• Shuffle
    Toggle On
    Toggle Off
  • Alphabetize
    Toggle On
    Toggle Off
  • Front First
    Toggle On
    Toggle Off
  • Both Sides
    Toggle On
    Toggle Off
  • Read
    Toggle On
    Toggle Off
Reading...
Front

Card Range To Study

through

image

Play button

image

Play button

image

Progress

1/62

Click to flip

Use LEFT and RIGHT arrow keys to navigate between flashcards;

Use UP and DOWN arrow keys to flip the card;

H to show hint;

A reads text to speech;

62 Cards in this Set

  • Front
  • Back
Explain the dose gradient of a CT
Figure 2. Dose gradient resulting from a full
360° exposure from a CT scan. The thicker lines
represent the entrance skin dose, which is much
larger than the dose at the inner radius, repre-
sented by the thinner lines. This difference re-
sults in a radially symmetric radiation dose gradi-
ent within the patient.
What has a greater dosage of radiation inner or outer?
Figure 3. Typical dose measurements in a 32-
cm-diameter (body) phantom from a single-de-
tector CT scan. Values measured at the center
and periphery (1 cm below the surface) positions
within a polymethyl methacrylate circular dosim-
etry phantom demonstrate a radial dose gradient
with a 2:1 ratio from periphery to center.
collimation.
What is a body phantom typically made out of
polymethyl methacrylate
How much radiation is obtained at the periphery compared to the center in a CT
center are typically about 50% of the measured
value obtained at one of the peripheral positions.
What is a consideration if the there is a smaller body part or phantum used
However, for a smaller-diameter
phantom—the 16-cm-diameter phantom referred
to as the head phantom—measured under the
identical exposure conditions, the center value
reading climbs to approximately 40 mGy, as do peripheral values
Are the radiation levels similar in the periphery compared to the center in a smaller object such as a head
yes
Are the values of absorbed dose size dependent
yes
What are Z-axis radiation variations
In addition to the variations within the scan
plane, there are variations along the length of the
patient or phantom.
What is does the Z-axis variation of a CT look like
Figure 5. Radiation profile of a full-rotation CT scan
measured at isocenter. This profile is the distribution of
radiation dose along the axis of the patient (the z axis)
and is known as D(z).
Explain why Z-axis variation occurs
When multiple adjacent scans
are performed, the tails of the radiation profiles
from adjacent scans can contribute to the ab-
sorbed dose outside of the primary area being im-
aged. If these tails are significant and are nonzero
at some distance from the location of the originat-
ing section, then these contributions can add up,
creating additional absorbed dose in the primary
area being imaged.
What influences the the z - axis radiation variation
The size of the contributions from adjacent
sections is very directly related to the spacing of
sections and the width and shape of the radiation
profile.
What is the MSAD (Multiple Scan Average Dose) used to account for (during one CT)
to account for the effect of multiple scans. This was used before CTDI
Why was CTDI made
Computed Tomogra-
phy Dose Index was created to computed the radiation dose, normalized to beam width,
measured from 14 contiguous sections:
Why was CTDI not used
only 14 sections could be measured
and one had to measure the radiation dose pro-
file—typically done with thermoluminescent do-
simeters (TLDs) or film, neither was very convienent
What was created to overcome the short fall of the CTDI
CTDI 100
What is different about the CDTI
This index relaxed the constraint
on 14 sections and allowed calculation of the in-
dex for 100 mm along the length of an entire pen-
cil ionization chamber

(pencil ionizing chambers are 100mm)
Does the amount of section size matter with the CTDI100
no, but the width does matter
Does there have to be a seperate calculation for the periphery and the center with CDTI 100
yes, to describe variation in the scan plan
Why was CTDIw created
CTDIw was created to represent a dose index
that provides a weighted average of the center an
peripheral contributions to dose within the scan plan
What was then created to acount for the pitch
CTDIvol
What is the pitch
pitch is defined as table distance traveled in
one 360° rotation/total collimated width of the
x-ray beam
What is DLP
DLP = CTDIvol x scan length.
What is another way of saying x-ray beam energy
kilovolt peak
What are some direct influences on radiation dose
beam energy
tube current
rotation or exposure time
section thickness
object thickness
pitch
dose reduction techniques
distance from x ray tube to isocenter
What are ways of reducing dosage in CT
tube current variation
modulation
How is the beam xray energy choosen
by the technologist who will choose a kilovolt peak
What are examples of the absorbed dose in the head and body with changes in kVp (with everything else held constant)
Dose current time product have a direct influence on radiation dosage
yes
What are the units of current time products when discussing CT
milliampere-second (mAs)
Is the radiation dose directly proportional to the miliampere-seconds value
yes
What has a greater increase in radiation increasing the kVp from 120 to 140 or the mAs from 100 to 200
the kVp
If all factors are held constant what happens to the radiation dosage if the pitch is increased
the radiation dose with decrease
How much the pitch influence radiation dosage if everything else is held constant
How much does collimation effect radiation sosage if everything else is held constant
Does collimnation have a small effect on radiation dosage
yes
How much affect on radiation occurs if the current is adjusted with everything else held constant
What has the most effect on radiation dosage
kVp
What has the least effect on radiation dosage
collimnation
What is the pitch
defined as
table distance traveled in one 360° rotation/total
collimated width of the x-ray beam
What is the only CTDI descriptor that takes pitch into account
CTDIvol (note that DLP also takes pitch into account)
Does thinner colimnation result in a greater degree of overlap and therefore a higher CTDI value
yes
Is the degree of radiation dose by using thinner colimmators greatly enhanced when using mulitidector scanners
yes (the previous table was a single detector scanner)
What causes less radiation; thick or thin collimation
thick
How does changes in collimation change radiation dose in a MULTIDETECTOR CT
Who generally receives more radiation; a small or large person
small person (important when it comes to pediatrics)
Why do smaller people generally recieve a greater amount of radiation
as the source moves around the patient.
For smaller patients, the exit radiation has been
attenuated by less tissue and therefore is closer to
the entrance radiation in its intensity, resulting in
a much more uniform dose distribution
How does the tube current modulation work
changes in tube cur-
rent based on the estimated attenuation of the
patient at a specific location. Thus, the tube cur-
rent will be programmed to a maximum value and
can be reduced when there is information that a
location along the patient is expected to be less
attenuating than the most attenuating location to
be imaged
Can tube current modulation vary by length of the patient (ie less for the chest than for a fat mans stomach)
yes
What happens to the signal to noise if thinner reconstruction thicknesses are used
decreased signal to noise
How do you "offset" the increased noise created by thin slice reconstruction
by increasing the kVp or mAs
What percent does a 50% decrease in mAs increase noise
41%
What is the trade off in increasing pitch
increasing the effective slice thickness which inturn will cause increased volume averaging (decreased signal)
What happens to the signal to noise if thinner reconstruction thicknesses are used
decreased signal to noise
What is a technique for reducing radiation to smaller people
size and weight based protocols
How do you "offset" the increased noise created by thin slice reconstruction
by increasing the kVp or mAs
What percent does a 50% decrease in mAs increase noise
41%
What is the trade off in increasing pitch
increasing the effective slice thickness which inturn will cause increased volume averaging (decreased signal)
What is a technique for reducing radiation to smaller people
size and weight based protocols
What is the Monte Carlo simulations
This hermaphro-
dite patient model uses geometric shapes (cylin-
ders, spheres, cones) to approximate the shape
and location of all radiosensitive organs in the
body;From these simulations, radiation doses for each
organ in each imaged section were calculated and
could be tallied to estimate the organ dose for
each organ. These organ doses could then be
combined with appropriate weighting factors to
estimate effective dose (or effective dose equiva-
lent)
How is the monte carlo simulation used by the tech
This work formed the basis for several software
programs that have taken the results of these monte carlo
simulations and put an interface on them to allow
users to input some technical parameters and cal-
culate an effective dose.
What is another type of simulation used to calculate effective dose
weltforschung (GSF) performed simulations on
two different mathematical sex-specific phan-
toms, “Adam” and “Eva”
Are there other methods
yes, but not sure I have to know this shit so go back and continue if this comes up "CT-radiation dosage of...."