What are important considerations when planning a functional processing area?
|
The height and overall dimensions of the work surface, plus a clear surrounding area
|
|
Noncritical"
|
Instruments, equipment, and materials that do not normally penetrate or contact mucous membranes but which are exposed to splatter, spray, or splashing of blood, or touched by contaminated hands. Items require intermediate-level disinfecton
|
|
To identify total resource s required to meet the guidelines of this directive and those of OSHA and the EPA, commands must use the ________
|
The planning, Programming, and Budeting System (PPBS)
|
|
Spray-Wipe-Spray
|
An acceptable method of cleaning and disinfecting
|
|
Review and revise all infection control policies and procedures at least _______
|
Annually
|
|
Biological indicators"
|
Monitor designed to assess whether sterilization actually occurred. Consists of bacterial endospores impregnated in paper strips or sealed in glass ampules or plastic vials
|
|
The safest and provides an effective cleaning process.
|
Automated Washer Processor
|
|
What are the types of Sterilizaton Monitors
|
Biological Monitors, Dosage indicators, and Process Indicators
|
|
What are the determinants of the space requirements for sterilization areas?
|
Size of and sufficient access for loading, unloading, and servicing the sterilizer.
|
|
When do you use Bead Sterilizers?
|
Only for intra-appointment sterilization of clean metallic instruments
|
|
In Chemical Vapor Sterilization a mixture of formaldehyde, water, and other chemicals heated under pressure forms a gas that can achieve sterilization. Sterilization requires ___ minutes at _____F
|
20 minutes at 270F
|
|
The ICO must ensure that infection control functions are addressed at least ________
|
quarterly
|
|
Dental Item Classification"
|
Dental items are classified as critical, semicritical, or noncritical based on the pathways through which cross contamination may occur and the location and technique of instrument use.
|
|
Ultrasonic Cleaning
|
Safer and more effective than manual scrubbing. Always use baskets in the ultrasonic unit to reduce instrument handling and improve cleaning efficiency.
|
|
Bioburden"
|
The number of micro-organism contaminating an object. AKA bioload or microbial load
|
|
True or False. Rinse and scrub contaminated instruments or materials in the patient treatment areas before taking it to the receiving section of the sterilization area.
|
False. Following the completion of a patient's treatment, take it directly to the receiving section of the sterilization area. Do not rinse, scrub, or unnecessarily handle contaminated instruments in the DTRs.
|
|
What kind of gloves should you wear while handling all potentially contaminated items?
|
Wear heavy, puncture proof gloves.
|
|
Semicritical"
|
Instruments, equipment, and materials that frequently contact mucous membrane but cannot be sterilized due to their design or inability to withstand heat. At minimum, items require high-level disinfection
|
|
Dosage Indicators"
|
Chemical dyes that chnge color when exposed to steam, dry heat, or chemical vapors for a specific period of time. Determines whether the conditions necessary for sterilization have been met.
|
|
True or False. Liquid Chemical Sterilization should be treated as high-level disinfectants rather than sterilant.
|
True
|
|
A positive biological monitoring occured
|
"1. Notify the ICO
|
|
Maintain a consolidated sterilization log for each sterilizer which must contain:
|
Identification number, sterilization date, duration and temperature of sterilization cycles operator's name, biological monitoring results, repair and preventive maintenance dates, and a synopsis of action taken.
|
|
Biological monitoring of the sterillization process at least _____
|
Weekly
|
|
For ethylene oxide sterilization, sterilization can be achieved in ___ to___ hours at ____F
|
2 to 3 hours at 120F. CO should not purchase new ethylene oxide equipment
|
|
True or False. Dosage and process indicators are replacements for biological monitoring
|
False. They are not replacements.
|
|
Bowie-Dick Test"
|
Diagnostic test of a prevacuum sterilizer's ability to remove air from the chamber and prevent air reentrant This is not a sterility assurance test.
|
|
Engineering Controls"
|
Equipment or methods which isolate or remove bloodborne pathogens from the workspace
|
|
Sterilizer (gravity displacement type)"
|
A type of sterilizer in which incoming steam displaces the residual air through a port or drain usually in or near the bottom of the sterilizer chamber. 30 minutes at 121-123C(250-254F) and 15 to 17 pounds per square inch or 15 minutes at 132-135C(270-274F) and 30 to 32 psi
|
|
All contaminated, reusable items must be decontaminated by immersion in an EPA-registered disinfectant unless_________
|
Items are cleaned in an ultrasonic cleaner with an EPA-registered disinfectant which is also approved as an ultrasonic cleaner
|
|
Critical Item"
|
Instruments and materials that penetrate the skin, mucus membranes, or bone. These items must be sterile before use
|
|
When receiving and cleaning during sterilization, physical separation from the remainder of the sterilization area is ideal. If physical separation is not obtainable, __________ and _________ are critical
|
proper outfitting and equipment selection
|
|
Work Practice Controls"
|
Controls that reduce the likelihood of exposure by altering the way one performs a task such as having patients brush their teeth, safe technique of recapping needles
|
|
What is the most preferable sterilization?
|
Heat sterilization
|
|
Manual Scrubbing
|
Time consuming and presents increased potential for contamination injury, but effective for cleaning instruments
|
|
True or False. To achieve sterility, moist heat under pressure must come in contact with all surfaces of allitems for the appropiate lenght of time.
|
True
|
|
Process Indicators"
|
Chemical Dyes that change color upon short exposure to sterilizing conditions
|
|
Perform biological monitoring ______
|
weekly
|
|
DTFs must have a central sterilization are because ________
|
Activity is safer, provides more efficient use of materials and personnel, and standardizes execution and monitoring procedures.
|
|
True or false. Liquid chemical disinfectant or sterilants cannot be biologically monitored
|
True
|
|
How do you sterilize Rubber products?
|
"Nonheat stable - ethylene oxide or chemical sterilization
|
|
Perform the _________ test weekly when using prevacuum sterilizers
|
Bowie-Dick test
|
|
True or false. In steam heat sterilization, do not allow sterile instruments and packs to cool before moving to storage areas.
|
False. Placing warm, wrpped, sterilized items on cool surfaces can induce condensate formation and result in contamination.
|
|
Chemical indicator"
|
Chemical dyes used to determine whether the conditions required for sterilization are met. AKA chemical monitor, dosage indicator, or process indicator
|
|
Sterilizer (prevacuum type)"
|
A type of sterilizer which relies on one or more pressure and vacuum exursions at the beginning or end of the cycle.Requires shorter cycle times at 132-135C
|
|
Who appoints the Infectious Control Officer?
|
Co and OIC
|
|
In dry heat sterilization, the typical dry heat cycle is ____ minutes at _____F, plus the time required tto bring the load up to sterilization temperature.
|
90 minutes at 320-345F
|
|
The preferred method for cleaning Burs and Diamonds is _______
|
Dry heat for 90 minutes at 320-345F. Clean in ultrasonic cleaner before sterilizing.
|
|
Biological Monitor"
|
A bacterial endospore test designed to assess whether sterilization has actually occurred. AKA biological indicator or biological spore test
|
|
What is Sterile Sterility?
|
Free from all living micro organisms.
|
|
What are Engineering Controls?
|
Equipment or methods which remove bloodbore pathogens from the workplace.
|
|
"Who directs continous review and upgrade of the sterilization and infection control syllabus at the Naval School of Dental Assisting and technology (NSDAT) for dental ""A"" school?"
|
BUMED (MED-54).
|
|
What is Saturated Steam Sterlization?
|
Steam heat under pressure for sufficient length of time to kill all forms of micro organisms
|
|
Dental Item are classified as?
|
Critical, Semi Critical and Non critical.
|
|
15 minutes at?"
|
"132-135 C 270-274 F
|
|
What is Bioburden?
|
"Number of Micro-organisms contaminating an object.
|
|
What is a Chemical indicator?
|
"Chemical dyes used to determine whether sterlization has been met. Also know as
|
|
Semicritical Items are?
|
Materials that frequently contact mucous membranes but cannot be sterlized due to their inability to withstand heat.
|
|
What is a Prevacuum type Sterilizer?
|
Sterilizer which relies on one or more pressure and vacuum excursions at the beginning or end of the cycle.
|
|
Infection Control function are addressed how often?
|
At least Quarterly.
|
|
Inital Infection control training is done how often after initial training?
|
Annually.
|
|
Non Critical Items are?
|
Materials that do not normally penetrate or contact mucous membranes but are exposed to splatter, spray or splashing of blood.
|
|
What is a Unit Dose?
|
Quantity of materials or supplies required to treat a single patient.
|
|
What is Chemical disinfection?
|
The destruction or inhibition of most viruses and bacteria while in their growth phase.
|
|
Asepsis is?
|
Process of preventing the access of micro-organisms
|
|
Commmands must document infection control training with?
|
Names of persons attending and conduction. Dates and a summary of the contents of the training.
|
|
How long are infection control manuals maintained?
|
For at least three years.
|
|
Critical Items are?
|
Materials that penetrate the skin, mucous membrane or bone.
|
|
What is a Gravity Displacement type sterilizer?
|
A Sterilizer in which incoming steam displaces the residual air through a port or drain in or near the bottom of the sterilizer chamber.
|
|
30 minutes at?"
|
"121-123 C (250-254 F)
|
|
Operating temperatures for the Prevacuum type Sterilizer are?
|
132-135 C.
|
|
What is the PPBS?
|
Planning, Programming and Budgeting system used to identify total resources required to meet Infection Control guidelines and those of OSHA and the EPA.
|
|
What is Dental Item Classification?
|
Dental Items are classified based on pathways through which cross contamination may occur and the location and technique of instrument use.
|
|
What is an invasive procedure?
|
A surgical entry into the tissues, cavities organs or repair of major tramatic injuries.
|
|
What is Barrier technique?
|
Use of fluid resistant materials to protect them from contamination. Rubber plastic, paper, foil.
|
|
Inital Infection control training is done?
|
within 90 days of reporting to a Command.
|
|
What is BUMEDINST 6600.10A?
|
Dental Infection control Instruction.
|
|
What is a Biological Monitor?
|
Bacterial endospore test which assess whether sterlization has occured. Also known as biological indicator or biological spore test
|
|
Medical testing for HBV and HIV are at what intervals?
|
6 weeks, 12 weeks and 6 months.
|