Dipole-Dipole Array Analysis

Improved Essays
Dipole-dipole array has wide using in resistivity/I.P. surveys because of the low E.M. coupling between the current and potential circuits. The arrangement of the electrodes is shown in Figure( ).,
Electrodes c1 and c2 are separated by a distance (a), and electrodes p1 and p2 are separated by a distance (b). the distance between the potential electrodes pair c1-c2 is the same as the distance between the potential electrodes pair P1-P2 and the distance between the centers of the respective pairs is (n+1) a, the “a” spacing is initially still fixed and the “n” factor is increased from 1 to 2 to 3 until 6 in order to increase the depth of investigation. the apparent resistivity determined by this arrangement is: pa = πa n(n+l)(n+2) . ( V/I)
…show more content…
The depth of this array depends on the “n” factor and the “a” factor . Dipole-dipole array has a shallower depth compared with the Wenner array. But , this array has a better horizontal data coverage than the Wenner (Figure ). One possible disadvantage of this array is the very small signal strength for large values of the “n” factor. The voltage is inversely proportional to the cube of the “n” factor. This means that for the same current, the voltage measured by the resistivity meter drops by about 200 times when “n” is increased from 1 to 6. to overcome this problem , we need to increase the “a” spacing between the C1-C2 and P1-P2 dipole pair to reduce the drop in the potential when the overall length of the array is increased to increase the depth of investigation. Figure 9 shows two different arrangements for the dipole-dipole array with the same array length but with different “a” and “n” factors. The signal strength of the array with the smaller “n” factor (Figure 9b) is about 28 times stronger than the one with the larger “n” …show more content…
With the proper field equipment and survey techniques, this array has been successfully used in many areas to detect structures such as cavities where the good horizontal resolution of this array is a major advantage. The plotting location of the corresponding datum point (based on the median depth of investigation) used in drawing the apparent resistivity pseudosection is also shown in Figure 8c. Note that the pseudosection plotting point falls in an area with very low sensitivity values. For the dipole-dipole array, the regions with the high sensitivity values are concentrated below the C1-C2 electrodes pair and below the P1-P2 electrodes pair. In effect, the dipoledipole array gives minimal information about the resistivity of the region surrounding the plotting point, and the distribution of the data points in the pseudosection plot does not reflect the subsurface area mapped by the apparent resistivity measurements. Note that if the

Related Documents

  • Improved Essays

    Nt1320 Unit 4 Analysis

    • 853 Words
    • 4 Pages

    There is nearly a linear correlation between conductivity and concentration of salt or other specific ions. In theory, the conductivity probe is measuring the concentration of those specific ions and their ability to produce an electric current between two electrodes. As the concentration of a specific ion increases so does conductivity. For most experiments, calibrating the conductivity probe is not required. Before using the probe set the range based on µS/cm.…

    • 853 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Nt1310 Unit 4 Lab Report

    • 435 Words
    • 2 Pages

    Finally, when measuring resistance it is a clear choice once again. The digital meter gave a much closer measurement to the expected value due to a better scale, no risk of parallax, and being significantly easier to…

    • 435 Words
    • 2 Pages
    Improved Essays
  • Superior Essays

    Nt1310 Unit 4

    • 1249 Words
    • 5 Pages

    The first part of the experiment was determining the resolution of the DAC in volts. Before a voltage was applied the noise was determined by measuring the DC offset, the reading of the DMM at an input of zero volts. This offset was removed from the recorded values in order to provide a more accurate reading of the resolution. These corrected values and their corresponding applied voltages can be found in Table 1 of the appendix. The lowest value recorded from the DMM was -9.968V and the highest value recorded was +9.968V making the range of the DMM about 19.94±0.01V. Using the equation for calculating resolution, the resolution was calculated for values of N from 10 to 18 in increments of 2 (recorded in Table 2).…

    • 1249 Words
    • 5 Pages
    Superior Essays
  • Improved Essays

    type of signals, as EEG, ECOG and EAP, that can be capture from recording blocks. All these signals have their particular properties. This particular work focuses on the recording of extracellular action potentials (EAPs). NRS usually implanted inside the brain, very near to the cortex part so that they can be close proximity of the neurons. Additional advantage for doing this is that it prevents the signal for catching noise and interferences from other…

    • 75 Words
    • 1 Pages
    Improved Essays
  • Improved Essays

    Unknown Lab Report

    • 750 Words
    • 3 Pages

    Conductivity was then tested with the solutions remaining in the beakers. A Go! Link by Vernier Software & Technology was utilized to connect the conductivity probe to Logger Pro in order to view real-time conductivity values. The electrode (epoxy-body graphite) end of the probe was first neutralized in a beaker of distilled water, and then placed in the solution of the unknown. When the unknown’s conductivity was measured, the probe was again placed in distilled water, and once again repeated with ammonium chloride.…

    • 750 Words
    • 3 Pages
    Improved Essays
  • Decent Essays

    Swarm Formal Lab Report

    • 362 Words
    • 2 Pages

    To start off with this experiment, we drew dot structures for each compound and decided whether the compound was polar, nonpolar, or ionic. We found that KCl and CaCl*2H2O were ionic compounds, so we used the electrical conductivity tester to insure us that they indeed conducted electricity. After testing these two compounds, we decided that KCl was the solution labeled “Wolves” and CaCl*2H2O was the solution labeled “Swarm”. We decided this because we knew that the CaCl*2H2O compound was a hydrate, so the compound had to have some kind of crystal-like structures within the compound and so therefore, we knew that it would have to be the solution called “Swarm”. We then took the remaining solutions that were in a liquid state, and mixed them…

    • 362 Words
    • 2 Pages
    Decent Essays
  • Decent Essays

    Distance Dye Lab

    • 156 Words
    • 1 Pages

    Activity 1. Box 4.1 Distance Dye Move Over Time 2-Measurements distance were done by classmates and me from the leading edge of the well to also the leading edge of the dye band. This experiment was done by using methylene blue. Furthermore, we measured distance of the bands that traveled toward the positive or negative pole by using potassium permanganate. We realized that Dyes moved to the positive electrode was the ones that absorbed potassium permanganate because the distance was increasing as the time did so as well.…

    • 156 Words
    • 1 Pages
    Decent Essays
  • Improved Essays

    Purpose The purpose of the experiment Thin Layer Chromatography was to analyze compounds, Aspirin, a mixture of Tylenol with ethanol, acetaminophen, a mixture of Excedrin with ethanol, Aspirin, and acetaminophen, Anthracene, Benzyl, and Triphenylmethanol by separating them with the help of silica plates. Procedure Three silica plates were taken and on the plastic side, they were all marked 1 cm above its end using a pencil. With the help of capillary tube, the first and the second plates were spotted with Aspirin on the left and Acetaminophen on the right. The middle of the first plate was then spotted with the mixture of Tylenol and ethanol but the middle of the second plate was spotted with the mixture of Excedrin and ethanol.…

    • 623 Words
    • 3 Pages
    Improved Essays
  • Improved Essays

    Beaker Experiment

    • 991 Words
    • 4 Pages

    Fill the flat beaker with water. 6. Place a plastic bag over the top of the cylinder and place it, upside down, in the flat beaker. Make sure the bottom of the beaker is always underwater. 7.…

    • 991 Words
    • 4 Pages
    Improved Essays
  • Decent Essays

    In this experiment, the student determined two significant sources of experimental errors: Drop in Volts. Most of the experiments are conducted on Genesys20 Spectrophotometer, which works from the electrical network; all of its components are calculated for certain number of Volt. If this machine has a lack or excess of power, the experiment results a possible failure from the “true” values. Unqualified manufacturing of graduation on the glassware surface.…

    • 149 Words
    • 1 Pages
    Decent Essays
  • Improved Essays

    Assay Lab Report

    • 349 Words
    • 2 Pages

    For part 2 of the experiment (please refer to next page for Figure 3), Figure 3 shows a graph comparing the absorbance readings of the Glutamate-pyruvate aminotransferase serum with the time in minutes. A similar rate was predicted to occur for both the assays done. This would be characterized by the readings showing two parallel lines on the graph. However, the two first readings did not demonstrate this. This could possibly be due to different and insufficient amounts of NADH+ being added to the serums, which would mean that the reaction would not occur as fast as normal and at different rates.…

    • 349 Words
    • 2 Pages
    Improved Essays
  • Improved Essays

    Ppar Rhy Lab Report

    • 585 Words
    • 3 Pages

    6. Quantitative determination of rat peroxisome proliferators activators γ (PPAR-γ): Rat PPAR-γ concentration was determined by ELISA; using a commercially available kit, MyBioSource, Inc., California, USA. The kit utilizing a monoclonal anti- PPAR-γ antibody and an PPAR-γ-HRP conjugate enzyme-linked immunosorbent process assay (ELISA) to determine the concentration of PPAR-γ in serum, plasma, urine, cell culture supernatant and tissue homogenates. Principle of the assay:…

    • 585 Words
    • 3 Pages
    Improved Essays
  • Decent Essays

    Fats And Carbohydrate

    • 302 Words
    • 2 Pages

    Questions: 1. Explain how RER is measured and how RER can be used to reflect usage of substrates. RER or Respiratory Exchange Ratio is measured by the VCO2 expired divided by the VO2 consumed. The RER can reflect what substrate is being used whether fats or carbohydrates based on what it is closest to. The range is .7 (fats) to 1.0 (carbohydrates).…

    • 302 Words
    • 2 Pages
    Decent Essays
  • Great Essays

    • The temperature in the room may have altered. A dependent variable is a variable you do not change during an experiment. Instead, it changes as a result of other changes you make.…

    • 2652 Words
    • 11 Pages
    Great Essays
  • Improved Essays

    Many people are concerned about the dangers that cell phones pose, and they are told that keeping it a greater distance from their heads and using devices such as headphones can diminish much of the radiation; all of which is true. Unfortunately, since the technology industry has made the invention of wireless headphones, if people use their cell phones in conjunction with these devices, the radiation dosage can be even greater than normal. Using Bluetooth, the main type of wireless communication between devices, can have just as much electromagnetic radiation as talking on the phone. One team realized this and began to test the power density rating and the specific absorption rate (SAR), both of which measure radiation dosages. This team…

    • 1625 Words
    • 7 Pages
    Improved Essays