Nt1320 Unit 4 Analysis

Improved Essays
Out of the 13 tests taken at each of the 4 sites, 4 tests will be chosen to compare and contrast the difference in values and reasoning behind those sites. Potassium ion, calcium ion, conductivity, and dissolved oxygen will be compared from Lake Mary, Colorado River Float, Lake Powell Marina, and Lake Powell Swimming locations. These 4 tests will be compared as it takes into account three popular ions found in most bodies of water (salt, calcium, and potassium), as well as the oxygen levels which is one of the most important factors to marine life.
For the potassium ion test, the probe must first be calibrated. To calibrate the potassium ion probe to a LabQuest tablet, first soak the membrane of the probe in the High Standard solution. Then
…show more content…
To calibrate, soak the electrode membrane into a High Standard solution for at least 24 hours prior. Then enter 1000 mg/L as the concentration for ‘reading 1’. When the value stabilizes hit ‘keep’ as it will save the max concentration. Next, remove from High to Low Standard solution and set ‘reading 2’ as 10 mg/L when the value stabilizes. After calibrating, remove the probe and rinse with distilled water, and blot with a paper towel. Dip the electrode membrane into desired testing solution and wait for the value to stabilize. This test, like the potassium ion, is calculating how many calcium ions are present in the solution related to the High and Low Standard solutions of pure calcium (“Calcium Ion-Selective Electrode.” O2 Gas …show more content…
There is nearly a linear correlation between conductivity and concentration of salt or other specific ions. In theory, the conductivity probe is measuring the concentration of those specific ions and their ability to produce an electric current between two electrodes. As the concentration of a specific ion increases so does conductivity. For most experiments, calibrating the conductivity probe is not required. Before using the probe set the range based on µS/cm. In (µS/cm) range 0-200 for low range, 0-2,000 for mid-range, 0-20,000 for high range. Rinse probes membrane with distilled water and slowly insert into desired testing solution and wait for the value to stabilize (“Conductivity Probe.” O2 Gas

Related Documents

  • Improved Essays

    It is important to place the dialysis tubing and the Conductivity Probe the same distance from one another for each test. With the stirring rod, stir the solution for 30 seconds, avoiding contact with the probe or tubing, before collecting data. Once the 30 seconds has passed begin collecting data by clicking on the “collect” box in the computer software. Continue to stir the solution for the entire two minutes of data collection. After two minutes have passed the data collection will automatically stop.…

    • 1263 Words
    • 6 Pages
    Improved Essays
  • Improved Essays

    Each test yielded its own unique set of observations, calculations and results. In order to determine these various values through calculation, the experimental data as well as the chemical reaction of calcium hydroxide was used for all calculations. The chemical equation is displayed below: The first experiment determined the molarity of OH- and Ca2+, the Ksp and the ΔG for saturated calcium hydroxide at 21°C. In order to determine these values, the experimental data as well as the chemical reaction of calcium hydroxide was used.…

    • 889 Words
    • 4 Pages
    Improved Essays
  • Superior Essays

    Enzyme Lab

    • 1376 Words
    • 6 Pages

    With the first test tube, 0 degrees, add 10 drops of the enzyme solution (the teacher will provide the enzyme). Pour the contents from test tube one into a clean 250mL Nalgene bottle and swirl it. Place the O2 Gas Sensor into the bottle making sure it is tight and nothing is able to leak out. After putting the sensor in the bottle, wait 30 seconds and then start collecting data. Data should be collected for 180 seconds.…

    • 1376 Words
    • 6 Pages
    Superior Essays
  • Great Essays

    Once the probe displayed a constant value, the value was recorded, and the probe was taken out of the creek and placed back in the storage solution. The conductivity meter was standardized by placing it into a standard conductivity solution of 10 ± 0.1 µS. The value was recorded, and then the meter was placed into the creek. Once both the conductivity value and the temperature reached a steady value, the data was recorded, and the conductivity meter was removed and put away. The nitrate concentration was measured in two steps.…

    • 1328 Words
    • 6 Pages
    Great Essays
  • Improved Essays

    How does a variation in the concentration of metal ions of the cathode in a Zinc-Copper electrochemical cell affect the electrical potential? Introduction: In a highly technological and quickly developing society, it is imperative to innovation that basic components are understood and explored to further expansion and advancement. In order to understand how everyday necessities such as cars, computers, and cell phones are constantly powered, it must first be understood what is in a battery as well as what happens to the ions within the battery.…

    • 1164 Words
    • 5 Pages
    Improved Essays
  • Improved Essays

    Water Testing Paper

    • 554 Words
    • 3 Pages

    Testing paper possibly suffered from many inaccurate readings due to human contact and vast exposure to air varying results each time. High conductivity presented in the sample of 1037.3 µs/cm suggests salty or mineralised sources of water. This is proven by the common Ions in water that conduct electrical currents including sodium, chloride, calcium and magnesium suggesting the measure of high salinity within the sample. This is due to the dissolved inorganic chemicals and salts which conduct electrical currents therefor the more conductivity; an increase in salinity of 9.6ppt is shown as well. High conductivity values suggest increased amounts of dissolved chemicals are situated in the water.…

    • 554 Words
    • 3 Pages
    Improved Essays
  • Great Essays

    Stiefel laboratories, a GSK company. Summer placement report. By: Shaunagh Linnane 03/06/2014 – 26/08/2014 Table of Contents Introduction & background 3 Training – materials and methods.…

    • 3688 Words
    • 15 Pages
    Great Essays
  • Improved Essays

    Using just the hand-drawn calibration graph to work out the concentrations of the diluted samples did not provide a very accurate result, so I also used the analysis toolpak on excel to work out the equation of the trend lines of the graph, using regression analysis. The concentrations of the diluted samples could then be worked out to a much greater degree of accuracy than by just interpolating them from the hand-drawn…

    • 1627 Words
    • 7 Pages
    Improved Essays
  • Great Essays

    Analytes are detected using a Flame Ionization Detector (FID). The amount of gas in the original liquid sample is directly proportional to the amount of gas in the headspace of a closed vessel. 3.2 Standards are prepared by saturating a solution of water with the gases, and making dilutions from that stock. 4.0 INTERFERENCES 4.1 Since these compounds are highly volatile, care must be taken during sampling to prevent agitation, creation of bubbles or headspace during storage. Transfer all samples quickly to the sampler vials while still cold, if possible.…

    • 1252 Words
    • 5 Pages
    Great Essays
  • Superior Essays

    Gravimetric Analysis Lab

    • 1150 Words
    • 5 Pages

    Gravimetric Determinations of BaCl2 from a mixed sample Dharmin Chauhan CHEM 1411 June, 25th, 2016 Introduction Gravimetric Determinations of BaCl2 from a mixed sample is the name of the experiment. This experiment also known as gravimetric analysis. Gravimetric analysis is the calculation of analyte concentration precipitation, isolation, and weighing (Filho, 2014). The two key concepts to note in this experiments are Gravimetric analysis and Mole Calculations.…

    • 1150 Words
    • 5 Pages
    Superior Essays
  • Improved Essays

    In this experiment, various electrochemical cells were measured and calculated to determine their electrical cell potential. In the electrochemical cell Zn (s) │Zn2+ (0.10 M) ││ Cu2+ (0.10 M) │ Cu(s), E cell potential was determined to be 1.040 V while the standard E˚ cell potential was 1.10 V. In the electrochemical cell Zn (s) │Zn2+ (0.10 M) ││ Cu2+ (1.0 M) │ Cu(s) E cell potential was determined to be 1.060 V while the standard E˚ cell potential was 1.10 V. In the electrochemical cell Cu (s) │Cu2+ (0.010 M) ││ Cu2+ (1.0 M) │ Cu(s), E cell potential was determined to be 0.032 V while the standard E˚ cell potential was 0.00 V. The E cell value in the electrochemical cell Cu (s) │Cu2+ ( M ) ││ Cu2+ ( M) │ Cu(s) was determined…

    • 1122 Words
    • 5 Pages
    Improved Essays
  • Improved Essays

    The steps should then be repeated for the other 4 temperatures. (for the other 4 test tubes the 5 water baths should be heated to these temperatures 25°C, 35°C, 65°C and room temperature). Section 2: Data Collection and Processing 2.1 Recording Raw Data Trial 24°C 35°C 45°C…

    • 959 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    Electronic materials is very useful in today 's world, they can be used almost in all field. There are a lot of electronic materials produced every day, like semiconductor, Piezoelectric and ferroelectric materials, magnetic material etc. This lab is more concentrate on the electronic resistivity with the changing of annealing time. And XRD can help people figure out the grain size of the sample, and find the rule of the electronic resistivity with the changing of annealing time.…

    • 916 Words
    • 4 Pages
    Improved Essays
  • Improved Essays

    10. Put the cuvettes in the colorimeter and record the value. Calibrate the colorimeter after every measurement. 11. Repeat daily.…

    • 1347 Words
    • 6 Pages
    Improved Essays
  • Improved Essays

    The electrode of pH meter was calibrated with standard buffer solutions of pH 4, 7 and 9. The electrode was then dipped in the sample and its pH was noted down. B.DETERMINATION OF ELECTRICAL CONDUCTIVITY: Conductivity was measured by using soil and water analyzed kit. The electrode was conditioned with KCl and calibrates with 0.01M KCl solution. The sample…

    • 1564 Words
    • 7 Pages
    Improved Essays