At what point in the xray circuit os the rectification circuit located?

Between the step up transformer and the xray tube

Where is the stator of the anode induction motor located?

Outside the glass envelope and outside the rotor.

The exposure switch on an xray machine is intended to?

Initiate exposure

The exposure switch on an xray machine is intended to?

Initiate the exposure

The exposure switch is found between the?

Auto transformer and the timer circuit

The electrical device used to adjust the mA selection a/an?

Variable resistor

The time control is found between the?

Exposure switch and the step-down transformer

The mA control is found between the?

Incoming line and the step-down transformer

The kvp control is found between the?

Incoming line and the exposure switch

All of the following. Are radiographer operated controls except?

Rotor speed

The diagnostic x-ray range is approximately?

10-1200 mA, 0.001 to 10 seconds, and 25-150 kvp

The key advantage of a curved-top radiographic table is the?

Smaller OID

When using AEC sat 80 kvp, back up times?

Must not exceed 600 mAs

Radiographic rooms equipped with a tilting table are primarily designed for performing ____ procedures.

Fluoroscopic

All of the following are X-ray tube support systems expect?

Fluoroscopic tower

To protect he radiographer during xray exposures...

He mA and kVp selections are on the low voltage side of the circuit.

Which tube suspension system offers the greatest multidirectional movement?

Overhead tube suspension system

The incoming line current is supplied?

In the form of a three-phase power cycle

Nearly all types of xray equipment operate from an incoming line of?

210 v

When full wave rectification is applied to three-phase current, a _____ ripple is produced.

Voltage

A three-phase, rectified current produces a voltage ripple of ___ pulses per HZ.

6

Ancillary equipment for tilting tables includes?

A footboard, shoulder supports, and handgrips

Al of the following are desired features of radiographic tables except?

Low abrasion surface to lessen patient slippage as they transfer

High frequency generators?

All of the above

Falling load generators?

Are designed to use the highlight allowable mA

Radiographing a thin oft part using automatic exposure control (AEC) may?

Produce inconsistent images due to AEC timer errors.