Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
10 Cards in this Set
- Front
- Back
BUMEDINST 6470.22A
|
Navy Radiological System Evaluation Program
|
|
Purpose?
|
provide guidance on the radiological safety management of all imagining systems in Navy medicine.
|
|
NAVMEDLOGCOM will notify facilities receiving new and replacement systems of prospective installation within?
|
10 working days of the equipment contract award.
|
|
All radiological systems shall be evaluated within?
|
30 days of installation and after major repairs.
|
|
Newly installed systems shall be evaluated when?
|
Prior to initial clinical use.
|
|
Medical x-ray units (fixed or mobile) ashore and on hospital ships are evaluated?
|
Annually
|
|
All fixed x-ray units afloat are evaluated?
|
Bi-annually. (24 months)
|
|
Dental (fixed and portable) units are evaluated?
|
36 - months.
|
|
How long are evaluation reports maintained?
|
3 years
|
|
How long are the structural radiation shielding designs kept?
|
For the lifetime of the facility.
|