• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/197

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

197 Cards in this Set

  • Front
  • Back
Full O2 or AIR =
1900 psi (660 L)
Full N2O =
745 psi (1590 L)
Pipeline Supply
(Intermediate Pressure)
Hospital supply enters at approximately 50-55 psig
Diameter Index Safety System
Supply hoses connect the wall outlets or column supply to the back of the anesthesia machine. Connects to the anesthesia machine
3 Standards organizations which govern the handling of gas cylinders include the
Compressed Gas Association, the Department of Transportation, and The National Fire Protection Agency.
and cylinder specifications
CGA= governs
manufacture, filling, shipping, storing, and handling; also dictate the colors, markings, labels, contents, and valve design
writes standards regarding the transport
DOT= writes standards regarding the transport of compressed gases as well as manufacturing and testing specifications; regulate the amount of gas that may be contained in the cylinder and require the specific markings on the cylinder
Cylinder Colors Oxygen
kelly green
Cylinder Colors Nitrous oxide
dark royal blue
Cylinder Colors Air
bright yellow
Cylinder Colors Carbon dioxide
gray
Cylinder Colors Helium
brown
__________ will identify the contents, manufacturing information and the hazard class of the gas
The label
Markings are permanently engraved on the cylinder and registered
Bureau of Explosives
Hazard Identification Green diamond
Non-flammable gas
Hazard Identification Yellow diamond
Oxidizer
Hazard Identification Red diamond
Flammable gas
Cylinders are attached ________ with ________ via a__________ and secured________
Cylinders are attached to the hanger yoke with the valve via a pin index safety system and secured with the T-handle
Pin index system
was designed to prevent misconnection of medical gas cylinders
Cylinder Valve is, contains, and has...
Conduit for the cylinder contents to reach the machine
Contains a safety relief valve
Pin index safety system
Necessary to prevent rupture of the cylinder under extreme conditions of exposure
Safety Relief Valve
Safety Relief Valve
This device may be of 3x
A fusible plug
A frangible disc
A spring loaded valve
Fusible plug is
Metal alloy, melt at either a low range of 157-170F &/or a high range 208-220F
Cannot be reset
Frangible disc is
Copper, burst when a certain high pressure is exceeded; the burst pressure for the cylinder must exceed the service pressure by at least a factor of two
Cannot be reset
Spring loaded valve is
Increase in pressure in the cylinder will force the spring open, excess pressure released & valve will close
Hangar Yoke 5 points (components)
Color-coded & labeled for each medical gas
Two pin index posts that correspond to the pin index holes on the cylinder valve
Gasket located between the gas outlet on the cylinder valve & the gas inlet on the anesthesia machine
Secure T-handle into seat above the safety relief valve after pins are correctly engaged
Includes filter for entrapment of particulate matter
Prevents reverse flow of gases
Check Valves
Check Valves 3 points
Free floating & respond to pressure
Prevent flow of gas from one cylinder to another when a second cylinder is turned on
Not leak proof
ASTM allows leak rate of
200 ml/minute
Safe Handling of Cylinders
4 points
Stored upright in holders in cool, dry, clean, and well-ventilated rooms with protective cap on
Placed in a free standing container when being moved
Secured from blunt trauma when being used for patient transport
Do NOT leave standing alone
Prior to Placement of o2 cylinder
“Cracked” briefly prior to placement in hanger yoke to clear dust, oil, or debris from gas outlet
Full E cylinder of compressed oxygen contains
660 L at a pressure of 1900 psig
Full E cylinder of compressed nitrous oxide contains
1590 L at a pressure of 745 psig
Full E cylinder of compressed air contains
625 L at a pressure of 1900 psig
Volume of N2O in a full cylinder is mostly
liquid state (small amount in gaseous state)
Nitrous Oxide
As gas changes from liquid to gas, the pressure indicated on the gauge remains
remains relatively constant until all of the liquid has been converted to gas
Nitrous Oxide
Once gauge drops below 745 psig,
, the cylinder contains approximately 300 L N2O
The critical temperature
temperature at and above which vapor of the substance cannot be liquefied, no matter how much pressure is applied.
Bourdon Gauge
Cylinder gauge displays the pressure of gas in the cylinder in psig
Indicates volume of gas remaining in the cylinder
First Stage Regulator are preset
at the factory.
First Stage Regulator reduce
reduced to 45 psig.
ASTM dictates that each gas supplied under high pressure must have at least one pressure reducing regulator.
First Stage Regulator Proper functioning
is confirmed when an O2 flow of 2L/M is restored to 2L/M within 2 seconds after each operation of the O2 flush valve.
Safety relief valve purpose
will vent excessive pressure out of the regulator
First Stage Regulator and flow
the cylinder flow to a intermediate pressure (45 psig) the anesthesia machine will preferentially seek flow from the pipeline supply because it is entering at a slightly greater pressure (50-55 psig.)
First Stage Regulators Also called
Cylinder Pressure Regulator
Bulk gas supply should conform to
should conform to the National Fire Protection Agency Standards (NFPA – 99)
Pipeline Supply converted to a vapor
Multiple vaporizer coils are used to warm the liquid to start the process
Results in expansion of the liquid and conversion to a gas
Pipeline supply enters the hospital at
50-55 psig
Pipeline Supply 3 Safeguards
actuating switches, shut off valves, and high pressure relief valves
Pipeline connections to the anesthesia machine contain a
diameter indexed safety system
diameter indexed safety system
3 points
Help prevent incorrect connections
Fitting for each gas has a different diameter
Each fitting is also internally keyed so that the operator cannot make a loose connection with a slightly larger fitting
4 Hazards from the pipeline supply
Fire
Explosion
Cross connection
Cross contamination
Auxiliary Oxygen Flowmeter 2 points
1. Used to administer O2 during local, MAC, or regional anesthesia
2. The auxiliary O2 flowmeter is operational as long as there is O2 pressure supplied to the machine, it bypasses the master on-off switch.
Oxygen Flush Valve 3 points
1. Operational when machine has a source of O2 pressure.
2 Inactivated, the intermediate pressure is retained behind the ball valve spring.
3 Once activated, the pin pushes the ball valve against the spring & O2 flows to the machine outlet.
Oxygen Flush Valves ASTM standards 4 points
1. Should be permanently marked
2. Type 3 O2 flush valve
manually operated with self-closing construction.
3. Does not pass through the vaporizers.
4 No allowable leak rate
5.Flow rate 35-75 psig
6 Minimal back pressure to vaporizers (max of 100cmH2O)
Master ON-OFF Switch sends
oxygen to ventilator connector
Master ON-OFF Switch
4 alarms switches which
warn against hypoxia are connected to this switch Prevents the anesthetists from delivering a hypoxic mixture without also turning on alarms to warn against hypoxia.
Oxygen supply failure alarm
Oxygen analyzer
Breathing pressure
Minute volume
2 Components not connected to the ON-OFF switch:
Oxygen flush valve
Auxiliary oxygen flowmeter
Master ON-OFF Switch 2 Problems
May resemble ventilator switches
Fail to engage fully or migrate off
4 ventilator alarm ASTM Standards
Electronic alarms be high, medium, or low priority
Each alarm has different visual & audible indicators
Alarms automatically reset
Silence a high or medium priority alarm for 120 seconds
primary driving gas of the ventilator.
The oxygen in intermediate pressure portion of the machine acts
Drager Narkomed vent driving device
uses oxygen supply to drive a venturi device, which augments the driving gas for the ventilator with entrained air
ASTM Standards require the ventilator connector to have a
check valve.
Ventilator Power Inlet Connector includes
DISS for O2 & spring and ball valve mechanism that should be leak tight.
Second Stage Regulator
5 points
1 Used on Ohmeda machines for O2 & N2O
2 O2 – takes intermediate pressure of 50 psig to lower pressure of 16 psig
3 N2O - reduced to 26 psig
4 Occurs prior to gas entering the flowmeters
5 Offer consistent supply of gas & diminish bobbing of flowmeter floats
Drager machines do not use
second stage regulators, their machines’ internal components exist at an intermediate pressure.
Pressure Sensor Shut Off Valve Linked to
master ON-OFF switch
Pressure Sensor Shut Off Valve Senses
O2 pressure and will shut off N2O or any other gases if there is inadequate O2 pressure.
N2O and AIR supply dependent on
adequate O2 supply.
Pressure Sensor Shut Off Valve senses
pressure only
the last flowing gas if supply lost
Pressure Sensor Shut Off Valve
Leaves O2 as the last flowing gas if oxygen supply lost
If O2 supply pressure decreases
, an internal alarm is activated.
O2 Supply Failure Alarm is Connected to
the master ON-OFF switch.
ASTM standards require that whenever O2 supply pressure decreases below a manufacturer specified threshold
, a medium priority alarm should sound within 5 seconds.
. (Usually a pressure below 25 psig).
O2 Supply Failure Alarm
Ohmeda-O2
Ohmeda-O2 pressure will fill the canister & close the relief valve by pushing against the diaphragm & closing the thrust pin
. A ball bearing in the inlet check valve rests in equilibrium as the pressure in the canister is equal to the pressure inside the machine.
If the pressure inside the machine decreases due to loss of O2 supply, the higher pressure in the canister pushes the ball bearing to close the inlet check valve, the diaphragm relaxes in the relief valve & the gas from the canister flows up to a metal reed, causing a loud 7sec whistle.
Flowmeters oxygen recieve flow rate
intermediate pressure (Drager 48 psig) or a low pressure (Ohmeda 16-26 psig)
Most delicate component of machine
O2 flowmeter
Standard for O2 placement
Standard for O2 flowmeter to be place to the right of other flowmeters
constructed of Pyrex glass
Thorpe tubes,
Tubes are pin indexed
to specific medical gas
Ohmeda- The float (3)
The float is an aluminum triangular device
Should be read at the top
‘Stops’ provide mechanism for minimum O2 flow when the machine is turned on
Drager The float (3)
The float is a sapphire coated ball type
Should be read in the middle
Uses a resistor device to provide a minimum O2 flow
Annular Space
The space that exists between the float & the internal surface of the Thorpe tube
Annular space increases as the float rises higher in the tube, as it would in high flows.
o2 Laminar Flow occurs
Low flows, less than 3L/M
Gas flows from a region of higher pressure to lower pressure
Gas flowing next to the sides of the flowmeter
will have a decreased velocity d/t frictional resistance of the tube
Gas flowing in the middle of the flow tube will have the greatest
velocity b/c it encounters less resistance to flow.
Turbulent Flow
Increased flows, greater than 4 L/M
Gas molecules begin to swirl and convert to turbulent flow
Critical velocity-
the point when laminar flow changes to turbulent flow
At higher flow the flowtube resembles
resembles an orifice
At turbulent flow the primary determinant of flow is
density of the gas (Graham’s Law)
4 Factors Decrease Accuracy of the float
Dust or dirt accumulating on the float
Static electricity
Changes in temperature
Changes in atmospheric pressure
ASTM Standards Require one, visible flow-control system .......
for each medical gas adjacent to the gas flowmeter it controls
Flowmeter knob should be marked
with the name of the gas and color coded appropriately
O2 knob must be
fluted, larger & protrude out farther than the round, smaller N2O and AIR knobs
Knobs should be designed to prevent
dis assembly when they are rotated
When supplying gas at 50 psig, flowmeters should be accurate to within
+/- 10 ml/M
Leakrate should be no more than
5 ml/M
Electronic Flow Devices
electronic flowmeters, electronic flow measurement systems, electronic mass flow sensors

Replaces the Thorpe tube flowmeter, reportedly are more accurate & allow the capture of data to be sent to an automated record keeper
Proportioning System
Components to prevent the delivery of a hypoxic mixture
O2 & N2O are
interfaced either mechanically or pneumatically with an electronic alarm so that the minimum O2 concentration at the common gas outlet is not less than 25%
The ratio of N2O to O2
should be approximately 3.5:1
Proportioning System Ohmeda is called ___________.
How does it work
Link – 25
Mechanical linkage of the O2 & N2O flow control knobs so that if O2 is decreased, the linkage will also pull the O2 up.
The O2 & N2O flowmeters must be adjacent to enable this system
Proportioning System Drager
and how it works
Oxygen Ratio Monitor Controller
Resistors in the O2 flowmeter create a back pressure to an O2 chamber depressing a diaphragm connected to the N2O control valve
If inadequate O2 the N2O control valve will close & no N2O will flow to the flowmeter
Proportioning System Does not protect completely against
against a hypoxic mixture.
The proportioning system does not protect completely against a hypoxic mixture.
name 4 poss failures
If a gas other than O2 is present in the O2 pipeline
If the mechanical linkage or the resistors and diaphragm are defective, the ratio may not be maintained
A leak downstream of these devices can exist
If a 3rd gas, such as helium, nitrogen, or carbon dioxide is administered, O2 may be diluted.
Vaporization – involves
the process of changing a liquid into a vapor
Vaporizer def
a container that holds the liquid agent and the gas flow from the flowmeter will either bubble through it or flow over it
Majority of deaths related to
to anesthetic overdose
Majority of deaths occur by
Most related to cardiac arrest
Keenan & Boyan study
Inadequate ventilation & overdose of agent
Morbidity and Mortality
Cooper’s critical incident studies
2 parts
Unintentional turning off of vaporizer
Technical error leading to overdose
AIMS
Morbidity and Mortality
Failed vaporizer
Caplan closed claims
3 failures
Valve failure
Leak
Wrong dial turned on
Anesthetic agents
4 points
Stored as liquid but converted to vapor
Higher molecular weight
Higher density
Much more complex than anesthesia gases (N2O)
Vapor density- density values obtained by
dividing the MW by 22.4
Vapor density- is
Ratio of the density of a gas to the density of a reference gas at the same pressure and temperature
< 1, vapor will rise
> 1, vapor will sink
Vapor specific gravity-
if the density is compared to air
Chloroform, ethyl chloride, & Trichloroethylene (TCE)
4 points
Disappeared because of their toxicity
TCE is toxic to multiple organs
Cardiovascular side effects
Damage to liver & kidney
Methoxyflurane
2 points
Mediated by cytochrome P450- catalyzed metabolism to toxic metabolites
Nephrotoxicity related to fluoride
Enflurane
neg side effect
Caused seizures with hyperventilation
Desflurane
3 initial problems with des
Initially produced using a potentially explosive synthesis that used elemental fluorine
No vaporizer
Decreased potency
Sevoflurane
1pos 1neg
Potential for nephrotoxicity
Very little airway irritability
Volatile liquids-
change readily and rapidly into vapor
Volatile liquids are defined by their
vapor pressure
Vapor pressure is a measure of a
liquid’s ability to evaporate
Higher the vapor pressure, the more
the more volatile the liquid, more readily the liquid gives off vapor
Vapor pressure- def
the amount of pressure the vapor of a volatile liquid agent exerts on the walls of a closed container at 20C and 760mmHg atmospheric pressure
Vapor pressure Desflurane
664
Vapor pressure Ether
400
Vapor pressure halothane·
243
Vapor pressure Isoflurane
238
Vapor pressure Enflurane
175
Vapor pressure Sevo
160
Vapor pressure , Water
18
Volatile liquid inside the vaporizer reacts to changes
changes in temperature and pressure
When the pressure or temperature changes in vaporizer
it will make changes to maintain equilibrium
Vapor pressure does not change with
atmospheric pressure
Boiling point changes with
atmospheric pressure
Vapor pressure numerical value changes
changes with temperature
Atmospheric pressure changes how easily the vapor can flow but
but does not change the vapor pressure of the agent
Vapor pressure at higher altitudes
Easier flow
Boiling point=
the temperature at which the vapor pressure equals atmospheric pressure
Lower atmospheric pressure (vapor)
less gas molecules put pressure on liquid
Takes less heat to turn the liquid into a vapor
Vapor pressure changes with changes
in temperature
As temperature rises, the vapor pressure rises…due to
kinetic energy
Latent Heat of Vaporization
Heat required to change the liquid into a gas during the period of constant temperature
No heat =
no vaporization
Vaporizer- a device that is designed to
accurately enrich the oxygen/nitrous oxide/ air mixture with a precise percentage of anesthetic agent.
early Copper Kettles &Vernitrols
Concentration-calibrated
Use temperature of kettle and temperature of room to calculate the percentage delivered
Measured flow vaporizer- operator determines how much gas should be bubbled through the anesthetic liquid by means of a formula
Contemporary vaporizers have internal device for
temperature compensation or supply heat to the vaporizer
Vaporizers Keep a portion of the mixed gas in contact with the anesthetic agent until
(the gas is saturated)
until the gas has absorbed all of the agent vapor it can hold at the current pressure and temperature
Concentration dial sets a value that
that splits the gas flow as it enters the vaporizer, a thermostatically controlled valve changes how much of the gas flows through the vaporizer chamber to give the correct output concentration
Vaporizers Constructed of materials that
transfer heat quickly to keep all of the vaporizer parts at the same temperature
Variable bypass-flow over vaporizer
‘concentration calibrated’
Vaporizing chamber- reservoir where liquid anesthetic is stored
Area above the liquid will be saturated with vapor molecules
Carrier gas enters chamber, flow over liquid and becomes saturated with vapor
Then carries anesthetic out of vaporizer to the patient
Variable bypass-flow over vaporizer
Method for regulating the output
is determined by the amount of gas flow that bypasses the vaporizing chamber as compared to the amount of carrier gas that flows over the vaporizing chamber
vaporizer Bimetallic strip
Metal with one side brass & one side nickel
Temperature increases brass side expands causing strip to bend = allows more carrier gas flow to bypass the vaporizing chamber & less to enter it (less anesthetic vapor out)
Temperature decreases the strip is not bent & the amount of carrier gas that bypasses the vaporizing chamber is diminished = allows more anesthetic vapor to be picked up
Bellows expansion element
Expands & deflates based on the temperature inside the vaporizer
Temperature increases, the bellows expand & allows more carrier gas flow to bypass the vaporizing chamber (less out)
Temperature decreases, bellows deflate, decrease the carrier gas flow thru the bypass tract & increase flow to vaporizing chamber (more out)
Wicks placed in the vaporizer
increase efficiency
Increases the surface area for gas to flow around allows more liquid into the vaporizer
Vaporizer Typically hold
hold 125mL of anesthetic agent
Baffles placed in vaporizer
3 points
1increase efficiency
2Take carrier gas flow closer to the liquid where it can pick up the agent at the liquid-vapor surface
3Give the gas a path to flow and decrease the incidence of spilling with tipping of vaporizer
Physical characteristics of Desflurane
Boiling point and Vapor pressure
close to room temperature (23.5C)
(664mmHg)
To store Desflurane in a closed container at room temperature
temperature requires pressure.
Desflurane bottle pressurized up to
80 psig
Desflurane Vaporizer heats to
39C
Desflurane Vaporizer Pressurized to __________ to keep in liquid form
More controlled & able to store in a closed container
1500mmHg
Vaporizer cannot be turned on until heated to
39C
Bellows expansion element
Expands & deflates based on the temperature inside the vaporizer
Temperature increases, the bellows expand & allows more carrier gas flow to bypass the vaporizing chamber (less out)
Temperature decreases, bellows deflate, decrease the carrier gas flow thru the bypass tract & increase flow to vaporizing chamber (more out)
Wicks placed in the vaporizer
increase efficiency
Increases the surface area for gas to flow around allows more liquid into the vaporizer
Vaporizer Typically hold
hold 125mL of anesthetic agent
Baffles placed in vaporizer
3 points
1increase efficiency
2Take carrier gas flow closer to the liquid where it can pick up the agent at the liquid-vapor surface
3Give the gas a path to flow and decrease the incidence of spilling with tipping of vaporizer
Physical characteristics of Desflurane
Boiling point and Vapor pressure
close to room temperature (23.5C)
(664mmHg)
To store Desflurane in a closed container at room temperature
temperature requires pressure.
Desflurane bottle pressurized up to
80 psig
Desflurane Vaporizer heats to
39C
Desflurane Vaporizer Pressurized to __________ to keep in liquid form
More controlled & able to store in a closed container
1500mmHg
Vaporizer cannot be turned on until heated to
39C
Desflurane vaporizer (if not warm)
Carrier gas flow turned on, flow through the vaporizer & encounters flow restrictor
Pressure transducer & pressure monitor that measures the pressure of the carrier gas flow & output from the
from Desflurane chamber
Pressure regulating valve is controlled by input from the
the pressure monitoring system to adjust pressure d/t Desflurane vapor until there is no difference between the Desflurane vapor & the pressure of the carrier gas flow
Desflurane vaporizer dial regulates the
actual quantity of Desflurane vapor available for mixing with the carrier gas flow
Desflurane vaporizer
Increasing concentration dial opens the rotary valve more, permits more
more Desflurane to enter the carrier gas flow & increases the percent output concentration
There is a zero pressure differential at the transducer system, the percent concentration is determined by the ratio of flow restrictions that are imposed by the fixed carrier gas flow restrictor & the concentration dial
Vaporizer Safety Features 2
Prevent possible overdose of anesthetic agents
Baffles to decrease chance agent will overflow
Vaporizer If tipped, what should you do
vaporizer must be flushed out using high flowrates (5-10 L/M) to dry the liquid from the outflow tract; concentration dial set to 4-5%; & flushed for 20 minutes
Vaporizer Interlock System
Interlock system to assure that only one vaporizer is turned on to prevent the contents of one vaporizer from contaminating the next
Integrity confirmed during the check out of the machine
Vaporizer Safety Features
Ohmeda
depends on the extension rods of the vaporizers being in contact with one another; if one turned on, the extension rods move in a manner that prevents any other vaporizer from being turned on
Vaporizer Safety Features
Drager system
is a set of metal rods that change conformation when one of the vaporizers is turned on
It can fail if any of the screws get loosened
Vaporizer cannot be turned on unless
unless it is properly locked in, this prevents major leaks from developing at the inlet and outlet of the vaporizer
To turn vaporizer ON, must
must push the ON button in and turn the dial
ASTM standards from 1989 for vaporizers include the following:
Location in the fresh gas circuit & concentration calibrated
Accept a total gas flow up to 15 L/M
Incorporate an isolation or interlock system
Fittings to demarcate correct flow through
Liquid level indicator
Agent specific filling device
Anesthetic concentration < .1% when OFF
Operator manual should state variation of ambient temperature, pressure, back pressure, input flowrates effect on vaporizer performance, accuracy of calibration and service intervals
Vaporizer Outlet check valve
Method to diminish the intermittent back pressure that may increase or decrease the percent concentration output of vaporizers
Maybe included in the vaporizer or as a separate component of the anesthesia delivery system
Outlet check valve
Drager-
Drager- employs a coil system inside the vaporizer; when intermittent back flow occurs, the gas does flow back into the vaporizer but encounters this coil and cannot reach the vaporizer chamber
Ohmeda vaporizers have a check valve
valve at the outflow tract or incorporate an outlet check valve distal to the vaporizers, but prior to the common gas outlet
Ohmeda-Negative pressure bulb must be utilized to
valve and evaluate if leaks exist behind it.
Vaporizer Pressure relief valve
Prior to the medical gases and anesthetic agents exiting the machine at the common gas outlet, there is a valve designed to open in the event of high pressure
Pressure relief valve Drager-
Drager- employs this inside the vaporizer and it is designed to crack at 18 psig
Ohmeda- Pressure relief valve
has a separate component in the piping system designed to open between 2.3-2.9 psig or at 5 psig depending on the particular machine
Common gas outlet
This must be patent and intact to deliver gas to the patient