
A Sterilizer in which incoming steam displaces the residual air through a port or drain in or near the bottom of the sterilizer chamber.
30 minutes at?"
"121-123 C (250-254 F)
Operating temperatures for the Prevacuum type Sterilizer are?
132-135 C.
What is the PPBS?
Planning, Programming and Budgeting system used to identify total resources required to meet Infection Control guidelines and those of OSHA and the EPA.
What is Dental Item Classification?
Dental Items are classified based on pathways through which cross contamination may occur and the location and technique of instrument use.
What is an invasive procedure?
A surgical entry into the tissues, cavities organs or repair of major tramatic injuries.
What is Barrier technique?
Use of fluid resistant materials to protect them from contamination. Rubber plastic, paper, foil.
Inital Infection control training is done?
within 90 days of reporting to a Command.
What is BUMEDINST 6600.10A?
Dental Infection control Instruction.
What is a Biological Monitor?
Bacterial endospore test which assess whether sterlization has occured. Also known as biological indicator or biological spore test
Medical testing for HBV and HIV are at what intervals?
6 weeks, 12 weeks and 6 months.