Process indicator (Class 1)
Ex: indicator type
This type CI is used to differentiate packages that have been processed in a sterilizer from those that have not been processes.
Indicators for specific tests Class 2
Are used in certain test procedures .
Ex: Bowie-Dick test for the air-removal performance of prevacuum sterilizer
Single-variable indicator Class 3
Are designed to respond to only one variable of the sterilization process that they are designed to monitor.

ex: glass ampule containing a chemical that melts when the sterilization temperature is achieved

uses in internal chemical indicators

Multi-variable indication class 4

Designed to react to two or more of the variable of the sterilization process that they are designed to monitor .

Integrating indicator class 5

Designed to react to all of the critical variables of the specific sterilization process that they are designed to monitor.




parallels to the performance of BIs.





Emulating indicator class 6

designed to react to all critical variables of a specific sterilization cycle



should only be used in specific cycle