ANZCA Guideline 1 |
All connections for gas cylinders must be PIN indexed |
Connections |
|
ANZCA Guideline 2 |
A reserve supply of oxygen must be attached |
Reserve |
|
ANZCA Guideline 3 |
A display of gas supply line and cylinder pressure must be provided |
Display |
|
ANZCA Guideline 4 |
An oxygen supply failure warning device must be present |
Failure |
|
ANZCA Guideline 5 |
There must be one gas flow control knob for each gas |
Knob |
|
ANZCA Guideline 6 |
The oxygen flow control know must be larger and fluted for tactile ID |
O2flow |
|
ANZCA Guideline 7 |
vaporiser dials must turn anti-clockwise to turn on |
Vaporisers |
|
ANZCA Guideline 8 |
Must have a protected master switch to prevent inadverdant operation |
Master switch |
|
ANZCA Guideline 9 |
Must have a 10 sec time delay when turned off |
Delay |
|
ANZCA Guideline 10 |
Must have a high pressure relief / blow off valve |
High pressure |
|
Oxygen failure device requirements 1 |
Must activate automatically when oxygen pressure falls below a predetermined critical level |
Critical |
|
Oxygen failure device requirements 2 |
Must cut off supply of all gases other than oxygen to the common gas outlet |
Cut |