BUMEDINST 6470.22A
Navy Radiological Systems Performance Evaluation Program
What provides guidance on the radiological safety managemnet of all imaging systems in Navy Medicine?
BUMEDINST 6470.22A
the Navy Rad Systems Perfomance Eval Program includes?
PQS, Structural shielding design, systems evals, reporting procedures, and criteria for prohibiting systems usage to ensure the safe and effective use of medical and dental radiological systems
When wil NAVMEDLOGCOM notify commands and facilities receiving new and replacement medical rad systems of the prospective installation date?
w/in 10 working days of the equipment contract award.
All rad systems shall be evaluated within _____ _____ of istallation and after major repairs, prior to inital use as well. This is who's overal responsibility?
30 days
CO's
Medical xray units (fixed or mobile) ashore and on hopital ships are eval'd?
Annually
All fixed x-ray units afloat are eval'd when?
q 24months
Dental fixed and portable x-ray units require evaluation every?
36 months
Mammography systems are evaluated every?
Periodically and uder the conditions req by manufacture
All system evaluation reports are kept for?
3 years
Structural radiation shielding designs are maintained?
for the lifetime of the facility
Report of Corrective actions is provided to the service command w/in _____ ______ of receipt of the x-ray survey report.
30 days
Radiological Equipment Survey Request Form
NAVMED 6470/14