Use LEFT and RIGHT arrow keys to navigate between flashcards;
Use UP and DOWN arrow keys to flip the card;
H to show hint;
A reads text to speech;
104 Cards in this Set
- Front
- Back
scatter radiation produced by what type of interaction? |
compton interaction |
|
compton interactions..... |
are of NO USE in demonstrating the structures of interest. |
|
scatter radiation reaching the film will increase/decrease the radiographic density?? |
increase (without improving the diagnostic value of the information) |
|
scatter radiation will reduce/improve the contrast of the radiographic image? |
reduce (adds more shades of grey) |
|
relationship between scatter and radiographic density? direct or indirect? |
direct (more scatter = more density) |
|
scatter related to contrast?? direct or indirect? |
indirectly proportional more scatter = low contrast (more grays) less scatter = high contrast (more b&w) |
|
increasing the kV will increase/decrease the penetrating ability of the x-ray photons?? |
increase |
|
if no other technical factor is changed other than kV what happens to the overall density of the radiograph? |
it will increase since more of the photons are exiting the patient and striking the film |
|
increasing kV will increase/decrease the penetrating ability of the beam? |
increase |
|
increasing kV causes an increasing number of what occurring within the patient?? |
compton interactions
(this increases the amount of scatter) |
|
Increase in kV will cause the number of photoelectric absorptions to increase/decrease? |
decrease
|
|
A decrease in PE absorptions will increase/decrease the patient dose? |
decrease |
|
increasing kVp and reducing mAs will ?? |
increase beam penetrability
maintain radiographic density
increase compton interactions causing more scatter
reduce image quality |
|
in irradiated material, as tissue volume increases the amount of scatter radiation increases/decreases?? |
increases |
|
high atomic # materials include: |
iodine, barium, and lead |
|
the higher the atomic # of the material stuck by the x-ray photon.... |
the more PE interactions... therefore reduce the amount of compton interactions (scatter) |
|
when using beam restrictions the number of x-ray photons reaching the patient will increase/decrease? |
decrease |
|
When using beam restriction, the need for an increase/decrease of technical factors? |
increase technical factors to compensate for overall reduction in x-ray photons |
|
three types of beam restrictors |
aperture diaphragms
cones/cylinders
collimators |
|
aperture diaphragm |
the simplest type of beam-restricting device
flat piece of metal (lead) with an opening to allow radiation to pass through |
|
what shape is the opening of an aperture diaphragm? |
the opening can be any shape. However, once it is cut, you can not re cut it. |
|
where is the aperture diaphragm located? |
placed on the bottom of the collimator or taped directly to the tube housing |
|
what is an advantage of an aperture diaphragm? |
they are inexpensive |
|
what type of machines is the aperture diaphragm used on? |
mammo machines. not used with modern diagnostic machines |
|
cones/cylinders |
circular aperture diaphragm with metal extensions.
cones flared at the end
cylinders are constant in diameter |
|
with adjustable cylinders, as cylinder length increases the amount of radiation reaching the patient and film increases/decreases?? |
decreases |
|
where are cylinders located?? |
they attach beneath and on the outside of the tube housing. |
|
increasing cylinder length will do what to the field size? |
decrease the field size |
|
how do you determine the cone field size?? |
projected image size = SID x lower diameter of cone/ distance from focal spot to bottom of cone |
|
using a 3" cylinder, 40" SID, the bottom of the cylinder is 12" below the focal spot. How big will the approximate field size be? |
40" x 3" / 12
120"/12
projected image size = 10" circle |
|
what is the most frequently used method of beam restriction? |
collimators |
|
collimators |
contain two sets of lead shutters to absorb radiation. knobs to control the longitudinal and horizontal fields. |
|
bottom set of shutters |
reduce penumbra along the periphery of the beam |
|
upper shutters |
prevent most of the off focus radiation from reaching the film |
|
what is the purpose of the mirror? |
acts as a filter and absorbs low energy x-rays.
simulates the path of the x-ray photons |
|
alignment accuracy of the light source and x-ray beam can be checked by??? |
doing a collimator test |
|
acceptable limits are determined by?? |
the department quality standards |
|
What is the U.S. Federal Standard? |
2% SID error |
|
Automatic Collimators also called ??? |
Positive Beam Limitation (PBL) device |
|
automatic collimators/PBL |
will adjust the dimension of the x-ray beam to the image receptor placed within the bucky |
|
ancillary devices |
lead blockers lead masks cut to a specific shape or size |
|
purpose of the grid |
used to improve the contrast of the image by absorbing the patient generated scatter radiation before it can strike the image receptor |
|
scatter creates _________ density on the image which reduces contrast? |
inaccurate |
|
grids are located __________ the patient and ________ the image receptor? |
after patient before image receptor
|
|
grids are used on what size body parts? |
thick 10 cm or larger |
|
grids are used on procedures utilizing what type of kVp settings? |
high 70 kVp or higher |
|
Grid Construction |
flat device radiolucent material separating thin, vertical radiopaque strips. |
|
The vertical strips are made of what? designed to do what? |
made of lead designed to absorb scatter radiation |
|
what are the radiolucent interspace materials designed to do? |
Allow the remnant photons to strike the film |
|
Rectangular grids, strip orientation |
strips running with the long axis of the rectangle |
|
Grid ratio is |
The relationship of the HEIGHT of the lead strips to the DISTANCE BETWEEN the strips. |
|
grid ratio equation |
grid ratio = h/D |
|
a/an _____________ relationship exists between the distance between the lead strips and grid ratio when the height of the grid strips remains the same. |
inverse |
|
which grid ratio stops scatter radiation better |
high grid ratio |
|
which grid ratio results in higher contrast images? |
high grid ratios |
|
Which require a higher adjustment in the radiographic technique to maintain the density of a non-grid technique? |
high grid ratios |
|
Grid Frequency |
The number of vertical strips per unit of grid width |
|
as the frequency per unit of distance increases the vertical strips... |
will be thinner |
|
grid patterns |
linear
cross hatched |
|
linear |
parallel strips of lead allows tube angulation if parallel with the lead strips |
|
Cross-hatched |
tow linear patterns criss-crossed
more efficient at removing scatter radiation
no tube angulation |
|
Focused Grids |
linear positioned lead strips
angled similar to the angulation of the primary beam as it diverges
has a point of convergence similar to the primary beam |
|
focused grids require proper SID to be efficient at reducing ______ and simultaneously allowing the remnant x-ray photons to pass. |
scatter |
|
what is extremely important when using focused grids? |
focal range is extremely important!! |
|
latitude |
freedom from the usually restraints, limitations, or regulations. |
|
high latitude |
larger allowable changes (forgiving) fewer restraints
|
|
grid latitude |
amount you can be off and still get an acceptable x-ray |
|
as grid ratio increases the need for proper SID _________. increases/decreases |
increases |
|
As grid ratio increases the latitude for SID _____________. increases/decreases |
decreases |
|
In high grid ratios small error in SID can cause.... |
a large amount of grid cutoff |
|
Low grid ratios have a ______ latitude regarding SID |
High |
|
low grid ratios regarding SID error |
small or large errors may not cause much grid cutoff |
|
grid cut off |
improper alignment of the grid will reduce the amount of remnant radiation from striking the image receptor. |
|
permanent grid location |
located within the potter-bucky diaphragm, above the image receptor location |
|
permanent grid lines |
will be parallel with the table length, or vertical direction of the wall bucky |
|
portable grids |
can be encased within a plastic holder into which the cassette is inserted
can be taped to an existing cassette
can be a permanent part of the cassette construction |
|
Relationship of grids and radiographic density
|
as grid ratio or frequency increases the radiographic density decreases
|
|
To avoid Moire artifacts grid frequency for digital imaging should be higher than |
80 lpi
(bucky device should also be used whenever possible) |
|
increasing grid efficiency will |
decrease radiographic density by reducing the number of x-ray photons striking the image receptor |
|
relationships of grids and radiographic contrast |
as grid ratio or frequency increases the radiographic contrast increases. |
|
Increasing grid efficiency will |
increase (create high) radiographic contrast
-by reducing the number of different energies of x-ray photons striking the image receptor. |
|
Grid cost |
increases with effectiveness increases with cassette size
they are hundreds of dollars |
|
which are used more frequently, focused or parallel grids??? |
focused |
|
grid conversions factors |
constants used to determine the change in technical factors needed for specific grid ratios. |
|
Because grids attenuate remnant primary beam, what must be adjusted when changing grid ratio? |
mAs |
|
To maintain density as grid ratio increases mAs must |
increase to offset the number of photons being attenuated by the grid |
|
Grid Conversions Factors |
see slide for equation and mAs increase |
|
increasing the grid ratio will _________ the amount of scatter radiation and the radiographic density if the mAs is not increased to maintain density |
decrease |
|
types of grid errors |
off level off center off focus upside down
|
|
off level grid error |
occurs when the tube is angled across the long axis of the grid strips OR the grid is unleveled in comparison to the tube
**portable grids |
|
off level grid appearance |
decrease in density across the entire image |
|
off center grid errors |
x-ray tube is not centered with the center of the focused grid.
**fixed or portable grids |
|
off center grid errors appearance |
decrease in density across the entire image. The greater the lateral off-centering, the greater the grid cut-off. |
|
off focus grid errors |
SID outside the acceptable range of the focused grid
**fixed or portable grids
|
|
off focus grid appearance |
decrease in density at the lateral margins of image |
|
upside down grid error |
center of the grid will allow the x-ray beam to pass unrestricted flipped upside down or not |
|
upside down grid appearance |
severe decrease in density at the lateral margins of image |
|
other scatter reduction methods |
collimating is still necessary with the use of grids
air gap technique
reverse cassette technique
|
|
air gap technique |
alternate method to the use of grid
large OID
the amount of scatter will stay consistent however the amount reaching the image receptor will be reduce with OID |
|
Reverse- Cassette technique |
flip the cassette so the remnant radiation strikes the back side of the cassette first
lead foil on the back side of the intensifying screen to prevent backscatter
foil acts as a filter stopping compton radiaton
must increase mAs to maintain density |
|
increasing grid ratio will ____________ radiographic density. |
decrease |
|
High grid ratio compared to low grid ratio |
high grid ratios have a narrow latitude conceding SID and alignment
High grid ratios have a larger GCF |
|
Grids do not CREATE ________ they ABSORB ___________ |
SCATTER |
|
What creates scatter? |
The patient |