ANZCA Guideline 1

All connections for gas cylinders must be PIN indexed

Connections

ANZCA Guideline 2

A reserve supply of oxygen must be attached

Reserve

ANZCA Guideline 3

A display of gas supply line and cylinder pressure must be provided

Display

ANZCA Guideline 4

An oxygen supply failure warning device must be present

Failure

ANZCA Guideline 5

There must be one gas flow control knob for each gas

Knob

ANZCA Guideline 6

The oxygen flow control know must be larger and fluted for tactile ID

O2flow

ANZCA Guideline 7

vaporiser dials must turn anti-clockwise to turn on

Vaporisers

ANZCA Guideline 8

Must have a protected master switch to prevent inadverdant operation

Master switch

ANZCA Guideline 9

Must have a 10 sec time delay when turned off

Delay

ANZCA Guideline 10

Must have a high pressure relief / blow off valve

High pressure

Oxygen failure device requirements 1

Must activate automatically when oxygen pressure falls below a predetermined critical level

Critical

Oxygen failure device requirements 2

Must cut off supply of all gases other than oxygen to the common gas outlet

Cut