Name the first five things used to optimize 2D Image
1
2
3
4
5
1. Gain
2. TGC
3. Depth
4. Focus
5. Dynamic Range
What controls affect the frame rate?

1
2
3
4
5
6
1. Depth
2. Write Zoom
3. Sector Size/Wide View
4. Multiple Focal pts
5. Frame Rate Function
6. Spatial Compounding
What 4 functions BESIDES magnification allow you to adjust the frozen image?

1
2
3
4
1. Gain
2. Dynamic Range
3. TGC
4. Gamma/Grayscale
Name and Explain the 3 components of Tri-phasic Flow.

1st Component:
2nd Component:
3rd Component:
1st: Forward Flow; systolic pumping of the heart
2nd:Reversal Flow; resistance caused by distal pressure
3rd:Low velocity forward flow; Elastic recoil of vessel during diastole
How do you avoid aliasing in spectral Doppler?

1
2
3
4
5
6
Increase PRF
Use a lower frequency Transducer
Increase Angle
Decrease Depth
Adjust Baseline
Use CW Doppler
Name 5 things you should address first when obtaining spectral waveform.
1
2
3
4
5
1. Good Long-Axis View
2. Adjust Cursor Parallel to flow, with proper gate size and location
3. Gain
4. Scale
5. Wall Filter
To adjust the angle so that it is < or = 60 degrees, parallel to flow or vessel wall, technologist can...
1. Heel/Toe Transducer
2. Steer Line of Insonation
3. Adjust Cursor Angle
What can you do to eliminate spectral broadening?

1
2
3
4
5
6
1. Increase Wall Filter
2. Increase Dynamic Range
3. Decrease Gain
4. Decrease Gate Size
5. Adjust line of insonation to get more parallel w/flow
6. Make sure vessel is in true long-axis, no salami
No velocities are detectable in a VERY tight stenosis. What can you do to demonstrate flow in the vessel?

1
2
3
4
5
1. Decrease Dynamic Range (Increase Compression)
2. Decrease Wall Filter
3. Decrease Scale
4. Increase Gate Size
5. Increase Gain
Spectrum is cut off, how can you eliminate this artifact?
1
2
3
4
5
6
1. Increase Scale/PRF
2. Increase Angle
3. Decrease Depth
4. Decrease Frequency
5. Adjust Baseline
6. Use CW doppler (unlimited PRF)
Doppler Signal is Audible, but spectral waveform cannot be seen, (Assume PW is on). What can you do to fix it?

1
2
3
1. Increase Gain
2. Decrease Dynamic Range
3. Adjust Gamma/Grayscale
You are imaging a healthy vessel with blood flow through it, no color can be seen. What can you do to improve it?

1
2
3
4
5
6
7
Increase Gain
Increase Power
Decrease Wall Filter
Decrease Scale
Decrease Frequency
Adjust color box angle to be parallel w/flow
Increase Data Number
Color Doesn't fill the entire vessel. What can you do to improve color filling?

1
2
3
4
5
6
7
Decrease Scale
Decrease CF Frequency
Decrease Wall Filter
Increase Data Number
Increase Gain
Increase Acoustic Power
Increase Frame interpolate
Color display shows aliasing. how can you elminiate this artifact?

1
2
3
4
5
Increase Scale
decrease Frequency
Decrease Depth
Increase Angle
Adjust baseline
The frame rate is too slow. What can be done to adjust it?

1
2
3
4
Increase FR Function
decrease Data Number
decrease Depth
decrease depth/width of box
Color image is noisy. what can you do to fix it?

1
2
3
4
5
Decrease Gain
Increase Filter
Increase Scale
Increase Time Smoothing
Increase persistance