Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
197 Cards in this Set
- Front
- Back
Full O2 or AIR =
|
1900 psi (660 L)
|
|
Full N2O =
|
745 psi (1590 L)
|
|
Pipeline Supply
|
(Intermediate Pressure)
Hospital supply enters at approximately 50-55 psig |
|
Diameter Index Safety System
|
Supply hoses connect the wall outlets or column supply to the back of the anesthesia machine. Connects to the anesthesia machine
|
|
3 Standards organizations which govern the handling of gas cylinders include the
|
Compressed Gas Association, the Department of Transportation, and The National Fire Protection Agency.
and cylinder specifications |
|
CGA= governs
|
manufacture, filling, shipping, storing, and handling; also dictate the colors, markings, labels, contents, and valve design
|
|
writes standards regarding the transport
|
DOT= writes standards regarding the transport of compressed gases as well as manufacturing and testing specifications; regulate the amount of gas that may be contained in the cylinder and require the specific markings on the cylinder
|
|
Cylinder Colors Oxygen
|
kelly green
|
|
Cylinder Colors Nitrous oxide
|
dark royal blue
|
|
Cylinder Colors Air
|
bright yellow
|
|
Cylinder Colors Carbon dioxide
|
gray
|
|
Cylinder Colors Helium
|
brown
|
|
__________ will identify the contents, manufacturing information and the hazard class of the gas
|
The label
|
|
Markings are permanently engraved on the cylinder and registered
|
Bureau of Explosives
|
|
Hazard Identification Green diamond
|
Non-flammable gas
|
|
Hazard Identification Yellow diamond
|
Oxidizer
|
|
Hazard Identification Red diamond
|
Flammable gas
|
|
Cylinders are attached ________ with ________ via a__________ and secured________
|
Cylinders are attached to the hanger yoke with the valve via a pin index safety system and secured with the T-handle
|
|
Pin index system
|
was designed to prevent misconnection of medical gas cylinders
|
|
Cylinder Valve is, contains, and has...
|
Conduit for the cylinder contents to reach the machine
Contains a safety relief valve Pin index safety system |
|
Necessary to prevent rupture of the cylinder under extreme conditions of exposure
|
Safety Relief Valve
|
|
Safety Relief Valve
This device may be of 3x |
A fusible plug
A frangible disc A spring loaded valve |
|
Fusible plug is
|
Metal alloy, melt at either a low range of 157-170F &/or a high range 208-220F
Cannot be reset |
|
Frangible disc is
|
Copper, burst when a certain high pressure is exceeded; the burst pressure for the cylinder must exceed the service pressure by at least a factor of two
Cannot be reset |
|
Spring loaded valve is
|
Increase in pressure in the cylinder will force the spring open, excess pressure released & valve will close
|
|
Hangar Yoke 5 points (components)
|
Color-coded & labeled for each medical gas
Two pin index posts that correspond to the pin index holes on the cylinder valve Gasket located between the gas outlet on the cylinder valve & the gas inlet on the anesthesia machine Secure T-handle into seat above the safety relief valve after pins are correctly engaged Includes filter for entrapment of particulate matter |
|
Prevents reverse flow of gases
|
Check Valves
|
|
Check Valves 3 points
|
Free floating & respond to pressure
Prevent flow of gas from one cylinder to another when a second cylinder is turned on Not leak proof |
|
ASTM allows leak rate of
|
200 ml/minute
|
|
Safe Handling of Cylinders
4 points |
Stored upright in holders in cool, dry, clean, and well-ventilated rooms with protective cap on
Placed in a free standing container when being moved Secured from blunt trauma when being used for patient transport Do NOT leave standing alone |
|
Prior to Placement of o2 cylinder
|
“Cracked” briefly prior to placement in hanger yoke to clear dust, oil, or debris from gas outlet
|
|
Full E cylinder of compressed oxygen contains
|
660 L at a pressure of 1900 psig
|
|
Full E cylinder of compressed nitrous oxide contains
|
1590 L at a pressure of 745 psig
|
|
Full E cylinder of compressed air contains
|
625 L at a pressure of 1900 psig
|
|
Volume of N2O in a full cylinder is mostly
|
liquid state (small amount in gaseous state)
|
|
Nitrous Oxide
As gas changes from liquid to gas, the pressure indicated on the gauge remains |
remains relatively constant until all of the liquid has been converted to gas
|
|
Nitrous Oxide
Once gauge drops below 745 psig, |
, the cylinder contains approximately 300 L N2O
|
|
The critical temperature
|
temperature at and above which vapor of the substance cannot be liquefied, no matter how much pressure is applied.
|
|
Bourdon Gauge
|
Cylinder gauge displays the pressure of gas in the cylinder in psig
Indicates volume of gas remaining in the cylinder |
|
First Stage Regulator are preset
|
at the factory.
|
|
First Stage Regulator reduce
|
reduced to 45 psig.
ASTM dictates that each gas supplied under high pressure must have at least one pressure reducing regulator. |
|
First Stage Regulator Proper functioning
|
is confirmed when an O2 flow of 2L/M is restored to 2L/M within 2 seconds after each operation of the O2 flush valve.
|
|
Safety relief valve purpose
|
will vent excessive pressure out of the regulator
|
|
First Stage Regulator and flow
|
the cylinder flow to a intermediate pressure (45 psig) the anesthesia machine will preferentially seek flow from the pipeline supply because it is entering at a slightly greater pressure (50-55 psig.)
|
|
First Stage Regulators Also called
|
Cylinder Pressure Regulator
|
|
Bulk gas supply should conform to
|
should conform to the National Fire Protection Agency Standards (NFPA – 99)
|
|
Pipeline Supply converted to a vapor
|
Multiple vaporizer coils are used to warm the liquid to start the process
Results in expansion of the liquid and conversion to a gas |
|
Pipeline supply enters the hospital at
|
50-55 psig
|
|
Pipeline Supply 3 Safeguards
|
actuating switches, shut off valves, and high pressure relief valves
|
|
Pipeline connections to the anesthesia machine contain a
|
diameter indexed safety system
|
|
diameter indexed safety system
3 points |
Help prevent incorrect connections
Fitting for each gas has a different diameter Each fitting is also internally keyed so that the operator cannot make a loose connection with a slightly larger fitting |
|
4 Hazards from the pipeline supply
|
Fire
Explosion Cross connection Cross contamination |
|
Auxiliary Oxygen Flowmeter 2 points
|
1. Used to administer O2 during local, MAC, or regional anesthesia
2. The auxiliary O2 flowmeter is operational as long as there is O2 pressure supplied to the machine, it bypasses the master on-off switch. |
|
Oxygen Flush Valve 3 points
|
1. Operational when machine has a source of O2 pressure.
2 Inactivated, the intermediate pressure is retained behind the ball valve spring. 3 Once activated, the pin pushes the ball valve against the spring & O2 flows to the machine outlet. |
|
Oxygen Flush Valves ASTM standards 4 points
|
1. Should be permanently marked
2. Type 3 O2 flush valve manually operated with self-closing construction. 3. Does not pass through the vaporizers. 4 No allowable leak rate 5.Flow rate 35-75 psig 6 Minimal back pressure to vaporizers (max of 100cmH2O) |
|
Master ON-OFF Switch sends
|
oxygen to ventilator connector
|
|
Master ON-OFF Switch
4 alarms switches which warn against hypoxia are connected to this switch Prevents the anesthetists from delivering a hypoxic mixture without also turning on alarms to warn against hypoxia. |
Oxygen supply failure alarm
Oxygen analyzer Breathing pressure Minute volume |
|
2 Components not connected to the ON-OFF switch:
|
Oxygen flush valve
Auxiliary oxygen flowmeter |
|
Master ON-OFF Switch 2 Problems
|
May resemble ventilator switches
Fail to engage fully or migrate off |
|
4 ventilator alarm ASTM Standards
|
Electronic alarms be high, medium, or low priority
Each alarm has different visual & audible indicators Alarms automatically reset Silence a high or medium priority alarm for 120 seconds |
|
primary driving gas of the ventilator.
|
The oxygen in intermediate pressure portion of the machine acts
|
|
Drager Narkomed vent driving device
|
uses oxygen supply to drive a venturi device, which augments the driving gas for the ventilator with entrained air
|
|
ASTM Standards require the ventilator connector to have a
|
check valve.
|
|
Ventilator Power Inlet Connector includes
|
DISS for O2 & spring and ball valve mechanism that should be leak tight.
|
|
Second Stage Regulator
5 points |
1 Used on Ohmeda machines for O2 & N2O
2 O2 – takes intermediate pressure of 50 psig to lower pressure of 16 psig 3 N2O - reduced to 26 psig 4 Occurs prior to gas entering the flowmeters 5 Offer consistent supply of gas & diminish bobbing of flowmeter floats |
|
Drager machines do not use
|
second stage regulators, their machines’ internal components exist at an intermediate pressure.
|
|
Pressure Sensor Shut Off Valve Linked to
|
master ON-OFF switch
|
|
Pressure Sensor Shut Off Valve Senses
|
O2 pressure and will shut off N2O or any other gases if there is inadequate O2 pressure.
|
|
N2O and AIR supply dependent on
|
adequate O2 supply.
|
|
Pressure Sensor Shut Off Valve senses
|
pressure only
|
|
the last flowing gas if supply lost
|
Pressure Sensor Shut Off Valve
Leaves O2 as the last flowing gas if oxygen supply lost |
|
If O2 supply pressure decreases
|
, an internal alarm is activated.
|
|
O2 Supply Failure Alarm is Connected to
|
the master ON-OFF switch.
|
|
ASTM standards require that whenever O2 supply pressure decreases below a manufacturer specified threshold
|
, a medium priority alarm should sound within 5 seconds.
. (Usually a pressure below 25 psig). |
|
O2 Supply Failure Alarm
Ohmeda-O2 |
Ohmeda-O2 pressure will fill the canister & close the relief valve by pushing against the diaphragm & closing the thrust pin
. A ball bearing in the inlet check valve rests in equilibrium as the pressure in the canister is equal to the pressure inside the machine. If the pressure inside the machine decreases due to loss of O2 supply, the higher pressure in the canister pushes the ball bearing to close the inlet check valve, the diaphragm relaxes in the relief valve & the gas from the canister flows up to a metal reed, causing a loud 7sec whistle. |
|
Flowmeters oxygen recieve flow rate
|
intermediate pressure (Drager 48 psig) or a low pressure (Ohmeda 16-26 psig)
|
|
Most delicate component of machine
|
O2 flowmeter
|
|
Standard for O2 placement
|
Standard for O2 flowmeter to be place to the right of other flowmeters
|
|
constructed of Pyrex glass
|
Thorpe tubes,
|
|
Tubes are pin indexed
|
to specific medical gas
|
|
Ohmeda- The float (3)
|
The float is an aluminum triangular device
Should be read at the top ‘Stops’ provide mechanism for minimum O2 flow when the machine is turned on |
|
Drager The float (3)
|
The float is a sapphire coated ball type
Should be read in the middle Uses a resistor device to provide a minimum O2 flow |
|
Annular Space
|
The space that exists between the float & the internal surface of the Thorpe tube
Annular space increases as the float rises higher in the tube, as it would in high flows. |
|
o2 Laminar Flow occurs
|
Low flows, less than 3L/M
Gas flows from a region of higher pressure to lower pressure |
|
Gas flowing next to the sides of the flowmeter
|
will have a decreased velocity d/t frictional resistance of the tube
|
|
Gas flowing in the middle of the flow tube will have the greatest
|
velocity b/c it encounters less resistance to flow.
|
|
Turbulent Flow
|
Increased flows, greater than 4 L/M
Gas molecules begin to swirl and convert to turbulent flow |
|
Critical velocity-
|
the point when laminar flow changes to turbulent flow
|
|
At higher flow the flowtube resembles
|
resembles an orifice
|
|
At turbulent flow the primary determinant of flow is
|
density of the gas (Graham’s Law)
|
|
4 Factors Decrease Accuracy of the float
|
Dust or dirt accumulating on the float
Static electricity Changes in temperature Changes in atmospheric pressure |
|
ASTM Standards Require one, visible flow-control system .......
|
for each medical gas adjacent to the gas flowmeter it controls
|
|
Flowmeter knob should be marked
|
with the name of the gas and color coded appropriately
|
|
O2 knob must be
|
fluted, larger & protrude out farther than the round, smaller N2O and AIR knobs
|
|
Knobs should be designed to prevent
|
dis assembly when they are rotated
|
|
When supplying gas at 50 psig, flowmeters should be accurate to within
|
+/- 10 ml/M
|
|
Leakrate should be no more than
|
5 ml/M
|
|
Electronic Flow Devices
|
electronic flowmeters, electronic flow measurement systems, electronic mass flow sensors
Replaces the Thorpe tube flowmeter, reportedly are more accurate & allow the capture of data to be sent to an automated record keeper |
|
Proportioning System
|
Components to prevent the delivery of a hypoxic mixture
|
|
O2 & N2O are
|
interfaced either mechanically or pneumatically with an electronic alarm so that the minimum O2 concentration at the common gas outlet is not less than 25%
|
|
The ratio of N2O to O2
|
should be approximately 3.5:1
|
|
Proportioning System Ohmeda is called ___________.
How does it work |
Link – 25
Mechanical linkage of the O2 & N2O flow control knobs so that if O2 is decreased, the linkage will also pull the O2 up. The O2 & N2O flowmeters must be adjacent to enable this system |
|
Proportioning System Drager
and how it works |
Oxygen Ratio Monitor Controller
Resistors in the O2 flowmeter create a back pressure to an O2 chamber depressing a diaphragm connected to the N2O control valve If inadequate O2 the N2O control valve will close & no N2O will flow to the flowmeter |
|
Proportioning System Does not protect completely against
|
against a hypoxic mixture.
|
|
The proportioning system does not protect completely against a hypoxic mixture.
name 4 poss failures |
If a gas other than O2 is present in the O2 pipeline
If the mechanical linkage or the resistors and diaphragm are defective, the ratio may not be maintained A leak downstream of these devices can exist If a 3rd gas, such as helium, nitrogen, or carbon dioxide is administered, O2 may be diluted. |
|
Vaporization – involves
|
the process of changing a liquid into a vapor
|
|
Vaporizer def
|
a container that holds the liquid agent and the gas flow from the flowmeter will either bubble through it or flow over it
|
|
Majority of deaths related to
|
to anesthetic overdose
|
|
Majority of deaths occur by
|
Most related to cardiac arrest
|
|
Keenan & Boyan study
|
Inadequate ventilation & overdose of agent
Morbidity and Mortality |
|
Cooper’s critical incident studies
2 parts |
Unintentional turning off of vaporizer
Technical error leading to overdose |
|
AIMS
Morbidity and Mortality |
Failed vaporizer
|
|
Caplan closed claims
3 failures |
Valve failure
Leak Wrong dial turned on |
|
Anesthetic agents
4 points |
Stored as liquid but converted to vapor
Higher molecular weight Higher density Much more complex than anesthesia gases (N2O) |
|
Vapor density- density values obtained by
|
dividing the MW by 22.4
|
|
Vapor density- is
|
Ratio of the density of a gas to the density of a reference gas at the same pressure and temperature
< 1, vapor will rise > 1, vapor will sink |
|
Vapor specific gravity-
|
if the density is compared to air
|
|
Chloroform, ethyl chloride, & Trichloroethylene (TCE)
4 points |
Disappeared because of their toxicity
TCE is toxic to multiple organs Cardiovascular side effects Damage to liver & kidney |
|
Methoxyflurane
2 points |
Mediated by cytochrome P450- catalyzed metabolism to toxic metabolites
Nephrotoxicity related to fluoride |
|
Enflurane
neg side effect |
Caused seizures with hyperventilation
|
|
Desflurane
3 initial problems with des |
Initially produced using a potentially explosive synthesis that used elemental fluorine
No vaporizer Decreased potency |
|
Sevoflurane
1pos 1neg |
Potential for nephrotoxicity
Very little airway irritability |
|
Volatile liquids-
|
change readily and rapidly into vapor
|
|
Volatile liquids are defined by their
|
vapor pressure
|
|
Vapor pressure is a measure of a
|
liquid’s ability to evaporate
|
|
Higher the vapor pressure, the more
|
the more volatile the liquid, more readily the liquid gives off vapor
|
|
Vapor pressure- def
|
the amount of pressure the vapor of a volatile liquid agent exerts on the walls of a closed container at 20C and 760mmHg atmospheric pressure
|
|
Vapor pressure Desflurane
|
664
|
|
Vapor pressure Ether
|
400
|
|
Vapor pressure halothane·
|
243
|
|
Vapor pressure Isoflurane
|
238
|
|
Vapor pressure Enflurane
|
175
|
|
Vapor pressure Sevo
|
160
|
|
Vapor pressure , Water
|
18
|
|
Volatile liquid inside the vaporizer reacts to changes
|
changes in temperature and pressure
|
|
When the pressure or temperature changes in vaporizer
|
it will make changes to maintain equilibrium
|
|
Vapor pressure does not change with
|
atmospheric pressure
|
|
Boiling point changes with
|
atmospheric pressure
|
|
Vapor pressure numerical value changes
|
changes with temperature
|
|
Atmospheric pressure changes how easily the vapor can flow but
|
but does not change the vapor pressure of the agent
|
|
Vapor pressure at higher altitudes
|
Easier flow
|
|
Boiling point=
|
the temperature at which the vapor pressure equals atmospheric pressure
|
|
Lower atmospheric pressure (vapor)
|
less gas molecules put pressure on liquid
Takes less heat to turn the liquid into a vapor |
|
Vapor pressure changes with changes
|
in temperature
|
|
As temperature rises, the vapor pressure rises…due to
|
kinetic energy
|
|
Latent Heat of Vaporization
|
Heat required to change the liquid into a gas during the period of constant temperature
|
|
No heat =
|
no vaporization
|
|
Vaporizer- a device that is designed to
|
accurately enrich the oxygen/nitrous oxide/ air mixture with a precise percentage of anesthetic agent.
|
|
early Copper Kettles &Vernitrols
Concentration-calibrated |
Use temperature of kettle and temperature of room to calculate the percentage delivered
Measured flow vaporizer- operator determines how much gas should be bubbled through the anesthetic liquid by means of a formula |
|
Contemporary vaporizers have internal device for
|
temperature compensation or supply heat to the vaporizer
|
|
Vaporizers Keep a portion of the mixed gas in contact with the anesthetic agent until
|
(the gas is saturated)
until the gas has absorbed all of the agent vapor it can hold at the current pressure and temperature |
|
Concentration dial sets a value that
|
that splits the gas flow as it enters the vaporizer, a thermostatically controlled valve changes how much of the gas flows through the vaporizer chamber to give the correct output concentration
|
|
Vaporizers Constructed of materials that
|
transfer heat quickly to keep all of the vaporizer parts at the same temperature
|
|
Variable bypass-flow over vaporizer
‘concentration calibrated’ |
Vaporizing chamber- reservoir where liquid anesthetic is stored
Area above the liquid will be saturated with vapor molecules Carrier gas enters chamber, flow over liquid and becomes saturated with vapor Then carries anesthetic out of vaporizer to the patient |
|
Variable bypass-flow over vaporizer
Method for regulating the output |
is determined by the amount of gas flow that bypasses the vaporizing chamber as compared to the amount of carrier gas that flows over the vaporizing chamber
|
|
vaporizer Bimetallic strip
|
Metal with one side brass & one side nickel
Temperature increases brass side expands causing strip to bend = allows more carrier gas flow to bypass the vaporizing chamber & less to enter it (less anesthetic vapor out) Temperature decreases the strip is not bent & the amount of carrier gas that bypasses the vaporizing chamber is diminished = allows more anesthetic vapor to be picked up |
|
Bellows expansion element
|
Expands & deflates based on the temperature inside the vaporizer
Temperature increases, the bellows expand & allows more carrier gas flow to bypass the vaporizing chamber (less out) Temperature decreases, bellows deflate, decrease the carrier gas flow thru the bypass tract & increase flow to vaporizing chamber (more out) |
|
Wicks placed in the vaporizer
|
increase efficiency
Increases the surface area for gas to flow around allows more liquid into the vaporizer |
|
Vaporizer Typically hold
|
hold 125mL of anesthetic agent
|
|
Baffles placed in vaporizer
3 points |
1increase efficiency
2Take carrier gas flow closer to the liquid where it can pick up the agent at the liquid-vapor surface 3Give the gas a path to flow and decrease the incidence of spilling with tipping of vaporizer |
|
Physical characteristics of Desflurane
Boiling point and Vapor pressure |
close to room temperature (23.5C)
(664mmHg) |
|
To store Desflurane in a closed container at room temperature
|
temperature requires pressure.
|
|
Desflurane bottle pressurized up to
|
80 psig
|
|
Desflurane Vaporizer heats to
|
39C
|
|
Desflurane Vaporizer Pressurized to __________ to keep in liquid form
More controlled & able to store in a closed container |
1500mmHg
|
|
Vaporizer cannot be turned on until heated to
|
39C
|
|
Bellows expansion element
|
Expands & deflates based on the temperature inside the vaporizer
Temperature increases, the bellows expand & allows more carrier gas flow to bypass the vaporizing chamber (less out) Temperature decreases, bellows deflate, decrease the carrier gas flow thru the bypass tract & increase flow to vaporizing chamber (more out) |
|
Wicks placed in the vaporizer
|
increase efficiency
Increases the surface area for gas to flow around allows more liquid into the vaporizer |
|
Vaporizer Typically hold
|
hold 125mL of anesthetic agent
|
|
Baffles placed in vaporizer
3 points |
1increase efficiency
2Take carrier gas flow closer to the liquid where it can pick up the agent at the liquid-vapor surface 3Give the gas a path to flow and decrease the incidence of spilling with tipping of vaporizer |
|
Physical characteristics of Desflurane
Boiling point and Vapor pressure |
close to room temperature (23.5C)
(664mmHg) |
|
To store Desflurane in a closed container at room temperature
|
temperature requires pressure.
|
|
Desflurane bottle pressurized up to
|
80 psig
|
|
Desflurane Vaporizer heats to
|
39C
|
|
Desflurane Vaporizer Pressurized to __________ to keep in liquid form
More controlled & able to store in a closed container |
1500mmHg
|
|
Vaporizer cannot be turned on until heated to
|
39C
|
|
Desflurane vaporizer (if not warm)
|
Carrier gas flow turned on, flow through the vaporizer & encounters flow restrictor
|
|
Pressure transducer & pressure monitor that measures the pressure of the carrier gas flow & output from the
|
from Desflurane chamber
|
|
Pressure regulating valve is controlled by input from the
|
the pressure monitoring system to adjust pressure d/t Desflurane vapor until there is no difference between the Desflurane vapor & the pressure of the carrier gas flow
|
|
Desflurane vaporizer dial regulates the
|
actual quantity of Desflurane vapor available for mixing with the carrier gas flow
|
|
Desflurane vaporizer
Increasing concentration dial opens the rotary valve more, permits more |
more Desflurane to enter the carrier gas flow & increases the percent output concentration
There is a zero pressure differential at the transducer system, the percent concentration is determined by the ratio of flow restrictions that are imposed by the fixed carrier gas flow restrictor & the concentration dial |
|
Vaporizer Safety Features 2
|
Prevent possible overdose of anesthetic agents
Baffles to decrease chance agent will overflow |
|
Vaporizer If tipped, what should you do
|
vaporizer must be flushed out using high flowrates (5-10 L/M) to dry the liquid from the outflow tract; concentration dial set to 4-5%; & flushed for 20 minutes
|
|
Vaporizer Interlock System
|
Interlock system to assure that only one vaporizer is turned on to prevent the contents of one vaporizer from contaminating the next
Integrity confirmed during the check out of the machine |
|
Vaporizer Safety Features
Ohmeda |
depends on the extension rods of the vaporizers being in contact with one another; if one turned on, the extension rods move in a manner that prevents any other vaporizer from being turned on
|
|
Vaporizer Safety Features
Drager system |
is a set of metal rods that change conformation when one of the vaporizers is turned on
It can fail if any of the screws get loosened |
|
Vaporizer cannot be turned on unless
|
unless it is properly locked in, this prevents major leaks from developing at the inlet and outlet of the vaporizer
|
|
To turn vaporizer ON, must
|
must push the ON button in and turn the dial
|
|
ASTM standards from 1989 for vaporizers include the following:
|
Location in the fresh gas circuit & concentration calibrated
Accept a total gas flow up to 15 L/M Incorporate an isolation or interlock system Fittings to demarcate correct flow through Liquid level indicator Agent specific filling device Anesthetic concentration < .1% when OFF Operator manual should state variation of ambient temperature, pressure, back pressure, input flowrates effect on vaporizer performance, accuracy of calibration and service intervals |
|
Vaporizer Outlet check valve
|
Method to diminish the intermittent back pressure that may increase or decrease the percent concentration output of vaporizers
Maybe included in the vaporizer or as a separate component of the anesthesia delivery system |
|
Outlet check valve
Drager- |
Drager- employs a coil system inside the vaporizer; when intermittent back flow occurs, the gas does flow back into the vaporizer but encounters this coil and cannot reach the vaporizer chamber
|
|
Ohmeda vaporizers have a check valve
|
valve at the outflow tract or incorporate an outlet check valve distal to the vaporizers, but prior to the common gas outlet
|
|
Ohmeda-Negative pressure bulb must be utilized to
|
valve and evaluate if leaks exist behind it.
|
|
Vaporizer Pressure relief valve
|
Prior to the medical gases and anesthetic agents exiting the machine at the common gas outlet, there is a valve designed to open in the event of high pressure
|
|
Pressure relief valve Drager-
|
Drager- employs this inside the vaporizer and it is designed to crack at 18 psig
|
|
Ohmeda- Pressure relief valve
|
has a separate component in the piping system designed to open between 2.3-2.9 psig or at 5 psig depending on the particular machine
|
|
Common gas outlet
|
This must be patent and intact to deliver gas to the patient
|